Reference and Troubleshooting Guide - GeneMapper Software Version 4.1

Page created by Casey Munoz
 
CONTINUE READING
Reference and Troubleshooting Guide - GeneMapper Software Version 4.1
GeneMapper® Software Version 4.1

                                   Process Quality
                                   Values and Basic
Reference and                      Troubleshooting

Troubleshooting Guide

                                   SNPlex™ System
                                   Troubleshooting

                                   Algorithms
Reference and Troubleshooting Guide - GeneMapper Software Version 4.1
Reference and Troubleshooting Guide - GeneMapper Software Version 4.1
GeneMapper® Software Version 4.1

                                    Process Quality
                                   Values and Basic
Reference and                       Troubleshooting

Troubleshooting Guide

                                   SNPlex™ System
                                    Troubleshooting

                                        Algorithms
Reference and Troubleshooting Guide - GeneMapper Software Version 4.1
For Research Use Only. Not for use in diagnostic procedures.
Information in this document is subject to change without notice.
APPLIED BIOSYSTEMS DISCLAIMS ALL WARRANTIES WITH RESPECT TO THIS DOCUMENT, EXPRESSED OR
IMPLIED, INCLUDING BUT NOT LIMITED TO THOSE OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE. TO THE FULLEST EXTENT ALLOWED BY LAW, IN NO EVENT SHALL APPLIED BIOSYSTEMS BE
LIABLE, WHETHER IN CONTRACT, TORT, WARRANTY, OR UNDER ANY STATUTE OR ON ANY OTHER BASIS FOR
SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH
OR ARISING FROM THIS DOCUMENT, INCLUDING BUT NOT LIMITED TO THE USE THEREOF, WHETHER OR NOT
FORESEEABLE AND WHETHER OR NOT APPLIED BIOSYSTEMS IS ADVISED OF THE POSSIBILITY OF SUCH
DAMAGES.
NOTICE TO PURCHASER: DISCLAIMER OF LICENSE
Purchase of this software product alone does not imply any license under any process, instrument or other apparatus,
system, composition, reagent or kit rights under patent claims owned or otherwise controlled by Applied Biosystems, either
expressly, or by estoppel.
GeneMapper Software has not undergone specific developmental validation for human identification applications. Human
identification laboratories analyzing single-source or parentage samples which choose to use GeneMapper Software for data
analysis should perform their own developmental validation studies.
The AFLP process is covered by patents owned by Keygene N.V.
TRADEMARKS:
Applied Biosystems, AB (Design), ABI PRISM, GeneMapper, GeneScan, Primer Focus, and SNaPshot are registered trademarks,
and GeneScan and SNPlex are trademarks of Applied Biosystems or its affiliates in the U.S. and/or certain other countries.
AFLP is a registered trademark of Keygene N.V.
This product includes software developed by the Apache Software Foundation (http://www.apache.org/). Copyright © 1999-2000
The Apache Software Foundation. All rights reserved.
This product includes software developed by the ExoLab Project (http://www. exolab.org/). Copyright 2000 © Intalio Inc. All
rights reserved.
JNIRegistry is Copyright © 1997 Timothy Gerard Endres, ICE Engineering, Inc., http://www.trustice.com.
Oracle is a registered trademark of Oracle Corporation.
All other trademarks are the sole property of their respective owners.
© Copyright 2009, Applied Biosystems. All rights reserved.

Part Number 4403673 Rev. A
04/2009

                                     GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Reference and Troubleshooting Guide - GeneMapper Software Version 4.1
Contents

                         Preface                                                                                          v
                         How to Use This Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . v
                         How to Obtain More Information . . . . . . . . . . . . . . . . . . . . . . . . . . vi
                         How to Obtain Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . viii

      Chapter 1          Process Quality Values and Basic
                         Troubleshooting                                                                                  1
                         Diagnosing and Resolving Basic Problems . . . . . . . . . . . . . . . . . . . 2
                              Common Troubleshooting Procedures . . . . . . . . . . . . . . . . . . . 5
                         ADO (Allele Display Overflow) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
                         AE (Allele Edit) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
                         AN (Allele Number) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
                         BD (Broad Peak) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
                         BIN (Out of Bin Allele) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
                         CC (Control Concordance) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
                         DP (Double Peak) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
                         GQ (Genotype Quality) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
                         LPH (Low Peak Height) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
                         MNF (Matrix Not Found) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
                         NB (Narrow Bin) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
                         OBA (One Basepair Allele) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
                         OS (Offscale) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
                         OVL (Overlap) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
                         PHR (Peak Height Ratio) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
                         SFNF (Sample File Not Found) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
                         SHP (Sharp Peak) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
                         SNF (Size Standard Not Found) . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                                                      iii
Reference and Troubleshooting Guide - GeneMapper Software Version 4.1
Contents

            SP (Split Peak) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
            SPA (Single Peak Artifact) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
            SPU (Spectral Pull-Up) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
            SQ (Sizing Quality) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
            XTLK (Cross Talk) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

Chapter 2   SNPlex™ System Troubleshooting                                                                     33
            Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34

Chapter 3   Algorithms                                                                                         35
            Genotyping Algorithms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
            Peak Detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
            Optimizing Peak Detection Sensitivity . . . . . . . . . . . . . . . . . . . . . 41
                 Example 1: Reducing Window Size . . . . . . . . . . . . . . . . . . . . 41
                 Example 2: Reducing Window Size/Increasing Polynomial Degree
                 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
                 Example 3: Extreme Settings . . . . . . . . . . . . . . . . . . . . . . . . . 43
            Slope Thresholds for Peak Start/End Parameters . . . . . . . . . . . . 44
            Slope Threshold Example . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
            Size-Matching/Size-Calling Algorithm . . . . . . . . . . . . . . . . . . . . . . 46
            Size-Calling Methods (Classic and Advanced Modes) . . . . . . . . . 47
                 Least Squares Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
                 Cubic Spline Interpolation Method . . . . . . . . . . . . . . . . . . . . . 49
                 Local Southern Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
                 Global Southern Method . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
            Allele-Calling Algorithms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
                 Microsatellite Analysis Methods . . . . . . . . . . . . . . . . . . . . . . . 54
                 SNPlex™ System Analysis Methods . . . . . . . . . . . . . . . . . . . . 55

            Glossary                                                                                           59

            Index                                                                                              67

iv                     GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Preface

How to Use This Guide
  Purpose of This        This guide describes the function of the Process Quality Values
           Guide         (PQV) for the supported analyses of the GeneMapper® Software,
                         explains the fundamental algorithms used by the software, and
                         provides basic troubleshooting techniques.

          Audience       This guide is intended for trained laboratory personnel. Applied
                         Biosystems is not liable for damage or injury that results from use of
                         this guide by unauthorized or untrained parties.

      Assumptions        This guide assumes that:
                          • You have installed GeneMapper® Software Version 4.1 as described
                              in the GeneMapper® Software Version 4.1 Installation and
                              Administration Guide (PN 4403614).
                          • You have a working knowledge of the Microsoft® Windows®
                              operating system.

