RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual

Page created by Ted Guerrero
 
CONTINUE READING
RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual
RAEGuard PID
    FGM-1000 Series
Photoionization Detector Transmitter

 Operation & Maintenance Manual
         p/n 033-4001-001 Rev B
RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual
READ BEFORE OPERATING
This manual must be carefully read by all individuals who have or
will have the responsibility of using, maintaining, or servicing this
product. The product will perform as designed only if it is used,
maintained, and serviced in accordance with the manufacturer’s
instructions.
                           CAUTION
 To reduce the risk of electric shock, turn off power before remov-
 ing the lamp housing cap and PID sensor for service. The lamp
 housing cap provides an electrostatic shield. Never operate the
 unit when the cap is removed. Remove the lamp housing cap and
 sensor block only in an area known to be NON-HAZARDOUS.

                          WARNING!
 The calibration of all newly purchased RAE Systems instruments
 should be tested by exposing the sensor(s) to known concentration
 calibration gas before the instrument is used or put into service.
 For maximum safety, the accuracy of the RAEGuard PID should
 be checked by exposing the sensor(s) to known concentration cali-
 bration gas.

                    AVERTISSEMENT!
 La calibration de toute instruments de RAE Systems doivent être
 teste en exposant l’instrument a une concentration de gaz con-
 nue par une procédure dietalonnage avant de mettre en service
 l’instrument pour la première fois. Pour une sécurité maximale,
 la sensibilité du RAEGuard PID doit être verfier en exposant
 l’instrument à une concentration de gaz connue par une procédure
 dietalonnage.
RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual
Contents
1. General Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  1
  Key Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2
  Applications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2
  Hazardous Location Classification . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2
  Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2
2. Operation of RAEGuard PID. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  3
2.1 Physical Dimensions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  5
2.2 Installation and Access Instructions. . . . . . . . . . . . . . . . . . . . . . . . . .  6
  Instrument Assembly Removal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6
  Electrical Wiring. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  8
  Instrument Assembly Instalation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  9
  Sample Gas Plumbing. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  10
2.3 Display and Alarm Functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
  System Initialization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
  Reading Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
  Alarm Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
  Access to Programming Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.4 Programming Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  12
2.5 Calibration Procedures. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  13
  Zero Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  14
  Span Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  15
2.6 Correction Factors for Other Gases . . . . . . . . . . . . . . . . . . . . . . . . .  16
  Programming the Correction Factor . . . . . . . . . . . . . . . . . . . . . . . . . .  17
  Modifying the Span Gas Value . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  17
  Manual Conversion. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  17
2.7 Calibration With Other Compounds. . . . . . . . . . . . . . . . . . . . . . . . .  17
2.8 4-20 mA Current Output. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  18
2.9 Duty Cycle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  19
3. Theory of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  19
4. Maintenance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  20
4.1 Cleaning the PID Sensor and Lamp. . . . . . . . . . . . . . . . . . . . . . . . .  21
4.2 Pump Replacement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  22
4.3 Filter Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  23
5. Troubleshooting Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  24
Appendix A: Correction Factors. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  26
Appendix B: Maximum Reading Capabilities and Indications . . . . . . .  29
Appendix C: Limited Product Warranty. . . . . . . . . . . . . . . . . . . . . . . . .  30
Appendix D: RAE Systems Contacts. . . . . . . . . . . . . . . . . . . . . . . . . . .  31

                                  www.raesystems.com                                                     1
RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual
1. General Information
The RAEGuard PID is a fixed photoionization detector (PID) that mea-
sures a broad range of volatile organic compounds (VOCs). It operates
on 9-36 VDC and provides an analog (4-20 mA) signal output in three
ranges of 0.01 - 20.00 ppm, 0.1-100.0 ppm, and 1-1000 ppm isobutylene
equivalent. Housed in an explosion-proof enclosure, the RAEGuard PID
is equipped with a local digital display of the gas concentration and func-
tion keys for performing calibration. This flow-through sampling unit is
ideal for remote sampling.

Key Features
    • 4-20 mA analog output
    • Resolution of 0.01 ppm for precision model (range to 20 ppm)
    • Explosion-proof enclosure for hazardous environment applications
    • Magnetic key interface eliminates need to open explosion-proof
      ­enclosure when making operation parameter adjustments
    • Long-life 10.6 eV ultraviolet lamp with 3D sensor
    • Can draw sample from up to 100 feet (30 meters)
    • Sampling pump duty cycling for lamp self-cleaning
    • LCD and status/error LEDs
    • Operation at 9 to 36 VDC
    • Two dry contacts (
RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual
Specifications
 Size                        5.0" L x 5.0" W x 4.5" H
                             127 mm x 127 mm x 115 mm
 Weight                      5.5 lbs (2.5 kg)
 Detector                    3D PID sensor std. 10.6 eV lamp
 Calibration                 2-point field calibration
 Enclosure Rating            US & Canada: UL, cUL Class I, Division
                             1 & 2, Groups B, C, D
                             Europe:       0575      II 2 G
                             Demko 03 ATEX 0229512
                             EEx d IIB T6
 Power                       9 to 36 VDC max. 125 mA at 24V
 Output                      4-20 mA w/12V compliance at 24V
 Sampling                    Internal diaphragm pump, >500 cc/min
 Sampling Distance           100 feet (30 meters) maximum
 Display                     7-segment, 4-digit screen with 4 color-
                             coded alarm LEDs
 Response Time               40 seconds to 90% of reading using isobu-
                             tylene and 40% duty cycle
 User Interface              Magnetically accessed keys for calibration
 Temperature                 -20º C to 55° C (-4º F to 131° F)
 Humidity                    0-95% relative humidity (non-condensing)
 Dry Contacts                High and low alarm levels Max 30V, 2A
                             each
 Mounting                    2 holes, 5.25" (133 mm) from center to
                             center
 Ranges                      0.01 - 20.00 ppm, 0.1 - 100.0 ppm, and
 (isobutylene equivalent)    1 - 1000 ppm
                             (Ranges are different product models, and
                             not user-selectable when operating unit.)

