Differences in reported backscatter factors, BSF, for low energy X-rays. A literature study
←
→
Page content transcription
If your browser does not render page correctly, please read the page content below
Institutionen för medicin och vård Avdelningen för radiofysik Hälsouniversitetet Differences in reported backscatter factors, BSF, for low energy X-rays. A literature study Carl A Carlsson Department of Medicine and Care Radio Physics Faculty of Health Sciences
Series: Report / Institutionen för radiologi, Universitetet i Linköping; 66 ISSN: 1102-1799 ISRN: LIU-RAD-R-066 Publishing year: 1991 © The Author(s)
-1- 1991-01-30 ISSN 0348-7679 Differences in reported backscatter factors, BSF, for low energy X-rays. A literature study. Carl A Carlsson Dept of Radiation Physics University of Linköping Report ULi-RAD-R-066 SSI P 587.90
-2- Differences in reported backscatter factors, nSF, for low energy X-rays. A literature study. Table of contents l. Introduction .... 3 2. Application of nSF in a code of practice • ••• 4 3. The concept backscatter factor, nSF • ••• 4 4. The concept hal f value layer, HVL • ..• 5 5. Determination of HVL • ••• 6 5.1 Calculations •••• 6 5.2 Measurements .... 7 5.3 Common errors in HVL-measurements · .•. 8 6. HVL as a measure of radiation quaiity .•. 10 7. Review of literature about backscatter factors •.. 12 7.1 BJR Suppl No 17 (1983) ... 12 7.2 IAEA No 277 (1987) ... 12 7.3 Grosswendt (1984) ... 12 7.4 Klevenhagen (1989) ... 13 7.5 Chan and Doi (1981) · .• 14 7.6 Harrison et al (1990) · .• 15 7.7 Grosswendt (1990) · .• 15 7.8 Knight and Nahum (1990) ... 16 8. Discussion · .. 16 9. Conclusions ... 19 10. Acknowledgement ... 20 11. References · •. 21
-3- 1. Introduction This work was ini tia ted by the large discrepancy in published values of the backscatter factor (BSF) as function of the half value layer (HVL) that exists in a new code of practice (IAEA, 1987) compared to the earlier, general ly used code (BJR, 1983). Change to the IAEA-code changes reported values of absorbed dose with up to 10%. These deviations apply for X-rays with HVL < 8 mm Al. Fig 1 shows BSF as function of field diameter with the HVL as a parameter. BSF HVL, mm Al 1.4 ------- === 4.0 // / / / / / / 1.3 / / / / / / / / / I 1.2 / 1.0 I I - - - - -------- -- --- --- / / ~ / /- // / // I 0.4 / 1.1 / / ------=-'--=-=---=-=---~, / , ----- I /1 ----::.-=:::- ------ / I , 0.1 /' ------------------------------ 1.0 '---- L-- L-- " ' -_ _- - - ' O 5 10 15 20 Field diameter, cm Fig 1: Backscatter factor as a function of field diameter and half value layer. From BJR (1983) with values from IAEA (1987) (Klevenhagen et al 1991) inserted (dotted lines).