Text Conventions         This guide uses the following conventions:
                          • Bold indicates user action. For example:
                              Type 0, then press Enter for each of the remaining fields.
                          • Italic text indicates new or important words and is also used for
                              emphasis. For example:
                              Before analyzing, always prepare fresh matrix.
                          • A right arrow bracket () separates successive commands you
                              select from a drop-down or shortcut menu. For example:
                              Select FileOpenSpot Set.
                              Right-click the sample row, then select View FilterView All.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                          v
Preface
How to Obtain More Information

     User Attention              Two user attention words appear in Applied Biosystems user
             Words               documentation. Each word implies a particular level of observation
                                 or action as described below:

                                 Note: Provides information that may be of interest or help but is not
                                 critical to the use of the product.

                                 IMPORTANT! Provides information that is necessary for proper
                                 instrument operation, accurate chemistry kit use, or safe use of a
                                 chemical.

                                 Examples of the user attention words appear below:

                                 Note: The size of the column affects the run time.

                                 Note: The Calibrate function is also available in the Control
                                 Console.

                                 IMPORTANT! To verify your client connection to the database, you
                                 need a valid Oracle user ID and password.

                                 IMPORTANT! You must create a separate Sample Entry Spreadsheet
                                 for each 96-well plate.

How to Obtain More Information
               Safety            See the GeneMapper® Software Version 4.1 Installation and
          Information            Administration Guide (PN 4403614) for safety information.

          Software               See the GeneMapper® Software Version 4.1 Installation and
      Warranty and               Administration Guide (PN 4403614) for warranty and licensing
           License               information.

vi                                GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Preface
                                                                       How to Obtain More Information

        Related          The following related documents are shipped with the software:
   Documentation          • GeneMapper® Software Version 4.1 Installation and
                             Administration Guide (PN 4403614) – Provides procedures for
                             installing, securing, and maintaining version 4.1 of the
                             GeneMapper Software.
                          • GeneMapper® Software Version 4.1 Getting Started Guides for
                             microsatellite analysis (PN 4403672), loss of hetereozygosity
                             (LOH) analysis (PN 4403621), AFLP® system analysis
                             (PN 4403620), SNaPshot® kit analysis (PN 4403618), and
                             SNPlex™ system analysis (PN 4403617) – Five guides that
                             explain how to analyze the application-specific example data
                             provided with the GeneMapper Software. The guides provide
                             brief, step-by-step procedures for the analysis of microsatellite,
                             LOH, AFLP® system, SNaPshot® kit, and SNPlex™ system data
                             generated by compatible Applied Biosystems electrophoresis
                             instruments and Data Collection Software. The guides are
                             designed to help you quickly learn to use basic functions of the
                             GeneMapper Software.
                          • GeneMapper® Software Version 4.1 Online Help – Describes
                             the GeneMapper Software and provides procedures for common
                             tasks. Access online help by pressing F1, selecting Help
                             Contents and Index, or clicking          in the toolbar of the
                             GeneMapper window.
                          • GeneMapper® Software Version 4.1 Quick Reference Guide
                             (PN 4403615) – Provides workflows for specific analysis types
                             and lists instruments, software, and analysis applications
                             compatible with the GeneMapper Software.
                          • GeneMapper® Software Version 4.1 Reference and
                             Troubleshooting Guide (PN 4403673) – Provides reference
                             information such as theory of operation and includes
                             troubleshooting information.
                         Portable document format (PDF) versions of this guide and the other
                         documents listed above are available on the GeneMapper® Software
                         Version 4.1 Documentation CD.

                         Note: For additional documentation, see “How to Obtain Support”
                         on page viii.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                             vii
Preface
How to Obtain Support

       Send Us Your     Applied Biosystems welcomes your comments and suggestions for
         Comments       improving its user documents. You can e-mail your comments to:
                                techpubs@appliedbiosystems.com

        Obtaining       The GeneMapper® Software features an online help system that
 Information from       describes how to use each feature of the user interface. To access the
   the Online Help      online help, click   in any window or dialog box (HelpContents
                        and Index if available) for more information.

How to Obtain Support
                        For the latest services and support information for all locations, go to
                        http://www.appliedbiosystems.com, then click the link for Support.
                        At the Support page, you can:
                         • Search through frequently asked questions (FAQs)
                         • Submit a question directly to Technical Support
                         • Order Applied Biosystems user documents, MSDSs, certificates
                             of analysis, and other related documents
                         • Download PDF documents
                         • Obtain information about customer training
                         • Download software updates and patches
                        In addition, the Support page provides access to worldwide telephone
                        and fax numbers to contact Applied Biosystems Technical Support
                        and Sales facilities.

viii                     GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1

                               Process Quality Values and
                                   Basic Troubleshooting

        Chapter 1
                           In this chapter:
  Process Quality Values     ■ Diagnosing and Resolving Basic Problems . . . . . . . . . . . . . . 2
           and
  Basic Troubleshooting     ■ ADO (Allele Display Overflow) . . . . . . . . . . . . . . . . . . . . . .                8
                            ■ AE (Allele Edit) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    8
                            ■ AN (Allele Number). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        9
                            ■ BD (Broad Peak) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      9
                            ■ BIN (Out of Bin Allele) . . . . . . . . . . . . . . . . . . . . . . . . . . . .          10
                            ■ CC (Control Concordance). . . . . . . . . . . . . . . . . . . . . . . . . . . 11
         Chapter 2
                            ■ DP (Double Peak). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        12
      SNPlex™ System
       Troubleshooting      ■ GQ (Genotype Quality) . . . . . . . . . . . . . . . . . . . . . . . . . . . .            13
                            ■ LPH (Low Peak Height). . . . . . . . . . . . . . . . . . . . . . . . . . . .             15
                            ■ MNF (Matrix Not Found) . . . . . . . . . . . . . . . . . . . . . . . . . .               16
                            ■ NB (Narrow Bin) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        18
                            ■ OBA (One Basepair Allele) . . . . . . . . . . . . . . . . . . . . . . . . .              18
                            ■ OS (Offscale) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    19
         Chapter 3
                            ■ OVL (Overlap) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .      20
         Algorithms
                            ■ PHR (Peak Height Ratio) . . . . . . . . . . . . . . . . . . . . . . . . . . .            21
                            ■ SFNF (Sample File Not Found) . . . . . . . . . . . . . . . . . . . . . .                 22
                            ■ SHP (Sharp Peak). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        23
                            ■ SNF (Size Standard Not Found). . . . . . . . . . . . . . . . . . . . . .                 23
                            ■ SP (Split Peak) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    24
                            ■ SPA (Single Peak Artifact). . . . . . . . . . . . . . . . . . . . . . . . . .            24
                            ■ SPU (Spectral Pull-Up) . . . . . . . . . . . . . . . . . . . . . . . . . . . .           25
                            ■ SQ (Sizing Quality) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .        25
                            ■ XTLK (Cross Talk) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .          32

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                                                    1
Chapter 1 Process Quality Values and Basic Troubleshooting
         Diagnosing and Resolving Basic Problems

Diagnosing and Resolving Basic Problems
      PQVs and the         The GeneMapper® Software has a system of Process Quality Values
    Troubleshooting        (PQVs) that provide the basis for troubleshooting fragment analysis
            Process        data using the software. PQVs are application-specific metrics, where
                           each evaluates the data for a specific quality that is consistent with a
                           problem associated with the type of analysis. In this way, the PQV
                           system can alert you to potential problems and provide you with a
                           starting point for investigation.

