Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
←
→
Page content transcription
If your browser does not render page correctly, please read the page content below
Demonstration Preparation The instruments and software listed in the table below are required to perform the demonstrations. Product Type Minimum System Requirement –– U2021XA or U2022XA Keysight U2020XA Series USB peak –– Firmware version A.01.00 or and average power sensors later Keysight USB 2.0 compliant cable, Mini-B connector with locking –– U2031A, U2031B, U2031C mechanism –– Firmware version A.01.80 or later Keysight N5182A MXG Signal Vector –– Frequency range option 503 or Analyzer 506 –– Baseband generator option 651 or 652 Keysight IO Libraries Suite Version 16.2.xx or later Keysight N1918A Power Analysis Power Analyzer (R03.05.xx or later) Manager To download and update the instrument firmware or software, please visit: www.keysight.com/find/usbpeaksensor Instructions for PC, U2021XA, and MXG Connectivity Procedures 1. Connect the Laptop (installed with the N1918A Power Analysis Manager, Power Analyzer) to the U2021XA USB power sensors via the USB Mini-B cable. 2. Connect the U2021XA USB power sensors to the RF Output of the N5182A MXG signal generator shown in Figure 1. 3. Connect the BNC-SMB cable from the N5182A MXG signal generator’s rear panel Event 1 to the Trig In socket on the U2021XA USB power sensor. 4. Launch the Keysight Connection Expert application (Keysight IO Libraries Suite) and verify the connection of the U2021XA USB power sensors. BNC-SMB cable Event 1 Trig In USB Mini-B N5182A MXG Signal Generator U2021XA Series Laptop cable USB Power Sensor Figure 1. Laptop, U2021XA and MXG signal generator setup diagram 1. Alternatively, you can substitute the N5182A MXG with the E4438C ESG Vector Signal Generator, firmware revision C.04.98 or later, frequency range option 503 or 506, and baseband generator option 601 or 602 2
Keysight N1918A Power Analysis Manager The Keysight Technologies, Inc. N1918A Power Analysis Manager, Power Analyzer is a Windows-based free license application software for the U2020XA Series USB Peak and Average power sensors. This application software enables you to perform advanced power analysis measurements in real time. For more information on the N1918A Power Analysis Manager, refer to the Keysight N1918A Power Analysis Manager Datasheet, Literature Number 5989-6612EN 3
Demonstration 1 Average Power Measurement in Compact Mode Display In this guide, keystrokes surrounded by [ ] represent front panel keys of the Objectives instruments, keystrokes surrounded by { } represent softkeys of the instrument, To demonstrate the capabil- and keystrokes in bold represent the softkeys of the software. ity of the Power Analyzer to perform power measurement and switch the measurement Instructions Keystrokes display to compact mode. On the MXG 1. Set instrument to its default Press [Preset] settings 2. Set the frequency and ampli- tude –– Press [Frequency] [1] {GHz} Frequency = 1 GHz –– Press [Amplitude] [0] {dBm} Amplitude = 0 dBm 3. Turn on the RF Output –– Turn [RF On/Off] to “On” In the Power Analyzer Double-click on the Power Analyzer application software shortcut on your desktop or access the program via the Windows Start menu. (Note: The latest version of Power Analyzer demonstrated in this demo 4. Run the Power Analyzer guide will start the measurement upon running of the application software) –– Click {channel} on the Instrument 5. Set the frequency to 1 GHz Navigator panel (See Figure 2) –– Set Frequency (Hz) to “1G” on the Instrument Properties panel –– Move the cursor to highlight 6. Switch the normal mode dsplay instrument measurement 1 to compact mode (See Figure 3) –– Click Switch { } the Power Analyzer toolbar 7. Switch the compact mode Click Switch { } on the Power Ana- to normal mode display (soft lyzer panel) 8. Stop the measurement Click Stop { } on the Power Analyzer 4
Figure 2. Average power measurement shown in normal mode display (soft panel) Figure 3. Compact mode display show the average power measurement 5