This product may be covered by one or more of the following U.S. Patents:
        5,393,979       5,561,344       5,773,833        6,225,633
        6,313,638       6,333,632       6,320,388

                       www.raesystems.com                              3
RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual
2. Operation of RAEGuard PID
The RAEGuard PID photoionization detector transmitter consists of a
UV light source, a detection chamber, a sample draw pump, and associat-
ed electronic driver and measurement circuits. The contents of the Lamp
Cleaning Kit (part number 081-0002-000) include: lamp cleaner, swabs,
finger cots, and lens paper. The contents of the Accessory Kit (part num-
ber 033-0912-000) include: sensor puller, Magnet Key, and screwdriver.
Prior to factory shipment, the RAEGuard PID is calibrated and tested
using isobutylene gas.

The calibration of all newly purchased RAE Systems instruments should be
tested by exposing the sensor(s) to a known concentration calibration gas
before the instrument is put into service for the first time. For maximum
safety, the accuracy of the RAEGuard PID should be checked at least
monthly when operated in a relatively clean environment. However, in a less
clean, highly humid, or variable environment, it may require more frequent
(such as once a week) accuracy checks for optimal performance. Recalibrate
the unit if readings are outside the limits defined by the user.

RAE Systems recommends the following procedure, based on the ISEA
(Industrial Safety Equipment Association) protocol, to establish the interval
between verification checks:
    1. The protocol defines the difference between a functional (bump) test
       and a full calibration. A functional (bump) test is defined as a means
       of verifying calibration by using a known concentration of test gas
       to demonstrate that an instrument’s response to the test gas is within
       acceptable limits. A full calibration is defined as the adjustment of an
       instrument’s response to match a desired value compared to a known
       concentration of test gas as defined in section 2.5 of this manual.
    2. During a period of initial use in the intended atmosphere, perform
       a daily functional (bump) test to ensure there is nothing degrading
       sensor performance. The period of initial use must be of sufficient
       duration to ensure that the sensor is exposed to all conditions that
       might have an adverse effect.
    3. Any instrument that fails a functional (bump) test must be adjusted
       by means of a full calibration procedure before further use.
    4. If the tests demonstrate that it is not necessary to make adjustments,
       then the time interval between checks may be lengthened. For
       maximum safety, this interval should not exceed 30 days.
4                             RAEGuard PID
RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual
2.1 Physical Description
Physical Dimensions
The design of RAEGuard PID allows it to be easily mounted and inter-
faced with a fixed-point gas monitoring system. The detector transmitter is
housed in a 5.0" L x 5.0" W x 4.5" H (127 mm x 127 mm x 115 mm) explo-
sion-proof case with two holes 5.25" (133 mm) apart from center to center.

3/4" - 14 NPT               3" [127.00]              3/4" - 14 NPT
        4.90" [124.46]

                                   ZERO SPAN

                              OK FAULT LOW HIGH
                                                      O 5/16"

                               +     MODE      -

                          O 4 1/2" [114.30]        3/4" - 14 NPT     1.08" [27.43]
                           5 1/4" [133.35]                            4.60" [116.84]

RAEGuard PID Standard Version Installation.

Sample Gas Flow
The RAEGuard PID has an internal diaphragm pump that pulls the
sample gas in and forces it out through its outlet. The gas is drawn in
through the bottom opening and pushed out through the upper left, while
the upper right hole is used for all electrical connections.

                         Outlet

                                                                            Sample
                                                                            Intake

RAEGuard PID flow-through intake and output
The sample pump runs intermittently according to the programmable
duty cycle. The default 40% duty cycle causes the pump to be on for 8
seconds and off for 12 seconds every 20 seconds.
                                   www.raesystems.com                                  5
RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual
2.2 Installation and Access Instructions
                             WARNING
    1. A minimum of 18 inches (45 cm) of explosion-proof conduit must
      be used at cable entry in group B atmospheres. CSA requires seals
      in conduit exceeding 5 feet (1.5 meters) in group C atmospheres.
      Note: For European applications, the installation must comply with
      the requirements of EN 60079-14.
    2. To prevent ignition of hazardous atmospheres, area must be free
      of flammable vapors and supply circuit must be disconnected
      before removing cover.
    3. Keep cover tight while circuits are alive.

Mounting
First, decide where the transmitter will be mounted. (Refer to installa-
tion drawing, below.) Drill two holes in the mounting surface, with the
centers of the holes spaced 5.25 inches (133 mm) apart. Mount the RAE-
Guard PID using suitable screws.

                               OK FAULT LOW HIGH

                                +   MODE    -

                            5.25" (133 mm)

Instrument Assembly Removal
       Prior to service, make sure power is OFF. Observe all Hazardous
       Location Safety procedures.

                             1. Unscrew the housing top from the housing
                                bottom.

6                            RAEGuard PID
RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual
2. Insert the small cylindrical end (key) of the RAEGuard PID Magnet Key
    (p/n 033-2032-000) into the hole on the left side of the support post.

 SIDE VIEW
              Magnet                                               Key
               End                                                 End
 TOP VIEW
Two views of RAEGuard PID Magnet Key.

