DRAWING CONTENTS C1 COVER SHEET GN GENERAL NOTES
←
→
Page content transcription
If your browser does not render page correctly, please read the page content below
FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. DRAWING CONTENTS C1 COVER SHEET GN GENERAL NOTES SECTION A A1 EQUIPMENT LAYOUT MEMORIAL HOSPITAL A2 MAGNETIC FIELD LAYOUT A3 DETAILS ST. FRANCIS A4 EQUIPMENT ELEVATIONS A5 EQUIPMENT ELEVATIONS SECTION S S1 STRUCTURAL LAYOUT / BASE DETAIL SECTION P P1 PLUMBING LAYOUT P2 WATER COOLING FACILITIES CONFIGURATION P3 PLUMBING / MEDICAL GAS DETAILS SECTION M M1 MECHANICAL LAYOUT M2 QUENCH PIPE DETAILS M3 QUENCH PIPE DETAILS M4 QUENCH PIPE DETAILS SECTION E E1 ELECTRICAL LAYOUT E2 ELECTRICAL SCHEMATIC E3 ELECTRICAL DETAILS E4 LINE FILTER PANEL / WALL CABINET DETAILS E5 POWER SYSTEM DIAGRAM / POWER QUALITY REQUIREMENTS E6 ELECTRICAL RISER DIAGRAM C1
MINIMUM SITE REQUIREMENTS CHECKLIST GENERAL NOTES HVAC REQUIREMENTS ELECTRICAL REQUIREMENTS FOR MRI SYSTEM WITH VRDU CUSTOMER/CONTRACTOR IS RESPONSIBLE FOR THE FOLLOWING UNLESS OTHERWISE NOTED. CUSTOMER TO PROVIDE THE NECESSARY HVAC REQUIREMENTS FOR THE TOSHIBA EQUIPMENT TO OPERATE PROPERLY. GENERAL ELECTRICAL NOTES CUSTOMER/CONTRACTOR IS RESPONSIBLE FOR THE FOLLOWING UNLESS OTHERWISE NOTED. FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. RF / MAGNETIC SHIELDING MEMORIAL HOSPITAL CODES AND PERMITS ST. FRANCIS NOISE SPECIFICATION SITE CONDITIONS NOTICE: CUSTOMER MUST COMPLETE ALL ITEMS ON THIS CHECKLIST BEFORE SCHEDULED DELIVER DATE FOR THE EQUIPMENT. IF CUSTOMER FAILS TO DO SO, DELIVERY MAY BE DELAYED. PLUMBING FURTHERMORE, THE EQUIPMENT WARRANTY MAY BE VOIDED. TRANSPORT REQUIREMENTS STRUCTURAL NOTES RF ROOM GROUNDING CUSTOMER/CONTRACTOR IS RESPONSIBLE FOR THE FOLLOWING UNLESS OTHERWISE NOTED. CUSTOMER/CONTRACTOR IS RESPONSIBLE FOR THE FOLLOWING UNLESS OTHERWISE NOTED. CEILING HEIGHT VIBRATION SPECIFICATION FLOOR LOADING PLUMBING NOTE GN
EQUIPMENT LEGEND FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. MEMORIAL HOSPITAL ST. FRANCIS EQUIPMENT LAYOUT ALL MATERIAL IN SCAN ROOM MUST BE NON-FERROUS SITE PLAN APPROVAL A1
MAGNETIC FIELD LAYOUT RF / MAGNETIC SHIELDING ST. FRANCIS MEMORIAL HOSPITAL A2 FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES.
FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. 1 FRINGE FIELD MEASUREMENTS (PLAN VIEW) 2 FRINGE FIELD MEASUREMENTS (ELEVATION VIEW) MEMORIAL HOSPITAL ST. FRANCIS FRONT SIDE MINIMUM CORRIDOR WIDTH 3 EFFECTS OF THE MAGNETIC FIELD 4 SERVICE CLEARANCE OF MAGNET 5 MAGNET ASSEMBLY FOR CARRYING IN 6 FOR MAGNET INGRESS A3
MAG GACC RFC PCH PLAN PLAN PLAN FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. FRONT SIDE FRONT SIDE SIDE FRONT GRADIENT POWER SUPPLY 1 3.0 TESLA MAGNET AND COUCH 2 AND CONTROL CABINET 3 RF CABINET TFR FPC1 FPC2 VRDU MEMORIAL HOSPITAL SHIM ST. FRANCIS PLAN PLAN PLAN PLAN FRONT SIDE FRONT SIDE FRONT SIDE FRONT SIDE FILTER PANEL COVER FILTER PANEL COVER VOLTAGE REGULATOR 4 TRANSFORMER CABINET 5 (SCAN ROOM SIDE) 6 (EQUIPMENT ROOM SIDE) 7 DISTRIBUTION UNIT A4
SUVS CON MFB RFG POSM HOSC PLAN PLAN PLAN PLAN FRONT SIDE FRONT SIDE FRONT SIDE FRONT SIDE FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. FRONT SIDE CONTROL CONSOLE PATIENT OBSERVATION SYSTEM 1 AND HOST CABINET 2 MAGNET FAN BOX 3 REFRIGERATOR CABINET 4 MONITOR 5 SUPERVISORY UNIT SWITCH OXMS OXMM ALRM IHE PLAN PLAN MEMORIAL HOSPITAL FRONT SIDE FRONT SIDE ST. FRANCIS FRONT SIDE OXYGEN SENSOR 6 AND OXYGEN MONITOR 7 ALARM BOX 8 INDOOR HEAT EXCHANGER A5
STRUCTURAL LEGEND ALL MATERIAL IN SCAN ROOM MUST BE NON-FERROUS FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. MEMORIAL HOSPITAL ST. FRANCIS STRUCTURAL LAYOUT 1 MAGNET FOOTPRINT AND COUCH LAYOUT S1
PLUMBING LEGEND ALL MATERIAL IN SCAN ROOM MUST BE NON-FERROUS FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. 1 SCAN ROOM SINK MEMORIAL HOSPITAL ST. FRANCIS PLUMBING LAYOUT P1
OUTSIDE EQUIPMENT ROOM REQUIREMENTS FOR COOLING WATER V7 CUSTOMER SUPPLIED CHILLER SYSTEM V9 V10 V8 Ο Ο Ο Ο Ο Ο V5 V6 Ο FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. Ο Ο Ο Ο Ο Ο Ο V4 ALL ITEMS OUTSIDE OF DASHED Ο Ο DRAIN AREA TO BE PROVIDED BY Ο CUSTOMER/CONTRACTOR Ο Ο INDOOR HEAT EXCHANGER FLOW SWITCH MEMORIAL HOSPITAL GRADIENT COIL ST. FRANCIS REFRIGERATOR F1 F2 V1 & V2 RF CABINET V3 - V11 BFP PR V1 V2 GRADIENT F1 POWER SUPPLY PR V3 CITY F2 BFP WATER IN WATER COOLING FACILITIES CONFIGURATION P2
MAGNET GANTRY FILTER MAGNET FAN BOX PANEL TRANSFORMER FLOW CABINET / SWITCH VACUUM PUMP CUSTOMER INDOOR HEAT SUPPLIED EXCHANGER REFRIGERATOR CHILLER AND MANIFOLD SYSTEM WATER FILTER FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. GRADIENT RF AMPLIFIER POWER SUPPLY SCAN ROOM SIDE COMPUTER ROOM SIDE 2 DEWAR - LHe REPLENISHMENT 3 LIQUID AND GAS HOSE CONNECTIONS EXAMPLE OF MEDICAL GAS THROUGH DUCT (PER CODE) MEMORIAL HOSPITAL ST. FRANCIS INSIDE SHIELD OUTSIDE SHIELD 1 TYPICAL SPRINKLER PENETRATION 4 TYPICAL MEDICAL GASES DETAIL (OPTIONAL) P3
MECHANICAL LEGEND ALL MATERIAL IN SCAN ROOM MUST BE NON-FERROUS ENTIRE EMERGENCY GHe DISCHARGE PIPE MUST BE MADE FROM STAINLESS STEEL OR ALUMINUM FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. MEMORIAL HOSPITAL ST. FRANCIS MECHANICAL LAYOUT M1
