LIGHTCYCLER MRSA ADVANCED TEST FOR USE WITH THE LIGHTCYCLER 2.0 INSTRUMENT

Page created by Danny Nichols
 
CONTINUE READING
LightCycler® MRSA Advanced Test
       For Use With The LightCycler® 2.0 Instrument
                                    FOR IN VITRO DIAGNOSTIC USE.

           ®
LightCycler MRSA Advanced Test                     LC MRSA               96 Tests         P/N: 05352894 190

LightCycler® Advanced Lysis Kit                                          96 Tests         P/N: 05205743 190
                                                   LYS ADV

INTENDED USE

The LightCycler® MRSA Advanced Test is a qualitative in vitro diagnostic test for the direct detection of
nasal colonization with methicillin-resistant Staphylococcus aureus (MRSA) to aid in the prevention and
                                                                                                    ®
control of MRSA infections in healthcare settings. The test is performed on the LightCycler 2.0
Instrument with nasal swab specimens from patients suspected of colonization, uses swab extraction and
mechanical lysis for specimen preparation followed by polymerase chain reaction (PCR) for the
amplification of MRSA DNA, and fluorogenic target specific hybridization probes for the detection of the
amplified DNA.

The LightCycler® MRSA Advanced Test is not intended to diagnose, guide or monitor treatment for MRSA
infections. Concomitant cultures are necessary to recover organisms for epidemiology typing or for further
susceptibility testing.

SUMMARY AND EXPLANATION OF THE TEST

Staphylococcus aureus is an opportunistic pathogen carried as a commensal organism on the skin and
nares of approximately 30% of the normal population, with the potential to cause a broad spectrum of
diseases, fatal within 30 days of infection in approximately 34% of cases1,2. S. aureus can rapidly adapt to
the selective pressure from antibiotic treatments, which has resulted in the emergence and spread of
methicillin-resistant S. aureus (MRSA) strains. The mecA gene is located on a large mobile genetic
element called the Staphylococcal Cassette Chromosome mec (SCCmec), and is responsible for resistance
to methicillin and to other β-lactam antibiotics. The mecA gene encodes for an altered penicillin binding
protein, PBP2a or PBP2’, which has a very low affinity for all β-lactam antibiotics allowing normal synthesis
of the peptidoglycan layer to occur, even in the presence of these antibiotics3. Numerous MRSA strains
have emerged and spread worldwide, and SCCmec has been acquired by different lineages of methicillin-
sensitive S. aureus4. To date, five SCCmec types (I–V) have been distinguished, and several variants of
these SCCmec types have been described5.

Healthcare associated infections (HAIs) caused by MRSA have recently become an important issue for
healthcare facilities worldwide due to high rates of infection, mortality, and high costs of treatment6-9. In
addition, community-associated MRSA (CA-MRSA) has spread in the past few years10, feeding the
pipeline of infection in hospitals,11, 12 and underscoring the need for comprehensive infection control
programs. In 2008, recommendations from published guidelines of several governmental, professional, and
public health organizations were summarized to help institutions prioritize and implement their MRSA
transmission prevention efforts13. Any program to prevent MRSA transmission must comprise a number of
key strategies, including active surveillance screening, hand hygiene, patient isolation, and other
transmission prevention activities. Real-time PCR detection of MRSA significantly reduces time to

The Document Revision Information section is located at the end of this document.
06369120001-01EN                                    1                                          Doc Rev. 1.0
detection (under 2 hours of laboratory time) compared with standard culture tests and, therefore,
improves the efficiency of control and management strategies.

PRINCIPLES OF THE PROCEDURE
Detection of MRSA with the LightCycler® MRSA Advanced Test relies on 3 major processes:
       •        Specimen preparation by mechanical lysis of the bacterial cell walls
       •        PCR amplification of target DNA and detection by specific hybridization probes
       •        Automated result generation after melting peak analysis
Specimen Preparation
                                                                                           ®
The mechanical lysis of the nasal swab specimens is performed by using the LightCycler Advanced Lysis
Kit and the MagNA Lyser Instrument. Swab heads are cut into lysis tubes and subjected to heating to
inactivate all bacteria. Subsequently lysis tubes are transferred in the MagNA Lyser Instrument where
mechanical lysis of bacterial cell walls occurs, resulting in crude lysate preparations. After a brief
centrifugation step to spin down glass beads and swab fibers, the processed specimen are subjected to
PCR analysis using the LightCycler® MRSA Advanced Test.
PCR Amplification
Target Selection
The primers and probes in the LightCycler® MRSA Advanced Test detect a proprietary sequence indicative
of the integration of the SCCmec cassette into the S. aureus chromosome, indicating the presence of
MRSA DNA.
Amplification
The processed samples and the amplification mixture containing hot-start Taq polymerase are placed in
LightCycler® Capillaries (20μL) in which PCR amplification will occur. Each LightCycler® MRSA Advanced
Test reaction contains an internal control, which is designed to control for specimen inhibition, and to
                                                                                                       ®
monitor reagent integrity. The AmpErase (uracil-N-glycosylase) enzyme included in the LightCycler
MRSA Advanced Test recognizes and catalyzes the destruction of DNA strands containing deoxyuridine,
but not DNA containing deoxythymidine. Since amplicons produced with the LightCycler® MRSA
Advanced Test contain deoxyuridine, potential amplicon contaminants are eliminated during a prolonged
heating step performed prior to the start of PCR amplification.
A target sequence in a plasmid is simultaneously amplified in the Positive Control. The Positive Control is
intended to monitor for reagent failure and is included into each run. Each run also includes a Negative
Control used to detect reagent or environmental contamination by MRSA DNA.
Specific detection of PCR products by Hybridization Probes
MRSA and Internal Control amplicons are detected by fluorescence using a specific pair of hybridization
probes. The probes attach to a specific internal sequence in the amplified fragment and are positioned in
a closed proximity to one another. Upon excitation, these bound probes emit a fluorescence signal of a
specific wavelength using a process called Fluorescence Resonance Energy Transfer (FRET). The emitted
light is measured by the LightCycler® 2.0 Instrument. MRSA or internal control specific amplicons are
detected in parallel in two different detection channels and thus can be differentiated.
After completion of the real-time PCR process, a melting peak analysis is performed automatically by the
            ®
LightCycler 2.0 Instrument. Single stranded DNA amplicons with bound hybridization probes are
subjected to increasing temperatures. When the PCR products reach a specific temperature one of the
two bound hybridization probes melts off, resulting in a loss of fluorescence signal. The decrease in the
fluorescence signal occurs at a specific temperature and results in melting peaks which are used to
identify and distinguish MRSA- and IC-specific amplicons.
Automated result generation after melting peak analysis
After a visual identification of the melting peaks is performed using TM Bars in the LightCycler® Software,
test results are transferred to a dedicated interpretation tool (the Micro Analysis Software) and a report is
generated.

06369120001-01EN                                       2                                       Doc Rev. 1.0
REAGENTS
LightCycler® Advanced Lysis Kit                         LYS ADV         96 Tests
(P/N: 05205743 190)

 LYS                                                                  96 x 0.6 mL
 (Lysis Reagent)
    Tris buffer
    EDTA
    < 15% Silica beads
    0.09% Sodium azide

LightCycler® MRSA Advanced Test                         LC MRSA         96 Tests
(P/N: 05352894 190)

 MRSA RXNMX                                                          3 x 0.352 mL
 (MRSA Reaction Mix)
   Tris buffer
WARNINGS AND PRECAUTIONS

A.    FOR IN VITRO DIAGNOSTIC USE. Use of this product should be limited to personnel trained in
      the techniques of PCR.

B.    This test is for use with nasal swab specimens only.

C.    Do not pipette by mouth.

D.    Do not eat, drink or smoke in laboratory work areas. Wear protective disposable gloves, laboratory
      coats and eye protection when handling specimens and kit reagents. Wash hands thoroughly after
      handling specimens and test reagents.

