Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide

Page created by Sharon Dominguez
 
CONTINUE READING
Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
Keysight Technologies
U2020 X-Series
USB Peak and Average Power Sensors

                                     Demo Guide
Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
Demonstration Preparation
                                              The instruments and software listed in the table below are required to perform
                                              the demonstrations.
                                               Product Type                            Minimum System Requirement
                                                                                           –– U2021XA or U2022XA
                                               Keysight U2020XA Series USB peak
                                                                                           –– Firmware version A.01.00 or
                                               and average power sensors
                                                                                              later
                                               Keysight USB 2.0 compliant cable,
                                               Mini-B connector with locking               –– U2031A, U2031B, U2031C
                                               mechanism
                                                                                           –– Firmware version A.01.80 or
                                                                                              later
                                               Keysight N5182A MXG Signal Vector           –– Frequency range option 503 or
                                               Analyzer                                       506
                                                                                           –– Baseband generator option 651
                                                                                              or 652
                                               Keysight IO Libraries Suite               Version 16.2.xx or later
                                               Keysight N1918A Power Analysis
                                                                                         Power Analyzer (R03.05.xx or later)
                                               Manager

                                              To download and update the instrument firmware or software, please visit:
                                              www.keysight.com/find/usbpeaksensor

                                              Instructions for PC, U2021XA, and
                                              MXG Connectivity
                                              Procedures
                                                1.    Connect the Laptop (installed with the N1918A Power Analysis Manager,
                                                      Power Analyzer) to the U2021XA USB power sensors via the USB
                                                      Mini-B cable.
                                                2.    Connect the U2021XA USB power sensors to the RF Output of the
                                                      N5182A MXG signal generator shown in Figure 1.
                                                3.    Connect the BNC-SMB cable from the N5182A MXG signal generator’s
                                                      rear panel Event 1 to the Trig In socket on the U2021XA USB power
                                                      sensor.
                                                4.    Launch the Keysight Connection Expert application (Keysight IO Libraries
                                                      Suite) and verify the connection of the U2021XA USB power sensors.

                                                          BNC-SMB cable
                               Event 1

                                                                                       Trig In

                                                                                      USB Mini-B
                 N5182A MXG Signal Generator                        U2021XA Series                           Laptop
                                                                                        cable
                                                                   USB Power Sensor
Figure 1. Laptop, U2021XA and MXG signal generator setup diagram

    1.    Alternatively, you can substitute the N5182A MXG with the E4438C ESG Vector Signal Generator, firmware
          revision C.04.98 or later, frequency range option 503 or 506, and baseband generator option 601 or 602

                                                                   2
Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
Keysight N1918A Power Analysis Manager
The Keysight Technologies, Inc. N1918A Power Analysis Manager, Power Analyzer is a Windows-based free license application
software for the U2020XA Series USB Peak and Average power sensors. This application software enables you to perform
advanced power analysis measurements in real time.

For more information on the N1918A Power Analysis Manager, refer to the Keysight N1918A Power Analysis Manager
Datasheet, Literature Number 5989-6612EN

                                                           3
Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
Demonstration 1
Average Power Measurement in Compact Mode Display

                               In this guide, keystrokes surrounded by [ ] represent front panel keys of the
         Objectives
                               instruments, keystrokes surrounded by { } represent softkeys of the instrument,
To demonstrate the capabil-    and keystrokes in bold represent the softkeys of the software.
ity of the Power Analyzer to
perform power measurement
and switch the measurement     Instructions                           Keystrokes
display to compact mode.       On the MXG
                               1.   Set instrument to its default
                                                                      Press [Preset]
                                    settings
                               2.   Set the frequency and ampli-
                                    tude                                –– Press [Frequency] [1] {GHz}
                                    Frequency = 1 GHz                   –– Press [Amplitude] [0] {dBm}
                                    Amplitude = 0 dBm
                               3.   Turn on the RF Output               –– Turn [RF On/Off] to “On”
                               In the Power Analyzer
                                                                      Double-click on the Power Analyzer
                                                                      application software shortcut on your
                                                                      desktop or access the program via the
                                                                      Windows Start menu.
                                                                      (Note: The latest version of Power
                                                                      Analyzer demonstrated in this demo
                               4.   Run the Power Analyzer            guide will start the measurement upon
                                                                      running of the
                                                                      application software)
                                                                        –– Click {channel} on the Instrument
                               5.   Set the frequency to 1 GHz             Navigator panel
                                    (See Figure 2)                      –– Set Frequency (Hz) to “1G” on the
                                                                           Instrument Properties panel
                                                                        –– Move the cursor to highlight
                               6.   Switch the normal mode dsplay          instrument measurement 1
                                    to compact mode (See Figure 3)      –– Click Switch {   } the Power
                                                                           Analyzer toolbar
                               7.   Switch the compact mode
                                                                      Click Switch {    } on the Power Ana-
                                    to normal mode display (soft
                                                                      lyzer
                                    panel)