-4- The aim of this work is to try to explain the discrepancy between the two sets of data in Fig 1. To reach this end, the concepts of BSF and HVL are first scrutinized and the experimental and calculational methods repor ted in the literature for their determination then critically reviewed. 2. Application of BSF in a code of practice In radiation therapy with X-rays, 10-300 kV, tabula ted values of central axis depth doses and BSF are mostly used instead of alaborious dose planning. Relative absorbed doses are given by BJR (1983) as [D(z)/D(O») • 100 as function of HVL and field diameter for different source to surface distances (8SD). The absorbed dose in tissue at depth z, D(z), is determined from a measurement of Ke,a i r at the en trance surface in the absence of scattering material (patient, phantom) and is given by D(z) = ~ Kc,alr . (1) D(O) where (~en)t~ssue p aIr is a weighted average of the ratio of the mass energy absorption coefficients for tissue and air (Carlsson and Alm Carlsson, 1990) . 3. The concept backscatter factor, BSF The backscatter factor can be defined as the quotient between the absorbed dose in the surface layer of a phantom to the absorbed dose in the same material in the absence of scattering material (the phantom), that is ~ f ('fE + 'fE ) ~dE p K O 'P ,s c'D + s + s BSF "'----=---..,,------- = (2) ro lJ en K D c,p P f 'f dE O E,p p
-5- Here is the energy fluence differentiated by photon energy E, indices p ~E Ilen and s stand for primary and secondary photons, respectively, and p is the mass energy absorption coefficient for tissue (water). The last step in Eq (2) is valid under charged particle equilibrium as then collision kerma is equal to absorbed dose (Alm Carlsson 1985). Grosswendt (1984) has expanded the definition to Il ~ )(~) dE E,s," p K BSF" Il (3) K ro J ~ (~) dE O E, P p K where" stands for phantom material and K for detector material. Corrections for determination of BSF = BSF~ from measured BSF~ are given. These corrections are small for nearly tissue equivalent materials as air and LiF. The determinations of BSF reviewed here have been made either experimentally (Klevenhagen 1989; Harrison et al 1990) or by Monte Carlo calculations (Chan and Doi 1981; Grosswendt 1984, 1990, Knight and Nahum 1990). In both experiments and calculations the detector should ideally be a point detector. The finite volume necessary to get sufficient energy impartation is preferably chosen as a thin plane detector as the dose gradient varies faster with depth than orthogonally to the normally incident central ray. 4. The concept half value layer, HVL BJR (1983) defines HVL as "The thickness of a specified material which attenuates the beam of radiation so that the exposure rate or absorbed dose rate at a point is reduced by half. The contribution of all scattered radiation, other than that which might be present in the beam initially, is deemed to be excluded". For calculation of HVL it is defined implicitly in Eq (4) ro Ilen HVL J (-p- )air ~E e- Il ' dE O (4) OJ }len J (--p--)air ~E dE O
-6- where ~ is the linear attenuation coefficient for the attenuator material. The other symbols are defined in Eq (2). Comment 1. In all publications cited here, HVL is defined for the quantities exposure X, air kerma K. aIr or air collision kerma Kc,alr .• These three HVL:s are practically identical. Comment 2. The radiation must be monodirectional. Otherwise the thickness of the attenuator is not well-defined. It is then necessary to modify the BJR- definition above and also exclude "scattered radiation present in the \Jeam ini tially". 5. Determination of HVL S.l Calculations With a known photon energy spectrum, RVL is calculated from Eq (4) by using different attenuation thicknesses, d, instead of RVL so that the quotients are both above and below 1/2. Interpolation gives HVL just as in measuring HVL. In early calculations of HVL it was common to regard the coherently scattered radiation as primary radiation. The coherently scattered photons were assumed to be strongly forward directed and to be detected as primary photons in measurements of HVL. In old tabulations of attenuation coefficients (Rubbeli 1969; Storm and Israel 1970) the total mass attenuation coefficient was tabula ted both with and without the contribution from coherent scattering. Calculations excluding the coherently scattered photons from the primary ones agreed better with measurements using narrow beam geometry (Carlsson 1963). Using modern cross sections for coherent scattering in liquid water derived from Morin (1982), Persliden and Alm Carlsson (1986) calculated the angular distribution of scattered photons behind water slabs and found a minimum of scattering at angles smaller than 2_S o • There is reas on to suppose that this effect is more evident in aluminium, supporting the idea that is possible to measure HVL with detection of a negligible amount of incoherently and coherently scattered photons.
-7- 5.2 Measurements To fulfil the requiremen ts of the definition of HVL, the measuremen ts mus t be made in narrow beam (good) geometry to avoid deteetion of seattered radiation from the attenuator (HVL-material). This eondition is best , o approaehed by a eollimation so that the opening angle, a, is less than ~ 4 and with the attenuator foils as elose to the souree as possible (Fig 2a). ej ,/ ,, - fT s ,, II M A °1 °2 Fig 2: Geometries used in HVL-measurements. S = radiation souree, M = monitor, D = diaphragm (eollimator), A = attenuator, a = opening angle of the beam, I = ionization ehamber (deteetor). a) Good geometry, minimal opening angle, minimal distanee souree to attenuator. b) Geometry propos ed by Trout et al (1960). Three positions of the attenuator - eollimator assembly is shown. e) Geometry propos ed by IAEA (1987).
