INSTI HIV-1 HIV-2 Antibody Test Kit - BioLytical Laboratories
←
→
Page content transcription
If your browser does not render page correctly, please read the page content below
Clinical Procedure Document Number: INSTI HIV-1 HIV-2 Antibody Test Kit – PROV-120 BioLytical Laboratories Version Number: 1 Approved Date: July 8, 2021 Approving Authority (Sponsor): Microbiology Discipline Leads Next Review Date: Clinical Standard Owner: Director – Clinical Quality Safety Logistics July 8, 2022 Contact for Interpretation: LQR@saskhealthauthority.ca Healthcare providers eligible to perform this function: Qualified Document Type: professionals as indicated by Lab License (May include: Medical Procedure – Laboratory Medicine Laboratory Professional, Registered Nurse, Licensed Practical Nurse, Respiratory Therapist, Paramedic, Registered Psychiatric Nurse, Physician, or a Nurse Practitioner and Physical Therapist) Key words: HIV, HIV POC, HIV POCT, INSTI Kit, Point of Care 1. PURPOSE To provide instruction on how to perform HIV Point of Care testing. A reactive INSTI test result should be considered a preliminary result, with appropriate counseling provided in point-of-care (POC) settings. Following a reactive, indeterminate or invalid rapid test result, a venous blood sample must be drawn and forwarded to the Roy Romanow Provincial Laboratory (RRPL) for HIV additional testing. 2. PRINCIPLES The INSTI™ HIV-1/HIV-2 Antibody Test is a single use, rapid, flow-through in vitro qualitative immunoassay for the detection of antibodies to Human Immunodeficiency Virus Type 1 and Type 2 in human EDTA whole blood, EDTA plasma, fingerstick blood or serum. The test is intended for use by trained personnel in medical facilities, clinical laboratories, emergency care situations, and physicians’ offices as a diagnostic test capable of providing results in less than one minute. Although suitable for near-patient or POC testing, the INSTI HIV-1/HIV-2 Antibody Test is not suitable for home testing. All required pre- and post-test counselling guidelines must be followed in each setting in which the INSTI HIV-1/HIV-2 Antibody Test is used. Page 1 of 19 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: 3. ROLES AND RESPONSIBILITIES HIV POC testing is beneficial where immediate knowledge of patient’s HIV statue is considered important. Criteria for testing can be found in The Guidelines for the Use of HIV Point of Care (POC) Test Kits in Saskatchewan on the RRPL Compendium of Tests: https://rrpl- testviewer.ehealthsask.ca/ or https://publications.saskatchewan.ca/api/v1/products/11997/formats/81578/download. 3.1 May include: Medical Laboratory Professional, Registered Nurse, Licensed Practical Nurse, Respiratory Therapist, Paramedic, Registered Psychiatric Nurse, Physician, or a Nurse Practitioner and Physical Therapist Perform and complete testing in facilities where testing has been implemented Participate and complete required training. See Appendix E: Laboratory Test or Instrument General Training Checklist Perform patient analysis, run external Quality Assurance, perform internal QC and follow QC procedures, troubleshooting and maintain competency, follow accreditation requirements Qualified employees will undergo training to perform INSTI HIV-1/HIV-2, including but not limited to reading the procedure and demonstration of test being performed During initial training, each employee will physically run controls, not just observe test being performed. Results of this initial control run will be recorded. Only after successful completion of training with control runs passing, an employee is able to test patients After competency has been established, each employee will observe the test being run at minimum every 6 months and perform kit controls or external quality controls from the College of Physicians and Surgeons every 12 months Document all observed and physical run controls by each employee with supervisor sign-off for proof of proficiency which can be produced during inspections, as requested. Refer to Appendix A: HIV POCT Quality Control Log 3.2 Manager/Supervisors Ensure that proficiency testing has been performed and reported Train all users. See Appendix E: Laboratory Test or Instrument General Training Checklist Retain documentation of staff training and competency of staff Manage supplies, controls and proficiency testing Ensure training for all new staff that would perform test Ensure competency of all staff that will perform test Sign-off all training and competency documentation Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 2 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: 4. SPECIMEN INFORMATION Specimen Type/Source EDTA-whole blood, EDTA-plasma, serum, fingerstick sample (capillary) Acceptable Collection Containers Lavender-top EDTA (anticoagulant tubes), red-top (no anticoagulant tube), yellow-top SST serum tubes (validated at RRPL) Minimum Requirements 50μl of whole blood, plasma or serum for testing