                                                           Support
                                                            Post
                                                         Lock     Open

                                                Magnet
                                                 Key

  3. Turn the Magnet Key 90° counterclockwise to rotate the support
    post to the “Open” position. You will feel the tension slacken, which
    indicates you have turned the support post far enough.

                                           Support
                                            Post
                                         Lock       Open

                                                                Magnet
                                                                 Key

Caution: Use only the provided Magnet Key to turn the support post.
Inserting a longer metal object (such as a screwdriver) may damage the
circuit board while rotating the support post.

                       www.raesystems.com                                7
RAEGuard PID FGM-1000 Series - Photoionization Detector Transmitter - Operation & Maintenance Manual
4. Carefully lift the entire instrument assembly halfway out of the
      RAEGuard PID bottom housing.
                                                    Instrument assembly
                                                    halfway out of bottom
                                                    housing with 8-pin
                                                    connector plugged in

                                             Detail of 8-pin
                                             electrical connector
                                             unplugged

    5. Unplug the white 8-pin electrical connector. Use tweezers to discon-
      nect the Tygon® tubing from the effluent (output) side of the pump.
    6. Remove the entire instrument assembly from the housing bottom.

Electrical Wiring
    1. Inside the housing bottom are two green terminal block plugs at-
      tached to the terminal block on the circuit board. Unplug the two
      terminal block plugs from the terminal block. Use tweezers, an IC
      puller, or needle-nose pliers and lift the terminal blocks straight out.

8                              RAEGuard PID
Note: The terminal block plugs accept 16 AWG to 28 AWG wire. Use
  16 AWG or 18 AWG wire for long wiring runs, which can be up to 1
  km (1,000 meters) long

 2. Lace the wires through the RAEGuard PID’s wire hole(s) and con-
   nect wires to the corresponding pin numbers of the terminal blocks:

Wire                                              Pin          Wire
                                                  Number       Color
Alarm Common                                      1            Yellow
High Level Alarm Contact                          2            Green
Low Level Alarm Contact                           3            Orange
4-20 mA Output                                    6            Blue
Power Supply Negative/Output Common               7            Black
Power Supply Positive (9-36VDC)                   8            Red
Not Used (troubleshooting only)                   4            Brown
Not Used (troubleshooting only)                   5            Purple

Instrument Assembly Installation
 1. Plug both green terminal block plugs back into the correct terminal
   block headers. Keep extra wires as close to the inside enclosure wall
   as possible.
 2. Plug the white 8-pin connector back into its socket.
 3. Carefully place the instrument assembly back into the housing bot-
   tom, sliding the hole in the bottom of the assembly over the support
   post. As you lower the assembly into the housing, make sure the sup-
   port post passes freely through the hole in the circuit board.

 Note: Be sure to reconnect the effluent Tygon® tubing before placing
  the assembly all the way down into the housing.

                      www.raesystems.com                                 9
4. Gently rock the assembly to ensure that it is properly seated and the
    hole in the support post is fully visible above the circuit board.
  5. Use the RAEGuard PID Magnet Key to rotate the support post 90º
    clockwise, which secures the instrument assembly in the housing.
  6. Tightly screw the housing top to the housing bottom.

  Caution: Damage to the LCD may result if the housing is screwed
   down before the assembly is seated all the way down. If you encoun-
   ter resistance, remove the housing top and check that the instrument
   assembly is properly seated.

  7. Ground the unit using the screw connection to the lower left of the
    sample inlet in order to reduce the chance of electrical shock.

Sample Gas Plumbing
Designed as a flow-through unit, the RAEGuard PID is designed to draw
a sample from up to 100 feet (30 meters), or against a vacuum of up
to 20 inches of water (38mm Hg). The pump can draw up to 40 inches
of water, but in this case, the flow will be reduced and the pump may
require more frequent maintenance. The typical flow rate without tubing
or a vacuum is 550 to 650 cc/min while the pump is on in the 20-sec-
ond cycle. See RAE Systems Technical Note 140 for details on tubing
pressure drop and time delays for extension tubing. If the gas supplied is
under pressure, throttle the flow so that it is within the range of 0.6 to 2.0
liters per minute.

Any tubing used must be inert to chemical reaction or physical adsorp-
tion. We recommend Teflon® or stainless steel tubing with an inner di-
ameter between 1/16" (1.5875 mm) and 3/16" (4.7625 mm). Do not use
Tygon® or other soft plastic tubing, as these absorb many organic vapors.

Note: Tubing of 1/16" (1.5875 mm) inside diameter may cause excessive
pressure drop and should be used only for sampling distances of ≤ 20 feet
(6 meters).

10                           RAEGuard PID
2.3 Display and Alarm Functions
System Initialization
When the RAEGuard PID system power is turned on, it initializes and an
“InIt” message appears on the display. As the transmitter is warming up,
the lamp and pump turn on as each component is checked, and the LEDs
blink all at once. The countdown timer appears on the screen for the 6-
second warm-up.

Reading Display
As the transmitter enters the Reading Display, it automatically starts test-
ing for errors and goes through a cycle of checking each alarm condition.
If there are no errors or alarm conditions, the green “OK” LED is lit and
the gas concentration is displayed.

If there is an error, the “Fault” LED blinks and an error message blinks
on the screen. Each alarm condition has a corresponding LED that blinks
an amber color when the readings are outside a specified range or limit.
Refer to Section 5: Troubleshooting Tips for alarm relay logic details and
Appendix B for Maximum Reading Capabilities and Indications.