-8- Trout et al (1960) made a thorough investigation of the performance of HVL- measurements and applied accurate corrections for the detected scattered radiation. A common practice to place the attenuator half-way between the source and detector originates presumably from this work. This, in turn, is based on the idea that the same area of the attenuator should be irradiated, independently of its position between source and detector (Fig 2b). With this presumption, the recommended position is correct. The presumption itself is, however, fundamentally wrong. The IAEA (1987) protocol has abandoned the idea of eons tant irradiated attenuator area (Fig 2c). The geometry is as in Fig 2a, but the introduction of a monitor, M in Fig 2c, means that the monitor may detect scattered radiation from the attenuator. In this case, the attenuator must be placed at a sufficient distance from both detector, I, and the monitor, M (Fig 2c). To get the same good result as with a measurement in the geometry in Fig 2a, the opening angle in Fig 2c has to be even smaller than in Fig 2a. Comment. If the monitor M in Fig 2c is in permanent use the discussion above is valid. On the other hand, if a monitor is introduced only for the HVL- measurement, it causes an extra filtration of the beam. The output and photon energy spectrum are changed as is the HVL. In this case, the monitor, M, should be placed outside the beam that reaches the detector, I (see Fig 2a). 5.3. Common er rors in HVL-measurements Fig 3 shows results from HVL-measurements in different geometries: (a) correct as in Fig 2a, (b) as in Fig 2b with scattered radiation detected by the ionization chamber, I, (c) as in Fig 2c with scattered radiation detected by the monitor M.
-9- Ke, air (d) Ke, air (O) 0.5 bJ aj ej 0.2 0.1 L - - - ' - - - - - - - ' - - - - - - ' - - - - 4 : - - O 1 2 3 Attenuator thickness, d, mm Fig 3: Results from HVL-measurements in different geometries. a) good geometry (Fig 2a) b) the ionization chamber, I, has detected scattered radiation (Fig 2b) c) the monitor, M, has detected scattered radiation (Fig 2c) In geometry b), HVL is overestimated while in geometry c) it is underestimated. The attenuator should ideally be perpendicular to all of the beam. In a divergent beam, the outer rays pass longer distances through the attenuator. They are thus more attenuated which should result in an underestimate of the HVL. This effect is, however, mostly negligible, and counteracts the effect of detection of scattered radiation by the ionization chamber. Detection of scattered radiation is generally the more severe problem in HVL-measurements. Influences of ionization chambers, purity of the attenuator material and measurements of their thickness on the accuracy of HVL-measurements are discussed in a recent paper by Wagner et al 1990.
-10- 6. HVL as a measure of radiation quality The same value of HVL can be obtained with different energy spectra. Two extremes are: a beam obtained with a thin window X-ray tube and no added filter and a monoenergetic beam with the same HVL. Fig 4 shows the energy distributions of the differential fluence, PE , and the air collision kerma, Kc,alr, . E' for a beam generated with 100 kV, 0.8 mm added aluminium filter (Seelentag et al 1979). The monoenergy (26.4 keV) that has the same HVL (1.61 mm Al) as the spectrum is inserted. dKc,air ,and dCP __ , relative units dE dE 1.0 ,-----.------..r,-r-,------.---.r----,,-----.----r-r-,---.----,--,---, r , I l C 66 r l I l I \ I \ 0.8 I I \ l I l I l , I \ I \ 0.6 I \ ,, I \ \ \ , I I \ 0.4 ., nll! I l \ \ \ ' , I III II I' \\ 'II" \ ~\ , , I 0.2 \ \ ... I l 'I ~ ...... l II - ",II. \ --- o L--=..L--L----LL.----l---L_L.----l---L_L-.:-y-_..1..---'------'-_-'--.-1 O 20 40 60 80 100 120 140 Photon energy, E, keY Fig 4: The relative energy distributions of the quantities fluence (solid line) and air collision kerma (broken line) for an X-ray beam 100 kV, 0.8 mm Al added filtration . (From Seelentag et al 1979). The monoenergy (26.4 keV) with the same HVL as the spectrum is inserted.
