A Theoretical Framework for Performance Characterization of Elastography: The Strain Filter

Page created by Bobby Fletcher
 
CONTINUE READING
164                 ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 44, no. 1, january 1997

       A Theoretical Framework for Performance
          Characterization of Elastography:
                   The Strain Filter
                                   Tomy Varghese, Member, IEEE, and Jonathan Ophir

   Abstract—This paper presents a theoretical framework
for performance characterization in strain estimation, which
includes the effect of signal decorrelation, quantization er-
rors due to the finite temporal sampling rate, and electronic
noise. An upper bound on the performance of the strain
estimator in elastography is obtained from a strain filter
constructed using these limits. The strain filter is a term
used to describe the nonlinear filtering process in the strain
domain (due to the ultrasound system and signal process-
ing parameters) that allows the elastographic depiction of
a limited range of strains from the compressed tissue. The
strain filter predicts the elastogram quality by specifying
the elastographic signal-to-noise ratio (SN Re ), sensitivity,
and the strain dynamic range at a given resolution. The
dynamic range is limited by decorrelation errors for large
tissue strain values, and electronic noise for low strain val-
ues. Tradeoffs between different techniques used to enhance           Fig. 1. The block diagram of the strain filter, indicating the filtering
elastogram image quality may also be analyzed using the               of the tissue strains by the strain filter to predict the dynamic range
strain filter.                                                        and respective SN Re at a given resolution in the elastogram. The
                                                                      tissue strain measured using the ultrasound system is obtained by a
                                                                      quasi-static tissue compression restricted by the mechanical bound-
                                                                      ary conditions. The contributions of the signal processing and ultra-
                      I. Introduction
                                                                      sound system parameters are indicated as inputs into the strain filter.
                                                                      Improvements in estimator performance due to other techniques are
     ltrasonic techniques for measuring the elasticity
U    of compliant tissue generally rely on the estimation of
the strain [1]–[5]. Elastography, a technique of estimating
                                                                      explained by the enhancement in the correlation coefficient (reduced
                                                                      signal decorrelation).

the axial strain using differential displacements of the tis-
sue elements due to tissue compression, was proposed by               range of elastic moduli. A distribution of strains in the
Ophir et al. [1].                                                     medium is caused by the quasi-static compression under
   This paper introduces the strain filter concept for the            certain mechanical boundary conditions. The input to the
performance characterization of the strain estimator in               strain filter is the actual tissue strain characterized by
elastography. The behavior of the upper bound of the                  an infinite dynamic range and SN Re , fine (denoted by
SN Re as a function of axial tissue strain forms a band-               where epsilon is a very small number) resolution and
pass filter in the strain domain. The strain domain refers            sensitivity. However, the interaction of the actual tissue
to the entire range of strains present in the compressed tis-         strain with the ultrasound system and signal processing
sue. The strain filter consists of a graphical and analytical         parameters corrupts the elastogram obtained, in the sense
representation of the allowable range of strain values and            that it now has a finite dynamic range, SN Re , sensitivity,
their resulting SN Re for a given elastographic resolution.           and resolution. The strain filter predicts the elastographic
The width of the strain filter specifies the dynamic range,           image quality (SN Re , sensitivity, and dynamic range at a
and its height the respective SN Re value of the estimated            given resolution) in terms of the signal processing and sys-
strain.                                                               tem parameters used to obtain the elastogram. The strain
   A block diagram elucidating the strain filter concept              filter therefore provides a quantitative assessment of the
is presented in Fig. 1. Elastography uses small mechani-              elastogram quality, in terms of the four parameters de-
cal compressions on soft biological tissue that have a wide           scribed above.
                                                                          In addition, the strain filter makes it possible to select
  Manuscript received November 27, 1995; accepted July 15, 1996.
This work was supported in part by NIH grants R01-CA38515, R01-       the appropriate ultrasound system and signal processing
CA60520, and P01-CA64579, and by a grant from Diasonics Corp.,        parameters to obtain the best possible elastogram under
Santa Clara, CA.                                                      given tissue conditions. In subsequent sections of this pa-
  T. Varghese and J. Ophir are with the Ultrasonics Laboratory, De-
partment of Radiology, University of Texas Medical School, Houston,   per we discuss the effects of the various ultrasound sys-
TX 77030 (e-mail: tomyv@msrad3.med.uth.tmc.edu).                      tem parameters such as pulse center frequency and band-

                                                   0885–3010/97$05.00 c 1997 IEEE
varghese and ophir: a theoretical famework for performance characterization of elastography                              165

width, and signal processing parameters (window size and         the CRLB at poor SNRs (
166                ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 44, no. 1, january 1997

