Procedure Guideline for Tumor Imaging with

Page created by Erin Lindsey
 
CONTINUE READING
Procedure Guideline for Tumor Imaging with
18F-FDG PET/CT 1.0*

Dominique Delbeke1, R. Edward Coleman2, Milton J. Guiberteau3, Manuel L. Brown4, Henry D. Royal5, Barry A. Siegel5,
David W. Townsend6, Lincoln L. Berland7, J. Anthony Parker8, Karl Hubner9, Michael G. Stabin10, George Zubal11,
Marc Kachelriess12, Valerie Cronin13, and Scott Holbrook14
1Vanderbilt University Medical Center, Nashville, Tennessee; 2Duke University Medical Center, Durham, North Carolina; 3Christus

St. Joseph Hospital, Houston, Texas; 4Henry Ford Hospital, Detroit, Michigan; 5Mallinckrodt Institute of Radiology, St. Louis, Missouri;
6University of Tennessee, Knoxville, Tennessee; 7University of Alabama Hospital, Birmingham, Alabama; 8Beth Israel Deaconess

Hospital, Boston, Massachusetts; 9University of Tennessee Medical Center, Knoxville, Tennessee; 10Vanderbilt University, Nashville,
Tennessee; 11Yale University, New Haven, Connecticut; 12Institute of Medical Physics, University of Erlangen-Nurnberg, Erlangen,
Germany; 13Mercy Hospital, Buffalo, New York; and 14Precision Nuclear, Gray, Tennessee

I. PURPOSE                                                                  available for several years, the readily apparent and doc-
                                                                            umented advantages of having PET and CT in a single device
   The purpose of these guidelines is to assist physicians in
                                                                            have resulted in the rapid dissemination of this technology
recommending, performing, interpreting, and reporting the
                                                                            in the United States. This Procedure Guideline pertains
results of 18F-FDG PET/CT for oncologic imaging of adult
                                                                            only to combined PET/CT devices.
and pediatric patients.

II. BACKGROUND INFORMATION AND DEFINITIONS                                  Definitions
   PET is a tomographic scintigraphic technique in which a                    A. A PET/CT scanner is an integrated device containing
computer-generated image of local radioactive tracer dis-                        both a CT scanner and a PET scanner with a single
tribution in tissues is produced through the detection of                        patient table and therefore capable of obtaining a CT
annihilation photons that are emitted when radionuclides                         scan, a PET scan, or both. If a patient does not move
introduced into the body decay and release positrons. 18F-                       between the scans, the reconstructed PET and CT
FDG PET is a tomographic imaging technique that uses a                           images will be spatially registered.
radiolabeled analog of glucose, 18F-FDG, to image relative                    B. PET/CT registration is the process of aligning PET
glucose use rates in various tissues. Because glucose use is                     and CT images for the purposes of combined image
increased in many malignancies, 18F-FDG PET is a sensi-                          display (fusion) and image analysis.
tive method for detecting, staging, and monitoring the                        C. PET/CT fusion is the combined display of registered
effects of therapy of many malignancies. CT is a tomo-                           PET and CT image sets. Superimposed data typically
graphic imaging technique that uses an x-ray beam to                             are displayed with the PET data color coded to the
produce anatomic images. This anatomic information is                            CT data in gray scale.
used to detect and to help determine the location and extent                  D. PET/CT acquisitions can include the whole body, an
of malignancies. Combined PET/CT devices provide both                            extended portion of the body, or a limited portion of
the metabolic information from 18F-FDG PET and the                               the body. These acquisitions are defined in Current
anatomic information from CT in a single examination. As                         Procedural Terminology 2005 as follows:
shown in some clinical scenarios, the information obtained                       1. Whole-body tumor imaging: from the top of the
by PET/CT appears to be more accurate in evaluating pa-                             head through the feet
tients with known or suspected malignancies than does the                        2. Skull base–to–midthigh tumor imaging
information obtained from either PET or CT alone or the                          3. Limited-area tumor imaging
results obtained from PET and CT separately but inter-                        E. Methods of attenuation correction:
preted side by side.                                                             1. CT transmission imaging with PET/CT scanners
   18F-FDG PET and CT are proven diagnostic procedures.
                                                                                 2. Transmission scanning with an isotopic source:
Although techniques for registration and fusion of images                           not commonly used with PET/CT scanners
obtained from separate PET and CT scanners have been

   Received Mar. 10, 2006; accepted Mar. 10, 2006.
   For correspondence or reprints contact: Dominique Delbeke, Vanderbilt    III. EXAMPLES OF CLINICAL OR RESEARCH
University Medical Center, 21st Ave. S. and Garland, Nashville, TN 37232-
2675.
                                                                            APPLICATIONS
   E-mail: dominique.delbeke@vanderbilt.edu
   *YOU CAN ACCESS THIS ACTIVITY THROUGH THE SNM WEB SITE
                                                                               Indications for 18F-FDG PET/CT include but are not
(http://www.snm.org/guidelines).                                            limited to the following:
A. Differentiating benign from malignant lesions                    a. For brain imaging, the patient should be in a quiet
  B. Searching for an unknown primary tumor when                         and dimly lit room for 18F-FDG administration
     metastatic disease is discovered as the first manifes-              and the subsequent uptake phase.
     tation of cancer or when the patient presents with a             b. For body imaging, the patient should remain seated
     paraneoplastic syndrome                                             or recumbent for 18F-FDG administration and the
  C. Staging known malignancies                                          subsequent uptake phase to avoid muscular uptake.
  D. Monitoring the effect of therapy on known malig-                 c. The blood glucose level should be checked before
      nancies                                                            18F-FDG administration. Tumor uptake of 18F-FDG

