DYEnamic ET Dye Terminator Cycle Sequencing Kit for MegaBace DNA Analysis Systems - Product Booklet GE Healthcare
←
→
Page content transcription
If your browser does not render page correctly, please read the page content below
GE Healthcare DYEnamic ET Dye Terminator Cycle Sequencing Kit for MegaBace DNA Analysis Systems Product Booklet Codes: US81090 US81095
Page finder 1. Legal 3 2. Handling 6 2.1. Safety warnings and precautions 6 2.2. Quality control 6 2.3. Storage 6 3. Components of the kit 7 4. Materials not supplied 9 5. Introduction 10 6. Important considerations for using this kit 12 7. Protocols 14 7.1. Preparation of sequencing reactions 15 7.2. Post-reaction cleanup 18 7.3. Resuspension of samples 20 7.4. Instrument setup and data analysis 21 7.5. Injection and run parameters 22 8. Appendixes 23 8.1. Appendix 1: Template DNA—general considerations 23 8.2. Appendix 2: Primers—general considerations 26 8.3. Appendix 3: Cycling conditions 27 8.4. Appendix 4: Considerations for post-reaction cleanup 29 9. Troubleshooting 31 10. References 36 2
1. Legal GE and GE monogram are trademarks of General Electric Company. AutoSeq, DYEnamic, MegaBACE, Sephadex, Sequenase, TempliPhi and Thermo Sequenase are trademarks of GE Healthcare companies. This kit is sold pursuant to Authorization from PE Applied Biosystems under one or more of the following U.S. Patents: 4,849,513; 4,855,255; 5,015,733; 5,118,800; 5,118,802; 5,161,507; 5,171,534; 5,242,796; 5,306,618; 5,332,666; and 5,366,860, and corresponding foreign patents and patent applications. The purchase of this kit includes limited non-transferable rights (without the right to resell, repackage, or further sublicense) under such patent rights to use this kit for DNA sequencing or fragment analysis, solely when used in conjunction with an automated instrument for DNA sequencing or analysis which have been authorized for such use by Applied Biosystems, or for manual sequencing. Purchase of this product does not itself convey to the purchaser a complete license or right to perform automated DNA sequence and fragment analysis under the subject patents. No other license is hereby granted for use of this kit in any other automated sequence analysis instrument. The rights granted hereunder are solely for research and other used that are not unlawful. No other license is granted expressly, impliedly, or by estoppel. Further information on purchasing licenses to perform DNA sequence and fragment analysis may be obtained by contacting the Director of Licensing at Applied Biosystems, 850 Lincoln Centre Drive, Foster City, California 94404. GE HEALTHCARE IS LICENSED AS A VENDOR FOR AUTHORIZED SEQUENCING AND FRAGMENT ANALYSIS INSTRUMENTS. 3
NOTICE TO PURCHASER ABOUT LIMITED LICENSE The purchase of this kit (reagent) includes a limited non-exclusive sublicense under certain patents* to use the kit (reagent) to perform one or more patented DNA sequencing methods in those patents solely for use with Thermo Sequenase II DNA polymerase purchased from GE Healthcare for research activities. No other license is granted expressly, impliedly or by estoppel. For information concerning avail- ability of additional licenses to practice the patented methodologies, contact GE Healthcare Bio-Sciences Corp, Director, Business Development, 800 Centennial Avenue, PO Box 1327, Piscataway, NJ 08855 USA. * US Patent numbers 4,962,020, 5,173,411, 5,409,811, 5,498,523, 5,614,365 and 5,674,716. Patents pending. † This product is sold under licensing arrangements with Roche Molecular Systems, F Hoffmann-La Roche Ltd and the Perkin- Elmer Corporation. Purchase of this product is accompanied by a limited license to use it in the Polymerase Chain Reaction (PCR) process for research in conjunction with a thermal cycler whose use in the automated performance of the PCR process is covered by the up-front license fee, either by payment to Perkin-Elmer or as purchased, i.e. an authorized thermal cycler. Energy Transfer dyes and primers—US Patent numbers: 5,654,419, 5,688,648, and 5,707,804. T7 Sequenase DNA polymerase—This reagent (kit) is covered by or suitable for use under one or more US Patent numbers: 4,795,699; 4,946,786; 4,942,130; 4,962,020; 4,994,372; 5,145,776; 5,173,411; 5,266,466, 5,409,811, 5,498,523 and 5,639,608. Patents pending in US and other countries. Thermo Sequenase II DNA polymerase—Patent pending. © 2006 General Electric Company – All rights reserved. 4
GE Healthcare reserves the right, subject to any regulatory and contractual approval, if required, to make changes in specification and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE Healthcare representative for the most current information and a copy of the terms and conditions. http//www.gehealthcare.com/lifesciences GE Healthcare UK Limited. Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA UK 5
2. Handling 2.1. Safety warnings Warning: This kit contains formamide. This protocol also and precautions requires the use of ethanol, a Warning: For research use flammable liquid. Gel reagents only. Not recommended may contain acrylamide, a or intended for diagnosis neurotoxin and suspected of disease in humans or carcinogen. Please follow animals. Do not use internally the manufacturer’s Material or externally in humans or Safety Data Sheet regarding animals. safe handling and use of these All chemicals should be materials. considered as potentially hazardous. We therefore 2.2. Quality control recommend that this product is All batches of DYEnamic ET Dye handled only by those persons Terminator Cycle Sequencing who have been trained in Kit for MegaBACE are assayed laboratory techniques and according to the recommended that it is used in accordance starting point protocol with the principles of good described in this booklet. laboratory practice. Wear The reactions are analyzed suitable protective clothing on MegaBACE sequencing such as laboratory overalls, instrument. Specifications safety glasses and gloves. for release are based on Care should be taken to avoid assessment of sequence by contact with skin or eyes. In length of read (> 500 bases), the case of contact with skin accuracy and signal quality. or eyes wash immediately with water. See material safety 2.3. Storage data sheet(s) and/or safety Store at -15°C to -30°C statement(s) for specific advice. 6