   In this paper, the strain variance is computed from the        the total tissue strain (st ) and the lower bound on the
TDE variance using the expression derived by Walker and           strain estimation standard deviation (σZZLB ) are substi-
Trahey [17], as long as the ZZLB coincides with the CRLB.         tuted in (3):
The decay in the value of the correlation coefficient with                                             st
tissue compression has been modeled using an analytic ex-                            SN ReU B =              .            (4)
                                                                                                  σZZLB
pression derived by Meunier and Bertrand in [19] to esti-            Incorporating the modified ZZLB expression for the
mate decorrelation effects due to rotation, translation, and      TDE variance (see Appendix B) into σZZLB using (2),
biaxial deformation of the elastic tissue elements. The the-      we obtain:
oretical development of the strain filter is presented in the
                                                                                    (sT )2
next section.                                                                      
                                                                                        ∆t ,            BT SN RC < γ 0
                                                                                   
                                                                                    T6Threshold
                                                                                   
                                                                                    2            γ < BT SN RC < δ 0
                                                                                                    0

                                                                         σZZLB ≥
                                                                          2           2σBB
                                                                                       T ∆t ,     δ 0 < BT SN RC < µ0     (5)
      III. Development of the Strain Filter                                        
                                                                                                    0                  0
                                                                                   
                                                                                    T  hreshold  µ    <  BT SN R   < η
                                                                                   
                                                                                    2σCRLB
                                                                                         2
                                                                                                                  C
                                                                                                    0
   Axial strain s is the displacement gradient, which may                               T ∆t  ,   η    <  BT SN RC

be estimated from two adjacent time delay estimates sepa-
                                                                  where η 0 , µ0 , δ 0 , and γ 0 are the modified thresholds (B-
rated by a time interval ∆t [1], assuming a constant speed
                                                                  2) scaled by the factor T 2∆t . (5) shows the three distinct
of sound in the tissue, viz:                                                                   2
                                                                  operating regions for σZZLB        , depending on the value of
                             τ2 − τ1                              BT SN RC . A distinct threshold region is observed between
                        s=           ,                     (1)
                               ∆t                                 the CRLB and the Barankin bound; however, the variance
                                                                  increases exponentially in this threshold region (see Ap-
where τ1 is the time delay estimate at time t, and τ2 is the      pendix B). Accurate estimation of the strain is possible
time delay estimate at time t + ∆t for a data window with         only within the CRLB.
duration T .                                                          The minimum variance of the time delay estimator is
   Since the strain estimate is obtained from a linear com-       given by the CRLB [13]–[17]. The CRLB for time delay
bination of two random variables (time delay estimates            estimation has been adapted for partially correlated sig-
separated by ∆t), the variance of the strain estimator de-        nals by Walker and Trahey in [17], and is given by:
pends on the variance of the time delay estimator. Assum-
                                                                                                                                
ing stationarity, the variance of the strain estimate (σS2 ) is                             3              1          1
                                                                       2
                                                                     σCRLB = 2∼                               (1 +         ) −1
                                                                                                                            2
expressed in terms of the variance of the time delay esti-                       2π T (B 3 + 12Bfo2 ) ρ2           SN R2
mates (σt2 ) in [20], and is given by:                                                                                            (6)
                                                                  where fo is the center frequency, B is the bandwidth, ρ
                                2στ2
                        σS2 ≥        .                     (2)    is the correlation coefficient, and the SNR term represents
                                T ∆t                              only the contribution due to electronic noise (SN RS ). This
(2) illustrates that for a given window size and overlap,         closed-form expression obtained for signals with a rectan-
strain variance is reduced when the variance in the time          gular spectrum assumes that the variance of the time delay
                                                                  estimate is bounded by the CRLB. The Barankin bound
delay estimate is minimized. The resolution in the elas-                                                      2
togram is reduced with an increase in T . Large overlap-          exceeds the CRLB by a factor of 12 fB0            [13],[14], and is
ping windows also generate correlated errors which bias           given by:
the strain estimate. In addition, an optimal window size                                           2
exists, where the strain estimation variance is minimum,                                  2         fo      2
                                                                                        σBB = 12          σCRLB                   (7)
with an increase in the variance observed as the window                                             B
size is increased or decreased [6, pp. 122, 21]. Therefore            The lower bound on the variance of the strain estimate
the strain estimation variance cannot simply be reduced              2
                                                                  (σZZLB   ) is obtained by substituting the bound on the vari-
in the limit as T → ∞, since στ2 also increases.                  ance of the TDE obtained using (6) and (7) into (5).
                                                                      The variation of SN ReU B (4) with tissue strain is de-
A. The Strain Filter in Elastography                              fined as the strain filter. Three distinct regions constitute
                                                                  the strain filter, which depends on the appropriate lower
    A measure of elastographic image quality was described                                      2
                                                                  bound that contributes to σZZLB      (5).
[6],[7] in terms of the mean to standard deviation ratio
(SN Re ) of the elastogram:                                       B. Effect of Decorrelation Due to Strain
                                  µs                              on the Correlation Coefficient
                       SN Re =       ,                     (3)
                                  σs
                                                                     Decorrelation errors increase with tissue strain, causing
where µs and σs are, respectively, the mean and standard          a decay in the values of the correlation coefficient. The cor-
deviation of the strain estimates in a region of uniform          relation coefficient with motion compensation due to ax-
elasticity. The upper bound of the SN Re is obtained when         ial deformation of elastic tissue for a 2-D Gaussian model
varghese and ophir: a theoretical famework for performance characterization of elastography                                        167

