Instruction Manual – Bedienungsanleitung

PerfectBlue™
Vertical Double Gel Systems /
Vertikale Doppelgelsysteme
Twin S, M, L & Twin ExW S
INSTRUCTION MANUAL - BEDIENUNGSANLEITUNG PERFECTBLUE VERTICAL DOUBLE GEL SYSTEMS / VERTIKALE DOPPELGELSYSTEME - TWIN S, M, L & TWIN EXW S - VWR
INSTRUCTION MANUAL - BEDIENUNGSANLEITUNG PERFECTBLUE VERTICAL DOUBLE GEL SYSTEMS / VERTIKALE DOPPELGELSYSTEME - TWIN S, M, L & TWIN EXW S - VWR
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

CONTENTS
SHORT INSTRUCTION                                                  2
WARRANTY                                                           4
SAFETY PRECAUTIONS                                                 4
PACKAGING LIST                                                     5
SYSTEM OVERVIEW                                                    6
    Technical properties                                           6
GENERAL INSTRUCTIONS                                               7
    Leak proof sealing of glass plates                             7
    Assembling the system and electrophoresis                      9
    Cleaning                                                      10
REQUIRED MATERIALS & RECIPES                                      11
    Required gel volumes                                          11
    Required buffer volumes                                       11
    Tris-Glycine SDS-PAGE                                         11
    Tris-Tricine SDS-PAGE                                         12
TROUBLESHOOTING                                                   13
TECHNICAL SUPPORT AND ORDERING INFORMATION                        15
    Power Supplies                                                15
    PerfectBlue™ Double Gel System Twin S                         15
    PerfectBlue™ Double Gel System Twin M                         16
    PerfectBlue™ Double Gel System Twin L                         16
    PerfectBlue™ Double Gel System Twin ExW S                     17

VWR_v0617_E                                                   1
INSTRUCTION MANUAL - BEDIENUNGSANLEITUNG PERFECTBLUE VERTICAL DOUBLE GEL SYSTEMS / VERTIKALE DOPPELGELSYSTEME - TWIN S, M, L & TWIN EXW S - VWR
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

SHORT INSTRUCTION

Systems overview
1. Upper buffer chamber (UBC)
   with side clamps and electrodes
2. Safety lid with power cords                                1
3. Lower buffer chamber (LBC)                                     2
4. Casting base

                                                         3

                                                                      4

Leak proof sealing of glass plates

Invert casting base, so the black silicone mats face down-
ward and the acrylic square upward.

Place UBC on inverted casting base. Arrange gel assem-
blies and place on each side of the UBC. It is important that
the bottom of the glass plates and the spacers sit flat on the
acrylic square of the inverted casting base. Tighten side
clamps. Tighten the blocking plate in a similar way at the
other side of the UBC in case of casting only one gel.

Invert casting base and turn out the cams.
Wet the silicone mats with e.g. water or
buffer to prolong their life span.
Place UBC on the gaskets and turn the
handles at the same time, half a turn.

VWR_v0617_E                                                   2
INSTRUCTION MANUAL - BEDIENUNGSANLEITUNG PERFECTBLUE VERTICAL DOUBLE GEL SYSTEMS / VERTIKALE DOPPELGELSYSTEME - TWIN S, M, L & TWIN EXW S - VWR
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

Assembling the system and electrophoresis

The gel can now be poured. Pour slowly to avoid bubbles.

After polymerization, lift UBC from casting base and
place into the LBC. Add buffer to the UBC. Load the
samples. Add buffer to the LBC.

Place safety lid onto the unit and start the run. After the run
is completed turn off the power supply and remove lid. Lift
UBC from LBC, loosen side clamps to remove gel assembly
for further analysis.

VWR_v0617_E                                                   3
INSTRUCTION MANUAL - BEDIENUNGSANLEITUNG PERFECTBLUE VERTICAL DOUBLE GEL SYSTEMS / VERTIKALE DOPPELGELSYSTEME - TWIN S, M, L & TWIN EXW S - VWR
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

WARRANTY
VWR guarantees that the vertical electrophoresis system you have received has been thoroughly tested
and meets its published specification.

However, immediately upon arrival, please check carefully that the shipment is complete and has not
been damaged in transit. For missing parts or to report any kind of damage, please contact VWR (see
'TECHNICAL SUPPORT AND ORDERING INFORMATION'). Please retain all packaging materials until
the delivery has been completely checked since this will speed up the return of goods if required and
reduce environmental impact. Any form of returns, replacements or credit notes must be agreed in
advance by VWR.

For the complete range of PerfectBlue™ electrophoresis and blotting systems, VWR guarantees a warran-
ty period of 36 months if the products have been used solely according to the instruction manual unless
a different warranty has been offered in writing. No liability is accepted for loss or damage arising from
incorrect use. VWR's liability is limited to the repair or replacement of the unit or refund of the purchase
price, at VWR's discretion. VWR is not liable for any consequential damages. After the warranty period
has expired VWR can offer repairs.

VWR reserves the right to alter the technical specifications of the PerfectBlue™ electrophoresis or blotting
systems without prior notice. This will enable us to implement developments as soon as they arise.

SAFETY PRECAUTIONS
  Please, read this Instruction Manual carefully before using the gel system.
  Only use a CE marked DC power supply.
  Always disconnect the system from the power supply before removing the safety lid.
  Always disconnect the gel system from the power supply when it is not in use or before moving it.
  Running conditions for this unit should not exceed the maximum operating voltage or current.
   The flow rate for cooling should not exceed 2 liter per minute if the chamber is connected to a water
   tab. If using a water circulator the flow should not exceed 12 liter per minute.

VWR_v0617_E                                                   4
INSTRUCTION MANUAL - BEDIENUNGSANLEITUNG PERFECTBLUE VERTICAL DOUBLE GEL SYSTEMS / VERTIKALE DOPPELGELSYSTEME - TWIN S, M, L & TWIN EXW S - VWR
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

PACKAGING LIST
Unless requested otherwise, the following items are included in shipment for the models Twin S, Twin M,
Twin L and Twin ExW S:

Twin S (PEQL45-1010-I)                                            Twin M (PEQL45-1614-I)
    Upper buffer chamber with electrodes                              Upper buffer chamber with electrodes
    Lower buffer chamber                                              Lower buffer chamber
    Safety Lid with power cords                                       Safety Lid with power cords
    4 blank glass plates                                              4 blank glass plates
    2 notched glass plates                                            4 notched glass plates
    2 notched alumina ceramic plates                                  4 clamp assemblies
    4 clamp assemblies                                                2 combs, 1.5 mm thick, 10 teeth
    2 combs, 0.8 mm thick, 10 teeth                                   2 combs, 1.5 mm thick, 15 teeth
    4 side spacers, 0.8 mm thick                                      8 side spacers, 1.5 mm thick
    2 bottom spacers, 0.8 mm thick                                    4 bottom spacers, 1.5 mm thick
    2 replacement gaskets                                             2 replacement gaskets
    4 replacement spring pegs                                         4 replacement spring pegs
    1 blocking plate for single gel operation                         1 blocking plate for single gel operation
    1 SpacerPlacer for easy spacer alignment                          1 SpacerPlacer for easy spacer alignment

Twin S (PEQL45-1010-C)                                            Twin M (PEQL45-1614-C)
as the PEQL45-1010-I but including casting base                   as the PEQL45-1614-I but including casting base

Twin ExW S (PEQL45-2010-I)                                        Twin L (PEQL45-2020-I)
    Upper buffer chamber with electrodes                              Upper buffer chamber with electrodes
    Lower buffer chamber                                              Lower buffer chamber
    Safety Lid with power cords                                       Safety Lid with power cords
    4 blank glass plates                                              4 blank glass plates
    4 notched glass plates                                            4 notched glass plates
    4 clamp assemblies                                                4 clamp assemblies
    2 combs, 0.8 mm thick, 15 teeth                                   2 combs, 1.5 mm thick, 15 teeth
    2 combs, 0.8 mm thick, 20 teeth                                   2 combs, 1.5 mm thick, 20 teeth
    8 side spacers, 0.8 mm thick                                      8 side spacers, 1.5 mm thick
    4 bottom spacers, 0.8 mm thick                                    4 bottom spacers, 1.5 mm thick
    2 replacement gaskets                                             2 replacement gaskets
    4 replacement spring pegs                                         4 replacement spring pegs
    1 blocking plate for single gel operation                         1 blocking plate for single gel operation
    1 SpacerPlacer for easy spacer alignment                          1 SpacerPlacer for easy spacer alignment

Twin ExW S (PEQL45-2010-C)                                        Twin L (PEQL45-2020-C)
as the PEQL45-2010-I but including casting base                   as the PEQL45-2020-I but including casting base

VWR_v0617_E                                                   5
INSTRUCTION MANUAL - BEDIENUNGSANLEITUNG PERFECTBLUE VERTICAL DOUBLE GEL SYSTEMS / VERTIKALE DOPPELGELSYSTEME - TWIN S, M, L & TWIN EXW S - VWR
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

SYSTEM OVERVIEW
PerfectBlue™ vertical double gel systems Twin S, M, L and ExW S are characterized by sophisticated
design and durable construction. With their unique arrangement of 3 cathodes and an extended anode
surrounding the complete base of the electrophoresis chamber, electrophoretic separations of excep-
tional high quality can be achieved. All of the systems are equipped with an alumina ceramic cooling
system for water circulation matching effective cooling and high migration rates without negatively
affecting separation results. Using the casting system developed for the PerfectBlue™ vertical double gel
systems up to 2 gels can get poured at once, conveniently and without leakage. Designed for conven-
ience, the glass plates do not have to be re-assembled after polymerization and electrophoresis can
performed straight away. This time saving step further eliminates potential sources of error.
PerfectBlue™ vertical double gel systems are available in 4 different formats, each one including a wide
range of accessories for increased versatility. The systems are constructed using solid parts, precisely
machined from 0.6 mm thick acrylic by CNC milling. The use of specialized adhesives neutralizes
mechanical forces and thermal expansions making the system durable and reliable. We are therefore
confident of our guarantee, a full 3 year warranty on the whole range of PerfectBlue™ electrophoresis
and blotting equipment.