     About Process         Each individual PQV displays the result of a unique algorithmic test
     Quality Values        that evaluates a specific property of the fragment analysis data. The
             (PQVs)        software performs the PQV tests in a specific sequence during the
                           analysis. With the exception of the Sizing Quality (SQ) PQV, the
                           software completes the analysis of each sample in a project even if a
                           sample fails one or more PQV tests.
                           The majority of PQV metrics yield numeric values between 0 and 1,
                           where 1 indicates that the related sample data or genotype completely
                           passed the associated test. Following the analysis, the software uses
                           the upper and lower thresholds for each PQV to translate the numeric
                           score into one of three symbols, displayed in the Samples or
                           Genotypes tab of the GeneMapper window.

                           Note: If the thresholds of a PQV can be customized, the software
                           displays the parameters in the Quality Flags tab of the analysis method.

                                                                                                  Default
                             Symbol                          Definition
                                                                                                  Range

                                         Pass: The sample or genotype passed the PQV            0.75 to 1.0
                                         test.
                                         Check: A possible problem exists for the sample        0.25 to 0.75
                                         or genotype.
                                         Low Quality/Fail: There is a strong possibility that   0.0 to 0.25
                                         a problem exists for the sample or genotype.

                           Note: Applied Biosystems recommends examining all samples that
                           produce       (Check) or        (Low Quality).

2                            GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                     Diagnosing and Resolving Basic Problems

                         Note: The Allele Display Overflow (ADO) and Allele Edit (AE)
                         PQVs of the Genotypes tab report their results as           instead of the
                         colored icons.

                         Note: The GeneMapper Software does not complete the analysis of
                         samples that display (Low Quality) for the Sizing Quality (SQ)
                         PQV test.

                         Adjusting the Threshold Settings of PQV Tests
                         Some PQV metrics include components that can be customized. In
                         those cases, the user-defined parameters for the PQV appear in the
                         Peak Quality tab of analysis methods for the applicable analysis type.

                         Adjusting the Weights of PQV
                         The majority of PQV contribute to the Genotype Quality (GQ) PQV, a
                         metric used to gauge the confidence of each genotype call. In those
                         cases, some PQV contain user-defined weights that determine how
                         significantly the associated PQV affect the GQ PQV calculation. For
                         those PQV, the user-defined weights appear in the Quality Flag tab of
                         analysis methods for the applicable analysis type. For more information
                         on the calculation of the GQ PQV, see “GQ (Genotype Quality)” on
                         page 13.

                         Note: You can configure a PQV so that it does not contribute to the
                         GQ (by setting the weight to 0). However, the PQV remains active.

                         Rules for PQV Columns
                            • Quality metrics with Pass/ Check values and no             Low
                              Quality values are warning flags. Analysis does not stop if problems
                              are detected with these properties, but Applied Biosystems
                              recommends examining results flagged as        (Check).
                            • Holding the cursor over a column header displays a tooltip that lists
                              the full name of the column (the default names are often acronyms).

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                                  3
Chapter 1 Process Quality Values and Basic Troubleshooting
            Diagnosing and Resolving Basic Problems

Table 1-1     PQV values by application

                                                                     Analysis
                     PQVs

                                                                                 SNaPshot®
                                                               Microsatelite
                                                                                                       See Page

                                                                                             SNPlex™
                                                      System

                                                                                             System
                                                      AFLP®

                                                                         Other
    Abb.                  Name

                                                                                 Kit
                                                                2n
    ADO     Allele Display Overflow                    ✔        ✔        ✔         ✔          ✔           8
    AE      Allele Edit                                ✔        ✔        ✔         ✔          ✔           8
    AN      Allele Number                                       ✔        ✔         ✔          ✔           9
    BD      Broad Peak                                 ✔        ✔        ✔         ✔          ✔           9
    BIN     Out of Bin Allele                                   ✔        ✔                    ✔           10
    CC      Control Concordance                                 ✔        ✔         ✔          ✔           11
    DP      Double Peak                                                            ✔                      12
    GQ      Genotype Quality                           ✔        ✔        ✔         ✔          ✔           13
    LPH     Low Peak Height                                     ✔        ✔         ✔          ✔           15
    MNF     Matrix Not Found                           ✔        ✔        ✔         ✔                      16
    NB      Narrow Bin                                                             ✔                      18
    OBA     One Basepair Allele                                 ✔                                         18
    OS      Off-Scale                                  ✔        ✔        ✔         ✔          ✔           19
    PHR     Peak Height Ratio                                   ✔        ✔         ✔          ✔           21
    SFNF    Sample File Not Found                      ✔        ✔        ✔         ✔          ✔           22
    SHP     Sharp Peak                                          ✔                                         23
    SNF     Size Standard Not Found                    ✔        ✔        ✔         ✔          ✔           23
    SP      Split Peak                                          ✔                                         24
    SPA     Single Peak Artifact                                ✔                                         24
    SPU     Spectral Pull-Up                           ✔        ✔        ✔         ✔          ✔           25
    SQ      Sizing Quality                             ✔        ✔        ✔         ✔          ✔           25
    XTLK    Cross Talk                                          ✔        ✔                    ✔           8

4                               GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                       Diagnosing and Resolving Basic Problems

Common Troubleshooting Procedures

         About the       Procedures commonly used to troubleshoot errors and irregularities
        Procedures       in fragment analysis data:
                           ■ Displaying Sample Information/Raw and EPT Data . see below
                           ■ Displaying Numeric PQV Metrics . . . . . . . . . . . . . . . . . . . . . . 7

Displaying Sample         1. In the GeneMapper window, select the Samples tab.
  Information/Raw
     and EPT Data         2. In the Navigation Pane of the GeneMapper window:
                                a. Click      to expand the contents of the project folder.
                                b. From the list of samples, select a sample that displayed
                                        (Check) or        (Low Quality).
                          3. Select the Info, Raw Data, and EPT Data tabs as needed to
                              display the sample information for the selected file.
                                • Info tab – Displays a summary of all information for the
                                   associated sample file (see Table 1-2 on page 6).
                                • Raw Data tab – Displays an electropherogram of spectral
                                   data collected during the run of the associated sample. The
                                   spectral data is displayed in relative fluorescent units (RFU).
                                • EPT Data tab – Displays the EPT (electrical, power, and
                                   temperature) data for the associated sample throughout the
                                   course of the run.