has been derived by Meunier and Bertrand in [19], and is
given by:
                     √                      2 (α−1)2
                    2 αβ           −1 f
          ρ= p                    e 2 ( σu ) α2 +1   (8)
               2(α2 + 1)(β 2 + 1)

where f is the spatial frequency in cycles/mm (f = 2fco
where c is the speed of sound in tissue = 1.54 mm/s), and
σf is the standard deviation of the Gaussian envelope in
                   2σ                1
cycles/mm (σu = cf , and σf = 2πσ      t
                                         , where σt is the spa-
tial standard deviation and σf is the standard deviation of
the Gaussian envelope in the frequency domain), α repre-
sents the axial compression where α = 1 − s, and s is the
tissue strain. The corresponding lateral expansion is de-
noted by β (with the incompressibility constraint αβ = 1).
The value of the correlation coefficient is substituted in the
expression for the CRLB (6) to model decorrelation effects.
   Expressions for the CRLB [13]–[17] have been derived
for flat bandlimited signal and noise spectra. However, the
                                                                  Fig. 2. The correlation coefficient and the standard deviation of the
correlation coefficient in [19] is derived for a Gaussian-        strain estimate for increasing strain values. The standard deviation
shaped spectrum rather than a rectangular spectrum. A             of the strain estimate has been normalized by the standard deviation
reasonable approximation was obtained by Céspedes et al.         value at 8% strain of 0.0058 µs to enable plotting both the standard
[11] using a rectangular spectrum centered at the Gaus-           deviation of the strain estimate and the correlation coefficient on the
sian center frequency with the same mean square ampli-            same graph.
tude value as the Gaussian spectrum. The equivalent noise
spectral bandwidth [22, pp. 141] is defined by:                   with the Barankin bound, the performance of the strain
                      R∞                                          filter drops sharply, with a further drop in performance
                          P (f ) df    √
                B= 0                = 2πσ f                 (9)   observed as the variance coincides with the constant vari-
                       P (f ) |max                                ance level. Note that the strain filter obtained using the
where B is the bandwidth of a rectangular spectrum with           ZZLB has a smaller dynamic range than the strain filter
the same total power and peak amplitude as the Gaussian           obtained under the optimistic assumption that the strain
pulse spectrum P (f ).                                            estimation variance is always bounded by the CRLB.
    Using the following typical signal parameters, T = 1              The range of strains that can be reliably estimated us-
mm (1.3 µs), fo = 5 MHz, B = 3 MHz (60% bandwidth),               ing the elastogram determines the dynamic range of the
SN RS = 100 (40 dB), and interval between strain esti-            strain filter. The dynamic range of the strain estimator in
mates ∆z = 0.5 mm (∆t = 0.66 µs). The lower bound                 decibels is defined as follows:
on the standard deviation of the strain estimator (normal-                                              
                                                                                                    smax
ized by its value at 8% strain) is plotted along with the                            DR = 20 log           ,             (10)
                                                                                                    smin
corresponding correlation coefficient in Fig. 2, for increas-
ing strain values. Note from Fig. 2 that the lower bound          where smax is the maximum strain and smin is the mini-
on the standard deviation (σZZLB ) increases dramatically         mum strain at a specified SN Re level in the strain filter.
for strain values >0.5%. This has been previously noted           The quantity smin also defines the sensitivity of the strain
by O’Donnell et al. [5] and Céspedes [6, pp. 115]. They          filter. A 1% tissue strain corresponds to a decibel value of
show that the standard deviation of the strain for a 1-D          −40 dB. The dynamic range estimated for a −3 dB cutoff
simulation increases dramatically due to decorrelation as-        level of the strain filter (SN Re level of 10) predicts a 30 dB
sociated with large strain values (>3%) for a fixed corre-        dynamic range (observed from Fig. 3) for the strain filter
lation integration time. For the same reason, past work in        obtained using the ZZLB bound, compared to the 40 dB
elastography has used small (≤ 2%) applied strains [1],[2].       dynamic range predicted using the CRLB. However, the
    Decorrelation errors increase with tissue strain, reduc-      dynamic range obtained experimentally for a single com-
ing the value of the correlation coefficient, and causing         pression agrees closely with the dynamic range predicted
  2
σZZLB    to move from the CRLB to the Barankin bound              using the ZZLB.
or the constant variance level as shown in (5). The three             The slope of the strain filter for low strain values is de-
distinct regions in (5) are observed in plots of the strain       termined primarily by electronic noise contributions, with
filter as shown by the curves in Fig. 3. Figure 3 shows           decorrelation noise contributing to the progressive flatten-
the strain filter obtained for a 3 MHz rectangular band-          ing of the curve. The plateau in the strain filter is caused
width using the ZZLB (CRLB for strains ≤10%, Barankin             by increasing decorrelation errors; however, the variance
bound for strains >10% and 30%). As the lower bound coincides        this region. At high strain values, the precipitous drop in
168                   ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 44, no. 1, january 1997