                         Upper buffer chamber (UBC)
                         with side clamps and electrodes             Safety lid with power cords

                      Lower buffer chamber (LBC)                     Casting base

Technical properties
PerfectBlue™ Cat. No.                     Gel size required buffer      maximal      max. current       typical
                                          (W x L)      volume           Voltage        per gel       running time
                                           10 x 10
Twin S           PEQL45-1010-I/C                        250 ml        up to 500 V       15 - 35 mA    30 - 90 min
                                                 cm
                                           16 x 14
Twin M           PEQL45-1614-I/C                        650 ml        up to 500 V       15 - 50 mA   60 - 120 min
                                                 cm
                                           20 x 20
Twin L           PEQL45-2020-I/C                       1250 ml        up to 500 V       15 - 75 mA   60 - 180 min
                                                 cm
                                           20 x 10
Twin ExW S       PEQL45-2010-I/C                        650 ml        up to 500 V       30 - 45 mA    30 - 90 min
                                                 cm

VWR_v0617_E                                                   6
INSTRUCTION MANUAL - BEDIENUNGSANLEITUNG PERFECTBLUE VERTICAL DOUBLE GEL SYSTEMS / VERTIKALE DOPPELGELSYSTEME - TWIN S, M, L & TWIN EXW S - VWR
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

GENERAL INSTRUCTIONS

Leak proof sealing of glass plates
When casting acrylamide gels, we recommend using the upper buffer chamber (UBC) for the fixation of
glass plates and the casting base for plate sealing. The casting base is included in delivery or can be
purchased separately. In addition the Cast-It™ gel caster is available for sealing glass plates and pour-
ing gels while the upper buffer chamber is used for electrophoresis. (see 'TECHNICAL SUPPORT AND
ORDERING INFORMATION').

1. The upper buffer chamber (UBC) is attached to the lower
   buffer chamber (LBC) with a safety interlocking pin and slot
   arrangement. The UBC may be removed for cleaning by
   sliding the pin out of the slot in the center fixture of the LBC
   and lifting it out. For casting the gel remove the UBC and
   loosen the black knobs and slide the side clamps sideways
   to the outermost position. The clamps will remain in place
   once positioned due to the spring design.

2. Invert the casting base so that the black silicone mats face
   downward. Inverting the casting base is essential for proper
   alignment of glass plates and spacers before clamping them
   inside the UBC assuring perfect sealing later on.

3. Place the UBC flat on the inverted casting base. Arrange a
   gel assembly including a notched glass plate, side spaces
   and blank glass plate onto the gasket of one side of the UBC,
   making sure to place the notched plate next to the gasket
   (this will allow buffer access to the gel during operation).
   Make sure that the bottom of the glass plates and the spacers
   sit flat on the acrylic squares of the inverted casting base. The
   usage of the Spacer Placer facilitates the arrangement of the
   spacers. Slide the side clamps over the assembly until the
   gasket of the UBC is slightly flattened by the glass plates (do
   not over tighten).
   Place a second gel assembly or the blocking plate on the oth-
   er side to create the upper buffer reservoir.

VWR_v0617_E                                                   7
INSTRUCTION MANUAL - BEDIENUNGSANLEITUNG PERFECTBLUE VERTICAL DOUBLE GEL SYSTEMS / VERTIKALE DOPPELGELSYSTEME - TWIN S, M, L & TWIN EXW S - VWR
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

4. After aligning and clamping the glass plate assem-
   bly the UBC will be removed and the casting base
   has to be turned around so that the black silicone
   mats are facing upward. Wet the silicone mats with
   e.g. water or buffer to prolong their life span.

5. Place the UBC onto the casting base.
   Insert the casting cams (with handles
   pointing up) into the holes in the UBC.
   Simultaneously turn cams one half turn
   to tighten the assembly onto the silicone
   gaskets of the casting base. The eccen-
   tric effect of the cams presses the UBC
   and therefore the lower angles of the
   glass plates downward onto the silicone
   mats of the casting base.

6. Once the glass plates are secured into the casting
   base, the gel may be poured. Take care to avoid air
   bubbles while pouring the gel. Immediately after the
   gel is poured insert an appropriate comb for the crea-
   tion of wells.

    There is no need to apply the casting base when using
    pre-cast gels. Since glass or plastic plates of pre-cast
    gels are often very thin take care not to over tighten the
    side camps when fixing the gel assembly into the UBC.

VWR_v0617_E                                                   8
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

Assembling the system and electrophoresis
1. To reassemble, place the upper buffer chamber (UBC)
   into the lower buffer chamber (LBC) on the open side of
   the slot in the center fixture of the LBC. Slide the UBC
   towards the center of the LBC until the pins lock into the
   slot. This interlocking pin and slot arrangement guaran-
   tees a tighter fit of the UBC in the LBC and allows a re-
   producibility of the electric conditions.

2. For cooling through a cold water tap or recirculator
   chiller, flexible lab tubing can get attached to each hose
   barb on the UBC marked as 'in' and 'out'. The inner di-
   ameter of the adapters of the integrated cooling system
   is 6 mm, the outer diameter is 9 mm. Note the marks
   indicating 'in' and 'out' at the adapters to assure the
   right flow direction. Water flow should not exceed 2 L
   per minute for the water tap and 12 L per minute for the
   recirculator.

3. Add electrophoresis buffer to the UBC to approximately
   5 mm from the top of the gel assemblies, making sure
   the level is above the gap formed by the 'ears' of the
   notched plate. Be sure that no buffer is leaking from the
   UBC to the LBC.

4. Remove comb and rinse wells with buffer. Load samples
   onto the gel.

5. Add buffer to the LBC to approximately 2-3 mm above
   the base of the gel. The open bottom end of the gel
   assembly should be in contact with buffer in order for
   electrophoresis to take place.

6. Set the safety lid onto the unit. Because of its asymmetrical
   design this will only work in the proper orientation, con-
   necting the electrode of the lower buffer reservoir (typical-
   ly the anode) with the red power cord and the electrodes
   of the upper buffer reservoir (typically the cathodes) with
   the black power cord.

VWR_v0617_E                                                   9
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

7. For cooling, turn on the water and allow it to circulate
   through the system. Water flow should not exceed 2 L
   per minute for the water tap and 12 L per minute for the
   recirculator.

8. When water has started to circulate through the system,
   connect the power cords to a power supply (red to red
   and black to black) and start the run by defining an
   appropriate voltage or current (see 'Technical proper-
   ties').

9. Turn off power supply and water after the electrophore-
   sis is complete.

10. Remove the lid by pushing on the acrylic thumb pegs
    protruding through the lid on the top of the unit. Slide
    and lift UBC from LBC and drain buffer chambers sepa-
    rately.

11. Loosen side clamps and remove gel assemblies to further proceed with analysis.

Cleaning
Lower buffer chamber, upper buffer chamber, silicone gaskets of the casting base, glass plates, spacers
and combs should get cleaned thoroughly under running water after use. Use a mild detergent to get rid
of any debris then rinse with distilled water to avoid formation of salt spots. It is recommended to allow
the chamber to air dry rather than drying with a towel to avoid damage to the electrode wires. Each
part should get completely dry before storage.

Do not use ethanol or other organic solvents to clean acrylic products, because organic solvents cause
acrylic to 'craze' or crack!

VWR_v0617_E                                                   10
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

REQUIRED MATERIALS & RECIPES

Required gel volumes
These volumes have been calculated for glass plates and spacers available at VWR.