3. Components of the kit Solutions included in DYEnamic™ ET Dye Terminator Cycle Sequencing Kit for MegaBACE™ DNA Analysis Systems have been carefully formulated for optimal sequencing results. Each reagent has been tested extensively and its concentration adjusted to meet GE Healthcare standards. It is strongly recommended that reagents supplied in the kit be used as described in this protocol. The following components are included in the kit: Kit component US81090 US81095 (500 rxns) (10 000 rxns) DYEnamic ET 4 x 1 ml 1 x 80 ml terminator reagent premix (MegaBACE) Ammonium 1 x 1 ml 1 x 20 ml acetate (7.5 M ammonium acetate) Control 1 x 200 μl Not included M13mp18 DNA (single-stranded, 0.2 μg/μl) Control primer 1 × 400 μl Not included (universal cycle primer 5’-GTTTTCCCAGTCACGACGTTGTA-3’) (2.0 pmol/μl) Loading solution 1 x 10 ml 1 x 200 ml (70% formamide, 1 mM EDTA) 7
Store these kits and their components at -15°C to -30°C (NOT in a frost-free freezer). When the reagents are not in a freezer, keep them on ice prior to use. For convenience, the kit can be stored at 2–4°C for up to three months with no loss of performance; however, this should be avoided if the reagents will not be completely consumed within three months. World Wide Web address http//www.gehealthcare.com/lifesciences Visit the GE Healthcare home page for regularly updated product information. 8
4. Materials not supplied Reagents • Water—Use only deionized, distilled water for the sequencing reactions. • Sequencing primers—Use primers appropriate for the template being sequenced. For most applications, 5 pmol of primer is sufficient. • Ethanol (95–100% and 70%)—For post-reaction cleanup. Note: Do NOT use denatured alcohol. • Electrophoresis matrix for MegaBACE—Long-read Matrix (US79676) for capillary electrophoresis. This is linear polyacrylamide (LPA) matrix. Equipment • Liquid-handling supplies—Vials, pipettes, centrifuge and vacuum centrifuge. All sequencing reactions should be run in plastic microcentrifuge tubes (typically 0.5 ml) or 96-well or 384-well plates suitable for thermal cycling. • Instrument—This kit is used with MegaBACE sequencing instrument. • Thermal cycler—For thermally cycled incubations between 50°C and 95°C (1–100 cycles). 9
5. Introduction DYEnamic ET Dye Terminator Cycle Sequencing Kit for MegaBACE DNA Analysis Systems is designed for sensitive and robust sequencing with MegaBACE sequencing system. Exploiting the capabilities of the instrument, the superior resolving properties of LPA long-read matrix, the sensitivity of DYEnamic ET terminators, and the robust performance of Thermo Sequenase™ II DNA polymerase, the kit provides a convenient and flexible dye terminator format for high throughput sequencing and industry-leading data quality. To use this product, a sequencing reaction premix is combined with template DNA and primer and thermally cycled. The reaction products are then precipitated with ethanol or isopropanol to remove unincorporated dye-labelled terminators. Samples are finally dissolved in a loading solution for separation and detection using MegaBACE sequencing instrument. Energy transfer dye terminator-based sequencing DYEnamic ET Terminator Kits are based on a modification of traditional dideoxynucleotide chain termination chemistry (1) in which terminators are labeled with fluorescent dyes for automated detection. In this case, however, each of the four dideoxy terminators—ddG, ddA, ddT and ddC—is labeled with two dyes— fluorescein and one of four different rhodamine dyes—rather than a single dye (2, 3). Fluorescein has a large extinction coefficient at the wavelength (488 nm) of the argon ion laser used in the sequencing instrument. Acting as the donor dye, fluorescein absorbs energy from incident laser light and transfers it to the rhodamine acceptor dye on the same terminator molecule. Each acceptor dye then emits light at its characteristic wavelength for detection, identifying the nucleotide that terminated extension of the DNA fragment. This energy transfer format (4) is more efficient than direct excitation of 10
the acceptor dye by the laser, and produces a sequencing method that is very sensitive and robust. The acceptor dyes used in the kits are the same standard rhodamine dyes—rhodamine 110, rhodamine-6-G, tetramethyl rhodamine, and rhodamine X—used in earlier Taq dye terminator methodologies, so the DYEnamic ET reaction products can be detected on any instrument that can monitor the original Taq dye terminator chemistry. The kit also features dITP, as well as Thermo Sequenase II DNA polymerase, a thermostable enzyme that efficiently incorporates dITP. By replacing dGTP with dITP, even very strong compression artifacts common to high GC-content DNA are resolved for more accurate data interpretation. Thermo Sequenase II DNA polymerase Thermo Sequenase II DNA polymerase is a thermostable DNA polymerase specifically engineered for cycle sequencing by GE Healthcare. The enzyme readily accepts dideoxynucleotide terminators (5) and generates bands of uniform intensity, much like T7 Sequenase™ DNA polymerase (6, 7). Its tolerance to high salt conditions, efficient utilization of dITP, high processivity, and excellent performance on GC-rich templates make it an efficient and robust sequencing enzyme. Cycle sequencing Thermostable DNA polymerases allow sequencing reactions to be cycled through alternating periods of thermal denaturation, primer annealing, and extension/termination to increase the signal levels generated from template DNA (8–13). This amplification process employs a single primer, so the amount of product increases linearly with the number of cycles. A cycling protocol is especially useful when the amount of template is limiting or the sensitivity of the detection system is low. 11