Alarm Contacts
The alarm contacts can be used to drive user-supplied external alarms
such as a light or buzzer. The low alarm contact closes when the display
reads over the programmed low alarm limit. When the reading exceeds
the high alarm limit, the low alarm contact opens and the high alarm
contact closes. When the display is over-range, the high alarm contact re-
mains closed and the output jumps to 22 mA to allow remote detection. A
fault condition can be detected by a drop to 2 mA output; no dry contact
closure occurs.

Access to Programming Functions
The RAEGuard PID has no external switches, but instead uses the mag-
netic end of the RAEGuard PID Magnet Key to activate switches built
into the unit. To use the RAEGuard PID Magnet Key, briefly touch the
glass above the MODE circle or the triangles labeled [+] and [-]. Then
remove the key straight out and away from the RAEGuard PID.

Access the Calibration menu from the Reading Display by keying [MODE],
or access the Advanced menu by keying [+], [-], and then [MODE].

                        www.raesystems.com                              11
2.4 Programming Menu
The programming menu lets you change the operating parameters using
the magnetic sensing key. To operate the key, briefly tap the display glass
directly above the triangles labeled [+] or [-] or the [MODE] circle.

Important! Do not drag the key sideways, or two functions may be activated.

To access the programming menu from the Reading Display, press [+],
[-], and then [MODE] in sequence. Then step through the parameters
shown below by keying [MODE]. Each tap of the key on [MODE] ad-
vances to the next item in sequence.

Note: The RAEGuard PID transmitter returns to the Reading Display
after 60 seconds of idle time.

                           Display Message
 Submenu Parameter FGM-1000 FGM-1001 FGM-1002
 Span Gas Value      C100       C10.0      C10.00
 Correction Factor   CF1.0      CF1.0      CF1.0
 Low Alarm Limit     L200        L20        L5
 High Alarm Limit    H500        H50        H12
 Pump Duty Cycle*     P40        P40        P40
* If the Pump Duty Cycle value is changed, the unit should be recalibrat-
  ed with zero and span calibration gases.

Note: Submenus only cycle through once. When [MODE] is keyed after
accessing the Pump Duty Cycle submenu, the transmitter returns to the
Reading Display.

Anytime a submenu screen is idle for more than 15 seconds, the transmit-
ter returns to the Reading Display.

To change a value, press [MODE] until the correct parameter appears. To
increase a value, press [+]. To decrease a value, press [-]. Press [MODE]
when you are finished changing the value.

12                          RAEGuard PID
If a value has changed, the new value blinks on the LCD. Press [-] to
discard changes and advance to the next submenu item. Press [+] to save
changes. The “SAUE” message appears on the screen to confirm changes
have been saved.

2.5 Calibration Procedures
The RAEGuard PID is calibrated using a two-point calibration process.
First, use a “zero gas.” Then use a “span gas” containing a known concen-
tration of a standard reference gas, to set the second point of reference.

Any gas can be used for calibration, but the greatest accuracy comes
from calibrating with the same gas as is being measured. However, if a
span gas cylinder of the measured gas is not available, isobutylene (IBE)
may be used to calibrate, and a Correction Factor (CF) may be applied.
See section 2.6 for Correction Factor procedures.The recommended
isobutylene span gas concentration depends on your RAEGuard PID’s
range. Below are the default values:

    RAEGuard PID Range		         Calibration Gas
    1000 ppm		                   100 ppm
    100.0 ppm		                  100 ppm
    20.00 ppm		                  10 ppm

                                   RAEGuard PID Inlet

                                       Connect calibration gas
                                       cylinder and then turn on
                                       gas.

RAEGuard PID connected to gas cylinder with calibration adapter.

                        www.raesystems.com                             13
WARNING
 The calibration of all newly purchased RAE Systems instruments
 should be tested with a known concentration calibration gas before the
 instrument is first put into service. For maximum safety, the accuracy
 of the RAEGuard PID should be checked with a known concentration
 calibration gas.

Note: Zero calibration must be performed before span calibration.

To perform a calibration, you need a cylinder of zero gas, a cylinder of
span gas, and tubing. The gas regulator should provide a flow of at least
600 cc/min.

Zero Calibration
  1. Connect the zero gas cylinder to the barbed gas inlet port on the
    RAEGuard PID with tubing. A charcoal tube may be connected to
    provide zero air.

  2. To access the Calibration menu, press [MODE] from the Reading
    Display. Zero calibration appears first, as indicated by the “ZErO”
    message on the screen.

     Optional: To advance to span calibration, press [MODE] a second
      time. “SPAn” appears on the screen. Skip to the Span Calibration
      instructions in the next section.

     Optional: To exit the calibration menu, press [-] to return to the
      Reading Display.

  3. Turn on the gas flow. Press [+] to start calibration. The “Zero” LED
    starts blinking, and the “ZErO” message alternates with a 60-second
    countdown timer.

     Optional: Before countdown reaches zero, you may press any key
      to interrupt zero calibration in order to advance directly to span
      calibration.

14                           RAEGuard PID
4. Once the countdown reaches zero, the “Zero” LED stops blinking,
   and the zero calibration data is saved.
 5. Turn off the gas flow from the calibration gas cylinder and remove
   it. Zero calibration automatically advances to span calibration when
   zero calibration is complete.

Span Calibration
 1. Connect the span gas cylinder to the RAEGuard PID’s barbed gas
   inlet port.

   Optional: To exit the calibration menu altogether, press [-] to return
    to the Reading Display. Alternatively, after 60 seconds of idle time,
    it automatically returns to the Reading Display.