PerfectBlue™ double gel system                  Twin S              Twin M           Twin ExW S   Twin L
0.8 mm Spacer                                    8 ml               18 ml            16 ml        32 ml
1.5 mm Spacer                                   15 ml               34 ml            30 ml        60 ml

Required buffer volumes

PerfectBlue™ double gel system                  Twin S              Twin M           Twin ExW S   Twin L
                                                250 ml              650 ml           650 ml       1250 ml

Tris-Glycine SDS-PAGE

10 x Tris-Glycine electrophoresis buffer

30 g/l Tris-Base [0.25 M]
144 g/l Glycine [1.92 M]
10 g/l SDS [1 %, w/v]
Adjust volume to 1 L using distilled H2O

Resolving gels for Tris-Glycine SDS-PAGE

                                            6%         8%          10 %      12 %    15 %
H2 O                                        5.3        4.6         4.0       3.3     2.3
30 % Acrylamide mix                         2.0        2.7         3.3       4.0     5.0
1.5 M Tris-HCl (pH 8.8)                     2.5        2.5         2.5       2.5     2.5
10 % SDS                                    0.1        0.1         0.1       0.1     0.1
10 % Ammonium persulfate (APS)              0.1        0.1         0.1       0.1     0.1
TEMED                                       0.008      0.006       0.004     0.004   0.004
All amounts in ml per 10 ml gel solution

5 % Stacking gels for Tris-Glycine SDS-PAGE

                                            1 ml       2 ml        3 ml      5 ml    10 ml
H2 O                                        0.68       1.4         2.1       3.4     6.8
30 % Acrylamide mix                         0.17       0.33        0.5       0.83    1.7
1.0 M Tris-HCl (pH 6.8)                     0.13       0.25        0.38      0.63    1.25
10 % SDS                                    0.01       0.02        0.03      0.05    0.1
10 % Ammonium persulfate (APS)              0.01       0.02        0.03      0.05    0.1
TEMED                                       0.001      0.002       0.003     0.005   0.01
All amounts in ml

VWR_v0617_E                                                   11
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

3 x Sample buffer for Tris-Glycine SDS-PAGE

1.75 ml 1 M Tris-HCl pH 6.8
1.5 ml Glycerine
5 ml SDS (10 %)
0.5 ml -Mercaptoethanol
1.25 ml Bromophenol blue

Tris-Tricine SDS-PAGE
1 x Tris-Tricine electrophoresis buffer

Cathode buffer             12.11 g/l Tris-Base [0.1 M]
                           17.92 g/l Tricine [0.1 M]
                           1 g/l SDS [0.1 %, w/v]
                           Adjust volume to 1 L using distilled H2O

Anode buffer               0.2 M Tris-HCl, pH 8.9

Resolving gels for Tris-Tricine SDS-PAGE
                                            6%         8%          10 %    12 %
H2 O                                        2.1        1.4         0.6     0.0
Glycerin                                    1.4        1.4         1.4     1.4
30 % Acrylamide mix                         2.1        2.8         3.5     4.2
3 M Tris-HCl (pH 8.45)                      4.3        4.3         4.3     4.3
10 % SDS                                    0.1        0.1         0.1     0.1
10 % Ammonium persulfate (APS)              0.1        0.1         0.1     0.1
TEMED                                       0.008      0.006       0.004   0.004
All amounts in ml per 10 ml gel solution

10 % Stacking gels for Tris-Tricine SDS-PAGE

                                            1 ml       2 ml        3 ml    5 ml    10 ml
H2 O                                        0.03       0.06        0.08    0.16    0.32
30 % Acrylamide mix                         0.32       0.65        1.0     1.62    3.24
3 M Tris-HCl (pH 8.45)                      0.62       1.24        1.86    3.10    6.20
10 % SDS                                    0.02       0.03        0.06    0.1     0.2
10 % Ammonium persulfate (APS)              0.005      0.01        0.015   0.025   0.05
TEMED                                       0.001      0.001       0.002   0.005   0.01
All amounts in ml

2 x Sample buffer for Tris-Tricine gels              3 x Sample buffer for native gel electrophoresis

1 ml Tris-HCl (pH 6.8)                               4.5 ml Glycerine
0.8 g SDS                                            0.31 g Bis-Tris-Base
2.4 ml Glycerine                                     adjust to pH 7 with HCL
0.31 g DTT                                           Adjust volume to 10 ml using distilled H2O
2 mg Coomassie Blue G-250
Adjust volume to 10 ml using distilled H2O

VWR_v0617_E                                                   12
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

TROUBLESHOOTING
Below are some possible solutions to potential problems. If these suggestions are unsuccessful, do not
hesitate to contact us (see 'TECHNICAL SUPPORT AND ORDERING INFORMATION').

 Problem                Cause                                 Solution

 Leakage of gel         Glass plates and spacers have For correct alignment of glass plate sandwich it is necessary
 solution               not been aligned correctly    to place the upper buffer chamber on top of the inverted
                                                      casting base. The bottom of the glass plates and the spacers
                                                      need to sit flat on the acrylic squares.

                        Glass plates or spacers are           Clean thoroughly.
                        dirty

                        Uneven edges of glass plates          Only use glass plates with polished edges.

                        Silicone gaskets of casting           Clean thoroughly with a mild detergent and rinse with
                        base are dirty                        distilled water, dry without smearing, i.e. hand dry with a lint
                                                              free towel.

                        Silicone gaskets of casting           Replace silicone gaskets which can be easily removed. New
                        base are brittle or worn out          gaskets will stick due to their own adherence. Also, wet
                                                              silicone mats with e.g. water or buffer before placing UBC
                                                              onto the casting base to prolong their life span.

 Upper buffer       Gaskets of the upper buffer               Remove gasket, wash in warm water to remove excess salts,
 chamber is leaking chamber are deformed, too                 and place the gasket back in the groove. If the gaskets are
                    old or dirty                              too old or brittle, they should be replaced. If the clamps have
                                                              been over tightened in the past, the gasket can be pushed too
                                                              far into the gasket groove and will not make a seal.

                        Leakage of buffer through gel         Do not move spacers after the gel polymerized. Make sure
                        sides/spacers                         that the glass plates are seated firmly against the gasket.
                                                              Always use combs and spacers of the same thickness.

 Run time is too        Buffer is too dilute                  After a certain running time the ions might be exhausted, so
 long                                                         the electric resistance increases and the current decreases at
                                                              constant voltage, so that the maximum voltage is not enough
                                                              to keep the current constant. Check buffer recipe and try
                                                              again with new buffer. Check if voltage and current of the
                                                              different runs is the same.

                        Upper buffer chamber is               See above 'Upper buffer chamber is leaking'.
                        leaking

                        Running at too low current or         Make sure you are running using the suggested running con-
                        voltage                               ditions for this unit (see 'Technical properties'). When running
                                                              at constant current, the current value is per gel.

 Running too fast       Buffers are too concentrated          Check buffer recipe; remake and try again. If voltage is lower
                                                              than usual when running at constant current, the buffer is
                                                              probably too concentrated.

                        Voltage or current set too            Turn down electrical settings.
                        high.

VWR_v0617_E                                                    13
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

 Problem                Cause                                 Solution

 Smiling of dye         Center of gel is running hotter Use the cooling option of the unit. Turn down the electrical
 front                  than the ends                   settings.

 Bands spreading        Diffusion of sample when              Make sure that the samples are loaded quickly and the
 outwards               loading                               power is applied as soon as possible after loading.

                        Diffusion of sample during run Increase percentage of stacking gel or increase current by
                        in the stacking gel            25 % when stacking.

                        Lower ionic strength of sample Match the ionic strength of sample with that of the gel.

 Bands are narrow- Ionic strength of sample is                De-salt the sample or use sample buffer of the same ionic
 er than sample    higher than of the gel                     strength as the gel.
 wells

 Broad lanes at         Will occur when adjoining      Use the same sample buffer for all of the samples loaded per
 bottom of gel          lanes are loaded with samples gel.
                        with different buffers. Normal
                        in gradient gels.

 Skewed bands           Gel has not polymerized               De-gas gel solution before casting and increase APS and
                        properly at wells.                    TEMED concentrations. The comb can be wiped with TEMED
                                                              just prior to casting to improve polymerization.

                        Salt concentration is too high        Desalt the samples by dialysis, desalting columns or precipi-
                        in sample.                            tation/resolving.

                        Resolving gel is uneven at            Overlay gel carefully using water saturated n-butanol and
                        top.                                  make sure casting stand is level.

                        The upper buffer chamber is           Check gel to make sure that it is a solid slab inside the glass
                        leaking either through the gel        and check the setup of the apparatus to ensure a seal with
                        or along the side spacers.            the gasket. Always use combs and spacers of the same
                                                              thickness.

 Streaked bands         Overloading of sample.                Use less protein or sample when next loading.

                        Sample has precipitated.              Centrifuge sample before adding sample buffer or use a
                                                              lower percentage acrylamide gel.

 'Frowning' of          Leakage of buffer along sides         Do not move spacer after polymerization of gel. Make sure
 outside lanes          or along spacers inside the           that the glass plates are pressed firmly against the gasket of
                        gel assembly.                         the UBC.

 Silicone mats          Sticky surface of the mats            Wet the silicone mats with e.g. water or buffer before
 wear out quickly       adheres too strongly to the           placing UBC onto the casting base to prolong their life
                        glass plates. Upon removal            span.
                        of UBC silicone mats are torn
                        apart.

 Double bands           Due to re-oxidation or                If using a reducing agent, prepare fresh sample buffer every
 ('doublets')           insufficient reduction of the         30 days. Increase the concentration of 2-mercaptoethanol
                        sample.                               or dithiothreitol in the sample.

VWR_v0617_E                                                    14
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

TECHNICAL SUPPORT AND ORDERING INFORMATION
For technical questions and more detailed information on VWR’s products please visit www.de.vwr.com
to find the respective contact person.

Power Supplies
Do not hesitate to contact us for advice on which Power Supply is most suitable for your application.