6. Important considerations for using this kit The reagent formulations and DNA polymerase used in this product differ from those in other sequencing kits. This DYEnamic ET Dye Terminator Kit (US81090 and US81095) is specifically designed to provide the longest reads and the greatest degree of success using MegaBACE sequencing systems. For optimal results, the following parameters should be noted: 1. For each 20 μl reaction volume, 8 μl of premix should be used. This ratio MUST be maintained for optimal results. If using a 384- well format, use 4 μl of premix for each 10 μl reaction volume. No other configuration is supported. 2. The DYEnamic ET terminator dye set is compatible with the standard MegaBACE sequencing filters and beam splitters. Table 1. MegaBACE filter and beam splitter assignments Filters Beam splitters MegaBACE 500 520DF20, 555DF20, 540DRLP and 595DRLP and 1000 585DF20, and 610LP MegaBACE 4000 520DF20, 555DF20, 570DRXR, 540DRLP 585DF20, and 610LP and 595DRLP 3. The ethanol/salt and isopropanol precipitation protocol recommended for post-sequencing cleanup has been carefully developed to provide an efficient and low-cost method and should be followed exactly as described for optimal results. 4. The metal ions in the enzyme reaction mix are optimized for the enzymes included in the premix. Therefore, template DNA and primer should be resuspended in either water (preferably) or in a buffer containing no more than 0.1 mM EDTA. If nmol quantities of Mg2+, EDTA or other metal ion chelators are introduced with 12
template or primer, increased failure rates, weak signals or short read-lengths may occur. 5. Prolonged denaturation steps (> 1 minute at 95°C) should be avoided during the cycling protocol since enzyme denaturation is likely with weak signals and failed reactions resulting. 6. Extension times < 4 minutes and extension temperatures > 60°C can be used with Thermo Sequenase II DNA polymerase. One minute at 60°C is suggested for extension. 13
7. Protocols Preliminary preparations and general handling instructions Thaw and maintain all kit reagents on ice prior to use. Whenever possible, cap the tubes to minimize evaporation of the small volumes of reagents used. Dispense reagents using disposable-tip micropipettes, and exercise caution to avoid contamination of stock solutions. Thoroughly mix reaction mixtures after each addition by “pumping” the solution two or three times with a micropipettor without creating air bubbles. Centrifuge briefly tubes/plates to collect the reaction mixtures at the bottoms of the vessels. With practice, reactions can be completed in 15–20 minutes. The protocol described below provides high-quality sequencing results using the control DNA and primer provided in the kit. However, this protocol should be regarded only as a starting point. Optimization of protocols might be necessary to obtain the best sequencing results for specific templates. Please refer to Appendixes 1–4 and the trouble-shooting section for additional information to help optimize the sequencing reactions. 14
7.1. Preparation of sequencing reactions Researchers who utilize 0.5 ml tubes should follow the sequencing reaction and post-reaction cleanup instructions specified for 96-well plates. 1. Assemble each sequencing reaction as follows: 96-well format Template DNA __ μl Primer __ μl Water __ μl Sequencing reagent premix 8 μl Total volume 20 μl Note: Adjust the amount of distilled water such that the total volume of DNA, primer and water is 12 μl. When combined with 8 μl of sequencing reagent premix, the total volume of the reaction mix should be 20 μl. 384-well format Template DNA __ μl Primer __ μl Water __ μl Sequencing reagent premix 4 μl Total volume 10 μl Note: Adjust the amount of distilled water such that the total volume of DNA, primer and water is 6 μl. When combined with 4 μl of sequencing reagent premix, the total volume of the reaction mix should be 10 μl. 15
Note: 0.1–1 μg (40–400 fmol) of single-stranded DNA or 0.2–2 μg (80–800 fmol) of double-stranded plasmid DNA and 5 pmol of primer are recommended for routine sequencing. The volumes of DNA and primer added to each reaction will depend on their concentrations. Dilute the DNA and the primer in deionized water or buffer containing no more than 0.1 mM EDTA. Do not use buffers containing > 0.1 mM EDTA since they may reduce the effective metal cofactor concentration in the reactions. For additional information concerning the amount of template and primer to use in the reaction, see Appendixes 1 and 2 on pages 21–25. Note: The most consistent results are obtained when sequencing reagent premix is used at full strength. No other configuration is supported. 2. Assemble the control reaction exactly as follows: 96-well format M13mp18 control template 1 μl Control primer 2.5 μl Water 8.5 μl Sequencing reagent premix 8 μl Total volume 20 μl 384-well format M13mp18 control template 1 μl Control primer 2.5 μl Water 2.5 μl Sequencing reagent premix 4 μl Total volume 10 μl 16