 2. Once the countdown reaches zero, the “Span” LED stops blinking.

 3. If the PID (photoionization detector) sensitivity is acceptable, the
   span data is calculated and saved.

 4. If the PID sensitivity is unacceptable, the screen alternately flashes a
   “SPAn” and “FAIL” message until any key is pressed.

   Note: If span failure occurs, it may be necessary to clean or replace
    the PID lamp.

 5. In case of span failure, you may restore the original calibration data
   by pressing [+] when “rStr” appears on the screen. Then the transmit-
   ter returns to Reading Display. Otherwise, press [-] or [MODE] to
   return to zero calibration to restart the entire calibration process.

 6. The calibration procedure is complete. After a few seconds, the
   RAEGuard PID returns to the Reading Display.

 7. Turn off the flow of gas from the calibration gas cylinder and re-
   move it.

                       www.raesystems.com                                  15
2.6 Correction Factors for Other Gases
          CAUTION: Calibrating the RAEGuard PID to a specific gas
          does not make it selective to that gas. A PID is a broadband
          detector that always responds to all the gases the lamp can ion-
          ize, and gives the readout in equivalent units of total VOC for
          the calibration gas.

Correction Factors have been determined that enable a large number of
chemicals to be quantified using only a single calibration gas, typically
isobutylene (IBE). When the instrument is calibrated with IBE, there
is no need to recalibrate it when other compounds are to be measured.
Rather, a Correction Factor (CF) can be applied to have the PID readout
in units of the new compound.

The CF is the ratio of the response to isobutylene over the response to a
particular compound. It is defined as:

  CF = True VOC Concentration
		          PID Reading

Once the CF has been established, the true concentration can be obtained
by multiplying the reading by the CF:

        True VOC Concentration = CF x PID Reading

For example, if the RAEGuard PID is calibrated to isobutylene and reads
10 ppm on a source of toluene in air (CF = 0.5), then true toluene con-
centration = 0.5 x 10 ppm = 5.0 ppm.

Correction factors are inverse to sensitivity. Compounds with CF greater
than 1.0 are less sensitive than isobutylene, and those with CF of less
than 1.0 are more sensitive than isobutylene.

Appendix A gives a short list of CFs. An extensive list is available in
RAE Systems Technical Note 106.

16                          RAEGuard PID
There are three ways to use these Correction Factors to measure the con-
centration of another gas:

A. Programming the Correction Factor
  1. Calibrate the RAEGuard PID using isobutylene gas as described in
    Section 2.5.
  2. Program the Correction Factor for the specific gas as described in
    Section 2.7.
  3. The unit now directly displays the concentration in ppmv (parts per
    million by volume) of the specific gas.

B. Modifying the Span Gas Value
  1. When programming the span gas value, multiply the true isobutylene
    concentration by the CF.
  Example: Benzene has a correction factor of 0.53 relative to isobutylene gas.
          Correction Factor x standard value = compensated value
  Example: 0.53 x 100 ppm = 53 ppm for benzene
  2. Enter a span gas value of 53 and calibrate with 100 ppm isobutylene
    as described in section 2.5.
  3. The unit now displays the concentration in ppmv of the specific gas
    (for example, benzene).

C. Manual Conversion
  1. Calibrate the RAEGuard PID unit using isobutylene gas as described
    in Section 2.5.
  2. Multiply the reading by the Correction Factor to obtain the true con-
    centration of the specific gas.

2.7 Calibration With Other Compounds
Note: Correction Factors introduce some error, and therefore it is always
preferable to calibrate directly with the gas to be measured, if available.
However, calibrating with compounds that have Correction Factors >1 may
cause a calibration “FAIL” error. This is because compounds with low sen-
sitivity generate less of a response signal than the RAEGuard PID expects.
In this case, you should modify the span value to a number equal to:
                  (calibration gas concentration) / CF
Then program in the CF for measurement.

                         www.raesystems.com                                17
For example, to calibrate using a standard of 100 ppm of hexane gas,
which has a CF of 4.3, do the following:

  1. Program the span gas concentration to:
  		 100 ppm hexane = 23 ppm
  			            4.3
  2. Program the CF to 4.3.
  3. Zero calibrate.
  4. Apply 100 ppm hexane and initiate span calibration.

At the end of the calibration countdown, the display should indicate 23
ppm. After exiting to the normal concentration display, the reading should
be close to 100 ppm.

2.8 4-20 mA Current Output
The current output correlates with the display reading as follows:
                   Display Reading
  Iout = 4 + (16 x Full Scale Range )

Thus, a display reading of 50 ppm on the 100 ppm model of the RAE-
Guard PID corresponds to an output of:
                    50
  I    = 4 + (16 x 100 ) = 12 mA
		 out
A reading of 50 ppm on the 1000 ppm model corresponds to:
                    50
  I    = 4 + (16 x 1000 ) = 4.8 mA
		 out

When a Correction Factor is programmed, it modifies both the display
reading and the current output. For example, if 25 ppm isobutylene is
applied to the 100 ppm model with no CF applied (that is, if CF=1.0), the
display reads 25 ppm and Iout = 8 mA. If the same gas is applied with a
CF of 2.0, the display reads 50 ppm and Iout = 12 mA.

Current outputs outside the 4-20 mA range (for example, 0 mA, 2 mA, or
22 mA) indicate failure modes (see Section 5).