-11- Fig 5 shows the result of applying Eq 4 to the spectrum in Fig 4 as weIl as to the monoenergetic beam with the same HVL. 0.5 , " "'" "',',, - HVL~", 0.2 ", ", , " "o - - HVL -·I-----HVL2·---~---+-1 Attenuator thickness, d, mm Fig 5: Attenuation curves obtained with the beams presented in Fig 4. Solid line: 26.4 keV monoenergy, dashed line: 100 kV energy spectrum with the same HVL. Second HVL, HVL 2 , is indicated for the spectral beam. From Figs 4 and 5 one can conclude that the low energy photons are decisive for the HVL-value of the beam while the high energy photons determine the long distance penetration. The influence of the spectral shape on the values of BSF will be discussed later.
-12- 7. Review of literature about backscatter factors 7.1. BJR Suppl No 17 (1983) This dosimetry protocol (code of practice) gives BSF and percentage central axis depth dose data for different HVL, SSD and field areas. The tables in BJR (1983) are, for the HVL-ranges discussed here, identical with those in its earlier editions (BJR Suppl Nos 11 and 10 1972 and 1961) and contain material from BJR Suppl No 5 (1953). The tabula ted values in BJR (1983) are in many cases weighted averages of results from several investigators. The values in the later editions (1983, 1972, 1961) are basedon the use of smaller (thinner) ionization chambers than in the earlier one (1953), a fact that motivated the revision. The revised BSF-values generally exceed the previous ones. BJR (1983) states a scatter of ± 15% of their data in the HVL-range 0.01 to 0.7 mm Al. This scatter is explained by variations in measuring technique, phantom material and energy spectra. The BJR code gives correlated values of depth doses and BSF. 7.2. IAEA No 277 (1987) In this dosimetry protocol two methods for absolute determination of absorbed dose in the range 10-100 kV are mentioned. a) Use of a plane-parallel ionization chamber directly on the phantom and b) Measurement of air kerma free in air. In the latter case the re is a need for knowledge of the BSF •. BSF-values are tabulated in the code but there are no depth dos e values given since "the main dosimetric task in this photon energy range is the determination of the absorbed dose at the surface qf the phantom ". 7.3. Grosswendt (1984) The BSF-values tabula ted by IAEA (1987) are derived from this work. Grosswendt has calculated BSF with a Monte Carlo method. The calculations are weIl explained and seem reliable. Perhaps the use of a collision density estimator could increase the efficiency of the calculations (Persliden and Alm Carlsson, 1986). Grosswendt used eight published energy spectra for his calculations. The spectra were chosen so that no overlapping HVL:s occurred. These spectra from Seelentag et al (1979) are presented later in Grosswendt
-13- (1990). The paper (Grosswendt 1984) is more methodological than quantitative. The results are presented in small graphs showing general trends only. 7.4. Klevenhagen (1989) Klevenhagen has presented careful experiments with a thin, plane-parallel ionization chamber. He uses increased added filtration with increased tube potential. As a result no HVL-s overlap. The results from Klevenhagen agree weIl with those of Grosswendt (1984). Significant differences appear at the smallest field diameter and the highest HVL (1-4 mm Al). This is explained as an effect of the lateral extension of the ionization chamber being less close to a point on the central ray than the sphere simulated in the calculations. Other attempts to analyze the reason for small differences are not equally successful. For example, Klevenhagen cites Trout et al(1960) and states that Grosswendt has not accounted for the variation