Note: The sole purpose of the control reaction is to confirm the performance of the sequencing premix under specified and tested conditions. It is crucial to assemble and perform the reactions exactly as described above. Customer data can then be compared with GE Healthcare quality control data if the performance of the sequencing premix is in doubt. 3. After dispensing all reagents, cap the tubes or seal the plates. Mix thoroughly by gentle vortexing or gentle pumping (to avoid bubbles) with a pipettor. Centrifuge briefly to bring contents to the bottom of the tubes or wells. 4. Place the tubes or plate into the thermal cycler. Run the following cycling program for 25 cycles: 95°C, 20 seconds 50°C, 15 seconds 60°C, 1 minute (Cycling is complete in about 1 hour) Note: For additional information concerning cycling conditions, see Appendix 3 on page 25. 5. After cycling is complete, centrifuge the tubes/plate briefly to collect the reaction mixtures at the bottoms of the tubes/wells. 17
7.2. Post-reaction cleanup Please refer to Appendix 4 for additional information. 1. Option 1—Ethanol precipitation 1.1. Add 2 μl (96-well plate) or 1 μl (384-well plate) of 7.5 M ammonium acetate to each reaction tube or well. 1.2. Add 55 μl (96-well plate) or 27.5 μl (384-well plate) of 100% ethanol or 60 μl (96-well plate) or 30 μl (384-well plate) of 95% ethanol to each reaction and mix by inverting the plate several times (do not vortex). The final concentration of ethanol should be 70%. It is not necessary to use cold ethanol nor is it necessary to incubate the samples at low temperature for precipitation. Note: This step is critical. Final ethanol concentrations < 65% produce weak signals while concentrations > 75% result in sequences with “blob” artifacts due to precipitation of unincorporated dye terminators. 1.3. Centrifuge tubes at either room temperature or 4°C in a microcentrifuge for 15 minutes at ~12 000 rpm. Centrifuge 96- well or 384-well plates for at least 30 minutes at 2 500 x g or greater. 1.4. Remove the supernatant from each microcentrifuge tube by aspiration. For plates, a brief inverted spin (1 minute at 300 x g) is sufficient for supernatant removal. Remove as much liquid as possible at this step to prevent dye blobs. 1.5. Wash the DNA pellets with 70% ethanol. Use as large a volume of 70% ethanol as the tube or well can accommodate safely. Centrifuge briefly. Note: Scientists at GE Healthcare routinely use 250–500 μl for 0.5-ml microcentrifuge tubes, 100 μl for 96-well plates, and 45 μl for 384- well plates. 18
1.6. Remove the supernatants by aspiration or by an inverted spin. Air-dry (preferably) or vacuum-dry (in a vacuum centrifuge) the pellets for 2–5 minutes. Do NOT overdry. 2. Option 2—Isopropanol Precipitation 2.1. Add 40 μl (96-well plate) or 20 μl (384-well plate) of 80% isopropanol to each reaction and mix well using a vortex mixer. Note: Good results are obtained using a final concentration of 40–65% isopropanol in the precipitation mix; 50–60% isopropanol is optimal. It is important to utilize an isopropanol solution that is less than 100% isopropanol because the addition of pure isopropanol, even to the same final concentration, produces dye blobs. These blobs are caused by very high local concentrations of isopropanol before and during mixing. Please see Appendix 4 for further details. 2.2. Centrifuge tubes at either room temperature or 4°C in a microcentrifuge for 15 minutes at ~12 000 rpm. Centrifuge 96- well or 384-well plates for at least 30 minutes at 2 500 x g or greater. 2.3. Remove the supernatant from each microcentrifuge tube by aspiration. For plates, a brief inverted spin (1 minute at 300 x g) is sufficient for supernatant removal. Remove as much liquid as possible at this step to prevent dye blobs. Note: DNA pelleted by isopropanol precipitation is less firm than DNA isolated after ethanol precipitation, and can be lost during wash and inverted spins at high relative centrifugal force. 2.4. Wash the DNA pellets with 70% ethanol; DNA precipitated by isopropanol should also be washed with 70% ethanol. Use as large a volume of 70% ethanol as the tube or well can accommodate safely. Centrifuge briefly. Note: Scientists at GE Healthcare routinely use 250–500 μl for 0.5–ml microcentrifuge tubes, 100 μl for 96-well plates, and 45 μl for 384- well plates. 19
2.5. Remove the supernatants by aspiration or by an inverted spin. Air-dry (preferably) or vacuum-dry (in a vacuum centrifuge) the pellets for 2–5 minutes. Do NOT overdry. 3. Alternatives to ethanol and isopropanol precipitation include the GE Healthcare AutoSeq™96 product line (See Appendix 4). For more information, contact your local GE Healthcare office or visit us at http//www.gehealthcare.com/lifesciences and search with the keyword, “Autoseq96”. 7.3. Resuspension of samples 1. Dissolve each pellet in 10 μl of MegaBACE loading solution and vortex vigorously for 10–20 seconds to ensure complete resuspension. Briefly centrifuge to collect the samples at the bottom of the well and to remove bubbles. General recommendations a) The DNA pellet MUST be completely dissolved at this step for optimal sequencing results. If a fixed angle rotor was used for centrifugation, the DNA pellet will be on the side of the well. This material must be washed to the bottom of the well to ensure that the entire reaction product is injected onto the MegaBACE. b) It is not necessary to heat samples prior to injection. Heating samples can cause excessive evaporation of the resuspension buffer and speed the breakdown of the dye-labeled sequencing products. 20
7.4. Instrument setup and data analysis 1. For instrument setup and data analysis, please refer to the instrument documentation supplied with the MegaBACE sequencing instruments. 2. Under the Plate Setup tab of the MegaBACE Instrument Control Manager (ICM), select New to create new instrument parameters. 3. Set up instrument parameters as follows: MegaBACE 500 and 1000 Matrix Fill / High Pressure Time: 200 seconds Matrix Fill / Relaxation Time: 20 minutes Prerun: 5 minutes Prerun Voltage: 9 kV Matrix Flush Time 1: 20 seconds Matrix Flush Time 2: 7 seconds Low-Pressure Time: 240 seconds User Input Time: 120 seconds Preinjection Time: 15 seconds MegaBACE 4000 Matrix Fill / High Pressure time: 120 seconds Matrix Fill / Relaxation Time: 1 minute Prerun: 5 minutes Prerun Voltage: 9kV Matrix Flush Time: 20 seconds Low-Pressure Time 1: 5 seconds Low-Pressure Time 2: 240 seconds User Input Time: 120 seconds Preinjection Time: 15 seconds 4. Proceed with the New Plate procedure as described in the instrument documentation. 21