Specimen Stability If plasma or serum is being used, separate from the blood cells by centrifugation Serum or EDTA-plasma is stable at 2-8°C for up to 5 days Do not heat or repeatedly freeze/thaw specimens Whole blood EDTA is stable at 2-8°C and should be tested within 48 hours Do not heat or freeze whole blood specimens Fingerstick sample must be tested immediately Storage Requirements Serum or EDTA-plasma - stored frozen at ≤ -20°C for 3 months, or stored frozen at ≤ -70°C for one year Whole blood can not be frozen or stored Fingerstick samples can not be frozen or stored Specimen Handling Do not dilute or pool samples prior to testing 5. EQUIPMENT & SUPPLIES, REAGENTS, FORMS & LABELS Equipment & Supplies Reagents Forms & Logs BioLytical Kit and components: INSTI Reagents Appendix A: Membrane unit, Sample Diluent, Membrane Unit HIV POCT Quality Color Developer, Clarifying Sample Diluent (solution 1) Control Log Solution, alcohol swab, lancet, Color Developer (solution 2) Appendix B: single-use capillary pipette HIV POCT Incident Log Clarifying Solution (solution 3) Personal protective equipment Appendix C: Biohazard waste containers HIV POCT Result Log INSTI HIV-1/HIV-2 Test Controls: Precision pipette (50μl) or single- Appendix D: Ordered separately use bulb pipette HIV Point of Care Absorbent cotton balls HIV-1 Supply Requisition SHA approved HIV-2 cleaner/disinfectants Negative For Venipuncture Blood Collection Venipuncture collection supplies Lavender-top EDTA or serum blood collection tubes Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 3 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: 6. REAGENT INFORMATION Membrane Unit Sample Diluent Color Developer Clarifying Solution (solution 1) (solution 2) (solution 3) Appearance White square plastic Clear Blue Clear unit with inner circle membrane inside a well Preparation Ready to use Ready to use. Sample Ready to use, Ready to use, no is added to this invert 2-3X mixing or container and mixed immediately preparation by inversion. before use. required Number of uses Single use only Single use only Single use only Single use only Kit Information Do not mix reagents from different lots INSTI components should be stored at 2-30oC Do not use reagents or kits beyond the stated expiration date Adding an excessive amount of specimen may cause the device to overflow or leak Do not use the Membrane Unit if the foil pouch has been previously opened or if the packaging integrity has been compromised. Once the Membrane Unit has been opened, it must be used immediately Controls are not contained in the kit and must be ordered separately 7. SAFETY PRECAUTIONS 7.1 Hazards Treat all patient specimens as highly infectious Puncture hazard working with venipuncture collection supplies and lancets 7.2 Safe Work Practices Use proper technique to not contaminate yourself, nor create aerosols All laboratory personnel shall don required personal protective equipment, as per standard precautions, prior to performing the task(s) outline in this standard operating procedure, as per Job Safety/Hazard Analysis Thoroughly wash hands after handling or performing this test Do not smoke, eat, or drink in areas where specimens or kit reagents are being handled Do not pipette by mouth Avoid contact with skin and eyes. If contact occurs, wash affected areas with water Avoid forming aerosols Dispose of all specimens and materials as biohazardous waste Spills should be cleaned up and decontaminated in accordance with the user facility’s established procedures for handling biohazardous spills Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 4 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: 8. QUALITY CONTROL Quality Control Internal Quality Control: The INSTI HIV-1/HIV-2 Antibody Test has a built-in IgG capture procedural control that demonstrates assay validity and adequate sample addition. A blue color in the control spot indicates that the proper specimen was added and that the assay procedure was performed correctly. The control spot will appear on all valid INSTI tests. (Refer to Interpretation of Results, below.) Control Criteria Three levels of Kit Controls: HIV-1, HIV-2 and Negative (supplied separately): The controls are used to verify test performance and interpretation of results. Kit controls must be run under the following circumstances: For new INSTI user verification When switching to new lot number of INSTI test kits Whenever a new shipment of kits is received (same or different lot number) When temperature during storage of the kit falls outside of 2-30°C When the temperature of the test area falls outside of 2-30°C If a site conducts >24 point-of-care tests per day, the controls should be run every day If a site conducts