Fig. 3. The strain filter illustrating the distinct regions of strain es-   Fig. 4. A group of three strain filters, showing the distinct regions of
timation obtained using (5), along with the strain filter obtained          strain estimation. These filters correspond to bandwidths of 2.5 MHz,
under the optimistic assumption that the strain estimation variance         3 MHz, and 3.5 MHz, respectively, with a 5 MHz center frequency
is always bounded by the CRLB adapted for partially correlated              and T = 1.3 µs. Note the improvement in SN Re and dynamic range
signals. Note that the dynamic range of elastography may be deter-          obtained with an increase in the signal bandwidth.
mined from the width of the strain filter, and the heights give the
respective SN Re at every strain level.
                                                                            resolution levels. In general, as resolution increases, the
                                                                            maximum value of the SN Re and the dynamic range of
performance is caused by decorrelation noise, which deter-                  the strain filter decrease. The dynamic range, sensitivity,
mines the slope of the graph in this region. Improvements                   and SN Re , along with the resolution, provide a complete
in the dynamic range can be obtained either by reduc-                       characterization of the noise properties in the performance
ing electronic noise errors for low values of the strain, or                of the strain estimator in elastography.
reducing the rate of decorrelation for large strain values.                    Simulation experiments in the next section are used to
Decorrelation errors can be reduced using a combination                     confirm the validity of the theoretical strain filter model
of small compressions that allow successful step-wise tem-                  developed in this section.
poral stretching. For example, temporal stretching of the
post-compression signal [6]–[8], MA [5],[9], and [10], and
SMA [9],[10] improves the dynamic range and SN Re of                                                 IV. Simulation
strain estimation without sacrificing resolution.
    The shape of the strain filter also depends on the vari-                   A simulation experiment to quantify the performance of
ation of T , f0 , and B terms in the denominator of (6). For                the strain estimator for a 1-D tissue mechanical model is
example, the performance of the strain filter for different                 presented in this section. The 1-D model used in the simu-
values of the bandwidth is shown in Fig. 4. With an in-                     lation accounts only for the decorrelation of the echo-signal
crease in the system bandwidth, there is an improvement                     due to the axial component of the strain. The theoretical
in the maximum SN Re value, sensitivity, and dynamic                        model can be adapted for the 1-D case by setting β = 1
range of the strain filter. In other words, signal decor-                   in (8). In the simulation of the ultrasound system, a 1-D
relation reduces with an increase in the bandwidth. The                     sampled array is used to insonify a 2-D point scatterer
strain filters obtained at different center frequencies are                 medium.
presented in Fig. 5. Note the improvement in the sensitiv-
ity, dynamic range, and SN Re obtained with increasing                      A. Method
pulse center frequency. The maximum attainable SN Re
is approximately proportional to the square of the center                      The pre- and post-compression echo-signals are gener-
frequency.                                                                  ated using a transducer with a center frequency of 5 MHz
    The resolution of elastography depends on T and ∆t,                     and −3 dB band-width of 3 MHz (pulse standard devi-
and is limited only by the correlation length of the ultra-                 ation of 0.27 µs). The A-scan was sampled at 50 MHz.
sound system. The resolution in the elastogram improves                     Cross-correlation analysis was performed using a 1 mm
with a decrease in the value of T or an increase in sys-                    (1.3 µs) overlapping window with 0.5 mm (0.66 µs) over-
tem bandwidth. All the performance plots of the strain                      lap between consecutive windows.
filter are plotted for a constant value of the resolution. A                   The transducer is modeled as a 1-D sampled aperture
family of performance curves can be obtained at different                   composed of point subtransducer elements equally spaced
varghese and ophir: a theoretical famework for performance characterization of elastography                                                169