PerfectBlue™ Double Gel System Twin S
Item                             Description                                                Cat. No.
Double gel system Twin S         system for gels 10 x 10 cm (W x L), without casting base   PEQL45-1010-I
                                 system for gels 10 x 10 cm (W x L), with casting base      PEQL45-1010-C
 Casting base S                  including silicone gaskets and cams                        PEQL45-1010-CB
 Cast-It™ S gel caster           complete system for sealing up to 2 gels                   PEQL45-1010-CU
 Silicone gaskets                2 replacement silicone gaskets for Casting base S          PEQL45-1010-CG
 Replacement gaskets             2 replacement gaskets for upper buffer chamber S           PEQL45-1010-GK
 Side clamps                     2 side clamps for fixation of glass plates                 PEQL45-1010-CL
 Wing screws                     2 wing screws for fixation of side clamps                  PEQL45-1010-WK
 Notched glass plate             clear, 10 x 10 x 0.24 cm                                   PEQL45-1010-10R
 Blank glass plate               clear, 10 x 10 x 0.24 cm                                   PEQL45-1010-10G
 Notched alumina plate           white, 10 x 10 x 0.10 cm                                   PEQL45-1010-10RA
 Standard combs                  1.5 mm                    6 teeth           230 µl*        PEQL45-1010-6D
                                 1.5 mm                    8 teeth           160 µl*        PEQL45-1010-8D
                                 1.5 mm                   10 teeth           120 µl*        PEQL45-1010-10D
                                 1.5 mm                   12 teeth            90 µl*        PEQL45-1010-12D
                                 0.8 mm                    6 teeth           120 µl*        PEQL45-1010-6C
                                 0.8 mm                    8 teeth            85 µl*        PEQL45-1010-8C
                                 0.8 mm                   10 teeth            60 µl*        PEQL45-1010-10C
                                 0.8 mm                   12 teeth            45 µl*        PEQL45-1010-12C
 Preparative combs               1.5 mm                    2 teeth    1500/100 µl*          PEQL45-1010-PD
                                 0.8 mm                    2 teeth      800/60 µl*          PEQL45-1010-PC
 SpacerPlacer                    Facilitates the correct alignment of Spacers               PEQL45-1010-SP
 Spacer Sets                     1.5 mm              0.7 cm wide            3 pieces        PEQL45-1010-SD
 (2 side & 1 bottom spacer)      0.8 mm              0.7 cm wide            3 pieces        PEQL45-1010-SC
* volumes are calculated for a well length of 14 mm

VWR_v0617_E                                                   15
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

PerfectBlue™ Double Gel System Twin M
Item                             Description                                                Cat. No.
Double gel system Twin M         system for gels 16 x 14 cm (W x L), without casting base   PEQL45-1614-I
                                 system for gels 16 x 14 cm (W x L), with casting base      PEQL45-1614-C
 Casting base M                  including silicone gaskets and cams                        PEQL45-1614-CB
 Cast-It™ M gel caster           complete system for sealing up to 2 gels                   PEQL45-1614-CU
 Silicone gaskets                2 replacement silicone gaskets for Casting base M          PEQL45-1614-CG
 Replacement gaskets             2 replacement gaskets for upper buffer chamber M           PEQL45-1614-GK
 Side clamps                     2 side clamps for fixation of glass plates                 PEQL45-1614-CL
 Wing screws                     2 wing screws for fixation of side clamps                  PEQL45-1010-WK
 Notched glass plate             clear, 16 x 14 x 0.32 cm                                   PEQL45-1614-14R
 Blank glass plate               clear, 16 x 14 x 0.32 cm                                   PEQL45-1614-14G
 Standard combs                  1.5 mm                   10 teeth            310 µl*       PEQL45-1614-10D
                                 1.5 mm                   15 teeth            180 µl*       PEQL45-1614-15D
                                 1.5 mm                   20 teeth            115 µl*       PEQL45-1614-20D
                                 1.5 mm                   24 teeth             85 µl*       PEQL45-1614-24D
                                 0.8 mm                   10 teeth            165 µl*       PEQL45-1614-10C
                                 0.8 mm                   15 teeth             95 µl*       PEQL45-1614-15C
                                 0.8 mm                   20 teeth             60 µl*       PEQL45-1614-20C
                                 0.8 mm                   24 teeth             45 µl*       PEQL45-1614-24C
 Preparative combs               1.5 mm                    2 teeth     3900/155 µl*         PEQL45-1614-PD
                                 0.8 mm                    2 teeth      2095/80 µl*         PEQL45-1614-PC
 SpacerPlacer                    Facilitates the correct alignment of Spacers               PEQL45-1010-SP
 Spacer Sets                     1.5 mm              1.0 cm wide            3 pieces        PEQL45-1614-SD
 (2 side & 1 bottom spacer)      0.8 mm              1.0 cm wide            3 pieces        PEQL45-1614-SC
* volumes are calculated for a well length of 20 mm

PerfectBlue™ Double Gel System Twin L
Item                             Description                                                Cat. No.
Double gel system Twin L         system for gels 20 x 20 cm (W x L), without casting base   PEQL45-2020-I
                                 system for gels 20 x 20 cm (W x L), with casting base      PEQL45-2020-C
 Casting base L                  including silicone gaskets and cams                        PEQL45-2020-CB
 Cast-It™ L gel caster           complete system for sealing up to 2 gels                   PEQL45-2020-CU
 Silicone gaskets                2 replacement silicone gaskets for Casting base L          PEQL45-2020-CG
 Replacement gaskets             2 replacement gaskets for upper buffer chamber L           PEQL45-2020-GK
 Side clamps                     2 side clamps for fixation of glass plates                 PEQL45-2020-CL
 Wing screws                     2 wing screws for fixation of side clamps                  PEQL45-1010-WK
 Notched glass plate             clear, 20 x 20 x 0.32 cm                                   PEQL45-2020-20R
 Blank glass plate               clear, 20 x 20 x 0.32 cm                                   PEQL45-2020-20G
 Standard combs                  1.5 mm                   10 teeth            405 µl*       PEQL45-2020-10D
                                 1.5 mm                   15 teeth            245 µl*       PEQL45-2020-15D
                                 1.5 mm                   20 teeth            165 µl*       PEQL45-2020-20D
                                 1.5 mm                   25 teeth            115 µl*       PEQL45-2020-25D
                                 0.8 mm                   10 teeth            215 µl*       PEQL45-2020-10C
                                 0.8 mm                   15 teeth            130 µl*       PEQL45-2020-15C
                                 0.8 mm                   20 teeth             85 µl*       PEQL45-2020-20C
                                 0.8 mm                   25 teeth             60 µl*       PEQL45-2020-25C
 Preparative combs               1.5 mm                    2 teeth     4915/155 µl*         PEQL45-2020-PD
                                 0.8 mm                    2 teeth      2620/80 µl*         PEQL45-2020-PC
 SpacerPlacer                    Facilitates the correct alignment of Spacers               PEQL45-1010-SP
 Spacer Sets                     1.5 mm              1.0 cm wide            3 pieces        PEQL45-2020-SD
 (2 side & 1 bottom spacer)      0.8 mm              1.0 cm wide            3 pieces        PEQL45-2020-SC
* volumes are calculated for a well length of 20 mm

VWR_v0617_E                                                   16
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

PerfectBlue™ Double Gel System Twin ExW S
Item                             Description                                                Cat. No.
Double gel system                system for gels 20 x 10 cm (W x L), without casting base   PEQL45-2010-I
Twin ExW S                       system for gels 20 x 10 cm (W x L), with casting base      PEQL45-2010-C
Casting base ExW                 including silicone gaskets and cams                        PEQL45-2010-CB
Cast-It™ ExW S gel caster        complete system for sealing up to 2 gels                   PEQL45-2010-CU
Silicone gaskets                 2 replacement silicone gaskets for Casting base ExW S      PEQL45-2010-CG
Replacement gaskets              2 replacement gaskets for upper buffer chamber ExW S       PEQL45-2010-GK
Side clamps                      2 side clamps for fixation of glass plates                 PEQL45-2010-CL
Wing screws                      2 wing screws for fixation of side clamps                  PEQL45-1010-WK
Notched glass plate              clear, 20 x 10 x 0.30 cm                                   PEQL45-2010-10R
Blank glass plate                clear, 20 x 10 x 0.30 cm                                   PEQL45-2010-10G
Standard combs                   1.5 mm                   10 teeth           285 µl*        PEQL45-2010-10D
                                 1.5 mm                   15 teeth           170 µl*        PEQL45-2010-15D
                                 1.5 mm                   20 teeth           115 µl*        PEQL45-2010-20D
                                 1.5 mm                   25 teeth            80 µl*        PEQL45-2010-25D
                                 0.8 mm                   10 teeth           150 µl*        PEQL45-2010-10C
                                 0.8 mm                   15 teeth            90 µl*        PEQL45-2010-15C
                                 0.8 mm                   20 teeth            60 µl*        PEQL45-2010-20C
                                 0.8 mm                   25 teeth            40 µl*        PEQL45-2010-25C
 Micro titer combs               1.5 mm                   18 teeth           135 µl*        PEQL45-2010-MTD
                                 1.5 mm                   36 teeth            55 µl*        PEQL45-2010-MTC
                                 0.8 mm                   18 teeth            70 µl*        PEQL45-2010-MT2D
                                 0.8 mm                   36 teeth            30 µl*        PEQL45-2010-MT2C
 Preparative combs               1.5 mm                    2 teeth    3575/110 µl*          PEQL45-2010-PD
                                 0.8 mm                    2 teeth     1905/60 µl*          PEQL45-2010-PC
 SpacerPlacer                    Facilitates the correct alignment of Spacers               PEQL45-1010-SP
 Spacer Sets                     1.5 mm              1.2 cm wide            3 pieces        PEQL45-2010-SD
 (2 side & 1 bottom spacer)      0.8 mm              1.2 cm wide            3 pieces        PEQL45-2010-SC
* volumes are calculated for a well length of 14 mm