18                          RAEGuard PID
2.9 Duty Cycle
The pump duty cycle can be varied between 30% and 70% of every 20-
second interval. Thus, a 70% duty cycle means that the pump is on for
14 seconds and off for 6 seconds. During the pump’s off time, the lamp
remains on to generate ozone and “burn off” heavy contaminants that
may build up on the lamp and sensor. The shortest duty cycle of 30%
gives the most lamp self-cleaning and thus the best sensor stability over
the long term. The longest duty cycles should be used only where rapid
response is crucial to the measurements. It is not currently possible to run
the pump continuously (that is, the duty cycle cannot be set to 100%).

3. Theory of Operation
The RAEGuard PID uses a discharge UV (ultraviolet) lamp as a high-en-
ergy photon source. After sample gas is drawn into the ionization cham-
ber by the pump, light emitted from the UV lamp ionizes a portion of the
VOC gas molecules. A pair of electrodes inside the ionization chamber
captures the ions and measures the resulting current. Because the mea-
surement is essentially non-destructive, the effluent gas can be used for
further analysis if desired.

The RAEGuard PID is a microcontroller-based instrument. After the
electrical signal from the PID sensor is conditioned and digitized, the mi-
crocontroller processes the data and then displays the results locally and
reconstructs the digital data into a standard 4-20 mA current signal.

The lamp failure detector monitors the UV light source, and the pump
current sensor monitors the pump operation. Failure conditions are im-
mediately indicated by LED, a message on the display, and a specified
electrical current signal.

The instrument has three magnetic sensing keys. Using the RAEGuard
PID Magnet Key, it is possible to calibrate the RAEGuard PID and
change the alarm level setting via the magnetic front-panel keypad.

The instrument is powered by a DC power supply located in a safe area.
The power supply on the printed circuit board contains a switching and

                        www.raesystems.com                              19
linear regulator that converts the main input voltage to 3.3V and 5V DC
to power the entire circuit.

Gas                           Sensor      Electrometer        A/D              Transmitter    4-20 mA
 In
                   Pump    UV
                          Lamp                                                  Display
Gas
Out                                         Power
                            Lamp Driver                  Microcontroller
                                            Supply
                                                                                Keypad
Flame-arresting                                                                                From
   metal frit                                                                                Controller

4. Maintenance
You should clean the PID sensor and the lamp when one of the following
occurs:

  1. The reading is inaccurate even after calibration.
  2. The reading is very sensitive to air moisture.
  3. A chemical liquid has been sucked into the unit and has damaged it.

                   Housing

                  Display

             Lock hole
        Support post

                  Pump                                                     PID sensor
  Terminal block
                                                                    Sensor gasket

                                                                           #005 O-ring

                                                                              Gas inlet
  Housing base

Exploded view of RAEGuard PID components.
20                                     RAEGuard PID
4.1 Cleaning the PID Sensor and Lamp
During the course of normal operation, a film of gas vapor may build up
inside the PID sensor. The rate of film build-up depends on the type and/
or the concentration of vapors being sampled. Cleaning the sensor helps
avoid a drifting response at high humidity. In clean ambient environ-
ments, clean the sensor once a year. Refer to Section 5 Troubleshooting
Tips for indications that the sensor needs cleaning.

                      If the lamp does not turn on or there is excessive
                      film build-up on the lamp window, the RAEGuard
                      PID produces a failure signal (2 mA) for current
                      output. This indicates the lamp requires cleaning.
                      Periodically clean the lamp window to remove film
                      deposits and restore lamp sensitivity. Exercise care
                      when cleaning the lamp window so that the surface
PID Sensor            is not damaged.

                      Follow this procedure for cleaning lamps or sensors:

                      1. Remove the instrument assembly from the hous-
                      ing body. (Refer to Section 2.2, Installation Instruc-
                      tions.) The instrument assembly may also be sent in
                      for repair. Refer to Chapter 5 Troubleshooting Tips.
Lamp
                      2. Loosen the two screws below the sensor, but do
   not completely remove it.
 3. Use the IC puller to remove the sensor. Pull it straight out with a
   slight rocking motion, using some force, since the sensor will be
   snug in its place. Do not touch the sensor’s electrode pins.
 4. Dip the entire PID sensor into GC Grade methanol. Cleaning the
   PID sensor in an ultrasound bath for about 15 minutes is highly rec-
   ommended. Do not disassemble the sensor for cleaning.
    Note: Never touch the window surface with fingers or anything that
    might leave behind film or dirt..
  5. Dry the sensor thoroughly. Never touch the electrode pins of the sen-
    sor with your hands.
  6. Examine the sensor for visible damage such as corrosion, bent pins,
    or any other mechanical damage. If the sensor is corroded or dam-
    aged, contact RAE Systems service for replacement.
  7. Carefully examine the sensor O-ring for cracks that may cause a
    leak. Replace as necessary.
                        www.raesystems.com                             21
8. Carefully pull out the lamp. Avoid contact with the flat window
   surface.
 9. Clean the lamp window surface by wiping it with anhydrous methanol*
   on a cotton swab. Rub in a circular motion with moderate pressure.
   * Caution: Water degrades window surfaces.
 10. After cleaning, hold the lamp up to the light at an angle to detect any
   remaining film. Repeat the process until the lamp window is clean.
 12. Re-insert the clean lamp, making sure the lamp’s 1/2" Teflon® O-
   ring is in place.
 13. Re-insert the clean sensor block. One of the four pins is thicker
   than the other three to act as a key and ensure proper alignment.
 14. Replace the instrument assembly into the housing body (refer to
   Section 2.2 Installation Instructions).
 15. Inspect the large O-ring on the body for damage and replace it if
   necessary. Then screw on the cover.