of beam quaiity (HVL) with changing field size. The measured HVL is said to decrease with decreasing field diameter. As has been discussed above, this effect is a systematic error in the HVL-measurement when using large field diameters (see Figs 2 and 3). There is, however, a variation of HVL over the field area due to attenuation of the bremsstrahlung in the anode. At the cathode side of the field, the HVL is lower and the photon fluence higher than on the central ray while the opposite is true (higher HVL and lower photon fluence) at the anode side (heel effect). HVL-measurements in different directions from a diagnostic X-ray tube have been repor ted by Carlsson (1963) and energy spectra in different directions by Svahn (1977). The heel effect is less pronounced using X-ray tubes with larger anode angles than those used in diagnostic units (10-20 0 ). The tube used for Grosswendt's calculations has an anode angle of 40 0 • Klevenhagen also tentatively states that his results are 1 - 1.5% higher than those of Grosswendt due to differences in their SSD:s used (Grosswendt 100 cm, Klevenhagen 15 and 25 cm). Ile claims that the more. divergent beam should give a higher BSF due to alarger phantom volume being involved in scattering interactions. On the contrary, it can be argued that the BSF should be lower with a shorter SSD as the same number of scatter interactions are then distributed over alarger volume. That is, the scattered photons have lo~ger distances to the point at zero depth on the
-14- central ray. Another possible explanation of these small differences could be that Klevenhagen in general produces his HVL:s with lower filtration and higher tube potential than Grosswendt. This will be further discussed below. Another criticism is that Klevenhagen discusses the energy distribution of scattered photons by means of effective energies. The effective energy, hveff , defined from ~(hveff) = ~~~ is a doubtful concept. The energy spectra of primary and scattered photons need be fully appreciated (Alm Carlsson and Carlsson, 1984; Carlsson and Alm Carlsson, 1990). 7.5. Chan and Doi (1981) Chan and Doi repor t Monte Carlo calculations of BSF for mammography units with both a Mo-anode (0.03 mm added Mo-filter) and a W-anode (1 mm added AI- filter). They used a geometry with all photons impinging perpendicularly on the phantom at the same point (pencil beam) and register all energy impartations throughout,a laterally infinite phantom. This approach is equivalent to calculate depth doses in the phantom irradiated with a broad beam of perpendicularly incident photons. The backscatter was received by extrapolating the depth doses to zero depth. They report BSF for four phantom materials, water, fat, 50% water + 50% fat, and Lucite. Their results for water agree weIl with those of Klevenhagen for a 20 cm field diameter. Chan and Doi repor t an interesting simulation of how a LiF-dosemeter (dimensions 3.2 x 3.2 x 0.89 mm 2 ) measures BSF. Tltey calculate the average absorbed dose in the dosemeter with and without the phantom present. The resulting BSF is significantly lower than that calculated without a dosemeter (up to 7% with a water phantom). These calculated "BSF:s" agree weIl with reported measurements using thermoluminescent LiF of the given dimensions. Their calculations of the average absorbed dose in the dosemeter are not presented in detail. It is somewhat astonishing that, in their calculations, use of a phantom equivalent dosemeter of the same dimensions as the LiF-dosemeter resuIts in approxilllately the same (only slightly . sIllaller) underestilIlate of the JlSF even if LiF has 2.5 to 3.5 higher density than the different phantom materials.