                               8.   Stop the measurement              Click Stop {     } on the Power Analyzer

                                                 4
Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
Figure 2. Average power measurement shown in normal mode display (soft panel)

Figure 3. Compact mode display show the average power
measurement

                                                                        5
Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
Demonstration 2
Enable Alert, Limit, and Recording using Power Analyzer

           Objectives                   Instructions                          Keystrokes
                                        (Continue from Demo 1)
  –– To demonstrate the alert func-
                                        On the MXG
     tions where the system will
     notify you the exact time and      1.   Set amplitude power to 0.1 dBm Press [Amplitude], [0.1], {dBm}
     date the power measurements
     of your DUT has exceeded           In the Power Analyzer
     the specified upper and lower
     limits. Alert functions are es-                                           –– Click Create Strip Chart {     } on
                                                                                  Power Analyzer toolbar.
     pecially useful for applications   2.   Add strip chart display
     such as aging and reliability                                             –– Select “Meas ID” and “Current
     text where long test times are                                               Active Tab”, then click “OK”
     required.
  –– To demonstrate the strip chart                                            –– Click Stop {     } on the Power
     function in the Power Analyzer                                               Analyzer toolbar
     application software, which                                               –– Move the cursor to highlight
     allows the user to monitor, re-                                              normal mode display
     cord, and playback the power       3.   Delete the normal mode display
     measurements tend for a long                                              –– Click {    } to delete normal
     period of time.                                                              mode display

                                                                               –– Click Run {     } to continue
                                                                                  the measurement
                                                                               –– Click {Measurements} on the
                                        4.   Enable limit and alert of            Instrument Navigator panel
                                             measurement (See Figure 4)        –– In Instrument Properties panel,
                                             Upper Limit = 0.1 dB                 check “Enable Alert”
                                             Lower Limit = –0.1 dB             –– Set Upper Limit to “0.1” and Lower
                                                                                  Limit to “–0.1”

                                                                               –– Click Stop {        } on the Power
                                                                                  Analyzer toolbar
                                                                               –– Move the cursor to highlight
                                                                                  normal mode display
                                                                               –– Click Recorder {      } on the Power
                                        5.   Enable recording function            Analyzer toolbar.
                                             (See Figure 6)                    –– The Recorder menu will prompt
                                                                                  up, select specific time for
                                                                                  recording and enter the recording
                                                                                  file name. In this example, select
                                                                                  “Immediately” and enter the
                                                                                  filename as “Texting” then click
                                                                                  “OK”

                                                          6
Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
Figure 4. Alert Limits textbox allows user to enter upper/lower limit as per their requirements

                                                                    Figure 5. Click Alert Indicator {   } to view the alert summary as shown here in order
                                                                    to identify the timestamp where the measurement was over the alert limit