                         Note: Table 1-2 lists only the information contained in the Info tab
                         that is relevant to troubleshooting. For a complete description of the
                         elements listed in the Info tab, see the GeneMapper® Software Online
                         Help.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                                   5
Chapter 1 Process Quality Values and Basic Troubleshooting
            Diagnosing and Resolving Basic Problems

Table 1-2     Info Tab information relevant to troubleshooting

     Group                                            Description/Information

Sample               Describes the source and status of the imported sample data.
Information
                     Sample Origin Path – Displays the path to the associated sample file at the time it
                     was imported (provided that the sample was added to the project from a sample
                     file).
                     Status Message – Displays information related to any events that occurred when
                     the sample was imported into the project.
Error Message        Displays any errors the software encounters during the analysis of the associated
                     sample. You can use the information in this group to verify the source of several
                     analysis problems.
Current Settings     Describes the analysis settings currently applied to the associated sample.
                     All of the data displayed in this group is useful for troubleshooting problems with
                     the GeneMapper Software. Although, most of the information can be viewed in
                     various parts of the software, the Current Settings group summarizes all the
                     information in a single location for easier access.
Run Information      Provides basic information about the configuration of the compatible
                     Applied Biosystems electrophoresis instrument and the run itself.
                     Instrument Name – Displays the name of the instrument used to run the sample.
                     The instrument name is important when diagnosing trends in analysis errors that
                     can be traced back to the instrument used to run the failed sample(s).
                     Data Collection Ver – Displays the version number of the Data Collection
                     Software used to run the sample.
Data Collection      Describes the configuration of the Data Collection Software at the time the sample
Settings             was run.
                     The data displayed in this group can be used to troubleshoot problems caused by
                     modifications to run modules. By comparing the Data Collection Settings
                     information from passing and failing runs, you can identify any changes made to
                     the run module (intentional or unintentional) that may have caused or contributed
                     to the failure.
                     Note: The data displayed in the EPT tab provides a log of the actual parameters
                     throughout the run.
Capillary            Provides the basic specifications of the capillary array used to run the sample.
Information
                     Capillary Number – Displays the number of the capillary used to run the
                     associated sample. Like the instrument name, the capillary number is important
                     when diagnosing trends in analysis errors that can be traced back to the capillary
                     used to run the failed sample(s).

6                               GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                     Diagnosing and Resolving Basic Problems

       Displaying        By default, the GeneMapper Software displays          (Pass),    (Check),
     Numeric PQV         or     (Low Quality) in some PQV columns to represent the numeric
          Metrics        score of the associated quality metric. When troubleshooting quality
                         errors, it is often more useful to configure the software to display
                         numeric representations of the quality values.
                         For example, the Sizing Quality (SQ) PQV evaluates the similarity
                         between the fragment pattern defined by the size standard definition and
                         the actual size standard peak distribution pattern in the sample data. The
                         Sizing Quality metric yields a value between 1 and 0 that represents a
                         combination of statistical measures for the size calling method used to
                         perform the analysis. Based on the PQV Threshold settings of the
                         Quality Flags tab, the software displays     (Pass),     (Check), or
                             (Low Quality) to indicate the result of the Sizing Quality calculation.

                            Sizing Quality Representation            Samples Table Example

                           Symbols
                           (default)

                           Numbers
                           (recommended for troubleshooting)

                         To display numerical representations of the quality metrics:
                          1. Select ToolsOptions, then select the Analysis tab.
                          2. In the Quality Metrics Display settings, select Numbers.

                          3. Click OK to apply the settings.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                                 7
Chapter 1 Process Quality Values and Basic Troubleshooting
          ADO (Allele Display Overflow)

ADO (Allele Display Overflow)
       Description/          The ADO PQV indicates that the number of alleles called for the
          Function           associated sample at the specified marker exceeds the Allele Setting in
                             the Genoptypes tab of the table setting. Because the software is
                             configured to display fewer alleles than are present, the data for the
                             additional allele is hidden from view.

                             Note: For each allele detected by the software, the Genotypes tab
                             displays six columns: name, size, height, area, mutation, and comments.

    Expected Values              – Indicates that the associated sample contains a number of
                             alleles at the specified marker that is greater than the user-defined
                             limit.

    Troubleshooting          Select the affected genotype, click   (AnalysisDisplay Plots),
                             then review the affected sample for miscalled peaks.

AE (Allele Edit)
       Description/          The AE PQV indicates whether or not a user modified the allele call
          Function           for the associated genotype.

                             Note: Allele calls can be modified in the Samples Plot, the
                             Genotypes Plot, and the Cluster Plot.

    Expected Values               – Indicates that the associated genotype call has been edited.

8                               GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                       AN (Allele Number)

AN (Allele Number)
       Description/      The AN PQV indicates that the associated sample contains either:
          Function        • A number of alleles at the specified marker that exceeds the
                             Max Expected Alleles setting (in the Peak Quality tab of the
                             analysis method)
                              or
                            • No alleles are present at the specified marker

 Expected Values             (Pass) or    (Check)

  Troubleshooting        Select the affected genotype(s), click    (AnalysisDisplay Plots),
                         then review the sample data at the affected marker for additional peaks
                         or for the absence of peaks.

BD (Broad Peak)
       Description/      The BD PQV indicates that the width of the peak for the associated
          Function       genotype exceeds the Max peak width setting (in the Peak Quality tab of
                         the analysis method).

                         Note: When the BD PQV is triggered, the software reduces the GQ
                         PQV by 50% because the default multiplier is 0.5.

 Expected Values             (Pass) or    (Check)

  Troubleshooting        Select the affected genotype, click    (AnalysisDisplay Plots),
                         then review the associated peak for irregularities.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                           9
Chapter 1 Process Quality Values and Basic Troubleshooting
        BIN (Out of Bin Allele)

BIN (Out of Bin Allele)
     Description/            The BIN PQV indicates that the apex of the peak for the associated
        Function             genotype is outside of the boundaries that define the associated bin.

                             Note: When the BIN PQV is triggered, the software reduces the GQ
                             PQV by 80% because the default multiplier is 0.8.

                             Note: For human identification (HID) analysis, the BIN PQV is
                             displayed as the OL (Off-Ladder Alleles) PQV.

 Expected Values                   (Pass) or     (Check)

 Troubleshooting             Select the affected genotype, click     (AnalysisDisplay Plots),
                             then review the allele(s) at the appropriate bin location.

     Symptom                            Possible Cause                               Solution

BIN PQV displays           After using Auto Bin to generate bins,     Correct the SPA flag by editing the
   (Check)                 the software did not create a bin for an   marker minimum fragment length,
                           allele peak because it considered the      then reanalyze and perform the Auto
                           peak to be a single peak artifact. The     Bin again.
                           SPA flag was triggered because the
                           software did not detect any stutter
                           peaks to the left of the allele peak; a
                           result of the minimum fragment length
                           for the marker being set too high
                           The GeneMapper Software detected           In the Samples tab of the
                           an allele peak that did not fit into any   GeneMapper window, set the Sample
                           of the defined bins because the bins       Type of the sample containing the
                           were not calibrated to the allelic         allelic ladder to Allelic Ladder.
                           ladder; a result of a sample file
                           containing an allelic ladder that is not
                           designated as an allelic ladder in the
                           Samples tab.
                           You generated bins using the Auto Bin      View the allele peak(s) for the marker
                           function but the GQ value for a marker     in the Genotypes Plot window.
                           was less than the Minimum Quality          Determine if the allele peaks(s) are
                           Value of 0.1 (as set in the Auto Bin       valid. If so, manually create bin(s) for
                           dialog box).                               the peak(s).