VWR_v0617_E                                                   17
Instruction Manual PerfectBlue™ Vertical Double Gel Systems

NOTES

VWR_v0617_E                                              18
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

INHALT
KURZANLEITUNG                                                       20
GARANTIE                                                            22
SICHERHEITSHINWEISE                                                 22
LIEFERUMFANG                                                        23
SYSTEMÜBERBLICK                                                     24
    Technische Merkmale                                             24
ALLGEMEINE BEDIENUNGSHINWEISE                                       26
    Auslaufsicheres Abdichten der Glasplatten                       26
    Zusammenbau des Systems und Elektrophorese                      28
    Reinigung und Pflege                                            29
BENÖTIGTE MATERIALIEN & REZEPTE                                     30
    Benötigte Gelvolumina                                           30
    Benötigte Puffer-Volumina                                       30
    Tris-Glycin SDS-PAGE                                            30
    Tris-Tricin SDS-PAGE                                            31
TROUBLESHOOTING                                                     32
TECHNISCHER SERVICE UND BESTELLINFORMATIONEN                        34
    Power Supplies                                                  34
    PerfectBlue™ Doppelgelsystem Twin S                             34
    PerfectBlue™ Doppelgelsystem Twin M                             35
    PerfectBlue™ Doppelgelsystem Twin L                             35
    PerfectBlue™ Doppelgelsystem Twin ExW S                         36

VWR_v0617_D                                                    19
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

KURZANLEITUNG

Systemübersicht
1. Mittelteil (obere Pufferkammer) mit
   Seitenklemmen und Elektroden
2. Sicherheitsdeckel mit Stromkabeln                               1
3. Untere Pufferkammer
4. Gießbasis                                                           2

                                                               3

                                                                           4

Auslaufsicheres Abdichten der Glasplatten

Invertieren Sie die Gießbasis, so dass die Gummimatten
nach unten und die Acrylglasquader nach oben gerichtet
sind.

Platzieren Sie das Mittelteil auf der invertierten Gießbasis.
Bauen Sie das Glasplattensandwich zusammen und schie-
ben Sie es zwischen die Seitenklemmen des Mittelteils.
Wichtig ist, dass Glasplatten und Spacer komplett auf den
Acrylglasquadern der invertierten Gießbasis aufsitzen.
Drehen Sie nun die Seitenklemmen fest. Spannen Sie auf
ähnliche Weise die Blindplatte auf der anderen Seite des
Mittelsteils ein, falls Sie nur ein Gel gießen wollen.

Drehen Sie die Gießbasis um und ziehen
Sie die Nockenhebel mit dem Griff nach
oben nach außen. Befeuchten Sie die
Dichtungen mit Wasser oder Puffer, um
deren Lebenszeit zu erhöhen. Setzen Sie
das Mittelteil samt Glasplatten mittig auf die
Dichtungen.
Schieben Sie die Nockenhebel nach innen
und drehen Sie sie gleichzeitig um eine
halbe Drehung nach unten.

VWR_v0617_D                                                    20
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

Zusammenbau des Systems und Elektrophorese

Das Gel kann nun gegossen werden. Gießen Sie langsam,
um Blasenbildung zu vermeiden.

Nach der Polymerisation wird das Mittelteil aus der Gießba-
sis entnommen und in die untere Pufferkammer eingesetzt.
Befüllen Sie das obere Pufferreservoir mit Laufpuffer, bela-
den Sie das Gel mit den vorbereiteten Proben und füllen Sie
danach Laufpuffer in die untere Pufferkammer.

Nach Aufsetzen des Sicherheitsdeckels kann der
Gellauf gestartet werden. Schalten Sie nach dem Lauf
den Power Supply ab und nehmen Sie den Sicherheits-
deckel herunter. Heben Sie das Mittelteil aus der unte-
ren Pufferkammer und lösen Sie die Seitenklemmen.
Das Gel kann nun von den Glasplatten gelöst und
analysiert werden.

VWR_v0617_D                                                    21
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

GARANTIE
VWR garantiert, dass das ausgelieferte System genauestens geprüft wurde und den geltenden Anforde-
rungen entspricht.

Bitte überprüfen Sie die Lieferung dennoch umgehend nach Erhalt auf Vollständigkeit und eventuelle
Transportschäden. Sollte die Lieferung beschädigt oder fehlerhaft sein, wenden Sie sich bitte umgehend
an den Technischen Service von VWR oder Ihren VWR-Außendienstmitarbeiter (siehe 'TECHNISCHER
SERVICE & BESTELLINFORMATIONEN'). Durch die Aufbewahrung des Verpackungsmaterials bis zur
vollständigen Prüfung der Lieferung wird die Umwelt geschont und eine evtl. Rückholung beschleunigt.
Alle Rücksendungen, Austauschlieferungen und Gutschriften müssen zuvor von VWR freigegeben
werden.

Auf sämtliche PerfectBlue™ Elektrophorese- und Blottingsysteme gewährt VWR 36 Monate Garantie,
sofern die Systeme ausschließlich der Bedienungsanleitung entsprechend verwendet wurden und keine
anderslautende Vereinbarung besteht. Ansprüche auf Ersatz oder Reparatur, die aus einer fehlerhaften
Verwendung entstanden sind, werden nicht erfüllt. Die VWR GmbH verpflichtet sich zur Reparatur oder
dem Ersatz des Gerätes bzw. der Rückerstattung des Kaufpreises nach ihren Bedingungen. VWR haftet
nicht für Folgeschäden, die aus der Verwendung des Systems entstanden sind. Nach Ablauf der Garan-
tiezeit können defekte PerfectBlue™ Elektrophorese- oder Blotting-Systeme durch VWR kostengünstig
repariert werden.

Um Neuentwicklungen zeitnah einführen zu können, behält es sich VWR vor, technische Details ohne
Vorankündigung zu ändern.

SICHERHEITSHINWEISE
  Bitte lesen Sie diese Bedienungsanleitung sorgfältig durch, bevor Sie das System in Betrieb nehmen.
  Verwenden Sie zur Stromversorgung eine CE-konforme Gleichstrom-Spannungsquelle.
   Trennen Sie das System von der Stromversorgung, bevor Sie den Sicherheitsdeckel entfernen oder
   das System bewegen.
  Trennen Sie das System bei Nichtbenutzung stets von der Stromversorgung.
   Die für das System definierte maximale Stromstärke und Spannung, welche auf dem Typenschild der
   Pufferkammer angegeben ist, darf nicht überschritten werden.
   Um die Entstehung eines Überdruckes zu vermeiden, darf die Fließrate des Kühlwassers durch das
   integrierte Kühlsystem 12 l/min nicht überschreiten. Bei Anschluss an einen Wasserhahn muss mit
   schwankendem Wasserdruck gerechnet werden, weshalb hier eine Fließrate von lediglich 2 l/min
   gewählt werden sollte.

VWR_v0617_D                                                    22
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

LIEFERUMFANG
Sofern nicht anders vereinbart bzw. auf dem Lieferschein angegeben, enthält der Lieferumfang der
Modelle Twin S, Twin M, Twin L und Twin ExW S folgende Komponenten:

Twin S (PEQL45-1010-I)                                              Twin M (PEQL45-1614-I)
    Mittelteil mit Platin-Elektroden                                    Mittelteil mit Platin-Elektroden
    Untere Pufferkammer                                                 Untere Pufferkammer
    Sicherheitsdeckel mit Stromkabeln                                   Sicherheitsdeckel mit Stromkabeln
    4 normale Glasplatten                                               4 normale Glasplatten
    2 ausgeschnittene Glasplatten                                       4 ausgeschnittene Glasplatten
    2 ausgeschnittene Aluminium-Ceramplatten                            4 Seitenklemmen
    4 Seitenklemmen                                                     2 Kämme, 1.5 mm dick, 10 Zähne
    2 Kämme, 0.8 mm dick, 10 Zähne                                      2 Kämme 1.5 mm dick, 15 Zähne
    4 Seitenspacer, 0.8 mm dick                                         8 Seitenspacer, 1.5 mm dick
    2 Bodenspacer, 0.8 mm dick                                          4 Bodenspacer, 1.5 mm dick
    2 Ersatzgummidichtungen                                             2 Ersatzgummidichtungen
    4 Ersatzfedern                                                      4 Ersatzfedern
    1 Blindplatte für Einzelgelläufe                                    1 Blindplatte für Einzelgelläufe
    1 'SpacerPlacer' zum Ausrichten der Spacer                          1 'SpacerPlacer' zum Ausrichten der Spacer

Twin S (PEQL45-1010-C)                                              Twin M (PEQL45-1614-C)
wie PEQL45-1010-I aber inklusive Gießbasis                          wie PEQL45-1614-I aber inklusive Gießbasis