  E. Determining whether residual abnormalities detected                 is reduced in hyperglycemic states. Most institu-
     on physical examination or on other imaging studies                 tions reschedule the patient if the blood glucose
     after treatment represent tumor or posttreatment fi-                level is greater than 150–200 mg/dL. Reducing the
     brosis or necrosis                                                  serum glucose level by administering insulin can
  F. Detecting tumor recurrence, especially in the pres-                 be considered, but the administration of 18F-FDG
     ence of elevated levels of tumor markers                            should be delayed after insulin administration (with
  G. Selecting the region of a tumor most likely to yield                the duration of the delay being dependent on the
      diagnostic information for biopsy                                  type and route of administration of insulin).
  H. Guiding radiation therapy planning                               d. For either a CT scan done for attenuation cor-
  I. Nononcologic applications, such as evaluation of                    rection/anatomic localization (AC/AL) or a diag-
      infection and atherosclerosis                                      nostic CT scan of the abdomen or pelvis, an
                                                                         intraluminal gastrointestinal contrast agent may
  18F-FDG   PET/CT is not equally effective for all malig-               be administered to provide adequate visualization
nancies, but other tracers are available; however, many of               of the gastrointestinal tract unless it is medically
these are not yet approved by the U.S. Food and Drug                     contraindicated or unnecessary for the clinical
Administration or reimbursable by the Medicare program.                  indication (see Section E.2.b.).
The scientific literature concerning its utility continues to
evolve rapidly.                                                 B. Information Pertinent to Performing Procedure
                                                                  See also the Society of Nuclear Medicine Procedure
                                                                Guidelines for General Imaging.
IV. PROCEDURE                                                     1. Focused history, including the type and site of malig-
A. Patient Preparation                                               nancy, dates of diagnosis and treatment (biopsy results,
  The optimum preparation for patients about to undergo              surgery, radiation, chemotherapy, and administration
PET/CT is evolving. The major goals of preparation are to            of bone marrow stimulants and steroids), and current
minimize tracer uptake in normal tissues, such as the                medications.
myocardium and skeletal muscle, while maintaining uptake          2. History of diabetes, fasting state, and recent infection
in target tissues (neoplastic disease). The following is a        3. Patient’s ability to lie still for the duration of the
commonly used protocol.                                              acquisition (15–45 min)
                                                                  4. History of claustrophobia
  1. Pregnancy and breast-feeding: see the Society of Nuclear     5. Patient’s ability to put his or her arms overhead
     Medicine Procedure Guidelines for General Imaging
  2. Before arrival
                                                                C. Precautions
        Patients should be instructed to fast and not con-
     sume beverages, except for water, for at least 4–6 h          See the Society of Nuclear Medicine Procedure Guide-
     before the administration of 18F-FDG to decrease           lines for General Imaging.
     physiologic glucose levels and to reduce serum insu-
     lin levels to near basal levels. Oral hydration with       D. Radiopharmaceutical
     water is encouraged. Intravenous fluids containing            With PET/CT, the radiation dose to the patient (Table 1)
     dextrose or parenteral feedings also should be with-       is the combination of the radiation dose from the PET
     held for 4–6 h.                                            radiopharmaceutical and the radiation dose from the CT
        When intravenous contrast material is to be used,       portion of the study. Radiation dose in diagnostic CT has
     patients should be screened for a history of iodinated     attracted considerable attention in recent years, in particular
     contrast material allergy, use of metformin for the        for pediatric examinations. It can be very misleading to
     treatment of diabetes mellitus, and renal disease.         state a ‘‘representative’’ dose for a CT scan because of the
     Intravenous contrast material should not be adminis-       wide diversity of applications, protocols, and CT systems.
     tered when the serum creatinine level is above 2.0         This caveat also applies to the CT component of a PET/CT
     mg/dL.                                                     study. For example, a body scan may include various portions
  3. Before injection                                           of the body and may use protocols aimed to reduce the
TABLE 1
                                       18F-FDG    Radiation Dosimetry for Adults and Children

                                                                  Organ receiving the largest radiation dose,        Effective dose,
       Patient          Intravenously administered activity                 mGy/MBq (rads/mCi)                     mSv/MBq (rems/mCi)

   Adult               370–740 MBq (10–20 mCi)                               Bladder, 0.16* (0.59)                     0.019 (0.070)
   Child (5 y old)     5.18–7.4 MBq/kg (0.14–0.20 mCi/kg)                    Bladder, 0.32y (1.2)                      0.050 (0.18)

   *Voiding interval, 3.5 h. Changes in bladder wall dose are approximately linear with changes in voiding interval; therefore, for a voiding
interval of 2.0 h, dose to bladder wall would change by a factor of 2/3.5.
   y
     Voiding interval, 2.0 h.
   Data are from International Commission on Radiological Protection. Radiation Dose to Patients from Radiopharmaceuticals. St. Louis,
MO: Elsevier; 2000:49. ICRP publication 80.