10                                GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                     CC (Control Concordance)

CC (Control Concordance)
       Description/      The CC PQV indicates that the associated control sample does not
          Function       exactly match the defined alleles for the related marker.

                         IMPORTANT! Applied Biosystems recommends running the control
                         sample at least once for every panel.

                         Note: The CC PQV serves primarily as an internal control for quality
                         assurance.

 Expected Values             (Pass) or      (Check)

  Troubleshooting        Select the affected control sample in the Samples tab of the GeneMapper
                         window, click      (AnalysisDisplay Plots), then review the positions
                         of the peaks relative to the bins.

      Symptom                     Possible Cause                                Solution

 CC PQV displays        The allele calls of the sample defined   Run the correct positive control and
    (Check)             as the Positive Control in the Samples   add the sample file to the project,
                        tab do not match the Positive Control    then define the sample as the
                        allele calls in the marker definition    Positive Control in the Samples tab
                        because the well contains the            of the GeneMapper window.
                        incorrect positive control sample.
                        The allele calls of the sample defined   Edit the Positive Control allele calls in
                        as the Positive Control in the Samples   the marker definition in the Panel
                        tab do not match the positive control    Manager.
                        allele calls in the marker definition
                        because the alleles were defined
                        incorrectly.
                        The sample defined as the Negative       Rerun the negative control and add
                        Control contains an “allele peak” due    the sample file to the project, then
                        to the presence of a spike caused by     define the sample as the Negative
                        dust or a gas bubble.                    Control in the Samples tab of the
                                                                 GeneMapper window.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                                    11
Chapter 1 Process Quality Values and Basic Troubleshooting
        DP (Double Peak)

DP (Double Peak)
     Description/           The DP PQV indicates that the peak for the associated genotype:
        Function             • Resides in a bin with another peak of the same dye color
                                  and

                               • The ratio of the peak height (minor/major peak height) is greater
                                 than the Double Peak setting in the Peak Quality tab of the
                                 Analysis Method

 Expected Values                (Pass) or      (Check)

 Troubleshooting            Select the affected genotype, click  (AnalysisDisplay Plots), then
                            review the sample data at the appropriate bin for additional peaks.

     Symptom                         Possible Cause                              Solution

DP PQV displays            A problem with the chemistry is          Check primer lengths and
   (Check)                 causing peaks from two different         electrophoresis conditions and
                           markers not to resolve, possibly         adjust as necessary.
                           because either of the primers are too
                           similar in length or the mobilities of
                           the two primer fragments are similar.

12                            GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                        GQ (Genotype Quality)

GQ (Genotype Quality)
       Description/      The GQ PQV provides a summary of the quality metrics for each
          Function       genotype. The GQ value is a calculated combination of the relevant,
                         weighted PQVs and the Marker Quality value for the genotype.

                         Calculation of the Genotype Quality (GQ) Metric
                         The formulas used by the GeneMapper Software to calculate the GQ
                         value are analysis-specific, and differ largely based on the PQVs
                         supported by each application. The following general formula
                         describes the genotype quality calculation:

                          GQ = MQ × ( ( 1 – BD ) × ( 1 – OS ) × … × ( 1 – SPU ) )
                         where the Marker Quality (MQ) value is modified by the user-defined
                         PQVs to generate the final GQ value, and the PQVs are weighted from 0
                         to 1. The actual value of each PQV in the equation is 1 minus the weight
                         assigned in the Quality Flags tab of the analysis method used to analyze
                         the data.

                           PQV Weight                   Net Effect on GQ Calculation

                                 0         No effect on the GQ calculation
                                           The initial value of 1 minus the weight of 0 yields a PQV of
                                           1. When used in the GQ calculation, the PQV has no effect
                                           since 1 ∞ MQ = MQ.
                                 1         Reduces the GQ value to 0
                                           The initial value of 1 minus the weight of 1 yields a PQV of
                                           0. When used in the GQ calculation, the PQV automatically
                                           causes the GQ to fail since 0 ∞ MQ = 0.
                              0
Chapter 1 Process Quality Values and Basic Troubleshooting
       GQ (Genotype Quality)

                         Calculation of the Marker Quality Metric
                         Figure 1-1 shows how the GeneMapper Software generates a Marker
                         Quality (MQ) value from sample peak data with assigned Allele
                         Quality (AQ) values. AQ values are a function of quality value
                         assignments for: sizing quality, allele calling quality, bin assignment
                         quality, and bin quality.

                         Note: When analyzing SNPlex™ System sample data, the GeneMapper
                         Software calculates GQ values depending on the method (Model or
                         Rules) selected to perform allele calling. The following figure
                         illustrates the derivation of GQ values using the Rules method.

                                    AQ
                                                                       AQ
                                           AQ
                                                                                    x (AQ) = MQ

                                                                       AQ

                                                              GR2119

                         Figure 1-1      Calculation of the Marker Quality metric

 Expected Values               (Pass),    (Check), or    (Fail)

                         Note: The software assigns the GQ PQV flags based on the PQV
                         threshold settings in the Quality Flags tab of the analysis method.

 Troubleshooting         Review the PQV for the affected genotype to determine the metric that
                         is causing the GQ PQV to fail.

                         Note: To better determine how individual PQV contribute to the GQ
                         PQV, configure the software to display the PQV numerically, as
                         explained in “Displaying Numeric PQV Metrics” on page 7.

14                         GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                    LPH (Low Peak Height)

LPH (Low Peak Height)
       Description/      The LPH PQV indicates that the height of the peak for the associated
          Function       genotype is lower than the associated heterozygous or homozygous
                         height limit that is specified in the analysis method. You can set
                         homozygous value (default is 200) and heterozygous value (default is
                         100) in the Peak Quality tab of the analysis method.

                         Note: When the LPH PQV is triggered, the software reduces the GQ
                         PQV by 50% (the default multiplier is “0.5”).

 Expected Values             (Pass) or    (Check)

  Troubleshooting        Select the affected genotype, click    (AnalysisDisplay Plots),
                         then review the associated peak for irregularities.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                        15
Chapter 1 Process Quality Values and Basic Troubleshooting
        MNF (Matrix Not Found)

MNF (Matrix Not Found)
     Description/         The MNF PQV indicates whether or not the GeneMapper Software
        Function          can access the matrix specified in the Matrix column of the Samples
                          tab for the associated sample.

                          IMPORTANT! Because recent models of Applied Biosystems
                          instruments save matrix data to the sample files they create, the MNF
                          flag is applicable only to sample files created by the ABI PRISM® 310
                          and 377 instruments.

 Expected Values                 (Pass) or    (Check)

 Troubleshooting          Determine the name of the missing matrix file(s) by reviewing the
                          sample information for the affected samples, as explained in
                          “Displaying Sample Information/Raw and EPT Data” on page 5.