Figure 6. Recording file saved as *.csv can be
viewed via MS Excel
                                                                              7
Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
Demonstration 3                       Instructions                          Keystrokes
                                      On the MXG
Pulse Power Measurement               1.   Set instrument to its default
                                                                            Press [Preset]
                                           settings
          Objectives                  2.   Set the frequency and ampli-
                                           tude                              –– Press [Frequency] [1] {GHz}
 –– To demonstrate the capability          Frequency = 1 GHz                 –– Press [Amplitude] [0] {dBm}
    of the U2021XA Series USB              Amplitude = 0 dBm
    power sensors to obtain
                                                                             –– {Pulse Period}, [40], {μsec}
    15-point pulse parameter          3.   Set the pulse signal
                                                                             –– {Pulse Width}, [10], {μsec}
    values from the trace graph,
    where the parameters values       4.   Activate the pulse signal        Toglle {Pulse On/Off} to “On”
    are automatically                 5.   Turn on the RF Output            Toglle [RF On/Off] to “On”
    calculated by the Power           In the Power Analyzer
    Analyzer application software
                                                                            Double-click on Power Analyzer
    whenever a trace pattern
                                                                            application software shortcut on your
    appears on the graph.             6.   Run the Power Analyzer
                                                                            desktop or access the program via the
 –– To demonstrate the capability                                           Windows Start menu.
    of the dual gates feature to
                                                                             –– Click {channel} on the Instrument
    measure the pulse width                                                     Navigator panel.
                                      7.   Set the frequency to 1 GHz
    and pulse period signal
                                           (See Figure 7)                    –– Set Frequency (Hz) to “1G” on the
    respectively.
                                                                                Instrument Properties panel.
 –– In this demo, pulse signal with
    pulse period 40 μs and pulse
                                                                             –– Click {     } on the Power
    width 10 μs will be generated
                                      8.   Add trace graph display              Analyzer toolbar.
    from the signal generator.
                                                                             –– Select “Meas ID” and “Current
    The U2021XA Series USB
    power sensors measures the                                                  Active Tab” then click “OK”.
    average, peak, and peak-to-
                                                                             –– Click Stop {   } on the Power
    average of the pulse signal via
                                                                                Analyzer toolbar
    the dual gated feature.
                                                                             –– Move the cursor to highlight the
                                      9.   Delete the normal mode display       normal mode display and click {
                                                                                    } to delete the normal mode
                                                                                display.
                                                                             –– Click Run {  } to continue the
                                                                                measurement.
                                      Measure Pulse Width Signal
                                                                            Move the cursor to highlight the trace
                                                                            graph display. Right-click on the trace
                                      10. Add dual gates to measure
                                                                            graph display and select “Apply Gate
                                          pulse width signal, 10 μs (See
                                                                            Option”, Gate 1. Move the dual gated
                                          Figure 8)
                                                                            (vertical line) to the highlighted pulse
                                                                            width duration.
                                      Measure Pulse Period Signal
                                                                            Move the cursor to highlight the trace
                                                                            graph display. Right-click on the trace
                                      11. Add dual gates to measure
                                                                            graph display and select “Apply Gate
                                          pulse period signal, 40 μs (See
                                                                            Option”, Gate 2. Move the dual gated
                                          Figure 9)
                                                                            (vertical line) to the highlighted pulse
                                                                            width duration.
                                                                            In the Measurement Panel (bottom of
                                                                            the trace graph display), click “Mea-
                                      12. Add 15-point pulse parameter
                                                                            surement Selector”, then check “Select
                                          values from the trace graph
                                                                            All” under Pulse Measurement menu,
                                          display (See Figure 10)
                                                                            and click
                                                                            “Apply”.

                                                          8
Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
Figure 7. Average power measurement of pulse signal (pulse period 40 μs and pulse width 10 μs)

Figure 8. Pulse width measurement (delta time, average, peak and peak-to-average) blue line shown in Pulse Analysis table

                                                                           9
Keysight Technologies U2020 X-Series - USB Peak and Average Power Sensors Demo Guide
Demonstration 2

Figure 9. Pulse period measurement (delta time, average, peak and peak-to-average power) orange color shown in Pulse Analysis table (on top of pulse
width measurement

Figure 10. Pulse Analysis table consists of 15-point pulse parameter values.