Twin ExW S (PEQL45-2010-I)                                          Twin L (PEQL45-2020-I)
    Mittelteil mit Platin-Elektroden                                    Mittelteil mit Platin-Elektroden
    Untere Pufferkammer                                                 Untere Pufferkammer
    Sicherheitsdeckel mit Stromkabeln                                   Sicherheitsdeckel mit Stromkabeln
    4 Glasplatten                                                       4 Glasplatten
    4 ausgeschnittene Glasplatten                                       4 ausgeschnittene Glasplatten
    4 Seitenklemmen                                                     4 Seitenklemmen
    2 Kämme, 0.8 mm dick, 15 Zähne                                      2 Kämme, 1.5 mm dick, 15 Zähne
    2 Kämme, 0.8 mm dick, 20 Zähne                                      2 Kämme, 1.5 mm dick, 20 Zähne
    8 Seitenspacer, 0.8 mm dick                                         8 Seitenspacer, 1.5mm dick
    4 Bodenspacer, 0.8 mm dick                                          4 Bodenspacer, 1.5mm dick
    2 Ersatzgummidichtungen                                             2 Ersatzgummidichtungen
    4 Ersatzfedern                                                      4 Ersatzfedern
    1 Blindplatte für Einzelgelläufe                                    1 Blindplatte für Einzelgelläufe
    1 'SpacerPlacer' zum Ausrichten der Spacer                          1 'SpacerPlacer' zum Ausrichten der Spacer

Twin ExW S (PEQL45-2010-C)                                          Twin L (PEQL45-2020-C)
wie PEQL45-2010-I aber inklusive Gießbasis                          wie PEQL45-2020-I aber inklusive Gießbasis

VWR_v0617_D                                                    23
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

SYSTEMÜBERBLICK
Die vertikalen PerfectBlue™ Doppelgelsysteme Twin S, M, L und ExW S überzeugen mit einer durch-
dachten Konstruktion und hochwertiger Verarbeitung. Durch die besondere Anordnung von 3 Kathoden
und einer langen, die komplette Basis der Elektrophoresekammer umlaufende Anode, werden mit ihnen
außerordentlich gute elektrophoretische Separationen erzielt. Sämtliche Kammern sind mit einem Was-
serzirkulationssystem ausgestattet, welches aufgrund der Verwendung von sehr gut Wärme leitendem
Aluminium-Ceram ein effektives Kühlen erlaubt. Dadurch können hohe Migrationsgeschwindigkeiten
erzielt werden, ohne das exzellente Laufverhalten negativ zu beeinflussen. Das zur Verfügung stehende
Gießsystem erlaubt das einfache und auslaufsichere Gießen von bis zu 2 Gelen gleichzeitig. Es ist so
konzipiert, dass die Elektrophorese nach der Polymerisation der Gele ohne einen aufwendigen Aus- und
Wiedereinbau der Gelplatten gestartet werden kann. Somit verkürzt sich die Arbeitszeit und minimieren
sich mögliche Fehlerquellen. Die vier Formate, in denen die vertikalen PerfectBlue™ Doppelgelsysteme
zur Verfügung stehen und das umfangreiche Zubehör schaffen beste Voraussetzungen für vielfältigste
Einsatzmöglichkeiten. Sämtliche PerfectBlue™ Gelkammern von VWR werden unter Verwendung einer
speziellen Klebetechnik aus einzelnen Acrylglas-Frästeilen zusammengesetzt. Dadurch werden mechani-
sche Spannungen durch z.B. Wärmeausdehnung und Handhabung ausgeglichen und führen nicht zur
Schädigung der Kammer. Unser auf langjährige Erfahrung gestütztes Vertrauen in die Haltbarkeit
unserer Gelsysteme veranlasst uns zur Vergabe einer 3-jährigen Garantie.

                    Mittelteil (obere Pufferkammer)
                    mit Seitenklemmen & Elektro-                    Sicherheitsdeckel mit
                    den                                             Stromkabeln

                     Untere Pufferkammer                            Gießbasis

Technische Merkmale
PerfectBlue™ Bestell-Nr.                 Gelgröße       benötigtes       maximale       Stromstärke    typische
                                          (B x L)     Puffervolumen      Spannung         pro Gel      Laufzeit
                                          10 x 10
Twin S           PEQL45-1010-I/-C                          250 ml        bis 500 V      15 - 35 mA     30 - 90 min
                                                cm
                                          16 x 14
Twin M           PEQL45-1614-I/-C                          650 ml        bis 500 V      15 - 50 mA    60 - 120 min
                                                cm
                                          20 x 20
Twin L           PEQL45-2020-I/-C                        1250 ml         bis 500 V      15 - 75 mA    60 - 180 min
                                                cm

VWR_v0617_D                                                    24
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme
                                           20 x 10
Twin ExW S       PEQL45-2010-I/-C                          650 ml   bis 500 V   30 - 45 mA   30 - 90 min
                                                cm

VWR_v0617_D                                                    25
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

ALLGEMEINE BEDIENUNGSHINWEISE

Auslaufsicheres Abdichten der Glasplatten
Für das Gießen von Gelen empfehlen wir die Verwendung des Mittelteils der Gelkammer zum Einspan-
nen der Glasplatten und die systemeigene Gießbasis zum Abdichten der Glasplatten. Letztere ist im
Lieferumfang inbegriffen oder kann nachträglich erworben werden. Darüber hinaus können mit den
optional erhältlichen Cast-It™ Gelgießständen Gele gegossen werden, während in der Gelkammer eine
Elektrophorese stattfindet (siehe 'TECHNISCHER SERVICE & BESTELLINFORMATIONEN').

1. Das Mittelteil ist über ein nützliches Stift/Nut-Prinzip
   verrutschsicher mit der unteren Pufferkammer verbunden. Es
   wird aus der unteren Pufferkammer entnommen, indem es kurz
   angehoben und seitlich nach oben aus der beidseitigen Nut
   der unteren Pufferkammer herausgeführt wird. Lösen Sie die
   schwarzen Flügelschrauben der Seitenklemmen und schieben
   Sie die Seitenklemmen jeweils in die äußerste Position. Die
   Seitenklemmen sind so konzipiert, dass sie aufgrund der spe-
   ziellen Federn in dieser Position bleiben.

2. Drehen Sie die Gießbasis um, so dass die Gummidichtungen
   nach unten gerichtet sind. Dies ermöglicht die nun folgende,
   exakte Positionierung der Glasplatten im Mittelteil, welche für
   das auslaufsichere Abdichten der Glasplatten essentiell ist.

3. Setzen Sie das Mittelteil auf die invertierte Gießbasis. Bauen Sie
   das Glasplattensandwich (normale & ausgeschnittene Glasplat-
   te, 2 Seitenspacer) zusammen und schieben Sie es von oben
   zwischen die Seitenklemmen des Mittelteils. Die ausgeschnittene
   Glasplatte muss hierbei nach innen gerichtet sein, damit das
   Gel später Kontakt zum Laufpuffer hat. Achten Sie darauf, dass
   sowohl die Glasplatten als auch die Spacer vollständig und
   eben auf den Acryl-Blöcken der invertierten Gießbasis aufsit-
   zen. Die Verwendung des SpacerPlacers erleichtert das Ausrich-
   ten der Spacer. Schieben Sie nun die Seitenklemmen nach innen
   und schrauben Sie die Seitenklemmen fest, bis die Glasplatten
   die Gummidichtungen des Mittelteils leicht flachdrücken. Vor-
   sicht: ein zu festes Anziehen der Seitenklemmen kann zum Bre-
   chen der Glasplatten führen!
   Setzen Sie nun auf der anderen Seite ebenfalls Glasplatten mit
   Seiten-Spacern oder eine Blindplatte ein, je nachdem ob Sie ein
   oder zwei Gele gießen möchten. Mit der Blindplatte oder dem
   zweiten Glasplattensandwich wird aus dem Mittelteil das später
   benötigte obere Pufferreservoir.

VWR_v0617_D                                                    26
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

4. Nachdem die Glasplatten justiert und fixiert wur-
   den, wird das Mittelteil nach oben abgenommen.
   Die Gießbasis wird nun herumgedreht, so dass die
   Silikongummidichtungen nach oben zeigen. Be-
   feuchten Sie die Dichtungen mit Wasser oder Puf-
   fer, um deren Lebenszeit zu erhöhen.

5. Platzieren Sie das Mittelteil auf den
   Gummidichtungen der Gießbasis. Die
   Nockenhebel müssen dazu mit dem
   Griff nach oben zeigend ganz nach
   außen gezogen werden. Anschließend
   werden die Nockenhebel nach innen, in
   die Aussparungen des Mittelteils ge-
   schoben und synchron um eine halbe
   Umdrehung gedreht. Die Exzenterwir-
   kung der Nockenhebel drückt das Mit-
   telteil und somit die untere Kante der
   Glasplatten nach unten auf die Silikon-
   gummimatten der Gießbasis.

6. Sobald die Glasplatten in der Gießbasis fest fixiert
   sind, kann das Gel gegossen werden. Gießen Sie die
   Gel-Lösung langsam zwischen die Glasplatten, um
   Luftblasenbildung zu vermeiden. Sobald das Gel ge-
   gossen ist, wird ein Kamm von oben zwischen die
   Glasplatten gesteckt, damit sich die Taschen bilden
   können.