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: Alternate QC Proficiency Testing in addition to QC Measures Storage of QC Data 2 years Initial HIV POC Initial set-up of becoming an HIV Point of Care location. location setup Instructions on How to Set Up an HIV Point of Care Site v4.0.pdf (ehealthsask.ca) 9. PROCEDURE Step Action 1 Obtain kit and bring to room temperature before performing the test. Note: All Test Units are single use only and must be used immediately once opened. All reagents should be dispensed evenly in the center of the well. Once test is completed, dispose components into appropriate biohazard container. 2 Gather support materials (swab, lancet, pipette), one sealed test pouch containing INSTI Membrane Unit, and one vial each of the Sample Diluent (solution 1), Color Developer (solution 2), and Clarifying Solution (solution 3) for each test to be performed. 3 Collect specimen If Then Sample Caution: The amount of sample (fingerstick blood) is critical. Fingerstick Blood To ensure that the proper amount of blood is achieved, follow these instructions carefully: o Label Sample Diluent vial and tab of membrane unit with control or patient information o Massage the finger to allow the blood to move to the surface (fingertip will become pink). Use heating pad if available to warm the hand. Hand must be positioned at waist level or lower. o Wipe the fingertip with the alcohol swab o As soon as the finger is dry, twist off the protective cap from the lancet, and then pull it straight out. Press the finger firmly at the point just below where the lancet will be applied. With the other hand, hold the lancet by the body and press the lancet body firmly against the puncture site to activate the device. Immediately dispose the used lancet into a proper sharps container. o As the blood bubbles up, hold the capillary pipette horizontally and touch the tip of the pipette to the blood sample. Capillary action automatically draws the sample to the fill line and stops. If very little blood trickles out of the puncture, gently apply intermittent pressure near the puncture site to obtain the required blood volume. If blood is inadequate, perform a second skin puncture using a new lancet. Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 6 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: Capillary Pipette CAUTION! Filling is automatic: Never squeeze the capillary pipette while sampling. o Transfer the blood held in the pipette to the Sample Diluent vial (solution 1). Align the tip of the capillary pipette with the Sample Diluent vial and squeeze the bulb of the capillary pipette to dispense the sample. Note: If the sample will not expel, hold the capillary pipette vertically and slide a finger over (without pressing) the vent hole, then squeeze the bulb. Recap the vial and mix by inversion. Proceed to step 4, below. Sampling Collect EDTA-whole blood, EDTA-plasma or serum specimens, follow normal EDTA Whole venipuncture blood collection procedures using lavender-top EDTA anticoagulant Blood, tubes (for whole blood and plasma), or red-top (no anticoagulant) or yellow-top SST serum, EDTA- tubes for serum. If plasma or serum is to be used, separate from the blood cells by plasma and centrifugation. Test Controls Bring specimens to room temperature and mix each specimen thoroughly prior to use. Do not heat or repeatedly freeze/thaw specimens. Label Sample Diluent vial and tab of membrane unit with control or patient information. Using a pipette, add 50μl using precision or bulb pipette of whole blood, serum, plasma, or kit controls (see Note) to the Sample Diluent (solution 1) vial. Recap the vial and mix by inversion. Adding an excessive amount of specimen may cause the device to overflow or leak. Proceed to step 4, below. Bulb Pipette Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 7 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: Note: In POC settings, for INSTI kit controls, it is important to use a 50μl precision or bulb pipette device to add the control material to the Sample Diluent vial. Do not use the disposable single-use capillary pipette provided for fingerstick blood collection. 4 Perform test Tear open the pouch and carefully remove the single use Membrane Unit without touching the center well. Place the unit on a level surface. For sample identification purposes, the tab of the Membrane Unit may be labelled with the control or patient’s identifier. NOTE: At this point, it is important that the following steps be performed immediately and in sequence Remix the Sample Diluent-specimen mixture and pour the entire contents to the center of the Membrane Unit well. (Note: Do this within 5 minutes after the specimen has been added to the Sample Diluent vial). The sample should be absorbed through the membrane in less than 30 seconds; however, absorption times will vary slightly depending upon sample type. Re-suspend the Color Developer by slowly inverting to mix the solution thoroughly. Continue this process until careful visual observation confirms that the reagent is evenly suspended. Open the Color Developer and add the entire contents to the center of the Membrane Unit well. The colored solution should