radiation dose to the patient or aimed to optimize the CT scan                    performing whole-body tumor imaging offers the
for diagnostic purposes. The effective dose could range from                      advantage of staging the entire body.
approximately 5 to 80 mSv (0.5–8.0 rems) for these options.                    c. For optimal imaging of the body, the arms should
It is therefore advisable to estimate the CT dose specific to the                 be elevated over the head if that position can be
CT system and protocol being used.                                                tolerated by the patient. Arms along the side may
    Pediatric and adolescent patients should have their CT                        produce beam-hardening artifacts over the torso.
examinations performed at milliampere-seconds settings                            However, for optimal imaging of the head and
appropriate for patient size, regardless of the CT protocol                       neck, the arms should be positioned along the
used, because radiation dose to the patient increases sig-                        side.
nificantly as the diameter of the patient decreases.                           d. The patient should void the bladder before the
                                                                                  acquisition of the images to limit the radiation
                                                                                  dose to the renal collecting system and bladder.
E. Image Acquisition                                                           e. Metallic objects should be removed from the
  See also the Society of Nuclear Medicine PET Tumor                              patient whenever possible.
Imaging Guidelines and the ‘‘Specifications of the Exam-                   2. Protocol for CT imaging
ination’’ and ‘‘Documentation’’ sections of the American                          The CT component of a PET/CT examination can
College of Radiology Practice Guideline for the Perfor-                       be performed either for AC/AL or as an optimized
mance of Computed Tomography of the Extracranial Head                         diagnostic CT scan. An AC/AL CT scan has not
and Neck in Adults and Children, the American College                         necessarily been optimized as a diagnostic CT
of Radiology Practice Guideline for the Performance of                        examination, whereas for a diagnostic CT scan, such
Pediatric and Adult Thoracic Computed Tomography                              optimization has been attempted. In some circum-
(CT), and the American College of Radiology Practice                          stances, both an initial CT acquisition for AC/AL
Guideline for the Performance of Computed Tomography                          (before the PET data acquisition) and diagnostic
(CT) of the Abdomen and Computed Tomography (CT) of                           CT (after the PET data acquisition) are performed.
the Pelvis.                                                                   Optimization of the CT technique used in PET/CT
  1. Field of view, positioning, and preacquisition preparation               continues to evolve.
      a. Skull base–to–proximal thigh imaging generally is                     a. If the CT scan is obtained for AC/AL, a low
          recommended to survey the body in the search for                        milliampere-seconds setting is recommended to
          areas of abnormal 18F-FDG accumulation for most                         decrease the radiation dose to the patient.
          tumor types. Such PET/CT scans typically are                         b. For an optimized diagnostic CT scan, standard
          acquired from the external auditory meatus to the                       CT milliampere-seconds settings are recom-
          midthigh region. For tumors with a high likelihood                      mended to optimize the spatial resolution of the
          of scalp, skull, or brain involvement or lower-                         CT scan. Tube current modulation may be used
          extremity involvement, whole-body tumor imag-                           to minimize radiation dose to the patient. In
          ing is performed.                                                       some cases, intravenous or oral contrast material
      b. Limited-area tumor imaging can be considered                             may be used. A separate CT acquisition may be
          when critical abnormalities are likely to be local-                     necessary to produce an optimized diagnostic
          ized in a known region of the body (e.g., solitary                      CT scan that is requested for a particular region
          pulmonary nodule, probable lung cancer, evalua-                         of the body. For many indications, the exami-
          tion of hilar lymph node involvement, diagnosis of                      nation is performed with intravenous contrast
          head and neck cancer, and monitoring of therapy                         material and appropriate injection techniques.
          of locally advanced breast cancer). However,                            High intravascular concentrations of intravenous
contrast material may cause an attenuation cor-                    imaging and is based on the administered activity,
      rection artifact on the PET image, but the impact                  patient body weight, and sensitivity of the PET
      usually is modest. This artifact is minimized                      scanner (as determined largely by detector com-
      on scanners by use of appropriate correction                       position and acquisition method). Typically, for
      factors.                                                           imaging skull to midthigh, the total acquisition
   c. For either a CT scan done for AC/AL or an                          time ranges from 15 to 45 min. The imaging time
      optimized diagnostic CT scan of the abdomen or                     typically is prolonged for the acquisition of brain
      pelvis, an intraluminal gastrointestinal noncaloric                images or for images of a limited region of interest.
      contrast agent may be administered to provide                   c. Semiquantitative estimation of tumor glucose me-
      adequate visualization of the gastrointestinal tract               tabolism by use of the SUV is based on relative
      unless it is medically contraindicated or unneces-                 lesion radioactivity measured on images corrected
      sary for the clinical indication. This agent may be                for attenuation and normalized for the injected dose
      a positive contrast agent (such as dilute barium),                 and body weight, lean body mass, or body surface
      an oral iodinated contrast agent, or a negative                    area. This measurement is obtained on a static
      contrast agent (such as water). Collections of highly              emission image typically acquired more than 45
      concentrated barium or iodinated contrast agents                   min after injection. The accuracy of SUV measure-
      can result in an attenuation correction artifact that              ments depends on the accuracy of the calibration of
      leads to a significant overestimation of the regional              the PET scanner, among other factors. The repro-
      18F-FDG concentration; other, dilute positive and                  ducibility of SUV measurements depends on the
      negative oral contrast agents cause less overestima-               reproducibility of clinical protocols, for example,
      tion and do not affect PET image quality.                          dose infiltration, time of imaging after 18F-FDG
   d. With regard to the breathing protocol for CT                       administration, type of reconstruction algorithms,
      transmission scanning, in PET/CT, the position                     type of attenuation maps, size of the region of
      of the diaphragm on the PET emission images                        interest, changes in uptake by organs other than the
      should match as closely as possible that on the CT                 tumor, and methods of analysis (e.g., maximum and
      transmission images. Although a diagnostic CT                      mean).
      scan of the chest typically is acquired during end-             d. Semiquantitative estimation of tumor metabolism
      inspiration breath holding, this technique is not                  can be based on the ratio of 18F-FDG uptake in a
      optimal for PET/CT because it may result in                        lesion to 18F-FDG uptake in internal reference
      substantial respiratory motion misregistration on                  regions, such as the blood pool, mediastinum,
      PET and CT images. Some facilities perform CT                      liver, and cerebellum.
      transmission scans during breath holding at mid-
      inspiration volume, and others prefer that the             F. Interventions
      patient continue shallow breathing during the CT             1. Intense urinary bladder tracer activity degrades image
      acquisition. Respiratory motion results in inaccu-              quality and can confound the interpretation of findings
      rate localization of lesions at the base and periph-            in the pelvis. Hydration and a loop diuretic, without or
      ery of the lungs, in the dome of the liver, or near             with bladder catheterization, may be used to reduce
      any lung–soft tissue interface and may result in                accumulated urinary tracer activity in the bladder.
      spurious standardized uptake value (SUV) deter-                 Bladder catheterization with a 3-way flushing catheter
      minations. Motion correction or respiratory gating              running a continuous bladder flush after injection until
      is recommended when available.                                  imaging has been used successfully to clear bladder
3. Protocol for PET emission imaging                                  activity.
   a. The radiopharmaceutical should be injected at a              2. Keeping the patient in a warm room for 30–60 min
      site contralateral to the site of concern. Emission             before the injection of 18F-FDG, particularly in cold
      images should be obtained at least 45 min after                 climates and air-conditioned environments, will help
      radiopharmaceutical injection. The optimal 18F-                 to minimize brown fat uptake. Lorazepam or diazepam
      FDG distribution phase is controversial. Many fa-               given before the injection of 18F-FDG may reduce
      cilities start the acquisition of the images at 60 or 90        uptake by brown adipose tissue or skeletal muscle.
      min after 18F-FDG administration. Some facilities               b-Blockers also may reduce uptake by brown fat.
      obtain a second set of images to assess the change in
      uptake over time. The 18F-FDG uptake time should           G. Processing
      be constant whenever possible and certainly when 2           1. PET reconstruction: PET emission data consist of the