E.    Avoid microbial contamination of reagents when removing aliquots from reagent bottles. The use of
      sterile disposable pipette tips is recommended.

F.    Do not pool reagents from different lots or from different bottles of the same lot.

G.    Do not mix reagents from different kit lots.

H.    Dispose unused reagents, waste and specimens in accordance with country, federal, state and local
      regulations.

I.    Do not use a kit after its expiration date.

J.    Material Safety Data Sheets (MSDS) are available on request from your local Roche office.

K.    Specimens and controls should be handled as if infectious using safe laboratory procedures such
      as those outlined in Biosafety in Microbiological and Biomedical Laboratories14 and in the CLSI
      Document M29-A315. Thoroughly clean and disinfect all work surfaces with a freshly prepared
      solution of 0.5% sodium hypochlorite in deionized or distilled water.

Note: Commercial liquid household bleach typically contains sodium hypochlorite at a
      concentration of 5.25%. A 1:10 dilution of household bleach will produce a 0.5% sodium
      hypochlorite solution.

L.    LYS, MRSA RXNMX, MRSA DETMX and MRSA (+) C contain sodium azide. Sodium azide may
      react with lead and copper plumbing to form highly explosive metal azides. While disposing of
      sodium azide-containing solutions down laboratory sinks, flush the drains with a large volume of
      water to prevent azide buildup.

M.    Wear eye protection, laboratory coats and disposable gloves when handling any reagent. Avoid
      contact of these materials with the skin, eyes or mucous membranes. If contact does occur,
      immediately wash with large amounts of water. Burns can occur if left untreated. If spills of these
      reagents occur, dilute with water before wiping.

N.    Contamination from positive controls and positive specimens can be avoided by good laboratory
      practices and careful adherence to the procedures specified in this package insert.

STORAGE AND HANDLING REQUIREMENTS
                            ®
A.    Store the LightCycler Advanced Lysis Kit (LYS) upright at 15 to 25ºC upon arrival.

B.    Store the LightCycler® MRSA Advanced Test Kit (MRSA RXNMX, MRSA DETMX and
      MRSA (+) C) at -15 to -25ºC upon arrival. Unopened, these reagents are stable until the expiration
      date indicated. Once opened, store the remaining MRSA RXNMX and MRSA DETMX at -15 to
      -25ºC for up to 30 days. MRSA RXNMX and MRSA DETMX can sustain up to 5 freeze-thaw
      cycles.

C.    Once opened, store the remaining MRSA (+) C at -15 to -25ºC or 2 to 8ºC for up to 30 days. The
      MRSA (+) C can sustain up to 22 freeze-thaw cycles.

D.    Protect the MRSA DETMX from light exposure.

E.    Working Master Mix (prepared by addition of MRSA RXNMX and MRSA DETMX) must be
      prepared fresh and utilized within 24 hours of preparation. If the Working Master Mix is not used

06369120001-01EN                                     4                                      Doc Rev. 1.0
immediately following preparation, the Working Master Mix must be stored away from light at 2 to
      8°C for up to 24 hours.

F.    Processed specimens and negative controls prepared from Liquid Stuart swabs are stable for up to
      1 day at 15 to 25ºC or 2 to 8°C or up to 8 months at -15 to -25°C. Processed specimens and
      negative controls prepared from Amies gel swabs (with or without charcoal) are stable for up to 5
      days at 15 to 25ºC or 2 to 8°C or up to 3 months at -15 to -25°C.

G.    Amplification should be started immediately after addition of the processed specimens and controls
      to the Working Master Mix. If this is not possible, store the filled capillaries away from light for a
      maximum 3 hours at 2 to 8°C.

MATERIALS PROVIDED
                   ®
A.    LightCycler Advanced Lysis Kit                                     LYS ADV
      (P/N: 05205743 190)

       LYS
       (Lysis Reagent)
                   ®
B.    LightCycler MRSA Advanced Test                                     LC MRSA
      (P/N: 05352894 190)

      MRSA RXNMX
      (MRSA Reaction Mix)

      MRSA DETMX
      (MRSA Detection Mix)

      MRSA (+) C
      (MRSA Positive Control)

MATERIALS REQUIRED BUT NOT PROVIDED

Instrumentation and Software
        •                    ®
                LightCycler 2.0 Instrument (P/N: 03 531 414 201)
                Note: For the USA, different catalog numbers do apply. Please contact the US sales
                office for details. (Serial numbers 1415001 and higher are approved for IVD use and use
                with this kit.)

         •       LightCycler® Software 4.1 (P/N: 04 898 915 001) or higher

         •                   ®
                 LightCycler MRSA Advanced Software Set v1.0 CE (P/N: 05 414 326 190) or higher

         •       LC Carousel Centrifuge 2.0 [P/N: 03 709 582 001 (230 V) or P/N: 03 709 507 001 (110V)]
                 Note: If you already have purchased a LC Carousel Centrifuge [P/N 12 189 682 001 (230
                 V) or P/N 03 030 512 001 (115V)], a LightCycler® Carousel 2.0 rotor set [P/N 03 724 697
                 001] is also required

         •       Or standard bench top microcentrifuge containing a rotor for 2.0 mL tubes, and LC
                 centrifuge adapters [P/N: 11 909 312 001]

         •       MagNA Lyser Instrument [P/N: 03 358 976 001 (230 V) or 03 358 968 001 (110 V)]

         •       Optional barcode reader [P/N: 05 536 421 001]

Reagents and Disposables

         •       LightCycler® Multicolor Compensation Set (P/N: 04 484 355 001)

         •                   ®
                 LightCycler Capillaries, 20 μL (P/N: 04 929 292 001)

06369120001-01EN                                    5                                      Doc Rev. 1.0
OTHER MATERIALS REQUIRED BUT NOT PROVIDED

         •        BD CultureSwab™ Liquid Stuart (BD P/N: 220099 or 220109), or Copan Venturi
                  Transystem™ Liquid Stuart (Copan Italia International P/N: 141C or 139C);

         •        Or BD CultureSwab™ plus Amies gel without charcoal (BD P/N: 220116 or 220117), or
                  Copan Venturi Transystem™ plus Amies gel without charcoal (Copan Italia International
                  P/N: 108C or 134C);

         •        Or BD CultureSwab™ plus Amies gel with charcoal (BD P/N: 220121 or 220122), or
                  Copan Venturi Transystem™ plus Amies gel with charcoal (Copan Italia International
                  P/N: 114C or 136C)

         •        Disposable gloves, powderless

         •        Gauze

         •        Ethanol (70 %)

         •        Adjustable Pipettors*: ( capacities: 10 μL, 100 μL and 1000 μL) with sterile, DNase-free,
                  filter-blocked or positive displacement micropipettor tips

         •        Dry heating block set at 95 ºC (± 2 ºC): VWR (P/N: 12621-088) or equivalent with 2.0mL
                  tube block : VWR (P/N: 12985-050)

         •        Microcentrifuge capable of delivering 8,000 - 20,000 x g centrifugal force with rotor
                  appropriate for 2.0 mL screw cap tubes: (P/N: Eppendorf model 5417C) or equivalent

         •        Sterile 1.5 mL Siliconized polypropylene microcentrifuge tubes: (P/N: Eppendorf order
                  Nr. 0030 108.051) or equivalent

         •        Vortex mixer: Fisher (P/N: 12-812) or equivalent

         •        Optional microtube tool: VWR (P/N: 20170-688) or Rainin (P/N: PJ-4A) or equivalent

* Pipettors should be accurate within 3% of stated volume. Aerosol barrier or positive displacement
  DNase-free tips must be used where specified to prevent specimen and amplicon cross-contamination.

SPECIMEN COLLECTION, TRANSPORT AND STORAGE

Note: Handle all specimens and controls as if they are capable of transmitting infectious agents.