7.5. Injection and run parameters 1. For maximum reproducibility, use a voltage < 5 kV for injection (2 or 3 kV is standard). Injection time can be varied widely during optimization with injections as short as 5 seconds or as long as 400 seconds being equally successful. A recommended starting point for injection from MegaBACE loading solution is 2 kV for 75 seconds. For optimal results, the product of the time and voltage of injection should be within the range of 100 to 200 kV seconds. If spin columns or gel filtration plates are used, it is convenient to leave samples in the eluent and directly inject. In this case, a recommended starting point is 3 kV for 75 seconds with optimal results obtained between 150 and 270 kV seconds. These parameters are suggested starting points that are suitable for the control template in the kit and for most samples. If signals are low, longer injection times might prove beneficial. If sample overloading is a problem, utilize shorter injection times as discussed within the troubleshooting section. 2. Using standard run conditions of 100 minutes at 9 kV, average read-lengths > 500 bases with 98.5% accuracy can be expected with the control template. To take advantage of the superior resolving power of the LPA matrix and to achieve the longest read-lengths, electrophoresis should occur for 200 minutes at 6 kV. Even with these recommendations, the quality and quantity of the template remain the most important factors affecting read length and success rate. 22
8. Appendixes 8.1. Appendix 1: Template DNA—general considerations Template amount This protocol typically produces optimal results using 100–200 fmol of template DNA, but these numbers should be considered as guidelines. In some cases, more or less template can be used due to the sensitivity and robustness of DYEnamic ET terminators. For example, scientists at GE Healthcare have obtained good results with the MegaBACE sequencing systems using 25–500 ng (10 fmol–200 fmol) of pure, (single-stranded) M13mp18 DNA. For routine sequencing, follow the guidelines described above. The following formula calculates the optimal mass (0.15 pmol) of double-stranded template to include in a sequencing reaction: Mass of template (ng) = Total length of DNA (in base pairs) x 0.1 For example, plasmid that is 3800 base pairs in total length (vector plus insert) should produce optimal data using ~ 380 ng in these protocols. The recommended range for template amount is 250–500 ng (100–200 fmol). These relationships are shown graphically in the Figures 1 and 2. The best sequencing results are obtained using quantities of template that fall within the ranges indicated by the dashed lines. 23
Short PCR products 140 120 Mass of DNA (in ng) 100 80 60 40 20 0 0 200 400 600 800 1000 Length of template (base pairs) Fig 1. Recommended mass of template DNA in sequencing reactions (PCR products) Plasmids and large PCR products 1400 1200 1000 Mass of DNA (in ng) 800 600 400 200 0 0 1000 2000 3000 4000 5000 6000 7000 8000 9000 10 000 Length of template (base pairs) Fig 2. Recommended mass of template DNA in sequencing reactions (plasmids, large PCR products) 24
Insufficient template DNA present in the sequencing reaction can produce low signal strengths (< 1 000) that can cause poor base- calling and short reads. In contrast, too much template can overload the capillaries or yield very high signal strengths (> 10 000), causing software miscalls. Excessive template DNA can also deplete the supply of nucleotides in the sequencing premix and lead to short sequence reads. This is especially problematic with Polymerase Chain Reaction (PCR) products where small mass amounts of DNA are required to provide the optimal picomole amount of template. Recommended buffer for dilution of DNA template Dilute the DNA template in water (preferably) or in a weakly buffered solution containing no more than 0.1 mM EDTA. A suitable buffer is 10 mM Tris-HCl (pH 8.5), 0.1 mM EDTA. This concentration of EDTA is lower than in typical TE buffers because excess EDTA in the template or primer resuspension buffer can inhibit sequencing reactions by reducing the effective magnesium concentration. Preparation of template DNA Template of suitable quality for use with DYEnamic ET terminator kits can be prepared using a variety of procedures and commercially available products. Single-stranded plasmid DNA Several published methods are available for preparing single- stranded DNA from clones in M13 vectors and hybrid plasmid-phage (phagemid) vectors (14, 15). Preparation of double-stranded plasmid DNA Sequencing double-stranded templates with this product requires no changes in the protocol—for example, alkaline denaturation is not required. For optimal results, use plasmid DNA purified by cesium chloride gradients, polyethylene glycol (PEG) precipitation, adsorption to glass, columns, and other common DNA purification methods. 25