                                                                               10
Demonstration 4
Burst Power Measurement

          Objectives                                                     4.613 ms timeframe duration
 –– To demonstrate the capability
    of U2021XA Series USB
    power sensors to measure
    burst power (average,
    peak and peak-to-average)
    signal via external/internal
    triggering methodology and
                                            577 μs timeslot duration
    automatically calculated by
    Power Analyzer application        Figure 11. GSM timeslot pattern with timeslot 0 on
    software whenever a burst
    waveform appears on the
                                       Instructions                                        Keystrokes
    graph.
 –– To demonstrate the capability      On the MXG
    of the dual gates function to      1.    Set instrument to its default
                                                                                           Press [Preset]
    measure the burst signal.                settings
 –– In this demo, GSM burst            2.    Set the frequency and ampli-
    signal with 4.613 ms duration            tude                                           –– Press [Frequency] [1] {GHz}
    (consists of 8 timeslots and             Frequency = 1 GHz                              –– Press [Amplitude] [0] {dBm}
    each timeslot with 577 μs                Amplitude = 0 dBm
    duration) will be generated                                                             –– Press [Mode], {Dual ARB},
    from the signal generator (See                                                             {Select Waveform}, {Waveform
    Figure 11).                        3.    Recall GSM modulation signal                      Segments}
 –– Using external triggering from                                                          –– Press [ ] or [ ] to select
    Event 1 of signal generator, to                                                            {GSM_BURST_WFM}
    synchronize with the U2021XA                                                            –– Press [Mode], {Dual ARB},
    USB power sensors via Trig                                                                 {Select Waveform}
    In port before starting to
                                                                                            –– Press [ ] or [ ] to highlight
    capture GSM waveform for           4.    Set GSM modulation signal                         {GSM_BURST_WFM}
    measuring the average, peak
                                                                                            –– Press {Select Waveform} and
    and peak-to-average power
    (see Figure 1).                                                                            toggle {ARB ON/Off} to “On”
                                                                                           Press [Mode], {Dual ARB} and [More],
                                       5.    Set external triggering for GSM
                                                                                           {Marker Utilities}, {Route Connectors},
                                             modulation signal
                                                                                           {Route to Event 1 BNC} and {Marker 2}
                                       6.    Turn on the RF Output                         Toggle [RF On/Off] to “On”

                                                            11
Instructions                           Keystrokes
In the Power Analyzer
                                       Double-click on the Power Analyzer
                                       application software shortcut on the
1.   Run the Power Analyzer
                                       desktop or access the program via the
                                       Windows Start menu
                                         –– Click {Channel} on the Instrument
                                            Navigator panel
2.   Set the frequency to 1 GHz
                                         –– Set Frequency (Hz) to “1G” on the
                                            Instrument Properties panel

                                         –– Click {     } on the Power
3.   Add trace graph display                Analyzer toolbar.
                                         –– Select “Meas ID” and “Current
                                            Active Tab” then click “OK”

                                         –– Press Stop {   } on the Power
                                            Analyzer toolbar.

                                         –– Move the cursor to highlight the
                                            normal mode display.
4.   Delete the normal mode display
                                         –– Click {  } to delete the normal
                                            mode display.

                                         –– Click Run {  } continue the
                                            measurement.
                                         –– Click {Channel} on the Instrument
5.   Configure trace display to cap-        Navigator panel
     ture GSM 4.613 ms waveform          –– Set Trace Start (s) to “–100μ” and
                                            Trace Length (s) to “5m”
                                         –– Click {Trigger}
6.   Configure power sensor to
                                         –– Set “Global Trigger Source” to
     external triggering mode
                                            “External”
Measure GSM Timeslot 0 Time-Gated Signal (Average, Peak, and Peak-to-Average
Power)
                                         –– Move the cursor to highlight the
                                            trace graph display
                                         –– Right-click on the trace graph
7.   Add dual gates to measure GSM          display and select “Apply Gate
     timeslot 0 duration (See Figure        Option”
     12).                                   and “Gate 1”
                                         –– Move the dual gated (vertical line)
                                            to highlight the GSM timeslot 0
                                            duration
Measure GSM Waveform Signal (Average, Peak, and Peak-to-Average Power)
                                         –– Move the cursor to highlight the
                                            trace graph display
                                         –– Right-click on the trace graph
8.   Add dual gates to measure GSM
                                            display and select “Apply Gate
     whole waveforms (See Figure
                                            Option”
     13).
                                            and “Gate 2”
                                         –– Move the dual gated (vertical line)
                                            to highlight the GSM waveform