A.    Specimen Collection

1.    Moisten the swab with two drops (approximately 50 μL) of sterile saline or use dry.

2.    Gently insert the swab inside the nostril approximately 0.5 inch or 1.3 cm (both swab heads at once
      in the case of a double headed swab) and rotate against the mucosa five times while applying light
      pressure on the outside of the nose (to help insure contact of the swab heads with the inside of the
      nose).

3.    Insert the same swab into the other nostril and repeat the procedure.

4.    Return the swab to its container and label appropriately.

B.    Specimen Transport and Storage

Transportation of specimens must comply with country, federal, state and local regulations for the
transport of etiologic agents16. Specimens can be transported at 15 to 25 ºC.

Specimens may be stored before processing (including time needed for transport) as defined below.

06369120001-01EN                                    6                                       Doc Rev. 1.0
•      Specimen taken with Liquid Stuart swabs is stable

       –       at 15 to 25 ºC for up to 4 days,

       –       at 2 to 8 ºC for up to 5 days,

       –       at -15 to -25°C for up to 1 month.

•      Specimen taken with Amies gel swabs with or without charcoal is stable

       –       at 15 to 25°C for up to 4 days.

INSTRUCTIONS FOR USE
Instrument Preparation

A.     Installation of the Micro Analysis Software (MAS)

For installation of updates, check whether additional information is provided with the product.
                                                     ®
1.     Login as an Administrator on the LightCycler Control Unit
                               ®
2.     Insert the LightCycler MRSA Advanced Software Set v1.0 CE CD (or higher) into the CD-drive of
       the LightCycler® Control Unit.

Note   If the program setup does not start automatically, start the installation of MAS manually by
       double clicking on the setup.exe file provided on the LightCycler® MRSA Advanced
       Software Set v1.0 CE CD (or higher).

3.     For first time installation:

       a.      if the Microsoft Data Access Components 2.8 is not already installed, you will be prompted
               to install it at the beginning of the MAS setup. The Microsoft Data Access Components 2.8
               must be installed, otherwise MAS will not run properly.

       b.      if the Microsoft .NET Framework 2.0 (x86) is not already installed, you will be prompted to
               install it at the beginning of the MAS setup. The Microsoft .NET Framework 2.0 (x86) must
               be installed, otherwise MAS will not run properly.

       c.      if the Microsoft Windows Installer 3.1 is not already installed, you will be prompted to install
               it at the beginning of the MAS setup. The Microsoft Windows Installer 3.1 must be installed,
               otherwise MAS will not run properly.

       d.      if the Microsoft Visual Studio 2008 Report Viewer is not already installed, you will be
               prompted to install it at the beginning of the MAS setup. The Microsoft Visual Studio 2008
               Report Viewer must be installed, otherwise MAS will not run properly.

       e.      if the Microsoft SQL Server 2005 Express SP2 (x86) is not already installed, you will be
               prompted to install it at the beginning of the MAS setup. The Microsoft SQL Server 2005
               Express SP2 (x86) must be installed, otherwise MAS will not run properly.

4.     If you are requested to reboot, confirm with .
                                            ®
Note: Do not remove the LightCycler MRSA Advanced Software Set CD during the rebooting
      process.

5.     After rebooting, follow the instructions of the installation wizard. Press  to start installation
       of MAS.

6.     Select  if only one user will work on the LightCycler® Control Unit, select 
       if more than one user will work on the LightCycler® Control Unit.

7.     Acknowledge that the MAS installation was successful by clicking on  in the
        window.

8.     Launch the application using the MAS icon on the computer’s desktop.

06369120001-01EN                                       7                                      Doc Rev. 1.0
9.   The data folder for *.ixo and *.pdf files  is automatically created upon start of
     the MAS.

B.   Verifying the installation of the Micro Analysis Software (MAS)
                                      ®                                                        ®
1.   Copy the proof files (LightCycler MRSA Advanced Proof File v1.0 CE.ixo and LightCycler MRSA
     Advanced Proof File v1.0 CE.pdf, or higher) from the LightCycler® MRSA Advanced Software Set
     v1.0 CE CD (or higher) into the  folder.

2.   If the MAS is not already open, double click on the MAS icon on desktop, and press .

3.   Select the file MRSA Advanced Proof File v1.0 CE.ixo (or higher) in the  folder.
                                                            ®
4.   Click on , indicating that the LightCycler             report does not contain the flag .

5.    Enter Lot No. Expiration Date and Lab Info into the appropriate fields of the MRSA Configuration
      Settings window.

                           Advanced Lysis Kit
                                   Lot No               11A00000
                              Expiration Date           12201812

                             MRSA Advanced
                                 Test
                                   Lot No               23A00000
                              Expiration Date           24201812

                            Personalized Info
                                  Lab Info              Roche Lab

6.    Select the report type  and proceed with . The file will be analyzed
      automatically and the data will be displayed.

7.    Print the report.

8.    Navigate to the  folder, open and print the MRSA Advanced Proof File v1.0
      CE.pdf (or higher).

9.    Insure that the printed MAS report and the printed PDF provide the same results. If not, please
      contact your local Roche office for technical assistance.

C.   Removal of the Micro Analysis Software (MAS)

1.   Login as an administrator on the LightCycler® Control Unit.

                                                        ®
2.   Click on  in the taskbar of the LightCycler Control Unit.

3.   Choose  /  /  in the bottom left corner of
     Windows desktop.

4.   A list of all currently installed programs will appear, click on .

5.   Choose  and confirm by clicking on .

06369120001-01EN                                   8                                    Doc Rev. 1.0
Note: The MRSA Advanced Software Set CD provides an uninstall and a repair feature. The
      uninstall feature can be used optional to the described process. The repair feature can be
      used to recover corrupted or missing files of the Micro Analysis Software.

D.     Installation of the LightCycler® MRSA Advanced Macros

1.     Install the *.lckit macros LightCycler® MRSA Advanced MCC Macro v1.0 (MRSA
                                                                 ®
       MCC_04484355001A_1.0.Ickit) or higher and LightCycler MRSA Advanced Macro v1.0 CE
       (MRSA_05352894190_1.0. Ickit) or higher from the LightCycler® MRSA Advanced Software Set v1.0
       CE CD (or higher).

2.     Please refer to the LightCycler® 2.0 Instrument Operator’s Manual A, “Executing a Roche Macro”
       chapter.
                                                                        ®
E.     Multicolor Compensation Experiment using the LightCycler Multicolor Compensation Set
                                   ®
       For Use with the LightCycler 2.0 Instrument

Use the following steps to create a new or to replace an expired Color Compensation file.

1.     Thaw vials 1,2, 3 and 5; vortex gently and spin down briefly.

2.     Prepare a 3 fold dilution of vial 2 by mixing together:

              a.      10 μL of solution contained in vial 2

              b.      20 μL of buffer from a tube of LYS without transferring glass beads

3.     Pipette 10 μL of each component into a separate LightCycler® capillary (20 μL).

              a.      Capillary #1: Vial 1

              b.      Capillary #2: dilution step E2

              c.      Capillary #3: Vial 3

              d.      Capillary #4: Vial 5

Note: Discard dilution of vial 2 after run is completed.

4.     Pipette 10 μL of buffer from a tube of LYS (without transferring glass beads) into each capillary to
       reach a total volume of 20 μL.

5.     Seal each capillary with a stopper using the Capping Tool.

6.     Centrifuge the capillaries using the LC Carousel Centrifuge 2.0 or a standard centrifuge and
       centrifuge adapters. The tubes must be in the following positions in the Carousel:

              a.      Carousel position 1: CCB (vial 1)

              b.      Carousel position 2: CC530 (vial 2)

              c.      Carousel position 3: CC610 (vial 3)

              d.      Carousel position 4: CC670 (vial 5)

Note: Do not reduce the volume used for each capillary.
                                  ®                                         ®
7.     Transfer the LightCycler Sample Carousel to the LightCycler 2.0 Instrument. If a regular
       centrifuge was used, transfer the capillaries into the LightCycler® Sample Carousel in the order
       described in step E6.

Note: Do not change the order of the reagents in the Carousel.