Fig. 5. A group of strain filters for center frequencies of 3.5 MHz, 5   Fig. 6. Plot of the theoretical strain filter superimposed on the strain
MHz, and 7.5 MHz, respectively, with a 60% bandwidth and T =             filters obtained using a 1-D simulation before ( o o o ) and after
1.3 µs. Note the improvement in SN Re and dynamic range obtained         ( x x x ) temporal stretching (motion compensation).
with an increase in the pulse center frequency. The maximum attain-
able SN Re is approximately proportional to the square of the center
frequency.                                                               as a percentage; SN Re is plotted, along the y-axis. The
                                                                         theoretical curve of the strain filter for the 1-D case is also
by λ/2. Each subtransducer element is modeled as a point                 shown in the figure.
source or receiver with a two-way Gaussian transfer func-                    Observe from Fig. 6 that both the strain filter curves ob-
tion. The scattering medium is modeled as a 2-D array of                 tained from the simulation are completely bounded by the
point scatterers. The scatterer density in the media was                 theoretical strain filter curve. The theoretical strain filter
set to 48 scatterers/pulse width. The elastic target is as-              curve forms the upper bound, which determines the attain-
sumed to have a Poisson’s ratio ≈ 0. This implies that                   able experimental or simulation performance. For strain
lateral and elevational decorrelation effects due to scat-               values >2%, decorrelation errors cause the drop in the
terer motion are ignored. The applied stress is assumed                  value of the estimated strain. The CRLB bound on the
to propagate uniformly so the localized stress is constant               variance is no longer applicable for strain values >2%, as
throughout the medium. The displacement of each scat-                    shown by the simulation experiment, since signal decorre-
terer is a function of the applied strain, and is modeled by             lation introduced due to tissue compression dramatically
considering an equivalent 1-D spring system described in                 increases the variance in the strain estimate.
[6]. The spring constant is a function of the Young’s mod-                   Inaccurate estimation of the time-delay (increased vari-
ulus of the tissue. The applied strain is the same as the                ance) is primarily due to the detection of false peaks and
tissue strain for a uniformly elastic homogenous medium.                 jitter [17]. The simulation results presented in this section
    The pre-compression A-line is obtained from the ran-                 are obtained using the basic normalized correlation coeffi-
domly distributed scatterers. Each scatterer location is                 cient function. Non-linear processing to remove phase am-
then changed depending on the compressive force and a                    biguities (contributing to the reduction of false peaks [17])
post-compression A-line generated. The stretched post-                   was not performed. Sophisticated algorithms that incorpo-
compression A-line is generated by applying a linear                     rate additional processing to remove false peaks can there-
stretch factor on the post-compression signal. The process               fore significantly improve the performance of the strain
is repeated for 28 different lateral locations in the sim-               estimator.
ulated phantom to obtain independent A-line pairs (lat-
eral step > beamwidth of the transducer). The strain val-
ues are computed from the individual A-line pairs. Time-                                          V. Conclusion
delay estimation is performed using the normalized cross-
correlation function, with the strains computed using (1).                  This paper presents a theoretical framework for quanti-
                                                                         tative assessment of the quality of elastograms. This frame-
B. Results                                                               work is described as a strain filter that is typically a band-
                                                                         pass filter in the strain domain. This filter allows only a
   Fig. 6 shows the mean SN Re value and its standard                    restricted range of strain values to be included in the elas-
deviation (error bars), derived from 28 independent sim-                 togram. The deviation of the strain filter from an ideal
ulations, before and after temporal stretching. The x-axis               all-pass characteristic in the strain domain is due to the
represents the total applied compressive strain expressed                ultrasound system parameters, the finite value of the sono-
170                ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 44, no. 1, january 1997