Demonstration 2 Enable Alert, Limit, and Recording using Power Analyzer Objectives Instructions Keystrokes (Continue from Demo 1) –– To demonstrate the alert func- On the MXG tions where the system will notify you the exact time and 1. Set amplitude power to 0.1 dBm Press [Amplitude], [0.1], {dBm} date the power measurements of your DUT has exceeded In the Power Analyzer the specified upper and lower limits. Alert functions are es- –– Click Create Strip Chart { } on Power Analyzer toolbar. pecially useful for applications 2. Add strip chart display such as aging and reliability –– Select “Meas ID” and “Current text where long test times are Active Tab”, then click “OK” required. –– To demonstrate the strip chart –– Click Stop { } on the Power function in the Power Analyzer Analyzer toolbar application software, which –– Move the cursor to highlight allows the user to monitor, re- normal mode display cord, and playback the power 3. Delete the normal mode display measurements tend for a long –– Click { } to delete normal period of time. mode display –– Click Run { } to continue the measurement –– Click {Measurements} on the 4. Enable limit and alert of Instrument Navigator panel measurement (See Figure 4) –– In Instrument Properties panel, Upper Limit = 0.1 dB check “Enable Alert” Lower Limit = –0.1 dB –– Set Upper Limit to “0.1” and Lower Limit to “–0.1” –– Click Stop { } on the Power Analyzer toolbar –– Move the cursor to highlight normal mode display –– Click Recorder { } on the Power 5. Enable recording function Analyzer toolbar. (See Figure 6) –– The Recorder menu will prompt up, select specific time for recording and enter the recording file name. In this example, select “Immediately” and enter the filename as “Texting” then click “OK” 6
Figure 4. Alert Limits textbox allows user to enter upper/lower limit as per their requirements Figure 5. Click Alert Indicator { } to view the alert summary as shown here in order to identify the timestamp where the measurement was over the alert limit Figure 6. Recording file saved as *.csv can be viewed via MS Excel 7
Demonstration 3 Instructions Keystrokes On the MXG Pulse Power Measurement 1. Set instrument to its default Press [Preset] settings Objectives 2. Set the frequency and ampli- tude –– Press [Frequency] [1] {GHz} –– To demonstrate the capability Frequency = 1 GHz –– Press [Amplitude] [0] {dBm} of the U2021XA Series USB Amplitude = 0 dBm power sensors to obtain –– {Pulse Period}, [40], {μsec} 15-point pulse parameter 3. Set the pulse signal –– {Pulse Width}, [10], {μsec} values from the trace graph, where the parameters values 4. Activate the pulse signal Toglle {Pulse On/Off} to “On” are automatically 5. Turn on the RF Output Toglle [RF On/Off] to “On” calculated by the Power In the Power Analyzer Analyzer application software Double-click on Power Analyzer whenever a trace pattern application software shortcut on your appears on the graph. 6. Run the Power Analyzer desktop or access the program via the –– To demonstrate the capability Windows Start menu. of the dual gates feature to –– Click {channel} on the Instrument measure the pulse width Navigator panel. 7. Set the frequency to 1 GHz and pulse period signal (See Figure 7) –– Set Frequency (Hz) to “1G” on the respectively. Instrument Properties panel. –– In this demo, pulse signal with pulse period 40 μs and pulse –– Click { } on the Power width 10 μs will be generated 8. Add trace graph display Analyzer toolbar. from the signal generator. –– Select “Meas ID” and “Current The U2021XA Series USB power sensors measures the Active Tab” then click “OK”. average, peak, and peak-to- –– Click Stop { } on the Power average of the pulse signal via Analyzer toolbar the dual gated feature. –– Move the cursor to highlight the 9. Delete the normal mode display normal mode display and click { } to delete the normal mode display. –– Click Run { } to continue the measurement. Measure Pulse Width Signal Move the cursor to highlight the trace graph display. Right-click on the trace 10. Add dual gates to measure graph display and select “Apply Gate pulse width signal, 10 μs (See Option”, Gate 1. Move the dual gated Figure 8) (vertical line) to the highlighted pulse width duration. Measure Pulse Period Signal Move the cursor to highlight the trace graph display. Right-click on the trace 11. Add dual gates to measure graph display and select “Apply Gate pulse period signal, 40 μs (See Option”, Gate 2. Move the dual gated Figure 9) (vertical line) to the highlighted pulse width duration. In the Measurement Panel (bottom of the trace graph display), click “Mea- 12. Add 15-point pulse parameter surement Selector”, then check “Select values from the trace graph All” under Pulse Measurement menu, display (See Figure 10) and click “Apply”. 8
Figure 7. Average power measurement of pulse signal (pulse period 40 μs and pulse width 10 μs) Figure 8. Pulse width measurement (delta time, average, peak and peak-to-average) blue line shown in Pulse Analysis table 9
Demonstration 2 Figure 9. Pulse period measurement (delta time, average, peak and peak-to-average power) orange color shown in Pulse Analysis table (on top of pulse width measurement Figure 10. Pulse Analysis table consists of 15-point pulse parameter values. 10