8.     Start and save the Color Compensation Experiment (CCE)

06369120001-01EN                                       9                                    Doc Rev. 1.0
®
             a.     For use of the Light Cycler MRSA Advanced MCC Macro please refer to the
                               ®
                    LightCycler 2.0 Instrument Operator’s Manual A, “Executing a Roche Macro”
                    chapter.

Note: Do not use the field . If you want to add the assay lot number to the
      experiment, insert information into the appropriate field in the sample editor.

             b.     Save the run using a short unique name (e.g. MRSA_CCE_YYMMDD-xx, Y=year,
                    M=month, D=day, xx=run number).

             c.     After the run is finished, a report is automatically created.

             d.     Print the report.

9.    Run Validation

      The report will not indicate, if the Color Compensation Experiment was successful. The success of
      the Color Compensation Experiment is only established upon applying the Color Compensation
      Object generated from the CCE to the initial MRSA Advanced run.
                                        ®                                           ®
             a.     If the LightCycler Software (LCS) report for the LightCycler Color Compensation
                    Experiment is flagged (e.g., ), then the
                    CCE is invalid. Repeat the Color Compensation Experiment (from step E1).

             b.     If the report of the LightCycler® Color Compensation Experiment does not contain a
                    flag, proceed to the next step.

10.   Create a Color Compensation Object (CCO)

             a.     Activate Color Compensation analysis module, and press button .

             b.     Create a short unique name using the mandatory naming convention for the Color
                    Compensation Object: MRSA_CCO_YYMMDD-xx (Y=year, M=month, D=day,
                    xx=run number).
                                                                                                ®
             c.     This object is used for further analysis of assays with the LightCycler         MRSA
                    Advanced Test.

Specimen and Control Preparation

A.    Specimen Preparation

1.    Label one LYS tube appropriately for each patient specimen.

Note: Do not mark tubes on the top of the cap.

2.    Remove the swab from the transport container.

3.    Place a single swab head in a LYS tube.

Note: If a single headed swab is used, a direct culture may be performed by streaking the swab
onto the plate before placing it into the LYS tube.

4.    Hold the swab by the stem near the rim of the tube, lift the swab a few millimeters from the bottom
      and bend the stem against the rim of the tube to break it.

Note: Use a gauze to minimize risk of contamination when breaking the swab. Use a new gauze
      for each swab.

5.    Close the LYS cap tightly.

06369120001-01EN                                       10                                Doc Rev. 1.0
B.    Negative Control Preparation

Note: The NC should be prepared at the same time as the samples to be tested.

1.    Label one LYS tube appropriately for negative control.

Note: Do not mark tubes on the top of the cap.

2.    Place a sterile swab into its transport container. Refer to Other Materials Required But Not Provided
      for a list of recommended swabs.

3.    Incubate at room temperature for at least 1 minute.

4.    Remove the swab from the transport container.

5.    Place a single swab head in a LYS tube.

6.    Hold the swab by the stem near the rim of the tube, lift the swab a few millimeters from the bottom
      and bend the stem against the rim of the tube to break it.

Note: Use a new gauze to minimize risk of contamination when breaking the swab.

7.    Close the LYS cap tightly.

C.    Specimen and Control Processing

1.    Heat specimens and negative control at 95 ± 2 ºC for 2 minutes in a dry heating block.

2.    Lyse specimen and negative control in the MagNA Lyser instrument for 70 seconds at a speed of
      5,000 rpm. Refer to the MagNA Lyser Operator’s Manual for details on how to use.

3.    Centrifuge specimen and negative control at 8,000 – 20,000 x g at room temperature for 1 minute.

Note: Processed specimens and negative controls prepared with Liquid Stuart swabs are stable
      for up to 1 day at 15 to 25ºC or 2 to 8°C or up to 8 months at -15 to -25°C. Processed
      specimens and negative controls prepared with Amies gel swabs (with or without charcoal)
      are stable for up to 5 days at 15 to 25ºC or 2 to 8°C or up to 3 months at -15 to -25°C.

Reagent Preparation

1.    Thaw MRSA RXNMX and MRSA DETMX and MRSA (+) C at room temperature. Ensure the
      reagents are completely thawed before use.

2.    Vortex MRSA RXNMX and MRSA DETMX and MRSA (+) C for 3-5 seconds and centrifuge
      briefly to collect material at the base of the tube.

3.    Prepare the Working Master Mix for the number of specimens and controls being tested into a
      sterile siliconized polypropylene microcentrifuge tube. (Use Table 1 as a guide). Each run must
      contain a minimum of one specimen, one PC, and one NC.

Note: The recommended volumes include 2 reactions for the negative control and the positive
      control and also an additional volume to account for dead volume and facilitate pipetting
      out of the mixing tube and into LC capillaries.

Note: Use Microtube Tool (VWR P/N 20170-688 or Rainin P/N PJ-4A) or equivalent to aid in
      opening MRSA RXNMX and MRSA DETMX if necessary.

06369120001-01EN                                   11                                      Doc Rev. 1.0
Table 1
                                      Preparation of Working Master Mix

                                        Volume of               Volume of
          Number of Samples          MRSA RXNMX            MRSA DETMX       Total Volume (μL)*
                                           (μL)                   (μL)

                    1                       40                     20              60

                    2                       50                     25              75

                    3                       60                     30              90

                    4                       70                     35              105

                    5                       80                     40              120

                    6                       90                     45              135

                    7                       100                    50              150

                    8                       110                    55              165

                    9                       120                    60              180

                    10                      130                    65              195

                    11                      150                    75              225

                    12                      160                    80              240

                    13                      170                    85              255

                    14                      180                    90              270

                    15                      190                    95              285

                    16                      200                    100             300

                    17                      210                    105             315

                    18                      220                    110             330

                    19                      230                    115             345

                    20                      240                    120             360

                    21                      260                    130             390

                    22                      270                    135             405

                    23                      280                    140             420

                    24                      290                    145             435

                    25                      300                    150             450

                    26                      310                    155             465

                    27                      320                    160             480

                    28                      330                    165             495

                    29                      340                    170             510

                    30                      350                    175             525
        *Includes volume for NC, MRSA (+) C, and dead volume.

06369120001-01EN                                      12                                   Doc Rev. 1.0
4.    Vortex Working Master Mix for 3-5 seconds and centrifuge briefly to collect material at the base of
      the tube. Working Master Mix must be prepared fresh daily. Discard any unused Working Master
      Mix within 24 hours of preparation.

Processed Specimen and Control Preparation

1.    Place the required number of LightCycler® capillaries into the LC Carousel or in the centrifuge
      adapters. One (1) capillary for the PC (position 1), 1 capillary for the processed NC (position 2), and
      1 capillary for each processed specimen (positions 3 and higher) are needed.
                               ®
Note: Handle the LightCycler capillaries by the reservoir portion of the capillary

2.    Pipette 15 μL of Working Master Mix into each capillary reservoir.

3.    Pipette 5 μL of clear supernatant from each processed specimen into capillaries # 3 and higher.
      Change the pipette tip after each sample.

4.    Pipette 5 μL of the processed negative control into the capillary # 2.

Note: Processed specimens and negative controls prepared with Liquid Stuart swabs are stable
      for up to 1 day at 15 to 25ºC or 2 to 8°C or up to 8 months at -15 to -25°C. Processed
      specimens and negative controls prepared with Amies gel swabs (with or without charcoal)
      are stable for up to 5 days at 15 to 25ºC or 2 to 8°C or up to 3 months at -15 to -25°C. If
      frozen processed specimens are used, thaw at room temperature, vortex for 10 sec, then
      centrifuge at 8,000 – 20,000 x g at room temperature for 1 minute.

5.    Pipette 5 μL of the thawed MRSA (+) C into capillary # 1.

6.    Seal each capillary with a stopper using the Capping Tool.

7.    Centrifuge the capillaries in the LC Carousel Centrifuge 2.0 or using a standard centrifuge and
      centrifuge adapters.