graphic SNR, and the effects of signal decorrelation. Signal      of the correlation coefficient (used in this paper) and a de-
decorrelation determines the largest value of strain that is      rating factor. The derating factor accounts for the decay in
accurately estimated, while SN RS determines the smallest         the correlation coefficient with increased window lengths.
measurable strain value. The dynamic range of the system          The strain filter described in this paper is therefore more
is thus limited on the low end by electronic noise effects,       optimistic in its prediction of the strain estimator perfor-
and on the high end by signal decorrelation, resulting in a       mance. In addition, the strain filters obtained at varying
bandpass filter in the strain domain.                             depths in tissue also would be different. The 2-D and non-
    The range of strains over which the CRLB specifies            stationary properties of the strain filter will be discussed in
the variance of the time delay defines the optimum per-           a later publication. It has also been demonstrated that un-
formance range for the strain estimator. The formulation          der certain conditions, i.e. optimal multicompression with
of the CRLB by Walker and Trahey does not address                 temporal stretching [23], the bandpass characteristic of the
the implications of decorrelation causing the deviation of        strain filter changes into a more desirable, high-emphasis
the time delay variance from the CRLB for large tissue            characteristic.
strains. Application of the ρ − SN Rρ relationship devel-
oped by Friemel [18] and Céspedes et al. [12] clearly de-
marcates the regions over which the CRLB, the Barankin                               Acknowledgments
bound, and the constant variance level are applicable. Sig-
nal decorrelation may be reduced using a combination of              The authors would like to thank Dr. Ignacio Céspedes
small compressions that allow successful temporal stretch-        for the many discussions, and Dr. Michel Bertrand and Dr.
ing, (e.g., temporal stretching described by Céspedes and        S. Kaiser Alam for their helpful comments.
Ophir [6]–[8], and SMA described by Huang et al. [9],[10]).
Temporal stretching is currently used to correct for decor-
relation only along the axial direction. Lateral and ele-                                 Appendix A
vational decorrelation errors are more difficult to correct,
since the scatterers move out of the beam for large strains.      A. Combining SN RS and SN Rρ to obtain SN RC
Multicompression averaging in conjunction with temporal
stretching of the strain estimates obtained from several             The expression for SN RC which is used to compute
small compressions, increases both the SN Re and the dy-          the post integration SNR and the thresholds in the ZZLB
namic range of strain estimation without sacrificing resolu-      is presented here. The correlation coefficient is converted
tion. The improvements in the elastogram obtained using           into an SNR measure [7],[18], that is given by:
these techniques may be quantitatively predicted using the
theoretical strain filter model developed in this paper.                                           ρ
                                                                                        SN Rρ =       .                    (A.1)
    The strain filter concept developed in this paper pro-                                        1−ρ
vides a graphical framework for characterizing elastogra-         The variation of SN Rρ with a linear variation in ρ is
phy. The strain filter quantified the elastogram image qual-      illustrated in Fig. 7. Note from (A-1) that for ρ = 1,
ity (−3 dB dynamic range 30 dB using the ZZLB and                 SN Rρ = ∞, dropping exponentially to an SN Rρ value of
maximum elastogram SN Re ≈ 23 dB for a single compres-            22.46 dB for ρ = 0.93. The composite value of the SN R,
sion), for the signal parameters specified in Section III. In     combining the contributions of SN RS and SN Rρ is given
a similar manner, different elastographic signal processing       by [13],[14]:
techniques can now be compared and their performance
quantified using the dynamic range, sensitivity, resolution,                                  SN RS SN Rρ
and maximum SN Re obtained from the strain filter. The                         SN RC =                       .             (A.2)
                                                                                           1 + SN RS + SN Rρ
shape of the strain filter defines the quality of the elas-
togram; a narrow, low filter will cause elastographic arti-       This expression for SN RC incorporates both the electronic
facts such as image noise and low dynamic range to plague         noise level and the decrease in SN R caused by increase in
the elastogram. Thus, the design of the optimal strain fil-       signal decorrelation with strain. From (A.2) we observe
ter for a given situation is of utmost importance to the          that SN RC will always be bounded by the smallest value
production of quality elastograms.                                of either SN RS or SN Rρ .
    The signal decorrelation parameter and shape of the
strain filter vary with changes in the cross-correlation win-
dow length (resolution) used to obtain the time-delay. A                                  Appendix B
family of strain filter curves can be obtained at different
resolution levels. The effect of the window length param-         B. Lower Bound on the Variance of
eter on signal decorrelation has not been analyzed in this        the Time Delay Estimator
paper. The dependence of the correlation coefficient on the
finite window length used in the cross-correlation analysis          The modified ZZLB derived by Weinstein and Weiss
is discussed in [21]. The expression for the effective correla-   [13],[14] gives the tightest lower bound on the TDE vari-
tion coefficient is presented as a product of the peak value      ance. We present here the results for bandpass signals with
varghese and ophir: a theoretical famework for performance characterization of elastography                                               171

                                                                         characterized by the constant variance level of (T )2 /12,
                                                                         which corresponds to the variance of a random variable
                                                                         uniformly distributed between −sT /2 < τ < sT /2. In the
                                                                         threshold region, the TDE variance increases exponentially
                                                                         with the post integration SNR.
                                                                            The thresholds given in (B.2) are computed for the fol-
                                                                         lowing typical signal parameters: T = 1 mm (1.3 µs), fo
                                                                         = 5 MHz, B = 3 MHz (60% bandwidth), and SN RS =
                                                                         40 dB. Assuming SN Rρ  40 dB, SN RC ≈ 40 dB. The
                                                                         numerical values of the thresholds η, µ, δ, and γ and in
                                                                         (B-2) are 13.52, 0.78, 0.59, and 0.46 (22.6, −1.1, −2.3, and
                                                                         −3.4 when expressed in dB), respectively, at a post inte-
                                                                         gration SNR value of 43.86 dB. The reader is referred to
                                                                         the papers by Weinstein and Weiss [13],[14] for a detailed
                                                                         description and derivation of the ZZLB and the thresholds
                                                                         for narrow and wide-band systems.

Fig. 7. Plot of the non-linear variation of SN Rρ with the correlation
coefficient.                                                                                        References

                                                                          [1] J. Ophir, E. I. Céspedes, H. Ponnekanti, Y. Yazdi, and X. Li,
a rectangular spectrum. The lower bound for the time de-                      “Elastography: a quantitative method for imaging the elastic-
lay estimation consists of two distinct threshold effects di-                 ity of biological tissues,” Ultrason. Imag., vol. 13, pp. 111–134,
viding the entire post integration SNR domain into three                      1991.
                                                                          [2] E. I. Céspedes, J. Ophir, H. Ponnekanti, N. Maklad, “Elastog-
disjointed segments (the CRLB, Barankin bound, and the                        raphy: Elasticity imaging using ultrasound with application to
constant variance level), as shown by Fig. 7 and (22) in                      muscle and breast in vivo, Ultrason. Imag., vol. 15, (2), pp. 73–
[14]. The distinct regions using (22) from [14] are given                     88, 1993.
                                                                          [3] M. Bertrand, M. Meunier, M. Doucet, and G. Ferland, “Ultra-
by:                                                                           sonic biomechanical strain gauge based on speckle tracking,”
                                                                             Proc. 1989 IEEE Ultrason. Symp., pp. 859–864.
               