flow through completely in about 20 seconds. Open the Clarifying Solution and add the entire contents to the center of the Membrane Unit well. This will lighten the background color and facilitate reading. Immediately read the result while the membrane is still wet. Do not read the results if more than 5 minutes has elapsed following the addition of Clarifying Solution. If more than 5 minutes have elapsed, repeat the test. 10. INTERPRETATION OF RESULTS AND EXPECTED VALUES Interpretation of Results: Do not read the results if more than 5 minutes has elapsed following the addition of Clarifying Solution. If more than 5 minutes have elapsed, repeat the test. If using the control samples provided by the Roy Romanow Provincial Laboratory, all positive controls must be reactive with INSTI and all negative controls must be non-reactive with INSTI. Controls that produce incorrect or invalid results must be re-tested with INSTI. If control results are still incorrect or invalid, inform Roy Romanow Provincial Laboratory immediately. Non-Reactive One blue dot that is clearly discernable above any background tint should appear on the membrane. This is the procedural Control Spot and shows that the test has been performed correctly. The Control spot location is indicated by the letter C. No reaction should be visible at the test spot, located below the control. A non-reactive result indicates that antibodies to HIV-1/HIV-2 were not detected in the specimen. Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 8 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: Reactive Two blue dots that are discernable above any background tint indicate that the specimen contains HIV-1/HIV-2 antibodies. One dot may be darker than the other. A sample giving this pattern is considered a preliminary reactive. Following a reactive rapid test result, a venous blood sample must be drawn and forwarded to the Roy Romanow Provincial Laboratory for HIV additional testing, see Section 11. Additional Testing Information. Invalid The test is invalid if any of the following occurs: There is no dot on the membrane The test dot appeared without the control dot Uniform tint across the membrane Only blue specks appear on the membrane Note: Invalid tests with fingerstick blood samples in POC settings should be repeated with a fresh sample using a new membrane unit, kit components and support materials. Invalid tests with EDTA whole blood, EDTA plasma or serum samples in laboratory settings should be repeated using a new membrane unit and kit components. Note: If a patient has two consecutive invalid INSTI test results, a venous blood sample must be drawn and forwarded to the Roy Romanow Provincial Laboratory for HIV additional testing, see Section 11. Additional Testing Information. Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 9 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: Indeterminate The test is indeterminate if a faint background ring appeared on the test area. Following an indeterminate INSTI test result, a venous blood sample must be drawn and forwarded to the Roy Romanow Provincial Laboratory for HIV additional testing, see Section 11. Additional Testing Information. Interpretation Notes: Depending on the antibody titer, a reactive specimen may be less intense in color than the procedural control, or vice versa. Only a blue spot of color discernibly darker than the background color should be interpreted as reactive or positive. Only tests exhibiting distinct fully formed blue test dot combined with a distinct fully formed blue control dot should be interpreted as reactive. Color intensity may be variable within or between the dots. The absence of a distinct control dot usually indicates that the sample volume was insufficient. A test resulting in a uniform blue tint across the entire membrane, thus obscuring the control and test spots, can occur when more than 60uL of whole blood is used and the flow through the assay membrane is obstructed. An individual who has a non-reactive result but was involved in HIV-risk is likewise recommended to obtain additional testing over the next months. 11. ADDITIONAL TESTING INFORMATION For reactive, indeterminate and two invalid INSTI test result, forward sample to Roy Romanow Provincial Laboratory for additional HIV testing. Sample collection, refer to the RRPL Compendium for additional details. o Serum (SST) - venous blood sample follow normal venipuncture blood collection procedures and centrifuge before shipping. o Lavender-top EDTA plasma - venous blood sample follow normal venipuncture blood collection procedures. Centrifuge and remove plasma from cells before shipping. Order HIV Screen test on requisition. Sample must be shipped Category B, UN3373 and in appropriate shipping containers. Only employees with valid TDG 6.2 certificates are able to ship samples. If HIV Viral Load testing is also required, send an additional requisition and frozen EDTA plasma, according to HIV Viral Load protocol. Remove plasma from cells and freeze before shipping. Samples must be sent and received frozen. Please refer to the RRPL Compendium for additional collection and shipment details. Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 10 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: 12. METHOD LIMITATIONS In some instances, samples may exhibit longer than normal flow times (from the time the Sample Diluent specimen mixture is poured in the membrane well to the time the Clarifying Solution has fully flowed through the membrane). This is due to variable factors such as cellular components, especially with whole blood. In instances of long flow times, a faint shadow in the form of a ring may appear at the test spot location, but th is should not be interpreted as a reactive result. This should be considered as an indeterminate result. In these instances, a venous blood sample must be drawn and forwarded to the Roy Romanow Provincial Laboratory for HIV additional testing, see Section 11. Additional Testing Information. The INSTI HIV-1/HIV-2 assay procedure and the interpretation of result must be followed closely when testing for the presence of antibodies to HIV in serum, plasma or whole blood. Absence of antibodies to HIV does not indicate that an individual is absolutely free of HIV-1 or HIV-2; HIV has been isolated from seronegative individuals prior to seroconversion. Insufficient data are available to interpret tests performed on other body fluids, pooled blood or pooled serum and plasma, or products made from such pools; therefore, testing of these specimens is not recommended. The INSTI HIV-1/HIV-2 assay has not been validated for detection of antibodies to HIV-1 Group N subtypes. Additional test limitations, warnings and performance characteristics of test can be located in the package insert. The INSTI HIV-1/HIV-2 assay detects antibodies to HIV-1/HIV-2 and is useful in establishing infection with HIV. Because a variety of factors may cause non-specific reactions, a patient found to be positive using the INSTI HIV-1/HIV-2 assay should have a sample drawn for laboratory-based additional testing. A person who has antibodies to HIV is presumed to be infected with the virus and appropriate counseling and medical evaluation should be offered. 13. MATERIALS MANAGEMENT BioLytical INSTI HIV-1/HIV-2 Antibody Test Kits and controls can be ordered through the Roy Romanow Provincial Laboratory. Refer to Appendix D: HIV Point of Care Supply Requisition. 14. PROCEDURE MANAGEMENT The management of this procedure including procedure communication, education, implementation, evaluation and audit is the responsibility of Laboratory Quality and Regulatory Manager, Regina Area. Renewal and amendment is the responsibility of the Laboratory Quality and Regulatory Manager, Regina Area and Laboratory Quality Safety and Logistics Director. Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 11 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: 15. REFERENCES BioLytical Laboratories INSTI HIV-1/HIV-2 Antibody Test package insert 51-1028N, 19-Sept- 2019 The Laboratory Biosafety Guidelines, 3rd Edition. Office of Laboratory Security, Health Canada. 3.1.2, 2004. Instructions on How to Set Up on HI Point of Care Site Instructions on How to Set Up an HIV Point of Care Site v4.0.pdf (ehealthsask.ca) Guidelines for the use of HIV Point of Care (POC) Test Kits in Saskatchewan, 2017 on the RRPL Compendium of Tests: https://rrpl-testviewer.ehealthsask.ca/ or https://publications.saskatchewan.ca/api/v1/products/11997/formats/81578/download RRPL Compendium of Tests: HIV Viral Load (NAAT) https://rrpl-testviewer.ehealthsask.ca/Home/Details?id=228 16. ASSOCIATED/SUPPORTING DOCUMENTS Document # Policy/Procedure/Form/Document Title SHA-02-002 Safety Reporting/Stop the Line PROV-15 Safety Practices during Pandemic MM003F V2 Doc HIV Point of Care Supply Requisition 1400-55 17. APPENDICES Appendix A: HIV POCT Quality Control Log Appendix B: HIV POCT Incident Log Appendix C: HIV POCT Result Log Appendix D: HIV Point of Care Supply Requisition Appendix E: Laboratory Test or Instrument General Training Checklist Appendix F: Repealed Policies, Procedures, Forms and Other Related Documents SHA accepts no responsibility for use of this material by any person or organization not associated with SHA. No part of this document may be reproduced in any form for publication without permission of SHA. Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 12 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