8.    Transfer the LightCycler® Sample Carousel to the LightCycler® 2.0 Instrument.
                                                                                                 ®
Note: If a regular centrifuge was used, transfer the capillaries into the LightCycler                Sample
      Carousel in the order described in step 1.

Note: Amplification should be started immediately after addition of the processed specimens and
      controls to the Working Master Mix. In case this is not possible, the filled capillaries can be
      stored for up to 3 hours, protected from light, at 2 to 8°C.

LightCycler® 2.0 Instrument Loading and Operation

1.    Log into the LightCycler® Software.

2.    Start the LightCycler® MRSA Advanced Macro (refer to the LightCycler® 2.0 Instrument Operator’s
      Manual A, chapter “Executing a Roche Macro” for details).

Note: Do not use the field . MAS will request reagent information prior to
      launching the result analysis process (the assay lot number can also be inserted into the
      appropriate field in the sample editor in the LightCycler® Software).

3.    Name the run using a unique identifier.

4.    Reduce the default loading list to the number of samples to be tested. If too many sample positions
      are accidentally deleted, start the procedure from step 2.

5.    Scan specimen barcode labels or manually enter specimen names in the sample name field, or
      leave the default settings. The  field can be used to enter data if required.

Note: Do not change the default name for Positive and Negative Controls.

6.    Follow the onscreen instructions and start the LightCycler® MRSA Advanced Test run.

06369120001-01EN                                    13                                      Doc Rev. 1.0
RESULTS

Peak Identification and Automatic Result Analysis:
                                                           ®
Note: Do not start data analysis while a LightCycler run is in process.

       •      Manual TM Calling requires manual operation in two TM Calling modules automatically
              selected.

       •      The Absolute Quantification modules are not used for the analysis of this test.

       •      In the melting peak window, all curves of the selected graphs are shown.

       •      The TM bar for each analysis module is predefined in the LightCycler® MRSA Advanced
              Macro.

       •      It is possible to select or deselect a TM bar without creating a flag.
                                                                                         ®
       •      The baseline of each analysis module is predefined in the LightCycler MRSA Advanced
              Macro according to the following table and must not be deselected or changed. If
              accidentally changed, type in the correct values from Table 2.

                                                 Table 2
                                              Baseline values

                                                       CH 610       CH 670
                                   Baseline value       0.085        0.100

       •      Peaks are defined as maxima above the baseline. Shoulders are not peaks.

       •      Examples for typical positive control (Graph 1, Channel 610) and Internal Control (Graph 2,
              Channel 670) peaks and peak with shoulder (Graph 3):

       Graph 1

       Graph 2

06369120001-01EN                                      14                                     Doc Rev. 1.0
Graph 3

Adjusting the TM bars:

1.     After the run has finished, the experiment wizard shows .

2.     Move the message window to the bottom of the screen in order to see the TM and Baseline values
       (do not press ).

3.     Click on the TM Calling 610 analysis module; an overview of all positions is displayed.

4.     Reduce the size of the graph display window by sliding the left handle bar of that window to the
       right; the “Baseline”, “TM1”, and “Height1” columns should now be visible.

Note: If more than one position is selected, the settings of the first selected capillary are
      displayed.

Note: Shifting of bars affects all activated and selected capillaries.

5.     Click on the positive control (#PC#) position to select it.

6.     If the TM bar is not positioned at the peak maxima (above the baseline), move it to the peak maxima,
       focusing on the specified TM ranges (see Table 3).

7.     If there are no peaks above the baseline, leave the TM bar at the default position. Ensure that bars
       do not mark slopes of any present peaks (e.g. outside the specified TM range).

8.     Ensure that values are displayed in the “TM1” and “Height1” columns for that sample (any values).
       If no values are observed (blank space), reposition the TM bar to the peak maxima; values should
       now be displayed in the “TM1” and “Height1” columns.

                                                  Table 3
                                            Specified TM ranges

                                                   Low TM (°C)       High TM (°C)
                                  MRSA
                                                      57.00             62.00
                                 (CH 610)
                              Internal Control
                                                      57.00             62.00
                                  (CH 670)

Note: In samples without peaks (e.g. in the Negative Control) auto scaling will adjust to the
      highest background value. The predefined baseline may not be visible in this case. If you
      want to make the baseline visible select this capillary position together with the positive
      control.

Note: Marked peaks outside the specified TM range will be reported in the MAS report.

9.     Proceed with peak detection as described in steps 5 to 8 above for all remaining positions (controls
       and specimens).

10.    Click on the TM Calling 670 analysis module.

06369120001-01EN                                      15                                     Doc Rev. 1.0
11.    Proceed with peak detection as described in steps 4 to 8 above for all positions (controls and
       specimens).

Finish TM calling:

1.     Move the experiment wizard window back to the center of the screen and press .
                     ®
2.     A LightCycler report is automatically generated and must be printed.

3.     Check the LightCycler® report for flags and whether  is displayed. If flags
       or other error messages are displayed, refer to the quality control and interpretation of
       results sections.

4.     The run is automatically saved.
                                                                               ®
Note: If changes are made to any sample name fields of the LightCycler Software after saving,
      the operator must refresh the database by selecting and clicking on the “Abs Quant 610”
                                                                            ®
      and “Abs Quant 670” icons and save the file again. Reprint LightCycler report.

5.     Export the *.ixo file by clicking on  \  to the folder .

Automatic result calling using the Micro Analysis Software (MAS):

1.     Double click on the MAS icon on the computer’s desktop, and press .

2.     Select the run to be analyzed from the  folder and load *.ixo file.
                                                       ®
3.     Answer the question “Does the LightCycler report contain the flag ?” appropriately, i.e. consistent with the LightCycler® report.

4.     Scan or manually enter the Lot Number and Expiration Date information supplied with the
       Advanced Lysis Kit and LightCycler® MRSA Advanced Test.

5.     Enter personalized information (e.g. operator's name) if desired.

Note: The latest information entered in MAS will be used from this point forward.

6.     Select an appropriate report format (tabular, grouped or combined).

7.     The MRSA Advanced experiment *.ixo file will be analyzed automatically, and the data will be
       displayed in the selected report format.

8.     Print the MAS report and/or export as *.pdf.

INTERPRETATION OF RESULTS

                                                       ®
1.     The results are calculated by the LightCycler Software from measured fluorescent signals and
       embedded calculation algorithms and presented with the Micro Analysis Software (MAS).

2.     Run, controls and sample validation is automatically done by the MAS.

Note: Flags and comments may be generated for RUN STATUS, CONTROLS STATUS and SAMPLE
      STATUS. The operator must check the MAS report printouts for flags or comments in all
      fields.
                                  ®
Note: A negative LightCycler          MRSA Advanced Test result does not preclude MRSA nasal
      colonization.

06369120001-01EN                                      16                                  Doc Rev. 1.0
3.     For a valid run, the results are interpreted as follows:

         CH 610              CH 670
                                              MRSA
          Result             Result                         Comment                 Interpretation
                                              Result
         (MRSA)               (IC)
                                                                            Specimen is positive for
                            POSITIVE or
            POSITIVE                         DETECTED             -         MRSA DNA. Presumed MRSA
                             NEGATIVE
                                                                            nasal colonization.
                                                                            Invalid result. Repeat PCR
                                                                            procedure     with     thawed
                                                                IC          processed specimen or new
            NEGATIVE         NEGATIVE         INVALID
                                                           not detected     specimen. If still invalid,
                                                                            report invalid result. Contact
                                                                            local Roche office.
                                                                            MRSA DNA not detected.
            NEGATIVE          POSITIVE     NOT DETECTED           -         MRSA nasal colonization
                                                                            unlikely.