Demonstration 4 Burst Power Measurement Objectives 4.613 ms timeframe duration –– To demonstrate the capability of U2021XA Series USB power sensors to measure burst power (average, peak and peak-to-average) signal via external/internal triggering methodology and 577 μs timeslot duration automatically calculated by Power Analyzer application Figure 11. GSM timeslot pattern with timeslot 0 on software whenever a burst waveform appears on the Instructions Keystrokes graph. –– To demonstrate the capability On the MXG of the dual gates function to 1. Set instrument to its default Press [Preset] measure the burst signal. settings –– In this demo, GSM burst 2. Set the frequency and ampli- signal with 4.613 ms duration tude –– Press [Frequency] [1] {GHz} (consists of 8 timeslots and Frequency = 1 GHz –– Press [Amplitude] [0] {dBm} each timeslot with 577 μs Amplitude = 0 dBm duration) will be generated –– Press [Mode], {Dual ARB}, from the signal generator (See {Select Waveform}, {Waveform Figure 11). 3. Recall GSM modulation signal Segments} –– Using external triggering from –– Press [ ] or [ ] to select Event 1 of signal generator, to {GSM_BURST_WFM} synchronize with the U2021XA –– Press [Mode], {Dual ARB}, USB power sensors via Trig {Select Waveform} In port before starting to –– Press [ ] or [ ] to highlight capture GSM waveform for 4. Set GSM modulation signal {GSM_BURST_WFM} measuring the average, peak –– Press {Select Waveform} and and peak-to-average power (see Figure 1). toggle {ARB ON/Off} to “On” Press [Mode], {Dual ARB} and [More], 5. Set external triggering for GSM {Marker Utilities}, {Route Connectors}, modulation signal {Route to Event 1 BNC} and {Marker 2} 6. Turn on the RF Output Toggle [RF On/Off] to “On” 11
Instructions Keystrokes In the Power Analyzer Double-click on the Power Analyzer application software shortcut on the 1. Run the Power Analyzer desktop or access the program via the Windows Start menu –– Click {Channel} on the Instrument Navigator panel 2. Set the frequency to 1 GHz –– Set Frequency (Hz) to “1G” on the Instrument Properties panel –– Click { } on the Power 3. Add trace graph display Analyzer toolbar. –– Select “Meas ID” and “Current Active Tab” then click “OK” –– Press Stop { } on the Power Analyzer toolbar. –– Move the cursor to highlight the normal mode display. 4. Delete the normal mode display –– Click { } to delete the normal mode display. –– Click Run { } continue the measurement. –– Click {Channel} on the Instrument 5. Configure trace display to cap- Navigator panel ture GSM 4.613 ms waveform –– Set Trace Start (s) to “–100μ” and Trace Length (s) to “5m” –– Click {Trigger} 6. Configure power sensor to –– Set “Global Trigger Source” to external triggering mode “External” Measure GSM Timeslot 0 Time-Gated Signal (Average, Peak, and Peak-to-Average Power) –– Move the cursor to highlight the trace graph display –– Right-click on the trace graph 7. Add dual gates to measure GSM display and select “Apply Gate timeslot 0 duration (See Figure Option” 12). and “Gate 1” –– Move the dual gated (vertical line) to highlight the GSM timeslot 0 duration Measure GSM Waveform Signal (Average, Peak, and Peak-to-Average Power) –– Move the cursor to highlight the trace graph display –– Right-click on the trace graph 8. Add dual gates to measure GSM display and select “Apply Gate whole waveforms (See Figure Option” 13). and “Gate 2” –– Move the dual gated (vertical line) to highlight the GSM waveform 12
Figure 12. Dual gate (Gate 1) press ON GSM timeslot 0, shows delta time, average power, peak power, peak to average power measurement Figure 13. Dual gate (Gate 2) press ON GSM waveform signal 4.613 ms duration, shows delta time, average power, peak power, and peak to average power measurement 13
Demonstration 5 Time-Gated CCDF and Waveform CCDF Measurements Complementary Cumulative Distribution Function (CCDF) Objectives –– To demonstrate the A CCDF curve is defined by how much time the waveform spends at or above a capability of U2021XA given power level. This is expressed in dB relative to the average power. Series USB power sensors to A CCDF curve is a plot of relative power levels versus probability where the measure and compute time- X-axis represents the dB above the average signal power, while the Y-axis gated CCDF and waveform represents the percent of time the signal spends at or above the power level CCDF measuremenet in trace specified by the X-axis. display view via the N1918A Power Analysis Manager, Power Analyzer. –– To demonstrate the capability of the dual gates function to measure and computer the measurement in time-gated CCDF (Burst CCDF) and waveform CCDF. Figure 14. CCDF plot shows the Y-axis that represents the percentage of time the signal power equals the power specified by the X-axis Waveform CCDF Conventional peak power meter/sensors provide a built-in function of waveform CCDF measurement in graphical and table format. Waveform CCDF calculates CCDF and peak-to-average ratio using the entire waveform, including portions where the data is zero (no signal). Time-Gated CCDF Time-gated CCDF or Burst CCDF calculates the entire waveform using only the actual signal instead of the whole waveform. Gated function of the U2020XA Series USB power sensor allows the user to allocate the gated (marker 1 and 2) on specific burst on duration where the signal is computed as a time-gated CCDF measurement. 14