      studies are being compared by use of semiquanti-                number of events along lines of response between
      tative parameters, especially the SUV.                          detector pairs. The emission data must be corrected
   b. The emission image acquisition time varies from                 for detector efficiency (normalization), system dead
      2 to 5 min or longer per bed position for body                  time, random coincidences, scatter, attenuation, and
sampling nonuniformity. Some of these corrections            H. Interpretation Criteria
   (e.g., attenuation) can be incorporated directly into the      1. Normal physiologic uptake of 18F-FDG can be seen to
   reconstruction process. Scanners with retractable                 some extent in every viable tissue, including the brain,
   septa can acquire data in both 2-dimensional (2D)                 myocardium (where the uptake is significant in some
   and 3-dimensional (3D) modes, whereas scanners                    patients despite prolonged fasting), breast, liver,
   without septa acquire data in the 3D mode only.                   spleen, stomach, intestines, kidneys and urine, mus-
   Datasets acquired in the 3D mode either can be                    cle, lymphoid tissue (e.g., tonsils), bone marrow,
   rebinned into 2D data and reconstructed with a 2D                 salivary glands, thymus, uterus, ovaries, testes, and
   algorithm or can be reconstructed with a fully 3D                 brown adipose tissue (see Section K).
   algorithm. Iterative reconstruction approaches are             2. For whole-body surveys, studies have shown that 18F-
   now widely available for clinical applications in both            FDG PET of the brain is relatively insensitive for
   2D and 3D modes, largely replacing the direct, filtered           the detection of cerebral metastases, because of high
   backprojection methods used previously. For a given               physiologic 18F-FDG uptake in the gray matter.
   algorithm, the appropriate reconstruction parameters           3. Increased uptake of 18F-FDG can be seen in neo-
   will depend on the acquisition mode, the type of                  plasms, granulation tissue (e.g., healing wounds),
   scanner, and the imaging task. It is considered good              infections, and other inflammatory processes.
   practice to archive reconstructions both with and              4. Although the pattern of 18F-FDG uptake and specific
   without attenuation correction to resolve issues arising          CT findings as well as the correlation with history,
   from potential artifacts generated by the CT-based                physical examination, and other imaging modalities
   attenuation correction procedure. The reconstructed im-           usually are the most helpful features in differentiating
   age volume can be displayed in transaxial, coronal, and           benign from malignant lesions, semiquantitative esti-
   sagittal planes and as a rotating maximum-intensity-              mates (e.g., SUV) also may be of value, especially for
   projection image.                                                 evaluating changes over time or with therapy.
2. CT reconstruction: CT sinograms are reconstructed by
   filtered backprojection at full field of view for the data   I. Reporting
   used for attenuation correction of the PET emission            See also the Society of Nuclear Medicine Procedure
   data and separately for CT interpretation with appro-        Guidelines for General Imaging.
   priate zoom, slice thickness and overlap, and recon-           1. Study identification
   struction algorithms for the particular region of the          2. Clinical information
   body scanned. The filtered backprojection can be                  a. Indication for the study
   either 2D after appropriate portions of the spiral CT             b. Relevant history
   data are collected into axial or tilted planes or fully           c. Information needed for billing
   3D. In addition to the reconstruction kernel that              3. Procedure description and imaging protocol
   adjusts in-plane features, such as spatial resolution             a. Radiopharmaceutical, including administered ac-
   and noise texture, longitudinal filtration (along the                tivity, route of administration, and 18F-FDG up-
   z-axis) is used to modify the z-resolution and the slice             take time
   sensitivity profiles. In addition, there are techniques           b. Other drugs administered and procedures performed,
   for emphasizing certain image features, for example,                 such as placement of intravenous line; hydration;
   bone, lung, or head algorithms. For attenuation cor-                 insertion of Foley catheter (size of catheter); admin-
   rection, only the standard kernels are used. Because                 istration of furosemide (amount and time), muscular
   CT volumes today are nearly isotropic, reslicing in                  relaxants, or pain medications; and sedation proce-
   coronal, sagittal, or even curved displays often is                  dures (briefly describe the procedure, state the type of
   preferred. Advanced display techniques, such as vol-                 medication and time of sedation in relation to the
   ume rendering and maximum- or minimum-intensity                      radiotracer injection, and state the patient condition
   projections applied to the complete volume or to                     at the conclusion of the PET study)
   thick, arbitrarily oriented sections, often are used.             c. Field of view and patient positioning: whole body,
   Organ- and task-specific automatic or semiautomatic                  skull base to midthigh, or limited area and position
   segmentation algorithms and special evaluation algo-                 of the arms
   rithms also are in routine use.                                   d. Baseline glucose level
3. Display: With an integrated PET/CT system, typically              e. CT transmission protocol (for AC/AL or diagnostic
   the software packages provide registered and aligned                 CT protocol with or without oral or intravenous
   CT images, 18F-FDG PET images, and fusion images                     contrast material and with the appropriate protocol
   in the axial, coronal, and sagittal planes as well as                for the clinical scenario and body region of interest)
   maximum-intensity-projection images for review in the             f. PET emission protocol: see the Society of Nuclear
   3D cine mode. 18F-FDG PET images with and without                    Medicine Procedure Guidelines for General
   attenuation correction should be available for review.               Imaging
4. Description of findings                                           a. Whenever possible, a precise diagnosis should be
    a. Quality of the study: for example, limited because               given
       of motion, muscular uptake, or hyperglycemia                  b. When appropriate, a differential diagnosis should
    b. Describe the location, extent, and intensity of                  be given
       abnormal 18F-FDG uptake in relation to uptake                 c. When appropriate, follow-up and additional diag-
       in normal comparable tissues and describe the                    nostic studies needed to clarify or confirm the
       relevant morphologic findings related to PET ab-                 impression should be recommended
       normalities on the CT images. An estimate of the
       intensity of 18F-FDG uptake can be provided by
       the SUV; however, the intensity of uptake may be         J. Quality Control
       described as mild, moderate, or intense or in rela-        1. Radiopharmaceuticals
       tion to the background uptake in normal hepatic                   See the Society of Nuclear Medicine Procedure
       parenchyma (average SUV: 2.0–3.0; maximum                     Guidelines for Use of Radiopharmaceuticals.
       SUV: 3.0–4.0). The integrated PET/CT report                2. Instrumentation specifications
       should include any detected incidental findings                   See also the ‘‘Equipment Specifications’’ and
       on the CT scan that are relevant to patient care. If          ‘‘Quality Control’’ sections of the American College
       the CT scan was requested and performed as a                  of Radiology Practice Guideline for the Performance
       diagnostic examination, then the CT component                 of Computed Tomography of the Extracranial Head
       of the study may be reported separately, if neces-            and Neck in Adults and Children, the American
       sary, to satisfy regulatory, administrative, or re-           College of Radiology Practice Guideline for the
       imbursement requirements. In that case, the                   Performance of Pediatric and Adult Thoracic Com-
       PET/CT report can refer to the diagnostic CT scan             puted Tomography (CT), and the American College of
       report for findings not related to the PET/CT                 Radiology Practice Guideline for the Performance of
       examination.                                                  Computed Tomography (CT) of the Abdomen and
    c. Limitations: when appropriate, identify factors               Computed Tomography (CT) of the Pelvis.
       that can limit the sensitivity and specificity of              a. Equipment performance guidelines
       the examination (e.g., small lesions or an inflam-                    For patient imaging, state-of-the-art scanners
       matory process)                                                   meet or exceed the following specifications:
    d. Clinical issues: address or answer any pertinent                      For the CT scanner: see the American College
       clinical questions raised in the request for the                  of Radiology guidelines mentioned above.
       imaging examination                                                   For the PET scanner: These specifications are
    e. Comparative data                                                  based on manufacturers’ data obtained according
           Comparisons with previous examinations and                    to National Electrical Manufacturers Association
       reports, whenever possible, should be part of the                 (NEMA) 2001 protocols; the uniformity estimate
       radiologic consultation and report. PET/CT                        is based on the NEMA 1994 protocol. For lutetium
       studies are more valuable when correlated with                    oxyorthosilicate–based scanners, a modified
       previous diagnostic CT, previous PET, previous                    NEMA protocol is defined to account for the
       PET/CT, previous MRI, and all appropriate                         natural radioactive background from lutetium
       imaging studies and clinical data that are relevant.              oxyorthosilicate.
           When PET/CT is performed for monitoring                       d In-plane spatial resolution: ,6.5 mm