Due to the very small quantity of template used in the reactions, even less-pure DNA samples can yield acceptable sequence data. Although there are many popular protocols for purifying plasmid DNA from 2 ml to 10 ml cultures, GE Healthcare scientists achieve consistent success using boiling (16, 17) and alkaline (18) mini-prep methods. TempliPhi™ Templates for sequencing can also be prepared using TempliPhi DNA sequencing template amplification kits manufactured by GE Healthcare (for more details contact your local GE Healthcare office or visit us at http//www.gehealthcare.com/lifesciences and search with the keyword, “TempliPhi”). A variety of templates can be amplified by rolling circle amplification using Phi29 DNA polymerase and sequencing quality DNA can be prepared within 4–6 hours directly from bacterial colonies (19). Microgram quantities of template DNA can be prepared at isothermal conditions from picogram amounts of starting material. Amplified DNA can be used directly for cycle sequencing without purification. 8.2. Appendix 2: Primers—general considerations Primer amount The optimal amount of primer for sequencing with these protocols is 5 pmol. If too little primer is used, signals may be weak. If too much primer is present, non-specific priming can occur, resulting in “noisy” sequences characterized by high background or double (superimposed) sequences. Excessive primer also can contribute to an artifact known as the “cliff effect” that typically appears as 50–200 bases of strong peaks in the beginning of the sequence abruptly followed by weak peaks. The likely cause of this artifact is the inadvertent generation of PCR products during cycle sequencing 26
which accumulate rapidly and deplete the nucleotide supply in the sequencing premix. Determine the concentration of your primer and include 5 pmol in each sequencing reaction (2 pmol for the control primer supplied in this kit). The concentration of the primer can be measured by the following method: Resuspend the primer in water (preferably) or in buffer containing no more than 0.1 mM EDTA, and determine its optical density at 260 nm (OD260). For primers containing N bases (measured in a cuvette with a 1 cm path length), the approximate concentration (pmol/μl) is given by the formula: Concentration (pmol/μl) = OD260/(0.01 x N) where N is the number of bases. Designing a sequencing primer The length and sequence of a primer determines its melting temperature and specificity. For cycling temperatures recommended in this protocol, the primer should be ~ 18–25 bases in length. The sequence of the primer should be checked for potential self- annealing or hairpin formation, especially at its 3’-end. Possible sites of false priming in the vector or other known sequences should also be identified, again stressing matches involving the 3’-end of the primer. 8.3. Appendix 3: Cycling conditions Of the three steps that comprise the cycling program (denaturation, annealing, and extension), denaturation is the most critical. While Thermo Sequenase II DNA polymerase has significant advantages over other DNA polymerases used for cycle sequencing, it is not as stable against thermal inactivation. The reaction buffer in the sequencing premix has been specially formulated to protect the stability of the enzyme and, with proper precautions, Thermo 27
Sequenase II DNA polymerase has ample stability for robust sequencing. Denaturation step Important! Do NOT use a denaturation temperature > 95°C or longer than 30 seconds. A long denaturation step prior to cycling is commonly employed in PCR, but is unnecessary and not recommended for cycle sequencing reactions. Extended denaturing can prematurely inactivate Thermo Sequenase II DNA polymerase and ultimately produce weak signals. Annealing step The appropriate annealing temperature varies with the length and sequence of the primer. In general, temperatures from 45 to 55°C are appropriate. An annealing step is usually required only with primers < 20 bases in length. Optimal annealing temperatures are up to 5°C higher in DYEnamic ET terminator reactions than with other dye terminator sequencing products. For primers with sufficiently high melting temperatures, the annealing step can be omitted, and a two-step cycling program, alternating between denaturation and extension temperatures, can be used. Extension step Extension at 60°C for 60 seconds is optimal. Thermo Sequenase II DNA polymerase incorporates dITP more rapidly than other DNA sequencing enzymes, hence there is no apparent advantage to increase the time or temperature of the extension step. Number of cycles Twenty five to thirty cycles are sufficient to sequence the recommended amounts of plasmids or PCR products. More cycles are usually not necessary and may lead to artifacts. Increasing the number of cycles might be appropriate when sequencing extremely large templates such as bacterial artificial chromosomes (BACs). 28
8.4. Appendix 4: Considerations for post-reaction cleanup This appendix provides a summary of the considerations for post- reaction cleanup. These recommendations are starting points for optimization—the duration of precipitation, length and speed of centrifugation, geometry of centrifuge rotor, and other parameters might need adjusting. AutoSeq96 plates Unincorporated dye terminator is efficiently removed using AutoSeq96 filtration plates (27-5340-10). AutoSeq96 is a 96-well spin plate containing prehydrated G-50 Sephadex™. Follow the instructions that accompany the plates. The purified sequencing product is recovered in approximately 20 μl of water. 384-well plates and reduced volume reactions Many researchers choose smaller reaction volumes for sequencing reactions performed in a 384-well microplate. In this case, the standard ethanol and isopropanol precipitation protocols can be scaled to match the desired reaction volume. For instance, if the total reaction volume is half of the recommended volume (5 μl instead of 10 μl), use half of the recommended volume of ammonium acetate (0.5 μl instead of 1 μl) and half of the recommended volume of 95% ethanol (15 μl instead of 30 μl). Success rates might be unacceptable with such small reactions. Alternately, isopropanol-mediated precipitation can be used. Isopropanol precipitation in 384-well plates Isopropanol has two advantages over ethanol: 1) Lower concentrations are required for precipitation, hence smaller total volumes are involved during cleanup and 2) It is unnecessary to add salt to the reaction. After cycling, add 1.5–2.5 volumes of 80% isopropanol. As discussed in detail in protocol step 2.3, the 29