-15- 7.6. Harrison et al (1990) In order to avoid the large perturbation caused by the displacement of phantom material by the relatively large air volume in an ionization chamber (Liden 1961), Harrison et al performed BSF-measurements with thermoluminescence (TL) dosemeters. They used chips of Li ZB40 with 7 dimensions 3.Z x 3.Z x 0.9 mm 3 which, they assumed, would not significantly perturb the radiation field. Their precision in measured BSF is low. The given standard deviation is about 4%. This is caused by alow-precision TL-technique. TL-measurements with a precision of 0.3% has earlier been repor ted (Carlsson, Mårtensson and Alm Carlsson 1968; Alm Carlsson 1973). It should also have been advantageous to use much thinner TL-dosemeters to avoid the perturbation effect repor ted by Chan and Doi (1981). The scatter of their results is in general large. For one of the beams used, they repor t measured and calculated HVL:s of 0.93 and 1.77 mm Al, respectively. In general, their BSF-measurements agree better with IAEA (1987) than with BJR (1983). 7.7. Grosswendt (1990) Here, Grosswendt presents BSF-values for a great number of radiation qualities, SSD:s and field diameters. Contrary to the author's earlier work (Grosswendt 1984) this work is quantitative. The BSF-values are systematicallY presented in a table, that is a valuable resource for everyone with questions about BSF. Besides an increased number of energy spectra, three monoenergies (ZOO, 400 and 661) are introduced. The BSF- values are restricted to water as both detector and phantom. Coll~sion kerma, Kc ' is used instead of kerma, K, in the definition of BSF. This is to prefer as Kc is identical to absorbed dos e in charged particle equilibrium and the absorbed dos e is the important quantity. A further advantage with this definition is that it uses mass energy absorption coefficients instead of mass energy transfer coefficients (Eq Z instead of Eq 3). The former are ~he ones given in modern tabulations. The Monte Carlo technique has als o been changed. Now the absorbed dose on the central ray at zero depth is . calculated byestimating the spectral fluence of photons traversing a small area on the surface. Earlier (Grosswendt 1984), spheres of various radii . were used and values extrapolated to zero radius. Eq Z in Grosswendt 1990
-16- 1 dE dE (A I cosa 1- d4i = dn ) (Sa) could more correctly be written dEdeda.L (Sb) where the area dal is perpendicular to the photon's direction of motion a. If da is an area element on the surface traversed by dN photons then da~ = dalcosal. In the estimat~ of ~: ' finite values of the differentials must be used and an integration over all angles a be performed. A curious detail is that the aU thor has here accepted the cd ticism by Klevenhagen that the HVL.should increase when the field area increases (see the above review, Section 7.4). Grosswendt uses, however, correct HVL-values calculated for narrow beam geometry (Eq 4). 7.8. Knight and Nahum (1990) These authors have with Monte Carlo technique calculated values of BSF for monoenergetic photons and derived BSF for spectra by weighting. Their results, repor ted in a graph in this abstract, agree weIl with those of Grosswendt 1990. 8. Discussion In summary, among modern and old literature about backscatter factors Grosswendt's (1984, 1990) results seem the most reliable. The experiments by Klevenhagen (1989) are also very convincing at least with large field diameters. Other calculations and experiments verify the results of these two authors. All authors use the HVL for describing radiation quaiity and are aware of its limitations. Nevertheless, they use a series of spectra that is produced by increasing added filtration and tube potential simultaneously. No one has used different spectra with nearly the same HVL. In Fig 6, results of Grosswendt (1990) and Klevenhagen (1989) are compared. Fig 6a shows the added filtrations, Fig 6b the tube potentials and Fig 6c the backscatter factors. All these quantities are presented aS function of HVL. From Fig 6 a) and b) it is evident that for the same HVL Klevenhagen
-17- uses less added filter material and instead higher tube potential than Gro.sswendt (at least for HVL > 1 mm Al). Fig 6 c shows that the BSF:s from Klevenhagen are larger than those from Grosswendt at the large field diameter (20 cm) and vice versa at smaller field diameters (5 and 2 cm). Addod Al filt or, mm a 4 3 2 . .0- 0'-' o 2 4 5 Tuba potontlal, kV b 140 120 100 80 60 40- [7 20 ,! Ol-_ _~_ _~ ~ ·