       therapy, a comparison of the extent and intensity                 d Axial resolution: ,6.5 mm

       of uptake may be summarized as metabolic pro-                     d Sensitivity (3D): .4.0 cps/kBq

       gressive disease, metabolic stable disease, meta-                 d Sensitivity (2D): .1.0 cps/kBq

       bolic partial response, or metabolic complete                     d Uniformity: ,5%

       response. The European Organization for Re-                          For the PET/CT scanner:
       search and Treatment of Cancer has published                      d Maximum co-scan range (CT and PET): .160

       criteria for these categories, although these criteria               cm
       have not yet been validated in outcome studies. A                 d Maximum patient weight: at least 157.50 kg

       change in the intensity of uptake with semiquan-                     (350 lb)
       titative measurements, expressed in absolute                      d Patient port diameter: at least 59 cm

       values and percent change, may be appropriate                           All currently available commercial PET/CT
       in some clinical scenarios. However, the technical                   scanners meet or exceed these specifications.
       protocol and analysis of images need to be con-                         The CT field of view should be at least 50 cm
       sistent in the 2 sets of images, as described in                     in diameter to minimize artifacts from CT-based
       Section E.3.                                                         attenuation correction because of mismatch be-
5. Impression (conclusion or diagnosis)                                     tween the CT and the PET fields.
A workstation with the capability to display             d Brown adipose tissue
            CT, PET, and fused images with different per-               d Thymus, especially in children
            centages of CT and PET blending also should be              d Lactating breast

            available. The workstation should allow multi-              d Areola

            planar display with linked CT and PET cursors.              d Skeletal and smooth muscles (e.g., neck or

            Postcollection registration of the PET and CT                 paravertebral; hyperinsulinemia)
            datasets and registration with other imaging stud-          d Gastrointestinal (e.g., esophagus, stomach, or

            ies, including nonrigid registration, are desirable.          bowel)
     b. Equipment quality control                                       d Urinary tract structures (containing excreted
                                                                          18F-FDG)
             PET performance monitoring should be in ac-
         cordance with the American College of Radiology                d Female genital tract (e.g., uterus during menses

         Technical Standard for Medical Nuclear Physics                   or corpus luteum cyst)
         Performance Monitoring of Nuclear Medicine                b.   Inflammatory processes
         Imaging Equipment.                                             d Postsurgical inflammation, infection, or hema-

             CT performance monitoring should be in ac-                   toma; biopsy site; or amputation site
         cordance with the American College of Radiology                d Postradiation (e.g., radiation pneumonitis)

         Technical Standard for Medical Physics Perfor-                 d Postchemotherapy

         mance Monitoring of Computed Tomography                        d Local inflammatory disease, especially granu-

         (CT) Equipment.                                                  lomatous processes (e.g., sarcoidosis, fungal
             The quality control procedures for PET/CT                    disease, or mycobacterial disease)
         should incorporate both CT procedures and PET                  d Ostomy site (e.g., trachea or colon) and drainage

         procedures according to the American College of                  tubes
         Radiology technical standards mentioned above.                 d Injection site

         The quality control procedures for CT should                   d Thyroiditis

         include air and water calibrations in Hounsfield               d Esophagitis, gastritis, or inflammatory bowel

         units for a range of kilovolts. The quality control              disease
         procedures for PET should include a calibration                d Acute and occasionally chronic pancreatitis

         measurement of activity in a phantom containing a              d Acute cholangitis and cholecystitis

         known concentration, generally as a function of                d Osteomyelitis, recent fracture sites, or joint

         axial position within the scanner field of view. A               prostheses
         daily check on the stability of the individual                 d Lymphadenitis

         detectors also should be performed to identify            c.   Benign neoplasms
         detector failures and drifts.                                  d Pituitary adenoma

