DIAGNOSTIC MEDICAL IMAGING - 3rd Part - Nuclear Magnetic Resonance Imaging
←
→
Page content transcription
If your browser does not render page correctly, please read the page content below
DIAGNOSTIC MEDICAL IMAGING 3rd Part – Nuclear Magnetic Resonance Imaging Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Physics of Magnetic Resonance 2 Magnetic resonance scanners use the property of nuclear magnetic Resonance (NMR) to create images All nuclei have positive charges (they are composed by protons and neutrons). A nucleus with either an odd atomic number or an odd mass number has an angular momentum – they have spin The nuclei of the hydrogen atoms (¹H) have spin ½ Rotating charges create an angular magnetic momentum, the rotating nuclei of the hydrogen atoms similar to little magnets Φ N + + Nucleus angular Microscopic + + momentum magnetization of + nucleus + + S Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Physics of Magnetic Resonance 3 In normal conditions individual spins of ¹H nuclei have a random orientation, has results no macroscopic magnetic field is produced If the nuclei of the hydrogen atoms (¹H) are subjected to a strong magnetic Field, B0 , they tend to align with the field (parallel and anti-parallel orientation); being the number of hydrogen atoms into the human body very high, this tendency results in a magnetization of the body A little more than half of the nuclei will be oriented in the same direction of the external magnetic field, “up” oreintation (having a lower energy content), while the others nuclei will be oriented in the opposite direction, “down” oreintation (having higher energy content). Random thermodynamic interaction between magnetic dipoles and the surrounding macro-molecules create a continuous change in spin orientation. The dynamic difference between “up” and “down” nuclei is strictly related the external field intensity. Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Physics of Magnetic Resonance 4 Macroscopic magnetisation grows as B0 grows A magnetic field of 1.5 T creates a difference between up and down spin oreintation of 6 p.p.m. 1 mmc of tissue 1019 hydrogen nuclei a significant magnetic magnetisation is produced Actually, nuclei spin precess around an axis along the direction of the field. This precession has a frequency, called Larmor frequency (proportional to B0 ), of the order of MHz (radiofrequency). As an example the precession frequency for an external field of 1.5 T is 64 MHz B0 Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Macroscopic Magnetisation 5 It i simportant to underline that there is no precession phase coherence between nuclei z The xy plane components of the microscopic magnetisation vectors are placed in every directions, their sum is LMM0≠0 zero. This component of the macroscopic magnetisation is called Transversal x y Macroscopic Magnetisation TMM0=0 The z plane components of the microscopic magnetisation vectors are placed both in z and –z directions. Their sum is positive along z direction. This component of the macroscopic magnetisation is called Longitudinal TMM0=0 y Macroscopic Magnetisation LMM0≠0 x Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Macroscopic Magnetisation 6 The Macroscopic Magnetisation produced by the external magnetic field is static and smaller than B0 it is not possible to measure it An external excitation has to be introduced in the system If a microscopic sample of nuclei is excited using a electromagnetic radiation having Larmor frequency, the radiation magnetic component interacts with nuclei magnetic moment A quantum of energy is absorbed changing the nuclei energy status from “up” to “down” The longer is the RF pulse duration the higher is the energy transfered to the system (the higher is the number of nuclei that changes their status from up to down) When these energy transitions occur, nuclei are resonant with applied radiation The Bloch equations are a set of coupled differential equations which can be used to describe the behavior of a MM vector Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Macroscopic Magnetisation 7 When the difference between up and down spin populations decreases their precession phase syncronisation grows It is possible to set the RF pulse duration to reach the condition of having the same “up” and “down” spin population that have complete precession phase coherence This RF is called “90° Pulse” The system subject to the 90° Pulse has LMM1=0 while the TMM1 reach its maximum, the TMM rotate in the xy plane with the nuclei precession frequency (the RF pulse one) z The MM, following a spiral movement with growing ray, deflects from its position along z axis to a position in the xy plane, so it is rotated by 90° TMM1 growing x y y x LMM1 decreasing Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Macroscopic Magnetisation 8 If the duration of the RF pulse is doubled (w.r.t. the one of the 90° pulse) It is possible to obtain the total “up” and “down” populations reversal (w.r.t. the initial situation), in this status there is no precession phase coherence between nuclei LMM2= -LMM0 , TMM2=0 This RF is called “180° Pulse” It is clear that it is possible to change RF pulse amplitude and duration so that different degrees of deflection of MM can be