                 12
Figure 12. Dual gate (Gate 1) press ON GSM timeslot 0, shows delta time, average power, peak power, peak to average power measurement

Figure 13. Dual gate (Gate 2) press ON GSM waveform signal 4.613 ms duration, shows delta time, average power, peak power, and peak to average
power measurement

                                                                         13
Demonstration 5
Time-Gated CCDF and Waveform CCDF Measurements

                                    Complementary Cumulative Distribution Function (CCDF)
         Objectives
 –– To demonstrate the              A CCDF curve is defined by how much time the waveform spends at or above a
    capability of U2021XA           given power level. This is expressed in dB relative to the average power.
    Series USB power sensors to     A CCDF curve is a plot of relative power levels versus probability where the
    measure and compute time-       X-axis represents the dB above the average signal power, while the Y-axis
    gated CCDF and waveform         represents the percent of time the signal spends at or above the power level
    CCDF measuremenet in trace      specified by the X-axis.
    display view via the N1918A
    Power Analysis Manager,
    Power Analyzer.
 –– To demonstrate the capability
    of the dual gates function to
    measure and computer the
    measurement in time-gated
    CCDF (Burst CCDF) and
    waveform CCDF.

                                    Figure 14. CCDF plot shows the Y-axis that represents the percentage of time the
                                    signal power equals the power specified by the X-axis

                                    Waveform CCDF
                                    Conventional peak power meter/sensors provide a built-in function of waveform
                                    CCDF measurement in graphical and table format. Waveform CCDF calculates
                                    CCDF and peak-to-average ratio using the entire waveform, including portions
                                    where the data is zero (no signal).

                                    Time-Gated CCDF
                                    Time-gated CCDF or Burst CCDF calculates the entire waveform using only the
                                    actual signal instead of the whole waveform. Gated function of the U2020XA
                                    Series USB power sensor allows the user to allocate the gated (marker 1 and
                                    2) on specific burst on duration where the signal is computed as a time-gated
                                    CCDF measurement.

                                                          14
Instructions                            Keystrokes
(Continue on from Demo 4)
In the Power Analyzer
                                         –– Move the cursor to highlight the
                                            trace graph display.
                                         –– Right-click on the trace graph
1. Delete GSM waveform dual gated           display and select “Apply Gate
   measurement                              Option” and “Gate 2”
                                         –– Gate 2 measurement will disap-
                                            pear and be removed from the
                                            trace display

                                         –– Click CCDF graph {     } on the
                                            Power Analyzer toolbar. Select
                                            “Meas ID” and “Current Active
                                            Tab”
2. Create time-gated CCDF graph
                                         –– Click “OK”
                                         –– In CDDF Type Selector menu, click
                                            “Gated CCDF”
                                         –– Click “OK”
Measure Time-Gated CCDF
                                         –– Move the cursor to highlight the
                                            Gated CCDF graph display.
3. Apply Gate 1 signal into Time-Gat-    –– Right-click the graph display,
   ed CCDF graph (See Figure 15)            select “Couple to Measurement
                                            Gate”
                                            and “Gate 1”.
Measure Waveform CCDF
                                         –– Move the cursor to highlight the
4. Move Gate 1 (marker 1 and 2) to          trace display graph.
   measure waveform CCDF graph           –– Move the “Gate 1” dual gated
   (See Figure 16).                         (vertical line) to highlight the GSM
                                            whole waveform signal.

                 15
Figure 15. Gate 1 (blue lines on the trace display) press ON the GSM timeslot 0 duration, to compute the time-gated CCDF measurement shown in table format

Figure 16. Gate 1 (blue lines on the trace display) press ON the entire waveform of GSM, including the burst off portion, and computes the waveform CCDF
measurement in table format

                                                                          16
Ordering Information
For further information, refer to Keysight U2020 X-series USB Peak
and Average Power Sensors Product Fact Sheet, Literature Number
5991-0308EN.