GENERAL QUENCH LINE DESIGN RULES CONTINUED QUENCH LINE DESIGN RULES CONTINUED QUENCH LINE OUTLET CONFIGURATIONS CONT. (FIG. 3) NOTE: THE ACOUSTIC DE-COUPLER SUPPLIED WITH MAGNET MUST BE INSTALLED QUENCH LINE CALCULATION NOMINALLY STRAIGHT, I.E., TO ALLOW FOR MINOR MISALIGNMENT. WARNING IT SHOULD TYPICALLY NOT EXCEED ±5mm, AND NEVER EXCEED THE DESIGN LIMIT OF ±20mm. KEEP AWAY. EXTREMELY COLD GAS MAY BE EXHAUSTED WITH NO WARNING. QUENCH LINE CALCULATION DO NOT WORK ON THE VENTING IF THE MAGNET IS AT FIELD. FIGURE 3: WARNING LABEL - QUENCH LINE OUTLET NO OTHER MATERIALS ARE PERMITTED SAFETY NOTES NOTE: WELDS MUST BE ONLY BE MADE BY COMPETENT WELDERS. (FIG. 16) WARNING: RISK OF ASPHYXIATION FAILURE TO OBSERVE THE FOLLOWING MAY RESULT IN DIZZINESS FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. AND LOSS OF CONSCIOUSNESS. -> DO NOT VENT HELIUM GAS DIRECTLY INTO THE MAGNET ROOM. -> DO NOT VENT EXHAUST GAS FROM THE QUENCH LINE INTO AN ENCLOSED SPACE. NOTE: THE AREA OF THE MESH MUST BE AT LEAST 2.5 TIMES THE CROSS-SECTION CALCULATOR QUENCH AREA OF THE QUENCH TUBE (FIG. 4-8). LINE ACCEPTANCE PROTOCOL QUENCH LINE CALCULATION NOTE: THE OPERATOR OF THE MRI SYSTEM MUST PREPARE AN "EMERGENCY PLAN". NOTE: THE DESIGN AND CONSTRUCTION OF THE QUENCH LINE MUST BE (FIG. 4) DOCUMENTED IN DRAWINGS AND CALCULATIONS, AND KEPT WITH THE NOTE: ONLY THE EXAMPLES OF THE TUBE IN THIS GUIDE MAY BE USED. "SYSTEM OWNER MANUAL". NOTE: PLANNING AND INSTALLATION OF QUENCH LINES MUST BE CONDUCTED FIGURE 6 - QUENCH LINE OUTLET TO ATMOSPHERE - HORIZONTAL OPTION 2 BY QUALIFIED PERSONNEL, AS APPOINTED BY THE PROJECT MANAGER. NOTE: COMPONENTS USED FOR OTHER TUBING, E.G., IN AIR CONDITIONING OR (FIG. 4) QUENCH LINE OUTLET CONFIGURATIONS ROOM VENTING, ARE GENERALLY NOT SUITABLE FOR QUENCH LINE CONSTRUCTION. (FIG. 5-8) NOTE: THE QUENCH LINE MUST BE IDENTIFIED WITH A MARKER TAPE ALONG THE COMPLETE LENGTH OF THE QUENCH LINE. THE CONTENT COULD CAUTION: ONLY USE THE QUENCH LINE OUTLET CONFIGURATIONS DESCRIBED BE, E.G.: "DO NOT CUT, QUENCH LINE EXHAUST LINE". IN THIS CHAPTER. IF THE QUENCH LINE IS NOT CONFIGURED CORRECTLY, SAFETY IS COMPROMISED. NOTE: NEW CUSTOMER SITES MUST HAVE THE QUENCH LINE INSTALLED AND AVAILABLE FOR IMMEDIATE USE BEFORE THE MAGNET SYSTEM ARRIVES TO ALLOW SUITABLE VENTING FOR THE MAGNET DURING INSTALLATION. (FIG. 2, 3, & 9) QUENCH LINE DESIGN RULES NOTE: WHERE WINDOWS ARE WITHIN THE RESTRICTED ACCESS AREA, THEY MUST BE SEALED AND PERMANENTLY CLOSED. MEANS OF OPENING THE WINDOWS MUST BE REMOVED. DESIGN CALCULATIONS 100 MBAR MEMORIAL HOSPITAL 450 MBAR (FIG. 11) ST. FRANCIS (FIG. 10-12) (FIG. 13) (FIG.13) (FIG. 4-8) CAUTION: NEVER REDUCE THE DIAMETER OF A TUBE IN THE DOWNSTREAM DIRECTION. THIS WILL CAUSE SHOCKWAVE REFLECTIONS AND LOCAL PRESSURE PEAKS WHICH MAY DAMAGE THE MAGNET. NON-MAGNETIC STAINLESS STEEL OR FIGURE 7 - QUENCH LINE OUTLET