APPENDIX A: HIV POCT QUALITY CONTROL LOG Quality Control Log – INSTI HIV-1 HIV-2 Antibody Test Kit Location _______________ Site_______________ Month_____________ Year______________ Date QC Material QC Lot # Test Kit Lot # Results Tester ID OK Observed or Supervisor Review run controls Page 13 of 19 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: APPENDIX B: HIV POCT INCIDENT LOG Incident Log - INSTI HIV-1 HIV-2 Antibody Test Kit Location _______________ Site_______________ Month_____________ Year______________ Date Type of Incident (QC failure, device Description Action Taken Date Resolved failure, EQA discrepancy, parallel testing Discrepancy Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 14 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: APPENDIX C: HIV POCT RESULT LOG Result Log - INSTI HIV-1 HIV-2 Antibody Test Kit Date test Lot # Test kit Patient ID Tester ID ID of staff Test results performed expiry date verifying results (if available) Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 15 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
APPENDIX D: HIV POINT OF CARE SUPPLY REQUISTION Page 16 of 19 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: APPENDIX E: LABORATORY TEST OR INSTRUMENT GENERAL TRAINING CHECKLIST Name: __________________________________________ Department: _____________________________________ Location: ________________________________________ Test or Instrument: ________________________________ (Circle any areas that you feel you did not completely understand or were not shown. Mark N/A if not applicable) (Please return completed form to your supervisor for filing) Documentation Trainee Trainer Additional Comments Date/Initial Date/Initial Trainers Date/Initial a. Read procedure b. Read package insert c. Watch training video d. Reviewed SDS e. Review Job Safety Analysis Performing Test a. Required supplies, reagents, labels and forms b. Reagent preparation or kit information c. Watch trainer perform test or operate instrument d. Trainee to perform test or operate instrument under direct observation e. Run parallel runs (trainee and trainer both run test and compare results) – document runs f. Quality Control parameters and procedure if out of range g. Calibrations h. Result entry and authorizing i. Any strict time parameters for test j. Abnormal result procedure or considerations k. Maintenance requirements l. Proficiency Panels m. Troubleshooting Safety Considerations a. PPE requirements b. Instrument hazards c. Sample hazards d. Other hazard or safety equipment (BSC) e. Reagent or sample disposal Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 17 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: Sample Processing Trainee Trainer Additional Comments Date/Initial Date/Initial Trainers Date/Initial a. Acceptable samples for testing (urine, serum or plasma) b. Proper preparation prior to testing & problems resulting from improper preparation c. Sample volume required d. Sample stability e. Sample storage requirements General Information and Prioritization a. Cleaning and restocking b. Kit or reagent re-order points c. Procedure for mislabeled or unlabeled specimens d. Review handling patient demographic discrepancies e. Priority of samples f. Expected turn-around times g. Workflow for instrument or test Additional a. Extra coaching required or training issues - document b. Competency assessment in 6 months Test or Instrument Specific Training Items a. b. c. d. e. Comments or Concerns a. b. c. d. e. Competency has been met Yes No (if no, state remedial plan to meet standards) Date Approving Supervisor/Designate Director/Manager/Designate (when required) Remedial Plan to Meet Standards (if competency has not been met) Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 18 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories (PROV-120, Version 1) Date Approved: July 8, 2021 Date Revised: Appendix F: Repealed Policies, Procedures, Forms and Other Related Documents This policy replaces or partially replaces the following former regional health authority policies, procedures, forms or other related documents including but not limited to: Policy Name (Policy Number) Full Repeals Name of former Policy/Procedure/Form/Document Title Document # health region Mamawetan HIV Quality Control Log Gen.3.1.A1 Mamawetan HIV Patient Result Log Gen.3.1.A3 Prince Albert HIV Screening VH-CH-145 Prince Albert HIV Screening VH-CH-10-52 Prairie North Policy and Procedure for Point of Care Testing PHU 002-2019 Prairie North HIV POC Quality Control Log INSTI HIV-1 Antibody Test Kit RLS.POC.19 Prairie North HIV POC Incident Log INSTI HIV-1 Antibody Test Kit RLS.POC.19 Regina RRPL INSTI HIV-1-HIV-2 Antibody Test Kit BioLytical Laboratories Doc 25000-00 IMM 900 V5 Policy Name (Policy Number) Partial Repeals Name of former Policy/Procedure/Form/Document Title Document # health region Mamawetan HIV POCT (Point of Care Testing) Gen.3.1 Prairie North INSTI HIV-1 Antibody Test Kit – BioLytical Laboratories RLS.POC.19 Regina HIV Point of Care (POC) Testing and Test Kits, Guidelines for Use Saskatoon HIV POCT CHEM-222 v1 Procedure: INSTI HIV-1 HIV-2 Antibody Test Kit – BioLytical Laboratories Page 19 of 19 Number/Version: PROV-120 V#: 1 Any PRINTED version of this document is only accurate up to the date of printing July 9, 2021; see Laboratory Quality Staff for current version.
You can also read