Related Literature
 –– Keysight U2020 X-Series USB Peak and Power Sensors, Data
    Sheet, Literature Number 5991-0310EN
 –– Keysight N1918A Power Analysis Manager, Technical Overview,
    Literature Number 5989-6613EN
 –– Keysight N1918 Power Analysis Manager, Installation Guide,
    Manual Part Number N1918-90002

                                                          17
18 | Keysight | U220 X-Series USB Peak and Average Power Sensors - Demo Guide

Evolving Since 1939                                                                         For more information on Keysight
                                                                                            Technologies’ products, applications or
Our unique combination of hardware, software, services, and people can help you             services, please contact your local Keysight
reach your next breakthrough. We are unlocking the future of technology.                    office. The complete list is available at:
From Hewlett-Packard to Agilent to Keysight.                                                www.keysight.com/find/contactus

                                                                                            Americas
                                                                                            Canada                              (877) 894 4414
                                                                                            Brazil                              55 11 3351 7010
                                                                                            Mexico                              001 800 254 2440
                                                                                            United States                       (800) 829 4444

                                                                                            Asia Pacific
                                                                                            Australia                           1 800 629 485
                                                                                            China                               800 810 0189
                                                                                            Hong Kong                           800 938 693
                                                                                            India                               1 800 11 2626
                                                                                            Japan                               0120 (421) 345
                      myKeysight                                                            Korea                               080 769 0800
                      www.keysight.com/find/mykeysight                                      Malaysia                            1 800 888 848
                      A personalized view into the information most relevant to you.        Singapore                           1 800 375 8100
                                                                                            Taiwan                              0800 047 866
                      http://www.keysight.com/find/emt_product_registration                 Other AP Countries                  (65) 6375 8100
                      Register your products to get up-to-date product information and
                      find warranty information.                                            Europe & Middle East
                                                                                            Austria                             0800 001122
                      Keysight Services                                                     Belgium                             0800 58580
                      www.keysight.com/find/service                                         Finland                             0800 523252
                      Keysight Services can help from acquisition to renewal across your    France                              0805 980333
                      instrument’s lifecycle. Our comprehensive service offerings—one-      Germany                             0800 6270999
                      stop calibration, repair, asset management, technology refresh,       Ireland                             1800 832700
                                                                                            Israel                              1 809 343051
                      consulting, training and more—helps you improve product quality
                                                                                            Italy                               800 599100
                      and lower costs.                                                      Luxembourg                          +32 800 58580
                                                                                            Netherlands                         0800 0233200
                      Keysight Assurance Plans                                              Russia                              8800 5009286
                      www.keysight.com/find/AssurancePlans                                  Spain                               800 000154
                      Up to ten years of protection and no budgetary surprises to ensure    Sweden                              0200 882255
                      your instruments are operating to specification, so you can rely on   Switzerland                         0800 805353
                      accurate measurements.                                                                                    Opt. 1 (DE)
                                                                                                                                Opt. 2 (FR)
                      Keysight Channel Partners                                                                                 Opt. 3 (IT)
                      www.keysight.com/find/channelpartners                                 United Kingdom                      0800 0260637
                      Get the best of both worlds: Keysight’s measurement expertise and
                      product breadth, combined with channel partner convenience.           For other unlisted countries:
                                                                                            www.keysight.com/find/contactus
                                                                                            (BP-9-7-17)

                      www.keysight.com/find/usbpeaksensor
                                                                                            DEKRA Certified
                                                                                            ISO9001 Quality Management System

                                                                                            www.keysight.com/go/quality
                                                                                            Keysight Technologies, Inc.
                                                                                            DEKRA Certified ISO 9001:2015
                                                                                            Quality Management System

                                                                                            This information is subject to change without notice.
                                                                                            © Keysight Technologies, 2017
                                                                                            Published in USA, December 1, 2017
                                                                                            5991-1221EN
                                                                                            www.keysight.com
You can also read