                (sT )2 /12,      BT SN RC < γ
               
               
                                                                          [4] H. E. Talhami, L. S. Wilson, and M. L. Neale, “Spectral tis-
                T hreshold, γ < BT SN RC < δ                                 sue strain: A new technique for imaging tissue strain using in-
          στ2 ≥ σBB2
                      ,       δ < BT SN RC < µ                                travascular ultrasound,” Ultrasound, Med. Biol., vol. 20, no. 8,
               
                                                                             pp. 759–772, 1994.
               
                T hreshold,  µ < BT SN RC < η         (B.1)
                2                                                        [5] M. O’Donnell, A. R. Skovoroda, B. M. Shapo, and S. Y.
                 σCRLB ,      η < BT SN RC ,                                  Emelianov, “Internal displacement and strain imaging using ul-
                                                                              trasonic speckle tracking,” IEEE Trans. Ultrason., Ferroelect.,
where the estimate of the time delay is confined to the                       Freq. Contr., vol. 41, no. 3, pp. 314–325, 1994.
interval −sT /2 < τ < sT /2, T is the length of the cross-                [6] E. I. Céspedes, “Elastography: Imaging of Biological Tissue
                                                       2                      Elasticity,” Ph.D. Dissertation, University of Houston, 1993.
correlation window, B is the bandwidth of the system, σBB                 [7] E. I. Céspedes and J. Ophir, “Reduction of image noise in elas-
                                     2
represents the Barankin bound, and σCRLB    represents the                    tography,” Ultrason. Imag., vol. 15, pp. 89–102, 1993.
CRLB. The quantity BT SN Rc is referred to as the post                    [8] M. Bilgen and M. F. Insana, “Deformation models and corre-
                                                                              lation analysis in elastography,” J. Acoust. Soc. Am., vol. 99,
integration SNR. The threshold points used in (B-1) are                       no. 5, pp. 3212–3224, 1996.
defined as follows:                                                       [9] Y. Huang, E. I. Céspedes, and J. Ophir, “Techniques for noise
                       2        2 2                                     reduction in elastograms,” Abstract, AIUM, 1995.
                    6 f0       −1   B                                    [10] T. Varghese, J. Ophir, and E. I. Céspedes, “Noise reduction in
              η= 2            ϕ                                               elastography using temporal stretching with multicompression
                  π    B           24f02
                        2                                                   averaging,” Ultrasound Med. Biol., vol. 22, no. 8 pp. 1043–1052,
                  2.76 fo                                                     1996.
             µ= 2                                                        [11] E. I. Céspedes, Y. Huang, J. Ophir, and S. Spratt, “Methods for
                    π    B                                                    estimation of subsample time delays of digitized echo signals,”
              δ = ζ/2                                                         Ultrason. Imag., vol. 17, pp. 142–171, 1995.
                                                                         [12] E. I. Céspedes, J. Ophir, and S. K. Alam, “The combined ef-
             γ ≈ 0.46 ,                             (B.2)                     fect of signal decorrelation and random noise on the variance
                                                                              of time delay estimation,” IEEE Trans. Ultrason., Ferroelect.,
where fo is the center frequency, ϕ−1 (y) ispthe in-                          Freq. Contr., vol. 44, no. 1, pp. 220–225, 1997.
                     R∞
verse of ϕ(y) = √12π y e−µ /2 dµ, and (ζ/2)ϕ( ζ/2) =
                            2                                            [13] E. Weinstein and A. Weiss, “Fundamental limitations in pas-
                                                                              sive time delay estimation- Part I: Narrow-band systems,” IEEE
(12π/BsT )2 , which has two solutions. The larger value of                    Trans. Acoust., Speech, Sig. Proc., vol. ASSP-31, no. 2, pp. 472–
ζ is used to compute the threshold.                                           485, 1983.
                                                                         [14] E. Weinstein and A. Weiss, “Fundamental limitations in pas-
   When η < BT SN Rc , the ZZLB coincides with                                sive time delay estimation- Part II: Wide-band systems,” IEEE
the CRLB, which is the ambiguity-free region. If δ <                          Trans. Acoust., Speech, Sig. Proc., vol. ASSP-31 (5), pp. 1064–
BT SN Rc < µ, the ZZLB coincides with the Barankin                            1078, 1984.
                                                                         [15] A. H. Quazi, “An overview of the time delay estimate in ac-
bound, where phase ambiguities increase the TDE vari-                         tive and passive systems for target localization,” IEEE Trans.
ance. Finally, when BT SN Rc < γ, the lower bound is                          Acoust., Speech, Sig. Proc., vol. ASSP-29, pp. 527–533, 1981.
172                  ieee transactions on ultrasonics, ferroelectrics, and frequency control, vol. 44, no. 1, january 1997