    Beim Einsatz von Fertiggelen kann auf die Verwendung
    der Gießbasis verzichtet werden. Da einige Fertiggele
    extrem dünne Glas- oder Plastikplatten besitzen, sollte
    hier in besonderem Maße ein zu festes Anziehen der
    Seitenklemmen vermieden werden, um eine Beschädi-
    gung des Fertiggels zu vermeiden.

VWR_v0617_D                                                    27
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

Zusammenbau des Systems und Elektrophorese
1. Setzen Sie das Mittelteil mit den eingespannten und polyme-
   risierten Gelen in die untere Pufferkammer ein. Dabei wird
   der Stift, der sich jeweils unten an den beiden kurzen Seiten
   des Mittelteils befindet, in die entsprechende Nut der unteren
   Pufferkammer eingeführt und das Mittelteil anschließend
   zentral abgesetzt. Dieses Stift/Nut-System garantiert den fes-
   ten Sitz des Mittelteils in der unteren Pufferkammer und ge-
   währleistet die Reproduzierbarkeit der elektrophoretischen
   Bedingungen.

2. Zur Kühlung können nun die beiden Anschlüsse des im
   Mittelteil integrierten Kühlsystems über flexible Wasser-
   schläuche und Schlauchklemmen mit dem Wasserhahn
   oder einer Kühlthermostat-Pumpe verbunden werden. Die
   Anschlüsse des integrierten Kühlsystems haben einen In-
   nendurchmesser von 6 mm und einen Außendurchmesser
   von 9 mm. Achten Sie beim Anschluss auf die Markierun-
   gen 'in' und 'out' an den Anschlüssen des Mittelteils, um
   die richtige Durchflussrichtung sicherzustellen. Um die Ent-
   stehung eines Überdruckes zu vermeiden, darf die Fließrate
   des Kühlwassers durch das integrierte Kühlsystem 12 l/min
   nicht überschreiten. Bei Anschluss an einen Wasserhahn
   muss mit schwankendem Wasserdruck gerechnet werden,
   weshalb hier eine Durchflussrate von lediglich 2 l/min ge-
   wählt werden sollte.

3. Geben Sie solange Elektrophoresepuffer in das Pufferreser-
   voir des Mittelteils, bis das Gel mit ca. 5 mm Puffer bedeckt
   ist. Achten Sie darauf, dass der Puffer bis über die kürzere
   bzw. ausgeschnittene Glasplatte reicht. Es darf kein Puffer
   vom Mittelteil in die untere Pufferkammer auslaufen.

4. Entfernen Sie den Kamm und spülen Sie die Taschen mit
   Puffer. Beladen Sie das Gel mit den vorbereiteten Proben.

5. Füllen Sie Elektrophoresepuffer in die untere Pufferkammer
   bis 2 – 3 mm über das untere Ende des Gels. Das offene
   Ende des Gels muss mit Puffer in Kontakt kommen, um eine
   vollständige Gelelektrophorese zu gewährleisten.

VWR_v0617_D                                                    28
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

6. Schließen Sie das System durch Aufsetzen des Sicher-
   heitsdeckels. Aufgrund des asymmetrischen Designs des
   Sicherheitsdeckels ist dies nur in der Orientierung mög-
   lich, in der das rote Stromkabel den Kontakt zur positiven
   Anode im unteren Pufferreservoir und das schwarze
   Stromkabel den Kontakt zur negativen Kathode (oberes
   Pufferreservoir) herstellt.

7. Starten Sie nun die Wasserzirkulation, falls eine Kühlung
   erwünscht ist. Um die Entstehung eines Überdruckes zu
   vermeiden, darf die Fließrate des Kühlwassers durch das
   inte- grierte Kühlsystem 12 l/min nicht überschreiten. Bei
   Anschluss an einen Wasserhahn muss mit schwankendem
   Wasserdruck gerechnet werden, weshalb hier eine Fließ-
   rate von lediglich 2 l/min gewählt werden sollte.

8. Verbinden Sie schließlich die Stromkabel mit den An-
   schlüssen des Power Supplies (rot mit rot, schwarz mit
   schwarz) und starten Sie die Elektrophorese indem sie ei-
   ne geeignete Spannung oder Stromstärke (siehe 'Techni-
   sche Merkmale') definieren.

9. Um die Elektrophorese zu beenden, schalten Sie den
   Power Supply und die Wasserversorgung ab.

10. Der Sicherheitsdeckel kann abgenommen werden, indem
    man leicht gegen die Acrylzapfen drückt, die vom Mittel-
    teil kommend durch den Deckel ragen. Heben Sie nun das
    Mittelteil aus der unteren Pufferkammer durch seitliches
    Schieben und Anheben des Mittelteils. Entleeren Sie die
    beiden Pufferreservoire separat.

11. Durch Lösen der Seitenklemmen können Sie den Gel-
    aufbau nun auseinander nehmen. Fahren Sie mit der
    Analyse des Gels fort.

Reinigung und Pflege
Mit Ausnahme des Sicherheitsdeckels sollten sämtliche Systemkomponenten nach der Benutzung unter
fließendem Wasser gereinigt werden. Salzrückstände und Polyacrylamidreste können mit einem milden
Reinigungsmittel entfernt werden. Ein anschließendes kurzes Abspülen mit destilliertem Wasser verhin-
dert die Bildung von Salzflecken. Um die Platinelektroden nicht zu beschädigen, empfehlen wir, das
Mittelteil an der Luft trocknen zu lassen und nicht mit Papiertüchern trocken zu wischen. Sämtliche Teile
dürfen nur vollständig trocken gelagert werden.

Zur Reinigung von Acrylglas dürfen weder Ethanol noch sonstige organische Lösungsmittel verwendet
werden, da sonst Risse im Material entstehen können!

VWR_v0617_D                                                    29
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

BENÖTIGTE MATERIALIEN & REZEPTE

Benötigte Gelvolumina
Diese Volumina wurden kalkuliert für die bei VWR erhältlichen Glasplatten und Spacer.

PerfectBlue™ Doppelgelsystem                   Twin S                Twin M           Twin ExW S   Twin L
0.8 mm Spacer                                   8 ml                 18 ml            16 ml        32 ml
1.5 mm Spacer                                  15 ml                 34 ml            30 ml        60 ml

Benötigte Puffer-Volumina

PerfectBlue™ Doppelgelsystem                   Twin S                Twin M           Twin ExW S   Twin L
                                               250 ml                650 ml           650 ml       1250 ml

Tris-Glycin SDS-PAGE

10 x Tris-Glycin Elektrophoresepuffer

30 g/l Tris [0.25 M]
144 g/l Glycin [1.92 M]
10 g/l SDS [1%, w/v]
ad 1 l mit H2O

Trenngele für Tris-Glycin SDS-PAGE

                                     6%               8%            10 %      12 %    15 %
H2 O                                 5.3              4.6           4.0       3.3     2.3
30 % Acrylamid-Mix                   2.0              2.7           3.3       4.0     5.0
1.5 M Tris (pH 8.8)                  2.5              2.5           2.5       2.5     2.5
10 % SDS                             0.1              0.1           0.1       0.1     0.1
10 % Ammoniumpersulfat (APS)         0.1              0.1           0.1       0.1     0.1
TEMED                                0.008            0.006         0.004     0.004   0.004
Alle Mengenangaben in ml für 10 ml Gellösung

5 % Sammelgele für Tris-Glycin SDS-PAGE

                                           1 ml       2 ml          3 ml      5 ml    10 ml
H2 O                                       0.68       1.4           2.1       3.4     6.8
30 % Acrylamid-Mix                         0.17       0.33          0.5       0.83    1.7
1.0 M Tris (pH 6.8)                        0.13       0.25          0.38      0.63    1.25
10 % SDS                                   0.01       0.02          0.03      0.05    0.1
10 % Ammoniumpersulfat (APS)               0.01       0.02          0.03      0.05    0.1
TEMED                                      0.001      0.002         0.003     0.005   0.01
Alle Mengenangaben in ml

VWR_v0617_D                                                    30
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

3 x Probenpuffer für die Tris-Glycin SDS-PAGE

1.75 ml 1 M Tris-HCl pH 6.8
1.5 ml Glycerin
5 ml SDS (10 %)
0.5 ml -Mercaptoethanol
1.25 ml Bromphenol-Blau

Tris-Tricin SDS-PAGE
1 x Tris-Tricin Elektrophoresepuffer

Kathodenpuffer             12.11 g/l Tris [0.1 M]
                           17.92/l g Tricin [0.1 M]
                           1 g/l SDS [0.1%, w/v]
                           ad 1 l mit H2O

Anodenpuffer               0.2 M Tris-HCl, pH 8.9

Trenngele für Tris-Tricin SDS-PAGE
                                     6%               8%            10 %    12 %
H2 O                                 2.1              1.4           0.6     0.0
Glycerin                             1.4              1.4           1.4     1.4
30 % Acrylamid-Mix                   2.1              2.8           3.5     4.2
3 M Tris-HCl (pH 8.45)               4.3              4.3           4.3     4.3
10 % SDS                             0.1              0.1           0.1     0.1
10 % Ammoniumpersulfat (APS)         0.1              0.1           0.1     0.1
TEMED                                0.008            0.006         0.004   0.004
Alle Mengenangaben in ml für 10 ml Gellösung