Additional comments for Positive Control, Negative Control and Specimen Result may be reported by the
MAS:

                        Comment                                           Interpretation
       Peak(s) outside Target TM range              Amplicons detected outside specified MRSA TM range
            Peak(s) outside IC TM range             Amplicons detected outside specified IC TM range

If the comment “Peak(s) outside Target TM range” is reported by MAS for a Specimen Result which is
either “Not Detected / Valid” or “Invalid / Invalid” (“MRSA Result” and “Status” columns on the MAS report,
respectively), the specimen shall be suspected of being positive for MRSA DNA. Collect another specimen
(or use leftover swab from a double-headed swab) and re-test using an alternate procedure such as
culture to determine the patient’s nasal colonization status.

Specimen Results which are “Detected / Valid” (“MRSA Result” and “Status” columns on the MAS report,
respectively), and for which the “Peak(s) outside Target TM range” is displayed by MAS shall be interpreted
as being positive for MRSA DNA (presumed MRSA nasal colonization).

The presence of these comments displayed in the Controls Status bar of the MAS report does not change
the Controls Status or the Run Status interpretation generated by MAS and displayed on the MAS report.

The following flags may be displayed by the MAS:

     Flag              Comments                Interpretation              Status            Corrective Action
                                                                                       Repeat LightCycler® run.
                                                                                       Check for correct macro in
Invalid macro          File could not                                       Run                          ®
                                          Wrong Macro applied                          the LightCycler Software
   version             be loaded                                          INVALID
                                                                                       and/or in the MRSA
                                                                                       Advanced Software Set.
                                          Critical parameters in the                   If additional flags are
                                          CCE and / or CCO and /                       displayed refer to the
    User                                  or MRSA Advanced *.ixo                       corrective action given for
Developed or                              files are different from the      Run        the respective flag.
                             -
  Modified                                reference *.ixo files of the    INVALID      If no additional flags are
Test Method                               MRSA Advanced Macros.                        displayed repeat
                                          Possible causes are                          experiment.
                                          described in the                             If this flag is repeatedly set,

06369120001-01EN                                            17                                       Doc Rev. 1.0
Flag        Comments             Interpretation         Status         Corrective Action
                              Operator's Manual of the                please call your local Roche
                                          ®
                              LightCycler 2.0                         Service.
                              Instrument (e.g. modifying
                              default settings of the
                              Macro)
  Lysis Kit                   LightCycler® Advanced          Run      Repeat LightCycler® run
                    -
  expired                     Lysis Kit expired            INVALID    with non expired reagents.
 MRSA Test                    LightCycler® MRSA              Run      Repeat LightCycler® run
                    -
  expired                     Advanced Test expired        INVALID    with non expired reagents.
                                                                      Prepare a new Color
                                                                      Compensation run and
CCO expired                   The Color Compensation                  create a new Color
                                                             Run
    on              -         file is older than six                  Compensation Object.
          1                                                INVALID                          ®
YYMMDD                        months                                  Repeat LightCycler run
                                                                      which had been performed
                                                                      with the expired CCO.
                                                                      Prepare a new Color
                                                                      Compensation run and
                                                                      create a new Color
                                                                      Compensation Object
 INVALID                      Color Compensation             Run      following the naming
                    -
CCO NAME                      Object incorrectly named     INVALID    convention stated in this
                                                                      document. Repeat
                                                                                    ®
                                                                      LightCycler run which had
                                                                      been performed with the
                                                                      incorrectly named CCO.
                                                                      Open MRSA experiment
                                                                                                   ®
                                                                      *.ixo file in the LightCycler
                              Different Color
                                                                      Software, select the correct
 MULTIPLE                     Compensation files             Run
                    -                                                 Color Compensation file.
 CCO USED                     applied in the analysis      INVALID
                                                                      Save the experiment, print
                              modules                                               ®
                                                                      LightCycler report and
                                                                      export again.
                                                                      Open MRSA experiment
                                                                      *.ixo file in the LightCycler®
                              One or more Color
                                                                      Software, select the correct
  NO CCO                      Compensation files are         Run
                    -                                                 Color Compensation file.
   USED                       missing in the analysis      INVALID
                                                                      Save the experiment, print
                              modules                                               ®
                                                                      LightCycler report and
                                                                      export again.
                              Wrong LightCycler® SW                   Open LightCycler®
INVALID LCS                                                  Run
                    -         version applied for                     Software and check for
  VERSION                                                  INVALID
                              procedure                               correct version number.
                                                                      Open MRSA experiment
                                                                      *.ixo file in the LightCycler®
                                                                      Software and correct
                              The baseline values might
  INVALID                                                    Run      baseline values according
                    -         have been changed
 BASELINE                                                  INVALID    to Table 2. Save the
                              accidentally
                                                                      experiment, print
                                                                                    ®
                                                                      LightCycler report and
                                                                      export again.
              •    NC:        •    Amplicon(s)
                   Target          detected in the         Controls
INVALID NC                                                            See section Quality Control.
                   peak(s)         MRSA TM range of        INVALID
                   detected        the Negative Control

06369120001-01EN                               18                                  Doc Rev. 1.0
Flag             Comments                  Interpretation         Status           Corrective Action
                       •    NC:              •    No IC detected in
                            IC not                Negative Control
                            detected
                       PC: PC not            No MRSA target detected     Controls
    INVALID PC                                                                        See section Quality Control.
                       detected              in the Positive Control     INVALID
                                                                                      Open MRSA experiment
                                                                                                                   ®
                                             Default name for                         *.ixo file in the LightCycler
INVALID NC                                                               Controls
                               -             Negative Control has                     Software and enter #NC#.
  NAME                                                                   INVALID                               ®
                                             been changed                             Save, print LightCycler
                                                                                      report and export again.
                                                                                      Open MRSA experiment
                                                                                                                   ®
                                                                                      *.ixo file in the LightCycler
                                             Default name for Positive
    INVALID PC                                                           Controls     Software and enter #PC#.
                               -             Control has been
      NAME                                                               INVALID      Save the experiment, print
                                             changed                                                ®
                                                                                      LightCycler report and
                                                                                      export again.
1
    The MAS provides a warning 30 days prior to reagent expiration.

QUALITY CONTROL
                               ®
1.         Each LightCycler MRSA Advanced reaction contains an Internal Control (IC). The IC is designed to
           control for specimen inhibition, and to monitor reagent integrity.

           a.      Internal Control (IC)

                   The IC must be detected positive in all MRSA negative specimens as well as in the Negative
                   Control (NC). If the IC is negative in a negative specimen, the specimen is invalid and PCR
                   procedure with freshly thawed processed specimen has to be repeated. If the IC is negative
                   in the NC, the entire run is invalid and must be repeated from frozen processed specimens
                   or new specimens. If the repeat frozen processed specimen run is still invalid, contact local
                   Roche office. The IC result is ignored in MRSA positive specimens and in a valid Positive
                   Control (PC) since MRSA amplification may compete with this control.

2.         At least 1 PC and 1 NC must be processed with each run. The PC is intended to monitor for reagent
           failure. The NC is used to detect reagent or environment contamination by MRSA DNA.
           Amplification reagents contain an enzyme to prevent contamination with MRSA amplicons. The
                       ®
           LightCycler Software automatically assigns the position of both controls in the instrument.

           a.      Negative Control (NC)

                   The NC must be negative while its internal control must be positive. If the NC does not meet
                   these criteria, the entire run is invalid and must be repeated from frozen processed
                   specimens or new specimens. If the repeat run is still invalid, contact your local Roche office.

           b.      Positive Control (PC)

                   The PC must be positive (with IC either positive or negative). If the PC does not meet this
                   criterion, the entire run is invalid and must be repeated from frozen processed specimen or
                   new specimen. If the repeat run is still invalid, contact your local Roche office.

                   Flags and comments may be generated by the Micro Analysis Software (MAS). The Operator
                   must check the run printouts for flags and comments to verify that the run is valid.