4.2 Pump Replacement
Follow this procedure for removing the pump:

 1. Remove the instrument assembly from
   the housing body. (Refer to Section 2.2,
   Installation Instructions.) The instrument
   assembly may also be sent in for repair.
   Refer to Chapter 5 Troubleshooting Tips.

 2. Disconnect the white 5-conductor con-
   nector that connects the pump and the
   circuit board.

22                          RAEGuard PID
3. Disconnect the tube from the left inlet on
   the pump. A small-blade screwdriver may
   be helpful in prying the hose off of the
   connector. Alternatively, you may use an
   IC puller or needlenose pliers.

 4. Hold the pump and remove the two Phillips screws from the base of
   the instrument assembly.
 5. Lift the pump out of the instrument assembly.

Replace the pump with a new pump (part number 033-3003-100). Follow
this procedure:

 1. Slide the pump into place inside the instrument assembly and locate
   its two screw holes over the matching holes in the instrument assem-
   bly’s base.
 2. Screw in the two Phillips screws until the pump is held snugly.
 3. Reconnect the hose to the left pump inlet.

   Important: Make sure to reconnect the tube between the effluent
   (output) port and the right port on the pump.

 4. Connect the 5-pin connector of the cable from the pump to the
   matching connector base on the circuit board.
 5. Replace the instrument assembly into the housing body (refer to
   Section 2.2 Installation Instructions).
 6. Inspect the large O-ring on the body for damage and replace it if
   necessary. Then screw on the cover.

4.3 Filter Adapter
The RAEGuard PID must not be operated without a filter in the inlet
path. Refer to RAE Systems Technical Note TN-187 for further details.

                       www.raesystems.com                               23
5. Troubleshooting Tips
Note: Before diagnosing measurement problems, perform zero and span
calibration.

Symptom         Reading abnormally high.
Reason          Dirty sensor block. Excessive moisture.
Solution        Clean or replace sensor block. Eliminate/control
                source of moisture and/or temperature difference.
Symptom         “Span Failure” signal.
Reason          Dirty sensor. Weak/defective lamp.
Solution        Clean sensor block. Replace sensor block. Re-
                place lamp.
Symptom         Reading abnormally low; reading jumping
                around randomly.
Reason          Dirty or weak lamp. Incorrect calibration. Sample
                tubing restricts flow or absorbs gas.
Solution        Clean or replace lamp. Recalibrate unit. Use inert
                or wider-bore tubing.
Symptom         No response during calibration.
Reason          No standard gas flow. Pump defective or blocked.
                Low sensitivity to calibration gas. Weak lamp.
Solution        Make sure standard gas flows into unit. Unblock
                flow path or replace pump. Use different calibra-
                tion gas. Replace lamp.
Symptom         Full-scale measurement in humid environment.
Reason          Dirty or wet sensor.
Solution        Clean sensor with methanol using an ultrasound
                bath, and then dry thoroughly.
Symptom         Low 4-20 mA output.
Reason          Power supply is lower than specified.
Solution        Check power supply voltage and connection.

24                       RAEGuard PID
Instrument Failure Table
Error & Reason   E001 Lamp failure.
Solution         Change the lamp.
Error & Reason   E002 Pump failure.
Solution         Make sure wires are not loose and that there
                 are no mechanical obstructions in the pump
                 motor. Replace pump assembly.
Error & Reason   E003 Calibration fails.
Solution         Perform zero and span calibrations. Clean
                 sensor. Clean or change lamp. Check gas flow
                 path. Check calibration gas.
Error & Reason   E004 Zero drift.
Solution         Perform zero and span calibration. Electri-
                 cally ground to housing using screw at lower
                 left of inlet.
Error & Reason   E005 Gas sensitivity beyond range of instru-
                 ment.
Solution         Use different calibration gas and use within
                 measurement range limits. Dilute gas.

                  www.raesystems.com                      25
Alarm Relay Logic
                       External         LED LCD               Analog
                       Alarm                                  Output
 Exceed Low            Low Alarm        Low     Reading       Based on
 Alarm Limit                                                  reading
 Exceed High           High Alarm       High    Reading       Based on
 Alarm Limit                                                  reading
 Over range            High Alarm       High    Flashing      22 mA
                                                8888
 ADC Saturated         High Alarm       Fault   Flashing      2 mA
 (max)                                          E005
 Sensor Fault,         Low Alarm        Fault   Flashing      2 mA
 Lamp Fail                                      E001
 Sensor Fault,         Low Alarm        Fault   Flashing      2 mA
 Pump Fail                                      E002
 Sensor Fault,         Low Alarm        Fault   Flashing      2 mA
 Calibration Fault                              E003
 Sensor Fault,         Low Alarm        Fault   Flashing      2 mA
 Sensor Drift                                   E004

Appendix A. Correction Factors
The following is a partial list of Correction Factors for a number of com-
monly used compounds for the PID sensor with 10.6 eV lamp. (Refer
to RAE Systems Technical note TN-106 at www.raesystems.com for
complete lists of Correction Factors).