disadvantage with isopropanol precipitation is that the pelleted DNA is more prone to loss during washings and inverted spins. An isopropanol solution that is less than 100% isopropanol must be used to avoid forming dye blobs, as explained in protocol step 2.1. 30
9. Troubleshooting Prior to diagnosing problems associated with the sequencing reaction chemistry, operation of the MegaBACE instrument should be verified for optimal performance by injecting a plate of MegaBACE M13 DNA Sequencing Standards (US79678) and carrying out electrophoresis according to the accompanying protocol. If the average overall read-length of this standard plate is < 500 bases (98.5% accuracy), routine instrument maintenance, such as capillary cleaning or focusing, might be required. For further details, contact GE Healthcare Technical Service for assistance. Note: Control reagents in the kit should always be run in parallel with test samples during optimization. Problem: Sequencing signals are weak. Weak signals with capillary sequencing can be difficult to troubleshoot, especially for researchers accustomed to slab gel sequencing. Weak signals can be the result of an unsuccessful sequencing reaction, or the inefficient injection or overinjection of reaction products. Possible causes/solutions 1. The ionic strength of the loading solution was too high. Electro- kinetic injection into capillaries is more efficient if the ionic strength of the loading solution is low. 2. Samples were overloaded. Overloaded samples frequently have low signals since the peaks are broad and diffuse, and it is common to misdiagnose overloading as insufficient signal. Under optimal conditions, detection of all samples should begin within a few minutes of each other. Samples with late starts and broad peaks are overloaded. With capillary sequencing, signal may often be improved by injecting less sample rather than more. 31
3. The injection conditions were not optimal. Confirm that the recommended injection conditions were used. Change the injection conditions by reducing and increasing the duration of injection three-fold. 4. Too much ethanol was used for precipitation. Excess ethanol will precipitate salts, buffers and contaminants in the template DNA. These will compete for the sequencing products and reduce the effective signal. Use the volumes of ethanol recommended in the protocol or calculate the volume that will yield a final concentration of 70% ethanol. 5. An inappropriate salt was used to precipite the reaction products, or the volume of salt used was incorrect. Use the ammonium acetate included in the kit since the protocol has been optimized with this salt. 6. The formamide loading solution was old. Aqueous solutions of formamide ionize over time to produce ammonium formate which increases the ionic strength of the buffer and reduces the efficiency of injection. Using a 100% low conductivity formamide stock, prepare a fresh solution of 70% formamide containing 0–1 mM EDTA and store at 4°C. 7. The DNA preparation was impure. Repeat the reaction using the Control DNA supplied in the kit. 8. The primer or template contained excess EDTA. Resuspend both primer and template in water or in dilute buffer containing < 0.1 mM EDTA. 9. Either the quantity of template DNA or the number of cycles used for amplification was insufficient. Increase either the amount of DNA used in the reaction or the number of cycles. 10. The annealing temperature was too high for the primer being used. Use a lower annealing temperature for cycling. 32
11. Too little primer was used. The recommended amount of primer is 5 pmol per reaction. 12. The sequence of the primer was inappropriate, forming dimers or hairpins which can interfere with annealing. Change the primer sequence. 13. The wrong volume of premix was used. The reagents are carefully formulated to work optimally with 8 μl of premix in a 20 μl reaction volume or 4 μl of premix in a 10 μl reaction volume. This ratio MUST be adhered to for optimal results. No other configuration is recommended or supported. 14. Residual salt was present in the samples. This can affect the ionic strength of the sample and interfere with electrokinetic injection. If products of the sequencing reation were purified using spin columns, confirm that they were eluted in water. Some preparations of size exclusion chromatography media are pre- swollen in a salt-containing buffer and must be washed several times with water to remove the salt. In some cases, it might be necessary to wash the dry media several times to remove residual ions that can interfere with injection. 15. The template DNA was of poor quality. Contaminants (salt, protein) can decrease the efficiency of electrokinetic injection. High quality DNA prevents downstream sequencing problems. Problem: Extensions appear short with read-length limited to < 350 bases. Possible causes/solutions 1. Too much template DNA was included in the sequencing reaction. In some cases, the use of too much template, especially PCR product DNA, can exhaust the supply of dye terminators in the reaction. If this occurs, the sequence will suddenly fade before reaching 350 bases in length. This problem is especially prevalent 33