             In addition, for PET/CT, a check on the align-             d Adrenal adenoma

         ment between the CT and the PET scanners should                d Thyroid follicular adenoma

         be performed periodically. Such a gantry align-                d Salivary gland tumors (e.g., Warthin’s tumor or

         ment check should determine any offset between                   pleomorphic adenoma)
         the CT and the PET scanners to be incorporated                 d Colonic    adenomatous polyps and villous
         into the fused image display to ensure accurate                  adenoma
         image alignment.                                               d Ovarian thecoma and cystadenoma

  3. Emergency procedures                                               d Giant cell tumor

        An emergency cart containing appropriate medi-                  d Aneurysmal bone cyst

     cations and resuscitation equipment must be readily                d Leiomyoma

     available to treat adverse contrast material reactions.       d.   Hyperplasia or dysplasia
                                                                        d Graves’ disease

K. Sources of Error                                                     d Cushing’s disease

   Processes other than malignancies may cause false-                   d Bone marrow hyperplasia (e.g., anemia or cy-

positive and false-negative results. The following list,                  tokine therapy)
although not all-inclusive, includes the most commonly                  d Thymic rebound hyperplasia (postchemotherapy)

encountered causes:                                                     d Fibrous dysplasia

   1. False-positive findings                                           d Paget’s disease

      a. Physiologic uptake that may lead to false-positive        e.   Ischemia
         interpretations                                                d Hibernating myocardium

         d Salivary glands and lymphoid tissues in the head        f.   Artifacts
            and neck                                                    d Misalignment between PET and CT data can

         d Thyroid                                                        cause attenuation correction artifacts. PET images
without attenuation correction and fusion images           their observations subsequently integrated, there is a need
           can be used to help identify these artifacts.              to define the training for people who can interpret and
        d Inaccuracies in converting from polychromatic               integrate both components of PET/CT scans. Regardless of
           CT energies to the 511-keV energy of annihila-             previous training, imaging experts interpreting PET/CT
           tion radiation can cause artifacts around metal            scans should have appropriate training in both PET and
           or dense barium, although these artifacts are less         CT. Ideally, a diagnostic radiologist who has not received
           common with newer conversion algorithms.                   training and experience in PET should have training and
  2. False-negative findings                                          experience in PET similar to that of a nuclear medicine
     d Small size (,2 times the resolution of the system)             physician, and a nuclear medicine physician should have
     d Tumor necrosis                                                 training and experience in CT similar to that of a diagnostic
     d Recent chemotherapy or radiotherapy                            radiologist, including treatment of contrast material reac-
     d Recent high-dose steroid therapy                               tions. In most instances, it is not feasible for a practicing
     d Hyperglycemia and hyperinsulinemia                             diagnostic radiologist to duplicate exactly the PET training
     d Some low-grade tumors (e.g., sarcoma, lymphoma,                that a nuclear medicine physician receives during a nuclear
       or brain tumor)                                                medicine residency, nor is it possible for a practicing
     d Tumors with large mucinous components                          nuclear medicine physician to duplicate the CT training
     d Some hepatocellular carcinomas, especially well-               obtained in a diagnostic radiology residency.
       differentiated tumors                                             An article summarizing discussions regarding issues
     d Some genitourinary carcinomas, especially well-                relating to imaging with PET, CT, and PET/CT recently
       differentiated tumors                                          was published by a collaborative working group with
     d Prostate carcinoma, especially well-differentiated             representatives from the American College of Radiology,
       tumors                                                         the Society of Nuclear Medicine (SNM), and the Society of
     d Some    neuroendocrine tumors, especially well-                Computed Body Tomography and Magnetic Resonance (J
       differentiated tumors                                          Nucl Med. 2005;46:1225–1239). These organizations agree
     d Some     thyroid carcinomas, especially well-                  that only appropriately trained, qualified physicians should
       differentiated tumors                                          interpret PET/CT images. Traditionally, appropriate train-
     d Some bronchioloalveolar carcinomas                             ing has been quantified by the number of continuing
     d Some lobular carcinomas of the breast                          medical education credits earned and the number of cases
     d Some skeletal metastases, especially osteoblastic or           interpreted. The training recommendations from the col-
       sclerotic tumors                                               laborative working group are summarized in Table 2.
     d Some osteosarcomas                                             Alternative approaches, such as determining the accuracy
                                                                      of each physician’s interpretation of images compared with
V. QUALIFICATION OF PERSONNEL                                         that of his or her peers by use of a workstation simulator
A. Physicians                                                         and a report generation and scoring system, may have equal
   The use of PET/CT technology is becoming common                    or greater validity.
practice. Because it is inefficient to have PET/CT images                In the future, the requirements of radiology and nuclear
interpreted by 2 different imaging experts and then to have           medicine residency training programs will include training

                                                        TABLE 2
                                    Summary of PET/CT On-The-Job Training Requirements

                                                                   No. of PET/CT          No. of CT
                                                                   interpretations     interpretations    PET/CT CME         CT CME
                 Training                  Board certification      (supervised)*       (supervised)y       credits           credits

   Nuclear medicine                          ABNM                       150                 500                  8             100
   Diagnostic radiology (recent CT)z         ABR                        150                                     35
   Nuclear radiology (recent CT)z            ABR                        150                                      8
   Radiology (recent CT)z                    ABR and ABNM               150                                      8
   Diagnostic radiology (no recent CT)       ABR                        150                 500                 35             100