     Symptom                     Possible Cause                             Solution

MNF PQV displays        The software could not          Locate the missing matrix file, then import it as
  (Check)               access the matrix file          explained in “Importing a Matrix File (Windows®
                        specified in the Matrix         Only)” on page 16
                        column for the associated
                                                        If the sample files for the matrix standards used
                        sample file.
                                                        to create the missing matrix are available, re-
                                                        create the matrix as explained in “Generating a
                                                        Matrix” on page 17.

     Importing a          IMPORTANT! You must convert matrix files created by Macintosh®
      Matrix File         computers before importing them. The conversion utility is free and
(Windows® Only)           available from the software support section of the Applied Biosystems
                          website (www.appliedbiosystems.com/support/software/).

                           1. Click          (ToolsGeneMapper Manager).
                           2. In the GeneMapper Manager, select the Matrices tab, then
                                  click Import.
                           3. In the Importing Matrix dialog box, navigate to and select the
                                  matrix file, then click Import.
                           4. Click Done to close the GeneMapper Manager, then analyze the
                                  project.
16                          GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                   MNF (Matrix Not Found)

      Generating a        1. Click      (ToolsGeneMapper Manager).
           Matrix
                          2. In the GeneMapper Manager, select the Matrices tab, then
                              click New.
                          3. In the Matrix Editor dialog box:
                                a. Type a name and description for the matrix.
                                b. In the Number of Dyes drop-down list, select the number of
                                   dyes present in the matrix (4 or 5).
                                c. Click B, navigate to and select the sample file for the blue
                                   matrix standard, then click Open.
                                d. Repeat step 3c for the remaining dyes in the matrix (Green,
                                   Yellow, Red, and Orange if applicable).
                                e. Click Create to create the matrix.
                                f. Click OK.

                                                                                Size standard
                                                                                name

                                                                                Size standard dye
                                                                                channel

                                                                                Fragment sizes
                                                                                for the size
                                                                                standard

                                                                                Generated matrix
                                                                                data

                          4. Click Done to close the GeneMapper Manager.
                          5. In the Matrix column of the Samples tab, select the new matrix,
                              then analyze the project.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                             17
Chapter 1 Process Quality Values and Basic Troubleshooting
        NB (Narrow Bin)

NB (Narrow Bin)
     Description/          The NB PQV indicates that the apex of the associated peak for the
        Function           associated genotype is present within 0.5 base pairs of a bin that does
                           not contain a peak. This PQV is designed to capture peaks that are
                           outside of bin boundaries because of incorrect bin definitions.

 Expected Values               (Pass) or      (Check)

 Troubleshooting           Select the affected genotype, click   (AnalysisDisplay Plots),
                           then review the peak at the appropriate bin location.

     Symptom                        Possible Cause                               Solution

NB PQV displays           You created a bin that is too narrow     In the Panel Manager, edit the bin
   (Check)                to contain its associated allele peak.   width and/or location so that it
                                                                   contains the allele peak.

OBA (One Basepair Allele)
     Description/          The OBA PQV indicates, for the associated genotype, that the apex
        Function           of the associated peak is present at a position within 1 base pair of
                           another peak.

 Expected Values               (Pass) or      (Check)

 Troubleshooting           Select the affected genotype, click     (AnalysisDisplay Plots),
                           then review the allele at the appropriate bin location for a
                           microvariant peak or an invalid allele call.

18                           GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                              OS (Offscale)

OS (Offscale)
       Description/      The OS PQV is displayed in both the Samples and Genotypes tabs of
          Function       the GeneMapper window, but the function of the OS PQV in each tab
                         differs in the following way:
                           • OS PQV for the Samples tab – The signal associated with the
                              size standard of the specified sample contains one or more peaks
                              that exceed the maximum detectable range.
                           • OS PQV for the Genotypes Tab – The signals associated with
                              the given sample contain one or more peaks that exceed the
                              maximum detectable range.

                         Note: When the OS PQV of the Genotypes tab is triggered, the
                         software reduces the GQ PQV by 50% (the default multiplier is “0.5”).

 Expected Values             (Pass) or    (Check)

  Troubleshooting         1. In the GeneMapper window, select the Samples tab.
                          2. In the Navigation Pane, click      to expand the project folder, then
                              select the sample that displays      (Check) in the OS column.
                          3. Select the Raw Data tab to display the electropherogram of
                              normalized spectral data collected during the associated sample run.
                              The spectral data is displayed in Relative Fluorescent Units (RFU).
                          4. Review the data for offscale peaks.

                                                                                        Offscale peak
                                                                                        (primer peak)

                          5. Use Table 1-3 on page 20 to determine an appropriate corrective
                              action.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                              19
Chapter 1 Process Quality Values and Basic Troubleshooting
            OVL (Overlap)

Table 1-3     OS PQV Troubleshooting

                  Symptom                     Possible Cause                  Solution

 • MNF PQV in the Samples tab displays      Too much size        No action necessary. The data
     (Check)                                standard injected    cannot be manipulated to remove
 • Raw data contains multiple off-scale     into the capillary   the oversized peaks.
   size standard peaks                                           Decrease the quantity of size
                                                                 standard used in subsequent runs.
                                                                 Also, make sure to use Hi-Di
                                                                 Formamide as the loading reagent.
                                                                 IMPORTANT! Water loading can
                                                                 produce artificially high signal and
                                                                 is not recommended.
 • MNF PQV in the Genotypes tab             Too much sample      No action necessary. The data
   displays  (Check)                        injected into the    cannot be manipulated to remove
 • Raw data contains multiple off-scale     capillary            the oversized peaks.
   peaks in the signal(s) associated with                        Decrease the quantity of sample
   the sample fragments                                          used in subsequent runs.

OVL (Overlap)
       Description/         The OVL PQV indicates that the peak for the associated genotype
          Function          has been called twice by the GeneMapper Software. If the ranges of
                            two bins overlap, a peak can reside in both bins and, therefore, be
                            called twice, once for each allele.

 Expected Values               (Pass) or     (Check)

  Troubleshooting           Select the affected genotype, click   (AnalysisDisplay Plots),
                            then review the peak and associated bins at the appropriate location.

20                           GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                     PHR (Peak Height Ratio)

PHR (Peak Height Ratio)
       Description/      The PHR PQV indicates that the apex of the peak for the associated
          Function       genotype is:
                           • Present at a position within 1 base pair of another peak
                              and
                            • The ratio of the height of the lower peak to that of the higher
                              peak is less than the Minimum Peak Height Ratio setting in the
                              Peak Quality tab of the analysis method.

                         Note: For LMS markers, the ratio is calculated based on the peak
                         heights of the called allele peaks.

                         Note: For SNaPshot® kit analysis, the ratios are calculated as they
                         are for microsatellite markers except that they span two different
                         colors, and only two peaks are used in the calculation.

 Expected Values             (Pass) or    (Check)

  Troubleshooting        Select the affected genotype, click   (AnalysisDisplay Plots),
                         then review the peaks at the appropriate location.