Table A: Correction Factors for PID with 10.6 eV Lamp
 Compound                       Correction Factor
 Acetylaldehyde                 5.5
 Acetic acid                    14
 Acetone                        1.1
 Acrylic acid                   12
 Allyl alcohol                  2.4
 Ammonia                        9.7

26                          RAEGuard PID
Table A: Correction Factors for PID with 10.6 eV Lamp
Continued
Benzene                    0.53
Butadiene                  1.0
Butyl acetate, n-          2.6
Butyl acrylate, n-         1.6
Carbon disulfide           1.2
Cyclohexane                1.4
Cyclohexanone              1.0
Dichloroethene, t-1,2-     0.45
Diesel Fuel                0.66
Diethylamine               0.97
Dimethylhydrazine, 1,1-    0.78
Epichlorohydrin            10
Ethanol                    12
Ethene                     10
Ethyl acetate              4.6
Ethyl acrylate             2.4
Ethyl benzene              0.52
Ethyl ether                1.1
Ethyl hexyl acrylate, 2-   1.1
Ethyl sulfide              0.51
Gasoline vapors            0.85
Gasoline, whole            1.0
Heptane, n-                2.6
Hexamethyldisilazane       0.24
Hexane, n-                 4.3
Hydrazine                  2.6
Hydrogen sulfide           4.1
Isobutane                  120
Isobutene                  1.00
Isobutyl acrylate          1.5
                                     Table continued on next page

                      www.raesystems.com                     27
Table A: Correction Factors for PID with 10.6 eV Lamp
Continued

 Isopropanol                     6.0
 Jet fuel JP-4                   0.99
 Jet fuel JP-5                   0.6
 Jet fuel JP-8                   0.6
 Methyl acrylate                 3.7
 Methyl ethyl ketone             0.86
 Methyl isobutyl ketone          1.2
 Methyl methacrylate             1.4
 Methyl t-butyl ether            0.91
 Nitric oxide                    5.2
 Octane, n-                      1.8
 Perchoroethene                  0.58
 Pinene, a-                      0.31
 Pinene, b-                      0.37
 Propene                         1.7
 Styrene                         0.42
 Tetrahydrofuran                 1.8
 Toluene                         0.5
 Trichloroethene                 0.52
 Vinyl chloride                  2.0
 Xylene, m-                      0.43
 Xylene, o-                      0.59
 Xylene, p-                      0.45
For more data, see RAE Systems Technical Note TN-106.

The VOC Correction Factors are measured relative to isobutylene gas at
Appendix B. Maximum Reading Capabilities and Indications
                     Model      Range              Saturation Analog        Saturation Analog       Over
                                                   Value      Output        Value      Output for   Range
                                                                                       Over Range   Display
                     FGM-1002 0.0 ppm - 20.0 ppm   20.62 ppm 4 mA - 20.0 mA 20.5 mA 22 mA           Flashing
                                                                                                    8888

                     FGM-1001 0.1 ppm - 100 ppm    103.1 ppm 4 mA - 20.0 mA 20.5 mA    22 mA        Flashing
                                                                                                    8888

                     FGM-1000 0.0 ppm - 1000 ppm   1031 ppm 4 mA - 20.0 mA 20.5 mA     22 mA        Flashing
                                                                                                    8888

www.raesystems.com
29
Appendix C. Limited Product Warranty
RAE Systems (RAE) warrants the RAE products to the original owner
purchased directly from RAE or an authorized RAE Reseller to be free of
defects in workmanship for a period of one year from the date of origi-
nal shipment to owner exclusive of consumables, including batteries,
filters, and calibration gases. Pumps and 10.6 eV electrodeless discharge
lamps are warranted for one year. RAE’s obligation under this warranty
is limited to replacing or repairing, at RAE’s option, any defective part if
returned to a RAE authorized factory repair center, with shipping charges
prepaid by the buyer, and which, upon inspection by RAE, shall prove to
have been defective in normal use and service.

To maintain warranty, Purchaser must perform maintenance and calibra-
tion as prescribed in the Operation and Maintenance manual. In the event
of defect, Purchaser will notify RAE designated factory repair center in
advance, and if trouble diagnosis procedures are unable to determine and
remedy the condition, a Return Material Authorization (RMA) will be
issued to assure proper repair and logistics tracking.

Warranty does not extend to any equipment malfunction or damage
which results from alteration, accident, theft, misuse, abuse, abnormal
use, improper or unauthorized repairs, or improper maintenance. RAE
neither assumes nor authorizes any other firm or person to assume on
RAE’s behalf any liability in any way connected with the sale of RAE
products.

This express warranty shall extend to the buyer of record only and not
to sales made by the buyer’s customers. Except for the warranty of title,
the foregoing express warranty is in lieu of any and all other warranties,
whether expressed or implied, including the implied warranties of fitness
for a particular purpose and merchantability. Seller’s liability under the
warranty provided herein exclusive of insurance process shall be limited
to a refund of purchase price.

30                          RAEGuard PID
Appendix D. RAE Systems Contacts
RAE Systems by Honeywell World Headquarters
3775 N. First St.
San Jose, CA 95134-1708 USA
Phone: 408.952.8200
Fax: 408.952.8480

E-mail: customerserv@raesystems.com
Web Site: www.raesystems.com

RAE Systems Technical Support
Monday through Friday, 7:00AM to 5:00PM Pacific Time
+1.888.723.4800 (toll-free)
email: tech@raesystems.com

RAE Systems Europe ApS
Kristinehøj 23 A
DK-2770 Kastrup
Denmark
Phone: +45 86 52 51 55, From 8.00 - 17.00
Fax: +45 86 52 51 77

RAE Systems (Hong Kong) Ltd.
Room 8, 6/F, Hong Leong Plaza
33 Lok Yip Road
Fanling, N.T. HONG KONG
Phone: +852.2669.0828
Fax: +852.2669.0803

                    www.raesystems.com                 31
RAE Systems by Honeywell
World Headquarters
3775 N. First St.
San Jose, CA 95134-1708 USA

Phone: 408-952-8200
Fax: 408-952-8480
E-mail: customerserv@raesystems.com

                              p/n 033-4001-001 Rev B, January 2009
                   www.raesystems.com
You can also read