06369120001-01EN                                                 19                                Doc Rev. 1.0
CH 610           CH 670
         Control         Result          Result        Interpretation
                        (MRSA)            (IC)
                         NEGATIVE        POSITIVE      Run valid, No action
                                        POSITIVE or    Run invalid, Run new specimens
                         POSITIVE
                                         NEGATIVE
         Negative
         Control                                       Run invalid, Repeat run from frozen
                                                       processed specimens (or new specimens).
                         NEGATIVE        NEGATIVE
                                                       If the repeat frozen processed specimen run
                                                       is still invalid, contact local Roche office.
                                        POSITIVE or    Run valid, No action
                         POSITIVE
                                         NEGATIVE
         Positive                                      Run Invalid, Repeat run from frozen
         Control                        POSITIVE or    processed specimens (or new specimens).
                         NEGATIVE
                                         NEGATIVE      If the repeat frozen processed specimen run
                                                       is still invalid, contact local Roche office.

                                                                                  ®
Note: If the Negative Control (NC) or the MRSA (+) C of the LightCycler MRSA Advanced Test
      is consistently invalid, the LightCycler® Multicolor Compensation file might be incorrect.
      Repeat the Multicolor Compensation run or contact your local Roche office for technical
      assistance.

External Positive Control (EPC)
                    ®
The MicroBioLogics KWIK-STIK™ MRSA positive control (P/N: 0158ROCHE) may be used for training,
proficiency testing and external Quality Control of the LightCycler® MRSA Advanced Test. An MRSA strain
representative of RE2 types is used for this external positive control. MRSA strains representing RE3 and
RE7 types, if available may be used as additional external positive controls to monitor assay primers and
probes not directly controlled in the assay. External controls may be used in accordance with local, state,
and federal accrediting organization, as applicable. Follow the steps in the KWIK-STIK™ package insert for
swab preparation, but do not place the swab onto culture medium. After saturation of the swab with the
hydrated material, follow the LightCycler® MRSA Advanced Test package insert instructions for specimen
preparation and testing of the swab.

The EPC must be positive (with IC either positive or negative). If the EPC does not meet this criterion, the
entire MagNa Lyser run is invalid and must be repeated from new specimen. If the repeat run is still invalid,
contact your local Roche office.

Note: To date, seven SCCmec types (I–VII) have been distinguished, and several variants of these
                                           5               ®
      SCCmec types have been described . The LightCycler MRSA Advanced Test is based on a
      proprietary detection system which is able to detect MRSA strains with different molecular
      sequences in the vicinity surrounding the right extremity (RE) junction of the SCCmec
      cassette with the orfX gene. The different RE types which are detected by the LightCycler®
      MRSA Advanced Test are referred to as RE2, RE3 and RE7 (proprietary nomenclature). In
      addition, sequence analysis of type RE1 has shown 100% homology to the primers and
      probes used in the LightCycler® 2.0 MRSA Advanced Test.

PROCEDURAL PRECAUTIONS

1.     As with any test procedure, good laboratory technique is essential to the proper performance of this
       assay.

PROCEDURAL LIMITATIONS
                                                           ®
1.     This product can only be used with the LightCycler 2.0 Instrument.

2.     This test has been validated for use with only human nasal specimens.

06369120001-01EN                                      20                                    Doc Rev. 1.0
3.      Reliable results are dependent on adequate specimen collection (number of organisms in the
        sample), transport, storage and processing procedures.

4.      Use of this product should be limited to trained personnel.
                                ®
5.      A positive LightCycler MRSA Advanced Test result does not indicate the presence of viable MRSA.
        It is however, presumptive for the presence of MRSA. Therefore, a positive result does not
        necessarily indicate intervention eradication failure since non-viable DNA may persist. However, a
        negative result following a previously positive test result may or may not indicate eradication
        success.

6.      Due to inherent differences between technologies, it is recommended that, prior to switching from
        one technology to the next; users perform method correlation studies in their laboratory to qualify
        technology differences.

7.      The presence of AmpErase enzyme in the LightCycler® MRSA Advanced Reaction Mix reduces the
        risk of amplicon contamination. However, contamination from MRSA (+) C and clinical specimens
        can be avoided only by good laboratory practices and careful adherence to the procedures
        specified in this Package Insert.

8.      Though rare, mutations or polymorphisms within the region of the bacterial genome covered by the
        LightCycler® MRSA Advanced Test primers and/or probe may impair detection. Rare forms of
        MSSA strains carrying a genetic variant of SCCmec cassette may lead to positive results with the
                   ®
        LightCycler MRSA Advanced Test.

9.      As with all PCR based in vitro diagnostic tests, extremely low levels of target below the LOD of the
        assay may be detected, but results may not be reproducible (refer to “Reproducibility” section for
        further details).

INTERFERING SUBSTANCES

Substances that may interfere with the detection of MRSA by the LightCycler® MRSA Advanced Test and
potentially generate invalid results include blood, excessive amounts of nasal secretions/mucus.

The following exogenous substances (see Table 4), which are components of decongestants and
substances used to relieve nasal dryness and/or irritation, have been shown not to interfere with the
                                    ®
detection of MRSA by the LightCycler MRSA Advanced Test.

                                             Table 4
                                                                      ®
     Exogenous Substances Tested For Interference With The LightCycler MRSA Advanced Test

          Trade name                          Active ingredient
                            ®
          InfectoPyoderm                      Mupirocin 20 mg/g (antibiotic)
                 ®
          Turixin                             Mupirocin calcium 21.5 mg/g (antibiotic)
                        ®
          Bepanthen nasal ointment            Dexpanthenol 50 mg/g (provitamin B5)
          Accumed 12-hour Nasal Spray         Oxymetazoline hydrochloride 0.05% (decongestant)
                    ®
          Otriven                             Xylometazoline 0.1 % (decongestant)

1,624 specimens were tested with the LightCycler MRSA Advanced Test. Overall, 10/1,624 (0.6%)
specimens had invalid results because of internal control (IC) failure following initial testing, and 7/1,624
(0.4%) specimens remained invalid because of IC failure following retesting. Blood was reported on the
nasal swab for 20/1,624 (1.2%) specimens, none of which had invalid results because of IC failure.

06369120001-01EN                                      21                                       Doc Rev. 1.0
EXPECTED VALUES
                   ®
In the LightCycler MRSA Advanced Test clinical trial, a total of 1,406 nasal swab specimens were
collected from 1,406 subjects at 5 sites across the United States. The study population was grouped into
subjects in hospital non-intensive care units (1,011; 71.9%), hospital intensive care units (67; 4.8%) nursing
homes/extended care facilities (255; 18.1%), and medical facility staff (73; 5.2%). The overall number of
positive MRSA samples by direct culture was 183 (13.0%).

PERFORMANCE CHARACTERISTICS
CLINICAL PERFORMANCE
                                               ®
Performance characteristics of the LightCycler MRSA Advanced Test were determined in a multi-site
                                                                                ®
prospective investigational study at 5 institutions by comparing the LightCycler MRSA Advanced Test
with a second FDA-cleared nucleic acid amplification test (NAAT), direct culture, and enrichment culture
(the most sensitive culture method). Subjects included individuals and medical staff at risk for nasal
colonization. Each subject was enrolled in the study only one time. Subjects who had received systemic or
topical-nasal antibiotics used to treat nasal colonization with MRSA from the day of sample collection and
up to one week prior to study enrollment, under 2 years of age, and/or had contraindication to nasal swab
collection were excluded from the study. Only those subjects meeting the inclusion and exclusion criteria
were enrolled.

A double-headed nasal swab was collected from each subject. One swab head was directly streaked onto
a quadrant of a chromogenic agar plate with cefoxitin, and then processed according to the package insert
                                ®
for testing with the LightCycler MRSA Advanced Test. The remaining quadrants of the chromogenic agar
were streaked using a sterile loop. The second swab head was directly streaked onto a separate
chromogenic plate with cefoxitin, and then processed for testing with the second FDA-cleared NAAT test
(according to the package insert). Thereafter, the second swab head was transferred into Trypticase Soy
Broth and incubated for 48 hours at 35-37ºC and then subcultured onto a chromogenic plate with cefoxitin
(enrichment culture). Chromogenic culture plates were incubated at 35-37ºC for 20–52 hours. Presumptive
MRSA colonies from all culture plates were confirmed by coagulase testing and Gram staining if found
after 44-52 hours of incubation. Each participating site performed all tests.