      Symptom                    Possible Cause                              Solution

 PHR PQV displays       The sample has undergone Loss of       Normal occurrence. No action
   (Check)              Heterozygosity (LOH). A difference     necessary.
                        in peak heights between alleles is
                                                               Further evaluate the sample for LOH
                        expected.
                                                               using the Report Settings Editor and
                                                               Report Manager.
                                                               For more information, see the
                                                               GeneMapper® Software Version 4.1
                                                               LOH Analysis Getting Started Guide
                                                               (PN 4403621).

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                             21
Chapter 1 Process Quality Values and Basic Troubleshooting
        SFNF (Sample File Not Found)

SFNF (Sample File Not Found)
     Description/         The SFNF PQV indicates whether or not the software can access the
        Function          sample file (*.fsa) shown in the Sample File column of the associated
                          sample. When the software adds a sample to a project from a sample file,
                          it retains a link to the original file. The software displays (Check) in
                          the SFNF column if the sample file is deleted, renamed, or moved.

 Expected Values              (Pass) or      (Check)

 Troubleshooting          Determine the name and location of the missing sample file:
                           1. In the GeneMapper window, select the Samples tab.
                           2. In the Navigation Pane of the GeneMapper window, click          to
                                expand the contents of the project folder, then select the sample
                                that display   (Check) in the SFNF column.
                           3. Select the Info tab, then note the name (Sample File) and
                                location (Sample Origin Path) of the sample.

     Symptom                Possible Cause                             Solution

SFNF PQV displays       Sample file has been     Search the local drives of the computer for the
   (Check)              renamed, moved, or       sample file, then do one of the following:
                        deleted.
                                                 • If you cannot find the file, no further action can
                                                   be taken to resolve the PQV flag.
                                                 • If you find the file, use the Associate Sample
                                                   feature to direct the software to the new
                                                   location as follows:
                                                    a. In the Samples tab of the GeneMapper
                                                       window, select the affected samples.
                                                    b. Select FileAssociate Samples.
                                                    c. In the Select Folder dialog box, select the
                                                       folder containing the missing files, then
                                                       click Select.

22                          GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                            SHP (Sharp Peak)

SHP (Sharp Peak)
       Description/      The SHP PQV indicates that the peak for the associated genotype is
          Function       part of a cluster of peaks with a large, narrow peak in the middle
                         whose width is 50% less than either of the neighboring peaks.

 Expected Values             (Pass) or      (Check)

  Troubleshooting        Select the affected genotype, click   (AnalysisDisplay Plots),
                         then review the peak at the appropriate location.

SNF (Size Standard Not Found)
       Description/      The SNF PQV indicates whether or not the GeneMapper Software
          Function       can access the size standard definition specified in the Size Standard
                         column for the associated sample.

 Expected Values             (Pass) or      (Check)

  Troubleshooting        Verify that the software does not contain the desired size standard:
                          1. In the Samples tab of the GeneMapper window, note the name
                               of the size standard assigned to the affected sample.
                          2. Click        (ToolsGeneMapper Manager).
                          3. In the GeneMapper Manager, select the Size Standards tab.
                          4. Verify that the Size Standard tab does not list the missing size
                               standard, or that it has not been renamed.

      Symptom                     Possible Cause                              Solution

 SNF PQV displays       (Autoanalysis only) The size             Do one of the following:
    (Check)             standard may have been set
                                                                 • If using an Applied Biosystems
                        incorrectly in the plate record of the
                                                                   size standard, click Import to
                        Data Collection Software.
                                                                   import the definition from the
                        The size standard definition has been      default Panels folder.
                        renamed, deleted, or does not exist.     • Click New to create a custom
                                                                   size standard of the same name.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                                23
Chapter 1 Process Quality Values and Basic Troubleshooting
        SP (Split Peak)

SP (Split Peak)
     Description/          The SP PQV indicates that the peak for the associated genotype is
        Function           part of a pair of overlapping peaks that are less than 0.25 base pairs
                           apart (the horizontal distance between two peak apexes).

 Expected Values               (Pass) or       (Check)

 Troubleshooting           Select the affected genotype, click   (AnalysisDisplay Plots),
                           then review the peak at the appropriate location.

SPA (Single Peak Artifact)
     Description/          The SPA PQV indicates that no peaks are present within a two-base-
        Function           pair range before the peak for the associated genotype. This PQV is
                           designed to detect the absence of stutter peaks that accompany
                           microsatellite peaks.

 Expected Values               (Pass) or       (Check)

 Troubleshooting           Select the affected genotype, click   (AnalysisDisplay Plots),
                           then review the peak at the appropriate location.

     Symptom                        Possible Cause                              Solution

SPA PQV displays          The GeneMapper Software did not           In the Panel Manager, edit the
   (Check)                detect any stutter peaks to the left of   marker minimum fragment length,
                          the allele peak because the               then reanalyze.
                          minimum fragment length for the
                          marker was set too high.

24                           GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                     SPU (Spectral Pull-Up)

SPU (Spectral Pull-Up)
       Description/      The SPU PQV indicates that the apex of the peak for the associated
          Function       genotype is at a position where the marker signal contains pull-up
                         peaks (also called bleed-through peaks). Pull-up peaks occur when
                         the peak height of the called allele is less than X% of the larger peak
                         that is within ±1 data point.

 Expected Values             (Pass) or     (Check)

  Troubleshooting        Select the affected genotype, click   (AnalysisDisplay Plots),
                         then review the peak at the appropriate location.

SQ (Sizing Quality)
       Description/      The SQ PQV reports the result of the Sizing Quality test, which gauges
          Function       the similarity between the fragment pattern defined by the size
                         standard definition, and the actual distribution of size standard peaks in
                         the sample data. The metric of the Sizing Quality test is a combination
                         of several values which measure the success of the algorithms that:
                           • Identify and eliminate primer peaks based on peak shape
                           • Perform size matching (ratio matching)
                           • Make a size-calling curve using the chosen sizing method
                         The Sizing Quality metric yields a value between 0 and 1. Based on the
                         PQV Threshold settings in the analysis method used to analyze the
                         data, the software translates the metric into the    (Pass),   (Check),
                         or     (Low Quality) flags to indicate the result of the test.

                         Note: The GeneMapper Software does not complete the analysis of
                         samples that fail the Sizing Quality test (samples that display      ).

 Expected Values             (Pass),     (Check), or     (Low Quality)

                         Note: When performing size calling using the Classic sizing method,
                         the software cannot determine Sizing Quality and, therefore, SQ is
                         always     (Check).

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                            25
Chapter 1 Process Quality Values and Basic Troubleshooting
      SQ (Sizing Quality)

 Troubleshooting            Review the data of the size standards that failed the SQ PQV:
                             1. In the Samples tab of the GeneMapper window, click
                                 (AnalysisLow Quality to Top) to sort the data so that the
                                 samples that produced errors appear at the top of the table.
                             2. In the Samples tab, select the rows for the sample(s) that display
                                    (Check) or      (Fail) in the SQ column.
                             3. Click     (AnalysisSize Match Editor) to view the sizing
                                 information for the selected sample(s).
                             4. In the Navigation Pane of the Size Match Editor, select a sample
                                 file to display the sizing data for the associated sample.
                             5. Review the data for the following qualities:
                                   • Signal Strength – The signal strength (peak height) of all
                                     peaks must exceed the Peak Detection Threshold defined in
                                     the analysis method used to analyze the data.
                                   • Correct Size Calls/Labels – All peaks must be correctly
                                     identified by the software. The labels above the peaks must
                                     be in sequential order from left to right, least to greatest.
                                   • Evenness of Signal Strength – All peaks should have
                                     relatively uniform signal strengths.
                                   • Sizing Quality – The sizing quality of each sample should
                                     be within the passing range for your chemistry application.
                                 Note: To magnify the plot of the Size Matches tab, drag the
                                 mouse cursor (    ) across a region of the x- or y-axis.