-18- An explanation of Fig 6c could be: because Klevenhagen's beams have both lower and higher energies than those of Grosswendt with the same IIVL, they produce seattered photons of higher energies and the BSF:s saturate at larger field diameters. If, on the other hand, the field area is small, the high energy scattered p~otons have a large probability for escaping th~ phantom outside the entranee field. The low energy photons «20 keY) eontribute little to the BSF (Persliden and Alm Carlsson 1984). Maybe, Klevenhagen's BSF-values for small field diameters are in fact eorrect and not too low, as he himself suspected because of the geometry of his ionization chamber (see Seetion 7.4). The differences in the BSF-values presented in Fig 6 for equal HVL are examples on the weakness of HVL as a measure of radiation quaiity. The spectral distribution of air collision kerma, Kc,a i r, E. from a known fluenee speetrum ~E is given by (ef Eq 4) Ilen Kc,alr, • E = (-p-)air • E • ~ E (6) Ilen The product (---).p all' . E contains two terms, one deereasing the other increasing with inereasing photon energy, E. As seen in Fig 7, the product deereases very fast with increasing energy at energies up to 30 keY, has a broad minimum around 70 keY and increases slowly above 120 keY. As a result Kc,a i r, E is more dominated by low energy photons than is ~E' as illustrated in Fig 4. Crudely speaking, the low energy photons of a diagnostic X-ray sp~ctrum determine the HVL and the high energy photons the penetration power and energy eontent of the beam (Carlsson and Alm Carlsson, 1990). Going back to Fig 1, the differences between BJR (1983) and IAEA (1987) could partly be explained as caused by using lower filtration and higher tube potentials for the BJR BSF-values than for those of rAEA. The former could be better than suggested by all the modern investigators. It is, of course, time to exchange these old BSF-values as they are compilations of many investigators and as relatively large ionization chambers have been used. BJR give both BSF and pereentage depth doses; IAEA means that depth doses are of little interest with these spectra. This is in accordance with the practice at many hospitals. A new code of practice, Klevenhagen et al (1991) presents BSF derived from Grosswendt (1990), supported by data from
-19- both Knight and Nahum (1990) and Klevenhagen (1989). These data are, in fact, used instead of those from IAEA (1987) for producing Fig 1. 2 Ilen air' E -1J1,-. keV lp '''9 10 9 8 7 6 5 4 3 2 1 oL_-=:::::::::=====::;:::======- O 50 keV 100 150 E, Fig 7: The product of the mass energy absorption coefficient, ~en/p, and photon energy, E, as function of photon energy. 9. Conclusions The new backscatter factors given by Grosswendt (1990) and Klevenhagen (1989) seem reliable taking into account the energy spectra they are derived from. With knowledge of filter, tube potential, HVL, anode material and anode angle one can try to find the actual energy spectrum from the spectral catalogue (Seelentag et al 1979). If possible, one can, for a given HVL, change filter and tube potential to the values given by Grosswendt (1990) and use his BSF-values with confidence.
-20- The use of BSF from Grosswendt or IAEA together with depth doses from BJR JI1~Y, however, be problema tic. It should be interesting to get BSF-values for beams with very broad spectra, e.g., from X-ray tubes with Be-windows and no added filter and also for the other extreme, monoenergetic radiation. The problem of the perturbation of a TL-dosemeter on the determination of BSF (Chan and Doi 1981) should benefit from further investigations. Experimental determinations of BSF suffer of insufficient knowledge of the energy distribution of the scattered radiation. Publication of such results from Monte Carlo calculations should be valuable. 10. Acknowledgement This work was initiated by Dr Jonas Karlberg and has been supported by grants from the Swedish National Institute of Radiation Protection, SSI.
-21- 11.· References Alm Carlsson G (1973). Dosimetry at interfaces. Acta Radiol Suppl 332, Stockholm. Alm Carlsson G (1985). Theoretical basis for dosimetry. In The dosimetry of ionizing radiation Vol I, Eds Kase, Bjärngard, Attix, Academic Press, Orlando, pp 1-75. Alm Carlsson G and Carlsson C A (1984). Effective energy in diagnostic radiology. A critical review. Phys Med Biol 29, 953-958. BJR (1953). Central axis depth dos e data. Brit J Radiol Suppl No 5. BJR (1961). Depth dose tables for use in radiotherapy. Brit J Radiol Suppl No 10. BJR (1972). Central axis depth dos e data for use in radiotherapy. Eds Cohen, Jones and Greene. Brit J Radiol Suppl No 11. BJR (1983). Central axis depth dose data for use in radiotherapy. Brit J Radiol Suppl No 17. Carlsson C (1963). Determination of integral absorbed dose from exposure measurements. Acta Radiol Ther, Phys Biol !' 433-458. Carlsson C A and Alm Carlsson G (1990). Dosimetry in diagnostic radiology and computerized tomography. In The dosimetry of ionizing radiation. Vol III. Eds Kase, Bjärngard and Attix. Academic Press, San Diego, pp 163-257. Carlsson C A, Mårtensson B K A and Alm Carlsson G (1968). High precision dosimetry using thermoluminescent LiF. Proc 2nd Int Gonf Lum Dosimetry. USAEG Rept CONF-680920. Chan H-P and Doi K (1981). Monte Carlo simulation studies of backscatter factors in mammography. Radiology 139, 195-199.