   *Supervision should be performed by qualified nuclear medicine physicians or diagnostic radiologists who have interpreted more than
500 PET/CT studies.
   y
     Supervision should be performed by qualified diagnostic radiologists as defined in American College of Radiology Practice Guidelines
for Performing and Interpreting Diagnostic Computed Tomography. CT cases should include reasonable distribution of head and neck,
chest, abdomen, and pelvis.
   z
     Includes radiologists or nuclear radiologists with recent experience in body CT (100 body CT cases/y for preceding 5 y).
   ABNM 5 American Board of Nuclear Medicine.
in the interpretation and supervision of integrated PET/CT         The Nuclear Medicine Technology Certification Board
studies. Certifying and recertifying examinations will in-      (NMTCB) has developed a PET specialty examination that
clude testing on CT, PET, and PET/CT. Eligibility for           is open to certified or registered nuclear medicine technol-
taking the recertification examinations will mandate par-       ogists, registered radiologic technologists, CT technolo-
ticipation in a maintenance of certification (MOC) program      gists, and registered radiation therapists, as long as they
and will include training in the interpretation of PET, CT,     have fulfilled the required prerequisites, as defined on
and PET/CT images. Some components of the MOC will              the NMTCB Web site (www.nmtcb.org). The American
include evaluation of the accuracy of each physician’s          Registry of Radiologic Technologists (ARRT) has adapted
interpretation of images compared with that of his or her       its CT certification examination and has allowed certified
peers by use of a workstation simulator and a report gen-       or registered nuclear medicine technologists who have
eration and scoring system. Performing and interpreting         met the required prerequisites to take this examination.
physicians should participate in and be able to show            Eligibility criteria are located on the ARRT Web site
evidence of participation in continuing medical education       (www.arrt.org).
in the techniques and interpretation related to the proce-         Licensure and regulation definitely are affecting the
dures discussed in these guidelines. Where MOC programs         opportunities that nuclear medicine technologists have for
exist, physicians should be able to show evidence of            obtaining the CT experience needed to take the ARRT CT
participation.                                                  examination as well as the opportunities that radiologic
   If no physicians with training in the interpretation of      technologists have for gaining the PET experience needed
integrated PET/CT studies are available, PET scans should       to take the NMTCB PET examination. The SNMTS is
be interpreted by nuclear medicine physicians with exper-       approaching these issues through both legislative and reg-
tise in PET, and CT scans should be interpreted by diag-        ulatory pathways. The SNMTS has been promoting the
nostic radiologists with expertise in CT. Strong opinions       Consumer Assurance of Radiologic Excellence bills pend-
have been expressed that a single report should be issued to    ing before the U.S. Congress. The Consumer Assurance of
avoid inconsistencies, confusion, and redundancy, although      Radiologic Excellence bills would establish minimum ed-
reimbursement issues still are being debated.                   ucation and credentialing standards for those who perform
                                                                medical imaging and therapeutic procedures. The second
B. Technologists                                                pathway recognizes the regulatory route in addressing these
   PET/CT technology presents similar practice issues re-       practice issues through a collaborative liaison relationship
garding the education, training, and certification of tech-     that has been established with the Conference of Radiation
nologists to become appropriately qualified and competent       Control Program Directors (www.crcpd.org), the profes-
to perform PET/CT. Additional issues arise with regard to       sional organization of state radiation regulators.
ensuring competency, standardizing the educational expe-
rience of technologists, and barriers placed by licensure and
regulation at the state level. The Society of Nuclear Med-      C. Qualified Medical Physicists
icine Technologist Section (SNMTS) and the American                A qualified medical physicist is an individual who is
Society of Radiologic Technologists (ASRT) have jointly         competent to practice independently one or more of the
developed a master plan and set into motion mechanisms to       subfields of medical physics. The SNM considers certifi-
sort out the practice issues regarding PET/CT. This master      cation and continuing education in the appropriate sub-
plan was crafted in July 2002 during a stakeholders’            field(s) to demonstrate that an individual is competent to
meeting known as the PET/CT Consensus Conference. The           practice one or more of the subfield(s) of medical physics
recommendations from this meeting can be found in a             and to be a qualified medical physicist. The SNM recom-
report of the PET/CT Consensus Conference (J Nucl Med           mends that the individual be certified in the appropriate
Technol. 2002;30:201–204) and are also accessible on the        subfield(s) by the American Board of Radiology (ABR) or
SNM Web site (www.snm.org).                                     the American Board of Science in Nuclear Medicine
   It is the responsibility of professional associations to     (ABSNM).
establish standards, delineate mechanisms for obtaining the        The appropriate subfields of medical physics are as
training necessary to promote a qualified and competent         follows:
workforce to perform these procedures, and partner with             d ABR: ‘‘Medical     Nuclear Physics’’ with initially at
organizations that can assist in sorting out practice issues.         least 15 h of continuing education credits in CT
To address educational needs, the ASRT and the SNMTS                  imaging physics, or
spearheaded the development of a PET/CT curriculum,                 d ‘‘Diagnostic Radiological Physics’’ with initially at

which was endorsed by numerous professional organiza-                 least 15 h of continuing education credit in PET
tions and distributed to each state radiation control board           imaging physics, or
and every program director in the United States; it is also         d ABSNM: ‘‘Nuclear Medicine Physics and Instrumen-