                             6. Use Table 1-4 on page 27 to determine an appropriate corrective
                                 action.
                             7. Repeat steps 4 through 6 for each sample file.

26                           GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                                SQ (Sizing Quality)

Table 1-4    SQ PQV Troubleshooting

            Symptom                          Possible Cause                           Solution

 • SQ PQV displays       or         The Size Standard Dye setting for     1. Verify that the correct size
 • Size Match Editor does not       the size standard definition is not      standard definition is in use.
   display peak data                set to the correct dye.
                                                                          2. Open the size standard
 • SQ PQV displays       or         The fragment sizes of the size           definition and verify that the:
 • Peaks do not contain size        standard definition do not match         – Dye setting is set to the
   labels                           the positions of the detected              correct dye
                                    peaks.                                   – Fragment sizes are correct
                                                                          3. Modify the size standard
                                                                             definition as necessary.
 • SQ PQV displays     or           Peak detection threshold              Adjust the analysis method so
 • One or more miscalled peaks      associated with the size standard     that the peak detection threshold
                                    is set too high or low.               associated with the size standard
                                                                          is greater than the height of the
 • SQ PQV displays       or         Peak detection threshold
                                                                          miscalled peak. See “Adjusting
 • Peaks are clear and              associated with the size standard
                                                                          Peak Detection Thresholds” on
   distinguishable, but have        is set too high or low.
                                                                          page 29 for more information.
   low signal strength
                                    Electrophoresis or pipetting error
 • SQ PQV displays      or          Insufficient cleanup step             Create and analyze the data
 • Size standard peaks occur                                              using a custom size standard
   within a primer peak                                                   that does not include the
                                                                          undetectable peak. See
                                                                          “Correcting Miscalled Peaks” on
                                                                          page 28 for more information.
 • SQ PQV displays       or         Incorrect concentration of size       Increase the concentration of
 • Size standard peaks are clear    standard in sample loading            size standard added to
   and distinguishable, but         reagent                               subsequent runs.
   consistently have low signal
   strength                         Incorrect injection settings          Review the injection settings of
                                                                          the run module for errors.
 • SQ PQV displays        or        • Electrophoresis or pipetting        See the user manual for your
 • Peaks are clear and                error                               Applied Biosystems
   distinguishable, but have        • Defective capillaries/arrays        electrophoresis instrument for
   low signal strength                                                    information on troubleshooting
 • Sizing failures occur in a                                             defective capillaries/arrays.
   regular pattern (the same
   wells fail repeatedly)
 • SQ PQV displays        or        Defective capillary
 • Size calling errors occur for
   different samples on the same
   capillary over multiple runs

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                                         27
Chapter 1 Process Quality Values and Basic Troubleshooting
      SQ (Sizing Quality)

      Correcting            You can use the Size Match Editor to correct peaks that are miscalled
 Miscalled Peaks            by the GeneMapper Software.

                            To correct a miscalled size standard:
                             1. In the Navigation Pane of the Size Match Editor, select the
                                sample file containing the miscalled peak.
                             2. Remove the label from the miscalled peak:
                                  a. Select the peak with the label by clicking inside the body of
                                     the peak.
                                  b. Select EditDelete Size Label (or right-click the peak,
                                     then select Delete).
                             3. Apply the label to the correct peak:
                                  a. Select the correct peak.
                                  b. Select EditAdd Size Label (or right-click the peak, then
                                     select Add).
                                  c. In the Select Size dialog box, double-click the label to
                                     apply to the selected peak.
                             4. Click    (ToolsCheck Sizing Quality) to verify that the
                                sample sizes correctly.
                             5. Click Apply to save the changes, then click OK.

                                IMPORTANT! You must click Apply to reanalyze the sample.

                                Note: Observe that the cell in the Status column for the sample
                                now displays     (Analysis Required).

                             6. Reanalyze the sample using the new setting to verify that the
                                problem is resolved.

28                           GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
Chapter 1 Process Quality Values and Basic Troubleshooting
                                                                        SQ (Sizing Quality)

   Adjusting Peak        You can resolve a significant number of sizing failures by adjusting
        Detection        the peak detection thresholds of the analysis method for a project.
       Thresholds        The software identifies peaks that exceed the threshold for each
                         associated dye channel, but it cannot identify peaks that fall below it.
                         Samples that exhibit low signal intensity (low peak heights) can
                         occasionally fail sizing because one or more peaks fall below the
                         threshold defined in the analysis method. By lowering the threshold
                         of the appropriate dye channel, the software can call the peak(s)
                         correctly.
                         In contrast, when the signals of the size standard peaks are very high,
                         the software may misidentify a shoulder preceding a peak as the main
                         peak (see Figure 1-2). Because the shoulder peak does not occur at
                         the correct position relative to the other peaks, sizing fails. By
                         adjusting the analysis method so that the threshold value is greater
                         than the height of the shoulder, you can achieve good sizing.

                                                                                  25-bp shoulder
                                                                                  incorrectly labeled
                                                                                  as a peak

                         Figure 1-2 Size standard with shoulder incorrectly labeled as a peak

                         To lower the peak detection thresholds of an analysis method:
                          1. In the GeneMapper window, click           (ToolsGeneMapper
                              Manager).
                          2. In the GeneMapper Manager, select the Analysis Methods tab.

GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide                              29
Chapter 1 Process Quality Values and Basic Troubleshooting
     SQ (Sizing Quality)

                           3. Do one of the following, to:
                              Modify the current analysis method, go to step 4.
                              Create a copy of the current analysis method:
                                a. Select the current analysis method.
                                b. Click Save As.
                                c. In the Save As dialog box, type a name for the new method,
                                   then click OK.
                           4. Select the analysis method you want to modify, then click Open.
                           5. In the Analysis Method Editor, select the Peak Detector tab.
                           6. Modify the appropriate Peak Amplitude Threshold settings as
                              needed. Ideally, you should set the threshold of the appropriate
                              dye channel to a value less than the signal intensity of the
                              shortest size standard peak.
                              Note: Applied Biosystems recommends using Peak Amplitude
                              Threshold settings of no less than 50 RFU.

                                                                           Peak Amplitude
                                                                           Threshold settings

                           7. Click OK to save the analysis method.
                           8. Click Done to close the GeneMapper Manager, then reanalyze
                              the samples using the new analysis method.

30                         GeneMapper® Software Version 4.1 Reference and Troubleshooting Guide
You can also read