obtained The displacement of MM from its longitudinal direction leads to a unstable system (no energetic equilibrium), having high energy and high phase nuclear synchronisation When the external RF ends the system goes back to initial conditions Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Relaxation 9 When the external RF ends, the rapidly rotating (and decreasing) TMM creates a radio frequency excitation within the sample that will in turn induce (Faraday induction) a voltage in a coil of wire located outside the sample. This signal is recorded for use in MRI. Transverse relaxation, also known as spin-spin relaxation, acts first to cause this received signal to decay. This relaxation is caused by the perturbations in the magnetic field due to other spins that are nearby. This interaction causes spins to momentarily speed up or slow down, changing their phases relative to other nearby spins (dephasing). The resulting signal is called free induction decay (FID), an exponential decay having a time constant called transverse relaxation time, T2 Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Free Induction Decay 10 FID amplitude is a function of the number of nuclei in the sample, the decreasing time is a function of the decreasing speed of TMM It is not easy to botain a good FID sampling, due to the fact that the transmitting and receiving coils are the same, moreover the FID signal decreases very quickly and is effected by non homogeneous magnetic field of the structure The RF pulses practically used in NMR imaging create spin echoes Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Relaxation 11 T2 is the time required to the TMM to reach the 37% of its initial value (reached when the RF pulse is applied) There are two factors causing the decreasing of phase coherence of resonant nuclei -The energy transitions of nuclei that changes their status from down to up - Interactions between spin and molecular micro-EM field, these interactions do not vary the energy of the system but create local variations of the static field, B0 ,inducing a change in precession velocity (phase displacement) Actually, local perturbations of B0 cause the received signal to decay exponentially with a time constant T2*, lower than T2 This effect is reversible using the technique of coherence refocusing thorugh the concept of echoes (that will be described later) Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Relaxation 12 The second relaxation mechanism is called longitudinal relaxation or spin-lattice relaxation.This process concerns the longitudinal magnetisation which recovers back to its equilibrium value as a rising exponential having a time constant called longitudinal relaxation time, T1 T1 is the time required to the LMM to reach the 63% of its equilibrium value T1 is a function of how fast the spin systems release energy to return from down to up position, this thermodynamic energy exchange involves the surrounding molecules (this process is a function of B0 ) T1 and T2 are different for various types of tissues and are responsible for generating contrast in MR images. For tissue in the body the relaxation times are in the ranges: 250 ms ≤ T1 ≤ 2500 ms , 25 ms ≤ T2 ≤ 250 ms Usually 5T2 ≤ T1 ≤ 10T2 and for all materials T2 ≤ T1 This is due to the fact that there are spin-spin interactions that cause phase coherence loss without changing up and down populations, therefore T1 relaxation gives contribution to T2 relaxation but not vice-versa Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Relaxation 13 Mz = Mo ( 1 - e-t/T1 ) MXY =MXYo e-t/T2 Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Relaxation 14 Proton density, PD, is the number of resonant nuclei per unit volume. The greater is PD the greater is the relaxation signal intensity It has to be outlined that not all the hydrogen nuclei in the tissues will give contribution to the MR signal, just the resonant ones (in particular the ones that are in “free water”). Solid structures that are made of a great number of nuclei, as bones, have a little percentage of mobile resonant nuclei therefore the MR signal is very low. Larmor frequency is a function of the applied magnetic field but it is also influenced by the electron clouds surrounding the nuclei that creates a small local EM field that ineracts with B0 Chemical shift is the variation of the resonance frequency in different substrates that is a function of the shielding effect produced by atom electrons on B0 Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Pulse Sequences 15 Generation of Spin Echoes: when the RF 90° pulse ends spins begin to point in different directions in the transverse plane, therefore there are “faster” and “slower” spins. If a RF 180° pulse is applied to the system the spin directions in the transverse plane is inverted and from this new position the fast spins “catch up” and the slow spins “fall back”. Therefore the signal generated by transverse spins recovering their coherence creates a spin echo that can be easily measured. The time interval from the initial 90° pulse to the formation of the spin echo is called echo time TE, the application time of the 180° pulse is TE/2 Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Pulse Sequences 16 Generation of Spin