Overall, 99.3% of specimens had concordant culture results for the two swab heads directly streaked onto
the chromogenic plates with cefoxitin. Only specimens with concordant results between the two swab
heads were considered in the analysis.
                                 ®
Performance of the LightCycler MRSA Advanced Test and the second FDA-cleared NAAT test were
calculated relative to the directly plated culture, and the enrichment culture results.

A summary of results for 1,389 eligible subjects (one nasal swab per patient) tested for MRSA by the
            ®
LightCycler MRSA Advanced test, a second FDA-cleared NAAT test, and directly culture is shown in
Table 5. For subjects tested, this resulted in a positive predictive value of 80.1% and a negative predictive
                                  ®
value of 99.6 %. The LightCycler MRSA Advanced Test correctly identified 173 of 178 of specimens
positive for MRSA by direct culture (positive percent agreement of 97.2%) and 1,172 of 1,215 of specimens
negative for MRSA relative to direct culture (negative percent agreement of 96.5%).

06369120001-01EN                                     22                                      Doc Rev. 1.0
Table 5
                                                         ®
                            Comparison Of The LightCycler MRSA Advanced Test,
                              The Second FDA-Cleared NAAT, And Direct Culture

                                                                                      Direct culture
                   ®                                         nd
    LightCycler MRSA Advanced Test                       2        NAAT       Positive           Negative           Total
                       Positive                           Positive               172                24              196
                       Positive                           Negative                1                 19               20
                       Negative                           Positive                1                 76               77
                       Negative                           Negative                4               1,092            1,096
                         Total                                Total              178              1,211            1,389
                                                                             ®                               nd
Note: Only evaluable specimens contributing both valid results in the LightCycler MRSA Advanced Test and the 2    NAAT
test results and concordant direct culture results are included in this summary table.

Further investigation (i.e., testing for methicillin resistance using CLSI disk diffusion methodology, fem- and
mecA-specific PCR, and sequencing of SCCmec regions) confirmed the presence of MRSA in 30 of 43
samples in which the LightCycler® MRSA Advanced Test gave a positive result but direct culture was
                                                                                               ®
negative. The presence of MRSA was confirmed in 4 of 5 samples in which the LightCycler MRSA
Advanced Test gave a negative result but direct culture was positive. The presence of MRSA was
confirmed in 13 of 76 samples in which the 2nd FDA-cleared NAAT gave a positive result but the
           ®                                                                          17
LightCycler MRSA Advanced Test and the direct culture gave a negative result .
                                        ®
Overall, 2 (1.7%) of 117 LightCycler runs performed during the investigational study were invalid because
of protocol deviations.

Performances of the LightCycler® MRSA Advanced Test, the second FDA-cleared NAAT, and the
enrichment culture relative to the direct culture are shown in Tables 6 and 7.

                                           Table 6
                                             ®
Positive Percent Agreement Of The LightCycler MRSA Advanced Test, The Second FDA-Cleared
        NAAT, and Enrichment Culture For Specimens Positive for MRSA by Direct Culture

                                     Positive Percent Agreement (95% exact CI)a
                                    ®
                        LightCycler MRSA
           Site           Advanced Test                      2nd NAAT                 Enrichment Culture
               1                96.8%                       96.8%                           93.5%
                       (30/31; 83.3% to 99.9%)     (30/31; 83.3% to 99.9%)         (29/31; 78.6% to 99.2%)
               2                97.7%                       97.7%                           97.7%
                       (42/43; 87.7% to 99.9%)     (42/43; 87.7% to 99.9%)         (42/43; 87.7% to 99.9%)
               3             100.0%                         100.0%                         100.0%
                    (14/14; 76.8% to 100.0%)       (14/14; 76.8% to 100.0%)       (14/14; 76.8% to 100.0%)
               4                96.2%                       96.2%                           97.5%
                       (76/79; 89.3% to 99.2%)     (76/79; 89.3% to 99.2%)         (77/79; 91.2% to 99.7%)
               5             100.0%                         100.0%                         100.0%
                    (11/11; 71.5% to 100.0%)       (11/11; 71.5% to 100.0%)       (11/11; 71.5% to 100.0%)
           Total              97.2%                     97.2%                     97.2%
                    (173/178; 93.6% to 99.1%) (173/178; 93.6% to 99.1%) (173/178; 93.6% to 99.1%)
           a
               CI = confidence interval.

06369120001-01EN                                              23                                          Doc Rev. 1.0
Table 7
        Negative Percent Agreement Of The LightCycler® MRSA Advanced Test, The Second FDA-
        Cleared Test, and Enrichment Culture For Specimens Negative for MRSA by Direct Culture

                                                                                a
                                  Negative Percent Agreement (95% exact CI)
                           ®
               LightCycler MRSA
Site             Advanced Test                      2nd NAAT                   Enrichment Culture
             97.5% (197/202; 94.3% to        92.0% (184/200; 87.3% to        97.5% (197/202; 94.3% to
    1
                     99.2%)                           95.4%)                          99.2%)
             96.2% (300/312; 93.4% to        82.1% (256/312; 77.3% to        96.7% (295/305; 94.1% to
    2
                     98.0%)                           86.1%)                          98.4%)
             97.6% (247/253; 94.9% to        96.8% (245/253; 93.9% to        98.8% (250/253; 96.6% to
    3
                     99.1%)                           98.6%)                          99.8%)
             93.9% (185/197; 89.6% to        93.4% (183/196; 88.9% to        98.0% (193/197; 94.9% to
    4
                     96.8%)                           96.4%)                          99.4%)
             96.8% (243/251; 93.8% to        97.2% (243/250; 94.3% to        98.8% (248/251; 96.5% to
    5
                     98.6%)                           98.9%)                          99.8%)
                       96.5%                           91.7%                           97.9%
Total
           (1,172/1,215; 95.3% to 97.4%)   (1,111/1,211; 90.0% to 93.2%)   (1,183/1,208; 97.0% to 98.7%)
a
    CI = confidence interval.

                                   ®
Performances of the LightCycler MRSA Advanced Test, the second FDA-cleared NAAT, and the direct
culture relative to the enrichment culture are shown in Tables 8 and 9.

                                           Table 8
Positive Percent Agreement Of The LightCycler® MRSA Advanced Test, The Second FDA-Cleared
        NAAT, and Direct Culture For Specimens Positive for MRSA by Enrichment Culture

                                                                               a
                                  Positive Percent Agreement (95% exact CI)
               LightCycler® MRSA
                                                        nd
Site             Advanced Test                      2        NAAT                   Direct Culture
    1      85.3% (29/34; 68.9% to 95.0%)   88.2% (30/34; 72.5% to 96.7%)   85.3% (29/34; 68.9% to 95.0%)
    2      88.5% (46/52; 76.6% to 95.6%)   86.5% (45/52; 74.2% to 94.4%)   80.8% (42/52; 67.5% to 90.4%)
    3      82.4% (14/17; 56.6% to 96.2%)   82.4% (14/17; 56.6% to 96.2%)   82.4% (14/17; 56.6% to 96.2%)
    4      97.5% (79/81; 91.4% to 99.7%)   96.3% (78/81; 89.6% to 99.2%)   95.1% (77/81; 87.8% to 98.6%)
    5      78.6% (11/14; 49.2% to 95.3%)   78.6% (11/14; 49.2% to 95.3%)   78.6% (11/14; 49.2% to 95.3%)
             90.4% (179/198; 85.4% to        89.9% (178/198; 84.8% to        87.4% (173/198; 81.9% to
Total
                     94.1%)                           93.7%)                          91.7%)
a
    CI = confidence interval.

06369120001-01EN                                     24                                    Doc Rev. 1.0
You can also read