posted on the SNM Web site (www.snm.org) and the ASRT                 tation’’ with initially at least 15 h of continuing
Web site (www.asrt.org).                                              education credits in CT imaging physics.
A qualified medical physicist must have at least 40 h of     E. Cohade C, Osman M, Nakamoto Y, et al. Initial experi-
practical experience providing physics support for both the       ence with oral contrast in PET/CT: phantom and
PETand the CT components in an established PET/CT facility.       clinical studies. J Nucl Med. 2003;44:412–416.
  A qualified medical physicist’s continuing education         F. Coleman RE, Delbeke D, Guiberteau MJ, et al.
should be in accordance with the American College of              Concurrent PET/CT with an integrated imaging system:
Radiology Practice Guideline for Continuing Education             intersociety dialogue from the Joint Working Group of
and should include at least 15 h in PET and CT physics            the American College of Radiology, the Society of
combined in a 3-y period.                                         Nuclear Medicine, and the Society of Computed Body
  A qualified medical physicist or other qualified scientist      Tomography and Magnetic Resonance. J Nucl Med.
performing physics services in support of a PET/CT facility       2005;46:1225–1239.
should meet all of the following criteria:                     G. Czernin J. PET/CT: imaging structure and function.
  1. Advanced training directed at the specific area of           J Nucl Med. 2004;45(suppl 1):1S–103S.
      responsibility (e.g., medical physics, health physics,   H. Dizendorf E, Hany TF, Buck A, et al. Cause and mag-
      or instrumentation)                                         nitude of the error induced by oral CT contrast agent in
  2. Licensure, if required by state regulations                  CT-based attenuation correction of PET emission
  3. Documented regular participation in continuing edu-          studies. Eur J Nucl Med. 2003;44:732–738.
      cation in the area of specific involvement to maintain   I. Donnelly LF. Lessons from history. Pediatr Radiol.
      competency                                                  2002;32:287–292.
  4. Knowledge of radiation safety and protection and of       J. Fearon T, Vucich J. Pediatric patient exposures from CT
      all rules and regulations applying to the area of           examinations: GE CT/T 9800 scanner. AJR. 1985;
      practice                                                    144:805–809.
                                                               K. Gambhir SS, Czernin J, Schimmer J, et al. A tabulated
VI. ISSUES REQUIRING FURTHER CLARIFICATION                        summary of the 18F-FDG PET literature. J Nucl Med.
                                                                  2001;42(suppl):1S–93S.
  A. Use of AC/AL CT, use of optimized diagnostic CT,
                                                               L. Kinahan PE, Hasegawa BH, Beyer T. X-Ray based
      or both may depend on the indication; the selection
                                                                  attenuation correction for PET/CT scanners. Semin
      of the CT protocol to be used with PET/CT is under
                                                                  Nucl Med. 2003;33:166–179.
      evolution
                                                               M. Nakamoto Y, Chin BB, Kraitchman DL, et al. Effects of
  B. Role of reregistration to correct for respiratory and
                                                                  nonionic intravenous contrast agents at PET/CT imag-
      other motion artifacts
                                                                  ing: phantom and canine studies. Radiology. 2003;
  C. Role of axial collimation
                                                                  227:817–824.
  D. Role of respiratory gating for PET, CT, or both
                                                               N. Osman MM, Cohade C, Nakamoto Y, Wahl RH.
  E. Optimal distribution time for 18F-FDG, scan time per
                                                                  Clinically significant inaccurate localization of lesions
     bed position, image regularization (smoothing), re-
                                                                  with PET/CT: frequency in 300 patients. J Nucl Med.
     construction algorithm, and best method for CT-based
                                                                  2003;44:240–243.
     attenuation correction
                                                               O. Paquet N, Albert A, Foidart J, Hustinx R. Within-
  F. Optimal methods for semiquantitative measurements
                                                                  patient variability of 18F-FDG: standardized uptake
     (e.g., SUV)
                                                                  values in normal tissues. J Nucl Med. 2004;45:784–788.
                                                               P. PET-CT Consensus Conference, SNMTS, American
VII. CONCISE BIBLIOGRAPHY                                         Society of Radiologic Technologists (ASRT). Fusion
A. Antoch G, Freudenberg LS, Egelhof T, et al. Focal              imaging: a new type of technologist for a new type of
   tracer uptake: a potential artifact in contrast-enhanced       technology. J Nucl Med Technol. 2002;30:201–204.
   dual-modality PET/CT scans. J Nucl Med. 2002;               Q. Thie J. Understanding the standardized uptake value, its
   43:1339–1342.                                                  methods, and implications for usage. J Nucl Med. 2004;
B. Antoch G, Freudenberg LS, Stattaus J, et al. Whole-            45:1431–1434.
   body positron emission tomography-CT: optimized CT          R. Townsend DW, Beyer T, Blodgett TM. PET/CT scan-
   using oral and intravenous contrast materials. AJR.            ners: a hardware approach to image fusion. Semin Nucl
   2002;179:1555–1560.                                            Med. 2003;33:193–204.
C. Antoch G, Jentzen W, Freudenberg LS, et al. Effect of       S. Weber WA. Use of PET for monitoring therapy and
   oral contrast agents on computed tomography-based              predicting outcome. J Nucl Med. 2005;46:983–995.
   positron emission tomography attenuation correction in      T. Yau YY, Chan WS, Tam YM, et al. Application of
   dual-modality positron emission tomography/computed            intravenous contrast in PET/CT: does it really introduce
   tomography imaging. Invest Radiol. 2003;38:784–789.            significant attenuation correction error? J Nucl Med.
D. Beyer T, Townsend DW, Brun T, et al. A combined                2005;46:283–291.
   PET/CT scanner for clinical oncology. J Nucl Med.           U. Young H, Baum R, Cremerius U, et al. Measurement of
   2000;41:1369–1379.                                             clinical and subclinical tumour response using [18F]-
fluorodeoxyglucose and positron emission tomography:          physician when considering the circumstances presented.
   review and 1999 EORTC recommendations. European               Therefore, an approach that differs from the guidelines is
   Organization for Research and Treatment of Cancer             not necessarily below the standard of care. A conscientious
   (EORTC) PET Study Group. Eur J Cancer. 1999;                  practitioner may responsibly adopt a course of action
   35:1773–1782.                                                 different from that set forth in the guidelines when, in his
                                                                 or her reasonable judgment, that course of action is indi-
VIII. DISCLAIMER                                                 cated by the condition of the patient, limitations on avail-
   The Society of Nuclear Medicine has written and ap-           able resources, or advances in knowledge or technology
proved these guidelines as an educational tool designed to       subsequent to publication of the guidelines.
promote the cost-effective use of high-quality nuclear              All that should be expected is that the practitioner will
medicine procedures in medical practice or in the conduct        follow a reasonable course of action based on current
of research and to assist practitioners in providing appro-      knowledge, available resources, and the needs of the patient
priate care for patients. These guidelines should not be         to deliver effective and safe medical care. The sole purpose
deemed inclusive of all proper procedures or exclusive of        of these guidelines is to assist practitioners in achieving this
other procedures reasonably directed to obtaining the same       objective.
results. They are neither inflexible rules nor requirements of      Advances in medicine occur at a rapid rate. The date of a
practice and are not intended nor should they be used to         guideline always should be considered in determining its
establish a legal standard of care. For these reasons, the       current applicability.
Society of Nuclear Medicine cautions against the use of
these guidelines in litigation in which the clinical decisions   IX. APPROVAL
of a practitioner are called into question.                         These guidelines were approved by the Board of Direc-
   The ultimate judgment about the propriety of any spe-         tors of the Society of Nuclear Medicine on February 11,
cific procedure or course of action must be made by the          2006.
You can also read