Echoes Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Contrast Mechanism 17 In MRI the ability to generate tissue contrast depends on both the intrinsic MNR properties of the tissue (PD, T1, T2) and the characteristics of the externally applied excitations. It is possible to control the tip angle, the echo time, TE, of the RF excitation and the pulse repetition interval, TR. This does not means that the brightness of the images are proportional to one of the three parameters but merely that the differences in intensity seen between different tisuses are largely determined by the difference in one of the three parameters Three time of weighted contrast can be used: • PD-weighted • T1-weighted • T2-weighted Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Contrast Mechanism 18 PD-weighted T2-weighted T1-weighted Tissue Type Relative PD T2 (ms) T1 (ms) White matter 0.61 67 510 Grey matter 0.69 77 760 Cerebrospinal fluid 1.00 280 2650 Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Contrast Mechanism 19 T1-weighted contrast: the image intensity should be proportional to T1. The differences in the longitudinal component of magnetisation must be emphasized. Short TE and TR has to be used. Short TE allows to image the echo signal when the phase coherence loss, dependant on T2, is small (so that the differences between tissues that are a function of T2 are minimised). Short TR, lower than the one of the tissue being imaged, allows the measured signal to be dependant on the LMM recovery time, i.e. short T1 tissues give rise to stronger signals. Usually TR is shorter than the T1 of the tissues having long T1 and of the order of T1 of the tissues having short T1. Intensity of tissues having long T1 will be very low (dark-gray colours), tissues having short T1 will look very bright (white-gray colours). In the brain-slice picture TR=600 ms, TE=17 ms and tip angle is 90° Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Contrast Mechanism 20 T1-weighted contrast Image Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Contrast Mechanism 21 T2-weighted contrast: the image intensity should be proportional to T2. Differences in the transverse relaxation times of different tissues must be apparent. Long TE and TR has to be used. Long TE allows to image the echo signal when the phase coherence loss is high (so that the differences between tissues that are a function of T2 are maximised). Long TR, greater than the T1 of the tissue being imaged, allows the measured signal to be not dependant on the LMM recovery time, so that the T1 differences between tissues are very negligible. For tissues having long T2, as fluids, the duration of phase cohernece is very high, so if the signal is analysed after a long TE the echoes intensity is still high. Intensity of tissues having short T2 will be very low (dark-gray colours), tissues having long T2 will look very bright (white-gray colours). In the brain-slice picture TR=6000 ms, TE=102 ms and tip angle is 90°. This value of TR is practically not usable because the images take too long to acquire. Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Contrast Mechanism 22 PD-weighted contrast: the image intensity should be proportional to the number of resonant hydrogen nuclei in the sample. Short TE and long TR has to be used. Starting form the sample in equilibrium a excitation RF pulse is applied then, before the signal has a chance to decay from T2 effects, the relaxation signal has to be imaged quickly. Therefore long TR (which allows the tissues to be in equilibrium, no dependance on T1) and either no echo or short TE (in order to minimize dependence from T2 decay) has to be used. The preferred tip angle is 90°, to obtain maximum signal. Intensity of tissues having low PD will be very low (dark-gray colours), tissues having high PD will look very bright (white-gray colours). In the brain-slice picture TR=6000 ms, TE=17 ms and tip angle is 90°. This value of TR is practically not usable because the images take too long to acquire. Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
MRI Scanner 23 An MRI scanner consist of five principal components: • The main magnet • A set of switchable gradient coils • RF coils • Pulse-sequence end receive elctronics, used to program timing of transmission and reception of signals • Console for viewing, manipulating and storing images. Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
MRI Scanner 24 The main magnet is commonly a cylindrical superconducting magnet having field strength ranging from 0.5 to 7 T The gradient coils produce the change in local magnetic field necessary to encode the spatial location of the MR signal (this will be discussed later) The RF coils, or resonators, both induce the RF signals to tip the magnetisation vector and have current induced in them by the spin systems. Therefore RF coils are used for the transmission and reception of RF pulses. Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Image Formation 25 How can we encode spatial position of MR signals? An MR scanner can create images at arbitrary location and orientation, for simplicity, the formation af axial images will be discussed The MR signal contributions coming from different voxels is obtained through the use of magnetic field gradients and the spatial position is encoded using frequency and phase (of the signal) Lets apply a field gradient on the x-axis to the external field. Due to the fact that the spin precession frequency is directly proportional to the field, the sample spins will be characterised by different precession velocities (along x-axis) Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Image Formation 26 Lets consider 8 water samples, and a External field intensity MR signal intensity magnetic field that crosses the samples Static B, the MR signal frequency is concentrated x axis Frequency If a gradient is applied, the External field intensity MR signal intensity MR signal frequency analysis can be used to identify the sum of the voxels signals along the planes perpendicular to the gradient x axis Frequency Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Slice Selection 27 In CT and ultrasound the energy used to image the selected slice is restricted to the slice itself, there is no other part of the body from which signal can arise. In SPECT and PET the whole body is a potential source but the observed signal is selected by collimation so that it belongs to a specific slice. In MRI both these approaches can be used, it is possible to excite only a selected slice or it is possible to excite the entire volume and then to extract images of selected slice. The first technique is called 2-D MR imaging, the second is called 3-D MR imaging. Only the first technique will be analysed. Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
Slice Selection 28 A magnetic field gradient along z-axis, High Gradient called selection gradient, is used to select the slice to be imaged The higher is the gradient the Low Gradient thinner is the slice, considering RF pulse a fixed bandwidth of the RF bandwidth pulse The RF pulse is a sinc function that has a rectangular Fourier transform able to excite a range of frequencies which in turn High Gradient excites a range of tissues layer selection Low Gradient layer selection Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
“Projection and Reconstruction” Method 29 Once the slice is selected another gradient has to be applied in order to encode the spatial position of the MR signal. This new gradient has a different position w.r.t. the selection gradient and it is located along x-axis and it is called readout gradient. The first storical method used to lacate MR signal is called projection and reconstruction After the application of the selection gradient, an echo signal is created through the use of a 90° RF pulse followed by a 180° RF pulse. During the creation and for all the duration of the echo a readout gradient is applied, for example along x-axis, so that it is possible to discriminate in the frequency domain the contributions coming from different “strips” perpendicular to x. A “projection” of the selected slice along x-axis is obtained. In order to reconstruct the image multiple projections have to be created. Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
“Projection and Reconstruction” Method 30 Different projections can be obtained in the same way described before but applying a combination of readout gradients along x-axis and y-axis so that to obtain different gradients along xy-plane A number of equally spaced projections between 0° and 180° are acquired The collected data are processed using one of the methods used to reconstruct CT images: • Forurier Method • Filtered Backprojection • Convolution Backprojection From the MR signals it is possible to obtain a map of multiple parameters, PD, T1 and T2 Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
“Projection and Reconstruction” Method 31 Monodimensional projection, having an Amplitude Slice selection, angle defined by Gx through Gz and RF and Gy Frequency Nuclear signal echo echo Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
2D DFT Method 32 In order to decrease the computational complexity the image reconstruction based on the bidimensional Fourier transform is used. For this method a series of projections is requested in order to reconstruct the image. After the application of the selection gradient, an echo signal is created through the use of a 90° RF pulse followed by a 180° RF pulse. During the creation and for all the duration of the echo a fixed readout gradient along x-axis is applied (the readout gradient is the same for every projection). The signal position is coded using for every projection a different Gy, applied right after the 90° pulse and before the 180° pulse in order to create a spin dephasing (along xy-plane) that is a function of the position along the y-axis of the element that generated the signal. Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
2D DFT Method 33 No gradient: Gradient on: Gradient of: spin in phase spin dephasing final phase remembered Gy time y0 Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