10 % Sammelgele für Tris-Tricine SDS-PAGE

                                           1 ml       2 ml          3 ml    5 ml    10 ml
H2 O                                       0.03       0.06          0.08    0.16    0.32
30 % Acrylamid-Mix                         0.32       0.65          1.0     1.62    3.24
3 M Tris (pH 8.45)                         0.62       1.24          1.86    3.10    6.20
10 % SDS                                   0.02       0.03          0.06    0.1     0.2
10 % Ammoniumpersulfat (APS)               0.005      0.01          0.015   0.025   0.05
TEMED                                      0.001      0.001         0.002   0.005   0.01
Alle Mengenangaben in ml

2 x Probenpuffer für Tris-Tricin-Gele              3 x Probenpuffer für Native Gelelektrophorese

1 ml Tris-HCl pH 6.8                               4.5 ml Glycerin
0.8 g SDS                                          0.31 g Bis-Tris
2.4 ml Glycerin                                    pH-Wert einstellen mit HCl auf pH 7
0.31 g DTT                                         ad 10 ml mit H2O
2 mg Coomassie Blue G-250
ad 10 ml mit H2O

VWR_v0617_D                                                    31
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

TROUBLESHOOTING
Hier finden Sie passende Lösungen zu möglichen Problemen. Sollten Sie weitere Fragen haben, wird
Ihnen das VWR-Service-Team gerne weiterhelfen (TECHNISCHER SERVICE & BESTELLINFORMATI-
ONEN).

 Problem               Ursache                            Lösung

 Gellösung läuft aus Glasplatten und Spacer               Beim Einspannen des Glasplattensandwiches muss das
                     wurden nicht korrekt justiert        Mittelteil unbedingt auf die invertierte Gießbasis gestellt
                                                          werden. Glasplatten und Spacer müssen vollständig auf den
                                                          Acrylglasquadern der invertierten Gießbasis aufsitzen.

                       Glasplatten und Spacer sind        Reinigen Sie Glasplatten und Spacer vor der Verwendung
                       verschmutzt                        gründlich.

                       Die Ränder der Glasplatten         Achten Sie darauf, dass insbesondere die untere Glasplatten-
                       sind uneben                        kante gerade und abgeschliffen ist, damit sie optimal durch
                                                          die Silikongummimatte der Gießbasis abgedichtet wird.

                       Silikongummimatten der             Die Silikongummimatten sollten gründlich mit einem milden
                       Gießbasis sind verschmutzt         Reinigungsmittel gereinigt, mit destilliertem Wasser abgespült
                                                          und fusselfrei getrocknet werden; z. B. an der Luft mit fussel-
                                                          freiem Papier oder Fensterleder.

                       Silikonmatten der Gießbasis        Erneuern Sie die Silikonmatten, welche lediglich aufgrund
                       sind spröde                        ihrer eigenen Klebkraft auf der Gießbasis sitzen und somit
                                                          ohne Aufwand ab- und aufgesetzt werden können. Bitte
                                                          beachten Sie, dass das Anfeuchten der Dichtungen mit
                                                          Wasser oder Puffer vor dem Aufsetzen des Mittelteils deren
                                                          Lebenszeit deutlich erhöht.

 Puffer läuft aus      Gummidichtungen des Mittel-        Waschen Sie die Gummidichtungen nach jeder Verwendung
 oberem Pufferre-      teils sind zu alt oder flachge-    mit fließendem Wasser ab, um Salzreste zu entfernen. Ist die
 servoir aus           drückt, um gut abdichten zu        Gummidichtung zu alt oder spröde, sollte sie ersetzt werden.
                       können                             Wurden die Seitenklemmen vorher zu fest angezogen, kann
                                                          das Herausnehmen der Gummidichtung und erneutes Einset-
                                                          zen in die Aussparung helfen.

                       Auslaufen des Puffers an den       Die Spacer sollten nach der Polymerisation nicht mehr
                       Seiten oder durch die Spacer       bewegt werden. Die Glasplatten müssen fest an die Gummi-
                       innerhalb des Gelaufbaus           dichtungen des Mittelteils gedrückt sein. Verwenden Sie
                                                          immer Kämme und Spacer gleicher Dicke.

 Laufzeit dauert zu    Puffer ist zu schwach              Evtl. sind ab einer gewissen Laufzeit die Ionen erschöpft, so
 lange                                                    dass der elektrische Widerstand steigt und die Stromstärke
                                                          bei konst. Spannung sinkt, bzw. die max. Spannung nicht
                                                          ausreicht, um die Stromstärke konstant zu halten. Überprüfen
                                                          Sie das Pufferrezept. Setzen Sie frischen Puffer an. Prüfen Sie
                                                          die eingestellten und tatsächlichen elektrischen Parameter.

                       Puffer läuft aus oberer Puffer-    Siehe oben: Problem 'Puffer läuft aus oberem Pufferreservoir
                       kammer                             aus'.

                       Spannung oder Stromstärke          Überprüfen Sie die Einstellungen des Power Supply. Die unter
                       ist zu niedrig                     'Technische Merkmale' angegebene konstante Stromstärke
                                                          bezieht sich auf jeweils ein Gel!

VWR_v0617_D                                                    32
Bedienungsanleitung PerfectBlueTM Vertikale Doppelgelsysteme

 Problem               Ursache                            Lösung

 Gel läuft zu schnell Puffer ist zu hoch konzentriert     Überprüfen Sie das Pufferrezept. Setzen sie frischen Puffer
                                                          an. Ist die Voltzahl niedriger als üblich bei gleicher Strom-
                                                          stärke, ist der Puffer wahrscheinlich zu hoch konzentriert.

                       Spannung oder Stromstärke          Reduzieren Sie die entsprechenden Parameter.
                       sind zu hoch

 'Smiling-Effekt' der Die Mitte des Gels erwärmt          Nutzen Sie die Kühloption des Gerätes.
 Lauffront            sich schneller als der Rand         Reduzieren Sie ggf. die Stromstärke.

 Banden dehnen         Diffusion der Proben beim          Die Proben sollten schnell nacheinander aufgetragen und der
 sich nach außen       Beladen                            Gellauf zügig gestartet werden.
 aus
                       Diffusion der Proben während Erhöhen Sie Konzentration des Sammelgels oder erhöhen Sie
                       des Laufes                   die Stromstärke während des Sammel-Laufes um 25%.

                       Niedrigere Ionenstärke der         Passen Sie die Ionenstärke der Probe der des Gels an.
                       Probe

 Banden liegen         Die Ionenstärke der Probe ist      Entsalzen Sie die Proben oder verwenden Sie einen Proben-
 enger zusammen        höher als die des Gels             puffer derselben Ionenstärke wie die des Gels.
 als die Taschen

 Breitere Spuren am Bei Verwendung unterschiedli- Verwenden Sie für alle Proben den gleichen Probenpuffer.
 Ende des Gels      cher Probenpuffer. Normal bei
                    Gradientengelen

 Verzerrte Banden      Gel war an den Taschen             Entgasen Sie die Gellösung vor dem Gießen und erhöhen
                       noch nicht vollständig             Sie die APS- und TEMED-Konzentrationen. Der Kamm kann
                       polymerisiert                      vorher mit TEMED abgewischt werden, um die Polymerisa-
                                                          tion zu verbessern.

                       Die Salzkonzentration der          Entsalzen der Probe durch Dialysieren, Entsalzungssäule
                       Proben ist zu hoch                 oder Präzipitation, Waschen und erneutes Lösen.

                       Keine gerade Oberkante des         Überschichten Sie das Trenngel im Anschluss an das Gießen
                       Trenngels                          vorsichtig mit N-Butanol oder Wasser.

                       Puffer des oberen Pufferre-        Das Gel muss vollständig auspolymerisieren. Das Glasplat-
                       servoirs läuft durch das Gel       tensandwich muss fest genug in den Seitenklemmen sitzen.
                       oder entlang der Spacer aus        Verwenden Sie immer Kämme und Spacer gleicher Stärke.

 Verschmierte          Überladung des Gels                Setzen Sie ein geringeres Protein- oder Probenvolumen ein.
 Banden/Spuren
                       Probe ist präzipitiert             Zentrifugieren Sie die Probe vor dem Beladen des Gels ab.

 Verzerrte Banden      Auslaufen des Puffers an den       Die Spacer sollten nach der Polymerisation nicht mehr
 in den äußeren        Seiten oder durch die Spacer       bewegt werden. Die Glasplatten müssen fest an die Gum-
 Spuren                innerhalb des Gelaufbaus           midichtungen des Mittelteils gedrückt sein.

 Silikondichtungen     Glasplatten kleben zu stark        Feuchten Sie die Silikondichtungen mit Wasser oder Puffer
 der Gießbasis         an den Dichtungen und              an, bevor Sie das Mittelteil auf die Gießbasis setzen. Dies
 nutzen sich           strapazieren diese beim            erhöht die Lebenszeit der Dichtungen deutlich.
 schnell ab            Abnehmen des Mittelteils

 Doppelbanden          Aufgrund von Reoxidation           Setzen Sie Probenpuffer mit reduzierendem Agens monat-
                       oder ungenügender Reduk-           lich frisch an. Erhöhen Sie die Konzentration von β-Mer-
                       tion der Probe                     captoethanol oder Dithiothreitol (DTT) in der Probe.

VWR_v0617_D                                                    33