[16] C. H. Knapp and G. C. Carter, “The Generalized correlation                                   Jonathan Ophir received his B.S.E.E.
     method for estimation of time delay,” IEEE Trans. Acoust.,                                   (summa cum laude), M.S.E.E., and Doctor
     Speech, Sig. Proc., vol. ASSP-24, pp. 320–327, 1976.                                         of Engineering degrees from the University of
[17] F. W. Walker and E. G. Trahey, “A fundamental limit on de-                                   Kansas, Lawrence Kansas in 1971, 1973, and
     lay estimation using partially correlated speckle signals,” IEEE                             1977, respectively. His Doctoral dissertation
     Trans. Ultrason., Ferroelect., Freq. Contr., vol. 42, pp. 301–308,                           describes an early implementation of digital
     1995.                                                                                        scan conversion techniques for diagnostic ul-
[18] B. H. Friemel, “Real-time ultrasonic two-dimensional vector ve-                              trasound. During 1976–1977 he worked as a
     locity estimation utilizing speckle-tracking algorithms: Imple-                              project engineer for Philips Ultrasound Inc.,
     mentation and limitations,” Ph.D Dissertation, Duke Univer-                                  where he developed a commercial version of
     sity, 1994.                                                                                  their first Digital Scan Converter. He then
[19] J. Meunier and M. Bertrand, “Ultrasonic texture motion analy-                                spent three years as an Assistant Professor of
     sis: theory and simulation,” IEEE Trans. Ultrason., Ferroelect.,     Radiology at the University of Kansas Medical School in Kansas City,
     Freq. Contr., vol. 14, pp. 293–300, 1995.                            where he was involved in the development of prototype sonographic
[20] E. I. Céspedes, M. F. Insana, and J. Ophir, “Theoretical bounds     contrast agents, ultrasound phantoms, and instrumentation for ul-
     on strain estimation in Elastography,” IEEE Trans. Ultrason.,        trasonic tissue characterization. From 1980 until the present time,
     Ferroelect., Freq. Contr., vol. 42, no. 5, pp. 969–972, 1995.        he has been with the University of Texas Medical School at Hous-
[21] T. Varghese, M. Bilgen, J. Ophir, and M. F. Insana, “Multireso-      ton, where he is currently Professor of Radiology. He is also Adjunct
     lution strain filter in elastography,” Submitted to IEEE Trans.      Professor of Electrical Engineering at the University of Houston. His
     Ultrason., Ferroelect., Freq. Contr., 1996.                          current field of interest is elastography, the imaging of the elastic
[22] J. S. Bendat and A. C. Piersol, Random Data: Analysis and            properties of soft tissues. Dr. Ophir has contributed more than 85
     Measurement, 2nd ed., New York: John Wiley, 1986.                    papers to peer-reviewed scientific journals, and holds 15 US and for-
[23] T. Varghese and J. Ophir, “Performance optimization in elas-         eign patents. He is a Fellow of the American Institute of Ultrasound
     tography: Multicompression with temporal stretching,” (in            in Medicine, past Chairman of their standards committee, and past
     press), Ultrason. Imag., 1996.                                       member of the board of governors. He is a member of the Editorial
                                                                          Board of Ultrasound in Medicine and Biology, and is Associate Edi-
                                                                          tor of Ultrasonic Imaging. In 1992 he received the Terrance Matzuk
                                                                          Award from the American Institute of Ultrasound in Medicine, and
                                                                          in 1995 he was recognized as Inventor of the Year by the mayor of
                        Tomy Varghese (S’92–M’95) received the            the city of Houston, Texas. In 1995 he also received an honorary
                        B.E. degree in instrumentation technology         commission as an Admiral in the Texas Navy by Governor George
                        from the University of Mysore, India, in 1988,    W. Bush.
                        and the M.S. and Ph.D. in electrical engineer-
                        ing from the University of Kentucky, Lexing-
                        ton, KY in 1992 and 1995, respectively. From
                        1988 to 1990 he was employed as an Engineer
                        in Wipro Information Technology Ltd., India.
                        He is currently a post-doctoral research asso-
                        ciate at the Ultrasonics Laboratory, Depart-
                        ment of Radiology, The University of Texas
                        Medical School, Houston. His current research
interests include detection and estimation theory, statistical pattern
recognition, tissue characterization using ultrasound, and signal and
image processing applications in medical imaging. Dr. Varghese is a
member of the IEEE and Eta Kappa Nu.
You can also read