2D DFT Method 34 If a Gy gradient is applied before the creation of the echo the nuclei located in a position with higher y coordinate precess faster than the ones having lower y coordinate. When the Gy gradient ends all the nuclei start precess at the same velocity. So it has been created a dephasing that is a function of the position along the y-axis. After the 180° pulse an echo rise up. Afterwards the Gx gradient is applied in order to change the precession frequency of the spins as a function of their position along x-axis, but these spins will maintain the dephasing acquired during the application of the Gy gradient that codes their position along y-axis. The Gy gradient is called phase encoding gradient. The Gz gradient select the slice while phase and frequency of the measured signal reflect its postion along y and x axes respectively Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
2D DFT Method 35 One monodimensional projection along x direction having phases Amplitude Amplitude coded by Gy Slice selection, through Gz and RF Frequency Frequency echo echo Nuclear signal Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
36 Example (1/5) Phase We record the phase change at change at the top of the knee during the the top of phase encoding process the knee phase encoding Rate of change each K-Space of phase signal is from the whole knee frequency encoding We started with a large positive phase change, and ended up with a large negative phase change. Over all the phase encoding steps, we have a rate of change of phase = frequency Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
37 Example (2/5) Phase change We record the phase change for for the section another section of the knee during of the knee the phase encoding process phase encoding Rate of change each K-Space of phase signal is from the whole knee frequency encoding We started with a medium positive phase change, and ended up with a medium negative phase change. Over all the phase encoding steps, we have a rate of change of phase = frequency Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
38 Example (3/5) No phase We record the phase change for the change for the central section of the knee during the central section phase encoding process of the knee! phase encoding each K-Space signal is from the whole knee frequency encoding During the phase encoding of the central section of the knee, the Gy phase encoding gradient is zero! Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
39 Example (4/5) Phase change We record the phase change for for the section another section of the knee during of the knee the phase encoding process phase encoding Rate of change each K-Space of phase signal is from the whole knee frequency encoding We started with a medium negative phase change, and ended up with a medium positive phase change. Over all the phase encoding steps, we have a rate of change of phase = frequency Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
40 Example (5/5) Phase change We record the phase change for for the bottom the bottom of the knee during the of the knee phase encoding process phase encoding Rate of change each K-Space of phase signal is from the whole knee frequency encoding We started with a large negative phase change, and ended up with a large positive phase change. The Fourier transform can separate out different frequencies (even ones that are made from a rate-of-change of phase over many phase encoding steps), we now have a way of determining the total signal from each of the rows! Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
2D DFT Method 41 All the acquired data are stored in the so called K-space Each row of the K-space is a monodimensional projection data; then, in order to reconstruct the image a bidimensional DFT is applied to the K- space The first DFT is applied to all the rows obtaining information on intensity of the frequency components and on their phases The second DFT is applied to all the columns obtaining information on the RM signal charcateristics This method is preferred with respect to the Projection and Reconstruction one due to the fact that is more efficient w.r.t. computational complexity; in fact FFT can be used to reduce complexity. Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
2D DFT Method 42 Two voxels having different magnetization Raw Data Two oscillations frequencies in the time domain & two oscillation frequencies in the phase domain Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
2D DFT Method 43 FFT in the frequency FFT in the phase encoding direction encoding direction Oscillations in the phase encoding direction Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
MR Image Quality 44 MR image quality is affected by intrinsic and extrinsic factors Intrinsic factors are a function of the examined tissues and affect signal Intensity, these factors are: Proton density, Spin-spin relaxation time, Spin-lattice relaxation time, Chemical shift, Movement, Temperature. Extrinsic factors are a function of the scanner, the installation site, environment and patient, these factors affect contrast and spatial resolution. Contrast is essentially a function of RF sequence, as previously explained, and external magnetic field intensity. Spatial resolution is a function of magnetic field homogeneity and gradients linearity; these two factors optimize the spatial location of the signal. Also RF Tx-Rx equipments affect spatial resolution. Moreover spatial resolution grows as pixel dimensions and slice tickness decrease and as signal to noise ratio, S/N, grows. Tommaso Rossi - Modulo di SEGNALI, a.a. 2019/2020
You can also read