-22- Grosswendt B (1984). Backscatter factors for X-rays generated at voltages between 10 and 100 kV. Phys Med Biol 29, 579-591. Grosswendt B (1990). Dependence of the photon backscatter factor for water on source-to-phantom distance and irradiation field size. ,Phys Med Biol 35, 1233-1245. Harrison R M, Walker C and Aukett R J (1990). Measurement of backscatter factors for low energy radiotherapy (0.1 - 2.0 mm Al HVL) using thermoluminescence dosimetry. Phys Med Biol 35, 1247-1254. Hubbell J H (1969). Photon cross sections, attenuation coefficients and energy absorption coefficients from 10 keY to 100 GeV. Report NSRDS-NBS 29. National Bureau of Standards, Washington. IARA (1987). Absorbed dose determination in photon and electron beams. An international code of practice. Techn Rep Ser No 277. International atomic energy agency, Vienna. Klevenhagen S C (1989). Rxperimentally determined backscatter factors for X- rays generated at voltages between 16 and 140 kV. Phys Med Biol 34, 1871-1882. Klevenhagen S C, Aukett R J, Burns J R, Harrison R M, Knight R T, Nahum A R and Rosser K R (1991). Report of the Institute of physical sciences in medicine working party on low- and medium energy X- ray dosimetry. Submitted to Phys Med Biol. Knight R T and Nahum A E (1990). Improved backscatter fac\ors for low energy X-rays. Abstract. Liden K V H (1961). Errors introduced by finite size of ion chambers in depth dos e measurements, pp 161-172 in Selected topies in radiation dosimetry (Proc lnt Con f Vienna), International atomic energy agency, Vienna. Morin LRM (1982). Molecular form factors and photon coherent scattering cross sections of water. J Phys Chem Ref Data ~, 1091-1098.
-23- Persliden J and Alm carlsson G (1984). Energy imparted to water slabs by photons in the energy range 5-300 keY. Calculations using a Monte Carlo photon transport model. Phys Med Biol 29, 1075- 1088. Persliden J and Alm Carlsson G (1986). Calculation of the small-angle distribution of scattered photons in diagnostic radiology using a Monte Carlo collision density estimator. Med Phys 13, 19-24. Seelentag V V, Panzer V, Drexler G, Platz L, Santner F (1979). A catalogue of spectra for the calibration of dosemeters. GSF Bericht 560. Gesellschaft fUr Strahlen - und Umweltforschung MBH, MUnchen. Storm E and Israel H I (1970). Photon cross sections from 1 keY to 100 MeV for elements Z = l to Z = 100. Nuclear Data Tables A7, 565-681. Academic Press, New York and London. Svahn G (1977). Diagnostic X-ray spectra. A study of the effects of high tension ripple, large x-ray tube currents, extra-focal radiation and anode angulation with Ge (Li)-spectrometry. Thesis, Radiation Physics Dept, University of Lund, 'X-221 85 Lund, Sweden. Trout D E, Kelley J P and Lucas A C (1960). Determination of half-value layer. Radiology 84, 729-740. Wagner L K, Archer B R and Cerra F (1990). On the mesurement of half-value layer in film-screen mammography. Med Phys 17, 989-997.
You can also read