TO ATMOSPHERE - HORIZONTAL OPTION 3 ALUMINUM NOTE: THE QUENCH LINE COMPONENTS OTHER THAN THOSE SUPPLIED WITH THE FIGURE 4 - QUENCH LINE OUTLET TO ATMOSPHERE - VERTICAL MRI SYSTEM (I.E. MAGNET ELBOW, HORIZONTAL ADAPTER, AND ACOUSTIC DE-COUPLER) MUST ONLY BE MADE OF STAINLESS STEEL OR ALUMINUM. MINIMUM OF 0.7mm (22 S.W.G) SQUARE SECTION IS NOT ALLOWED ACOUSTIC DE-COUPLER SUPPLIED BY NOTE: THE QUENCH LINE INSULATION MUST EXTEND UP TO THE QUENCH VALVE CUSTOMER/CONTRACTOR ARE TO BE FITTED AT A MINIMUM OF EVERY 10 METERS ON THE MAGNET. NOTE: THERE MUST BE CLEARANCE BETWEEN THE FINISHED INSULATION AND THE MAGNET COVERS. NOTE: FLEXIBLE LINES AND ACOUSTIC DE-COUPLERS SUPPLIED BY CUSTOMER/CONTRACTOR MAY ONLY BE MADE OF STAINLESS STEEL OR ALUMINUM. (FIG. 2) WARNING VENT PIPE FOR EXTREMELY COLD HELIUM GAS ONLY AUTHORIZED PERSONNEL TO WORK ON VENT PIPE FIGURE 2: WARNING LABEL - QUENCH LINE ENSURE THAT THE QUENCH VALVE DOES NOT CARRY ANY LOAD FROM THE WEIGHT OF THE QUENCH LINE FIGURE 5 - QUENCH LINE OUTLET TO ATMOSPHERE - HORIZONTAL OPTION 1 FIGURE 8 - QUENCH LINE OUTLET TO ATMOSPHERE - HORIZONTAL OPTION 4 M2
QUENCH LINE OUTLET CONFIGURATIONS CONT. QUENCH LINE WATER DRAINS (AS REQUIRED) RF WAVEGUIDE QUENCH LINE CONFIGURATIONS CONTINUED SECTIONS WHERE WATER MIGHT COLLECT ARE TO BE AVOIDED, BUT IF THEY CANNOT BE AVOIDED, AN AUTOMATIC WATER DRAIN MUST BE FITTED AT THE BOTTOM OF THE SECTION. SEE (FIG. 14 & 15) FOR DIMENSIONS AND THE METHOD OF FITTING TO THE QUENCH PIPE. A FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. FIGURE 14 - DRAWING OF AUTOMATIC WATER DRAIN (SCALE: 3:1) FIGURE 9 - EXAMPLE OF QUENCH LINE OUTLET POSITION REQUIREMENTS TO THE OUTSIDE (NOT TO SCALE) QUENCH LINE ELBOWS FIG. 20 FIG. 20 FIGURE 15a - WATER DRAIN INSTALLATION (SCALE: 2:1) MEMORIAL HOSPITAL FIGURE 17 - RF WAVEGUIDE FIGURE 19 - EXAMPLE OF VERTICAL ACOUSTIC DE-COUPLERS AND CABIN FEED-THROUGH ST. FRANCIS QUENCH LINE CONFIGURATIONS ACOUSTIC DE-COUPLER FIGURE 10 - QUENCH LINE SMOOTH ELBOW FIGURE 11 - QUENCH LINE SEGMENTED ELBOW FIGURE 15b - WATER DRAIN INSTALLATION (NOT TO SCALE) DIFFUSER ACOUSTIC DE-COUPLER GALVANIC SEPARATION FIG. 20 FIG. 20 FIGURE 13 - DIFFUSER FIGURE 16 - GALVANIC JOINT AT ACOUSTIC DE-COUPLER (NOT TO SCALE) FIGURE 18 - EXAMPLE OF HORIZONTAL ACOUSTIC DE-COUPLERS AND CABIN FEED-THROUGH FIGURE 20 - ACOUSTIC DE-COUPLER FOR 3.0 TESLA MAGNETS M3
QUENCH LINE CONDENSATION QUENCH LINE AND EMERGENCY VENTILATION FAN FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. ³ FIGURE 21 - QUENCH LINE CONDENSATION ( NOT TO SCALE) MEMORIAL HOSPITAL ST. FRANCIS SIDE VIEW FRONT VIEW M4
ELECTRICAL LEGEND FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. MEMORIAL HOSPITAL ST. FRANCIS ELECTRICAL LAYOUT ELECTRICAL DUCT LEGEND SCAN ROOM EMERGENCY LIGHTS E1 ALL MATERIAL IN SCAN ROOM MUST BE NON-FERROUS