Instructions Keystrokes (Continue on from Demo 4) In the Power Analyzer –– Move the cursor to highlight the trace graph display. –– Right-click on the trace graph 1. Delete GSM waveform dual gated display and select “Apply Gate measurement Option” and “Gate 2” –– Gate 2 measurement will disap- pear and be removed from the trace display –– Click CCDF graph { } on the Power Analyzer toolbar. Select “Meas ID” and “Current Active Tab” 2. Create time-gated CCDF graph –– Click “OK” –– In CDDF Type Selector menu, click “Gated CCDF” –– Click “OK” Measure Time-Gated CCDF –– Move the cursor to highlight the Gated CCDF graph display. 3. Apply Gate 1 signal into Time-Gat- –– Right-click the graph display, ed CCDF graph (See Figure 15) select “Couple to Measurement Gate” and “Gate 1”. Measure Waveform CCDF –– Move the cursor to highlight the 4. Move Gate 1 (marker 1 and 2) to trace display graph. measure waveform CCDF graph –– Move the “Gate 1” dual gated (See Figure 16). (vertical line) to highlight the GSM whole waveform signal. 15
Figure 15. Gate 1 (blue lines on the trace display) press ON the GSM timeslot 0 duration, to compute the time-gated CCDF measurement shown in table format Figure 16. Gate 1 (blue lines on the trace display) press ON the entire waveform of GSM, including the burst off portion, and computes the waveform CCDF measurement in table format 16
Ordering Information For further information, refer to Keysight U2020 X-series USB Peak and Average Power Sensors Product Fact Sheet, Literature Number 5991-0308EN. Related Literature –– Keysight U2020 X-Series USB Peak and Power Sensors, Data Sheet, Literature Number 5991-0310EN –– Keysight N1918A Power Analysis Manager, Technical Overview, Literature Number 5989-6613EN –– Keysight N1918 Power Analysis Manager, Installation Guide, Manual Part Number N1918-90002 17
18 | Keysight | U220 X-Series USB Peak and Average Power Sensors - Demo Guide Evolving Since 1939 For more information on Keysight Technologies’ products, applications or Our unique combination of hardware, software, services, and people can help you services, please contact your local Keysight reach your next breakthrough. We are unlocking the future of technology. office. The complete list is available at: From Hewlett-Packard to Agilent to Keysight. www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 myKeysight Korea 080 769 0800 www.keysight.com/find/mykeysight Malaysia 1 800 888 848 A personalized view into the information most relevant to you. Singapore 1 800 375 8100 Taiwan 0800 047 866 http://www.keysight.com/find/emt_product_registration Other AP Countries (65) 6375 8100 Register your products to get up-to-date product information and find warranty information. Europe & Middle East Austria 0800 001122 Keysight Services Belgium 0800 58580 www.keysight.com/find/service Finland 0800 523252 Keysight Services can help from acquisition to renewal across your France 0805 980333 instrument’s lifecycle. Our comprehensive service offerings—one- Germany 0800 6270999 stop calibration, repair, asset management, technology refresh, Ireland 1800 832700 Israel 1 809 343051 consulting, training and more—helps you improve product quality Italy 800 599100 and lower costs. Luxembourg +32 800 58580 Netherlands 0800 0233200 Keysight Assurance Plans Russia 8800 5009286 www.keysight.com/find/AssurancePlans Spain 800 000154 Up to ten years of protection and no budgetary surprises to ensure Sweden 0200 882255 your instruments are operating to specification, so you can rely on Switzerland 0800 805353 accurate measurements. Opt. 1 (DE) Opt. 2 (FR) Keysight Channel Partners Opt. 3 (IT) www.keysight.com/find/channelpartners United Kingdom 0800 0260637 Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. For other unlisted countries: www.keysight.com/find/contactus (BP-9-7-17) www.keysight.com/find/usbpeaksensor DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System This information is subject to change without notice. © Keysight Technologies, 2017 Published in USA, December 1, 2017 5991-1221EN www.keysight.com
You can also read