if excess primer is also present. Use < 1 pmol of template DNA and 5 pmol of primer for each sequence. By using less template, the concentration of any potential contaminant is also reduced. 2. The run voltage was too high. Limit the run voltage to ≤ 9 kV. 3. The extension step incubation period was too short. Increase the duration of the extension step in the cycling program to 2–4 minutes. Problem: Late signal-starts and broad, poorly resolved peaks are prevalent in the sequences. Capillary overloading that disrupts capillary current most often causes late appearance of the primer peak and poorly resolved sequencing fragment peaks. Overloaded samples frequently produce low signals because the peaks are broad and diffuse, and it is common to misdiagnose overloading as insufficient signal. Under optimal conditions, detection of all samples should begin within a few minutes of each other. Possible causes/solutions 1. Excessive template DNA was used in the sequencing reaction and carried over into the capillary upon electrokinetic sample injection. Template DNA molecules compete with sequencing products for injection, resulting in late starts and poorly resolved peaks. Use less template in the sequencing reaction. To determine the optimal amount of template, perform a titration of template over a 50-fold range (0.2, 0.5, 1, 2, 5, and 11 μl, for example). This titration can be accomplished easily in a single run with several templates and control DNA. 2. The injection conditions were not optimal. Confirm that the recommended injection conditions were used. Change the injection conditions by reducing the duration of injection three- fold. 34
3. The injection voltage was too high. Reduce the voltage to 2–3 kV. 4. Insufficient loading solution was used. Resuspend sequencing reaction products in a larger volume of loading solution, e.g. 20, 50 or 100 μl. 5. If the sequencing reaction products are in water, evaporate the samples to dryness, resuspend in loading solution and then inject. Problem: Localized broad peaks or very tall early peaks— terminator blobs—are prevalent in the sequences. Possible causes/solutions 1. Residual terminators were not eliminated from the samples. Carefully follow the protocol (Step 2) for post-reaction cleanup. Problem: Peak spacing changes during the run giving rise to the “accordion effect”. Possible causes/solutions 1. Samples were near the limits of overloading. Follow the suggestions to avoid overloading described above within the section “Late signal starts and broad, poorly resolved peaks are prevalent in the sequences”. Problem: Sequences are noisy or double sequences are present. Possible causes/solutions 1. The annealing temperature of the sequencing reaction was too low. Either increase the annealing temperature or eliminate it completely for cycling between 95°C and 60°C. The effective annealing temperature of primers is higher with DYEnamic ET Dye Terminator Cycle Sequencing Kit for the MegaBACE DNA Analysis Systems than with other terminator sequencing products. If problems persist, please contact GE Healthcare’s Technical Service for assistance. 35
10. References 1. Sanger, F. et al., Proc. Nat. Acad. Sci. USA 74, 5463–5467 (1977). 2. Prober, J. M. et al., Science 238, 336–341 (1987). 3. Lee, L. G. et al., Nucleic Acids Research 20, 2471–2483 (1992). 4. Ju, J. et al., Proc. Nat. Acad. Sci. USA 92, 4347–4351 (1995). 5. Tabor, S. and Richardson, C. C., Proc. Nat. Acad. Sci. USA 84, 4767–4771 (1987). 6. Tabor, S. and Richardson, C. C., J. Biol. Chem. 264, 6447–6458 (1989). 7. Tabor, S. and Richardson, C. C., Proc. Nat. Acad. Sci. USA 92, 6339–6343 (1995). 8. Huibregtse, J. M. and Engelke, D. R., DNA and Protein Engineering Techniques 1, 39–41 (1988). 9. McMahon, G. et al., Proc. Nat. Acad. Sci. USA 84, 4974–4978 (1987). 10. Carothers A. M. et al., Biotechniques 7, 494–496, 498–499 (1989). 11. Murray, V., Nucleic Acids Research 17, 8889 (1989). 12. Levedakou, E. N. et al., Biotechniques 7, 438–442 (1989). 13. Lee, J. S., DNA Cell Biol. 10, 67–73 (1991). 14. Messing, J., Methods in Enzymology 101, 20–78 (1983). 15. Mead, D. A. and Kemper, B. in Vectors: A Survey of Molecular Cloning Vectors and Their Uses, Butterworth Publishers, Massachusetts USA (1986). 16. Dente, L. et al., Nucleic Acids Research 11, 1645–1655 (1983). 17. Holmes, D. S. and Quigley, M., Anal. Biochem. 114, 193–197 (1981). 36
18. Birnboim, H. C. and Doly, J., Nucleic Acids Research 24, 1513–1523 (1979). 19. Lizardi, P. et al., Nat. Genet. 19:225–32 (1998). 37
38
39
GE Healthcare offices: GE Healthcare France Portugal regional office Tel: 01 6935 6700 Tel: 21 417 7035 GE Healthcare Bio-Sciences AB contact numbers: Fax: 01 6941 9677 Fax: 21 417 3184 Björkgatan 30 751 84 Uppsala Asia Pacific Germany Russia & other C.I.S. Sweden Tel: + 85 65 6 275 1830 Tel: 0800 9080 711 & N.I.S Fax: +85 65 6 275 1829 Fax: 0800 9080 712 Tel: +7 (495) 956 5177 GE Healthcare Europe GmbH Greater China Fax: +7 (495) 956 5176 Munzinger Strasse 5 D-79111 Australasia Freiburg Tel: + 61 2 8820 8299 Tel:+852 2100 6300 Spain Germany Fax: +61 2 8820 8200 Fax:+852 2100 6338 Tel: 902 11 72 65 Italy Fax: 935 94 49 65 GE Healthcare UK Limited Austria Amersham Place Tel: 01 /57606 1613 Tel: 02 26001 320 Sweden Little Chalfont Fax: 01 /57606 1614 Fax: 02 26001 399 Tel: 018 612 1900 Buckinghamshire Japan Fax: 018 612 1910 Belgium HP7 9NA Tel: 0800 73 890 Tel: +81 3 5331 9336 Switzerland UK Fax: 02 416 82 06 Fax: +81 3 5331 9370 Tel: 0848 8028 10 GE Healthcare Bio-Sciences Korea Fax: 0848 8028 11 Canada Corp. Tel: 1 800 463 5800 UK Tel: 82 2 6201 3700 800 Centennial Avenue Fax: 1 800 567 1008 Tel: 0800 515 313 P.O. Box 1327 Fax: 82 2 6201 3803 Fax: 0800 616 927 Central, East, & South Piscataway Latin America East Europe USA NJ 08855-1327 Tel: +55 11 3933 7300 Tel: +43 1 972720 Tel: +1 800 526 3593 USA Fax: + 55 11 3933 7304 Fax: +43 1 97272 2750 Fax: +1 877 295 8102 GE Healthcare Bio-Sciences KK Middle East & Africa Denmark Sanken Bldg. 3-25-1 Tel: +30 210 9600 687 Tel: 45 70 25 24 50 Hyakunincho Shinjuku-ku Fax: +30 210 9600 693 Fax: 45 16 24 24 Tokyo 169-0073 Netherlands Japan Eire Tel: 0800 82 82 82 1 Tel: 1 800 709992 Fax: 0800 82 82 82 4 Fax: 0044 1494 542010 Norway Finland & Baltics Tel: +47 815 65 777 Tel: +358-(0)9-512 39 40 Fax: 47 815 65 666 Fax: +358 (0)9 512 39 439 http://www.gehealthcare.com/lifesciences GE Healthcare UK Limited Amersham Place, Little Chalfont, Buckinghamshire, HP7 9NA UK imagination at work US81090PL Rev B 2006
You can also read