CONDUIT SCHEDULE CONTRACTOR CONDUIT REFERENCE CABLE REFERENCE RUN CONDUIT CONDUIT CONDUIT CONDUIT CABLE CABLE LENGTH CABLES NO. (POINT TO POINT) (ROUTING) (DIAMETER) (MAX. LENGTH) (POINT TO POINT) (USABLE) (SUPPLIED BY) FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. MEMORIAL HOSPITAL ST. FRANCIS NOTE: A. CONDUITS SUPPLIED/INSTALLED BY CUSTOMER/CONTRACTOR. B. ALL CONDUIT RUNS MUST TAKE THE SHORTEST MOST DIRECT ROUTE POSSIBLE. ELECTRICAL LAYOUT C. CONDUITS MAY HAVE A MAXIMUM OF (3) 90° BENDS. D. CONDUIT IS NOT TO BE RUN IN SUCH A MANNER THAT WILL ALLOW CABLE POINT TO POINT LENGTHS TO BE EXCEEDED AS SHOWN IN CONDUIT LEGEND. (PROVIDED FOR REFERENCE PURPOSES ONLY) E. "EPO" CONNECTS TO "VRDU" SHUNT TRIP CIRCUIT BREAKER WITHIN HOSPITAL FACILITY DISTRIBUTION PANEL (SEE SHEET E5, DETAIL 1). F. CUSTOMER SUPPLIED CABLES INTO SCAN ROOM MUST RUN THROUGH CUSTOMER/CONTRACTOR SUPPLIED RF FILTERS. REFER TO DETAIL #1, SHEET E5. G. ALL GROUND WIRES NEED TO BE INSULATED/ISOLATED. H. CONTRACTOR TO PROVIDE PULL STRINGS IN EACH CONDUIT. ALL MATERIAL IN SCAN ROOM MUST BE NON-FERROUS E2
FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. 1 CONNECTION OF "OXMM" & "OXMS" 2 TYPICAL CABLE LADDER ELEVATION 3 "VRDU" ELECTRICAL ELEVATION MEMORIAL HOSPITAL ST. FRANCIS TYPICAL WALL DUCT DETAIL WITH 4 TYPICAL SCAN ROOM GROUNDING 5 FLOOR DUCT / J-BOX / VERTICAL RISER 6 PATIENT MONITORING CAMERA E3
PLAN FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. ELEVATION ELEVATION ASSEMBLY ISOMETRIC (SCAN ROOM SIDE) SIDE (EQUIPMENT ROOM SIDE) (FOR REFERENCE ONLY) INSTALLATION SURFACE 1 TYPICAL INSTALLATION AREA OF THE FILTER PANEL 2 OF THE FILTER PANEL MEMORIAL HOSPITAL ST. FRANCIS AIR HOSE WAVE GUIDE (SCAN ROOM SIDE) WAVE GUIDE HOLES (SCAN ROOM SIDE) SIDE VIEW 3 TYPICAL FILTER PANEL WAVE GUIDES 4 AIR HOSE WAVE GUIDE E4
POWER QUALITY REQUIREMENTS TITAN SCANNER WITH VRDU ³ FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. STANDARD POWER QUALITY NOTES MEMORIAL HOSPITAL ST. FRANCIS 1 POWER SYSTEM DIAGRAM 2 POWER REQUIREMENTS WITH VRDU E5
SPECIAL RUN POINT POINT CABLE NO. USABLE MAX PART NUMBER* TYPE* BSM43-1014-3 BSM43-1184-3 SPECIAL BSM43-1245-3 CM49-05567-1 CM49-05565-1 SPECIAL CM43-08177-1 SPECIAL SM49-06196-1 CM49-05585-3 SPECIAL CM49-05625-3 CM49-02353-1 SPECIAL CM40-01876-1 FAN FAN DRIVE DISTRIBUTION SWITCH RELAY BOX PANEL BOX OXYGEN SENSOR OXYGEN MONITOR SPEAKER EMERGENCY DISCHARGE FAN LCD MONITOR PATIENT MONITOR CONTROL ROOM FOR REFERENCE ONLY. NOT TO BE USED FOR CONSTRUCTION PURPOSES. CONTROL PAD HOST ALARM CABINET CONTROL BOX SCAN ROOM DOOR SWITCH SPECIAL SPECIAL SPECIAL PM49-06204-1 SPECIAL SPECIAL SPECIAL MAGNET MAGNET FAN BOX RF CABINET SHIM MEMORIAL HOSPITAL COUCH AMP GRADIENT POWER SUPPLY / ST. FRANCIS CONTROL CABINET POWER SYSTEM SUPERVISORY SPEAKER TRANSFORMER UNIT SWITCH CABINET / INNER VACUUM PUMP VISION FLOW SWITCH SPECIAL EQUIPMENT CAMERA 1 REFRIGERATOR ROOM LINE FILTER PANEL LINE FILTER PANEL (SCAN ROOM) (EQ. ROOM) 1 CABLE RISER DIAGRAM E6
You can also read