Determination of Anions in Fracking Flowback Water From the Marcellus Shale Using Automated Dilution and Ion Chromatography

Page created by Charlotte Cobb
 
CONTINUE READING
Determination of Anions in Fracking Flowback Water From the Marcellus Shale Using Automated Dilution and Ion Chromatography
Determination of Anions in Fracking

                                                                                                                                              Te ch ni cal N ote 1 3 9
                   Flowback Water From the Marcellus
                   Shale Using Automated Dilution
                   and Ion Chromatography
                   Carl Fisher, Richard Jack, and Linda Lopez
                   Thermo Fisher Scientific, Sunnyvale, CA, USA

Key Words
Dionex AS-AP Autosampler, Dionex IonPac AS18, Hydraulic Fracturing,
In-line Conductivity

Goal
Demonstrate the use of in-line conductivity measurement to trigger
automated dilution prior to separation and quantification of inorganic
ions and organic acids in Marcellus Shale fracking flowback samples
using ion chromatography.

                   Introduction
                   The rapid increase in U.S. natural gas production in
                   recent years has been propelled by the extensive use of
                   hydraulic fracturing (also known as fracking). This
                   process extracts natural gas by drilling into bedrock
                   (primarily shale) and then injecting fluid under high
                                                                                 Organic acids, such as acetate and formate, are added to
                   pressure causing cracks in bedrock, thereby releasing
                                                                                 fracking fluid to adjust the pH.6 The pH of fracking fluid
                   trapped gas that can be captured.1 Fracking fluid contains
                                                                                 is maintained within a narrow range to ensure the
                   approximately 85% water and 13% sand (proppant), the
                                                                                 effectiveness of surfactant additives that facilitate the
                   latter to prop open cracks, which facilitates the flow of
                                                                                 penetration of fracking solution into bedrock. These
                   gas. The remainder consists of chemical additives such
                                                                                 organic acids can act as a carbon source to promote the
                   as friction reducers, anti-bacterial agents, and corrosion
                                                                                 growth of anaerobic bacteria, resulting in the production
                   inhibitors. 2
                                                                                 of corrosive and toxic hydrogen sulfide. Additionally,
                   Fracking has been in use since the 1940s primarily for        because not all of the fracking solution is recovered,
                   oil extraction, however recent refinements to this process    bacteria can grow within the interstitial cracks following
                   have enabled the exploitation of energy resources that had    a fracking event and block gas flow over the long term,
                   previously been inaccessible. 3 While fracking provides       inhibiting well production. Thus, the concentration of
                   financial benefits to both local and national economies,      organic acids should be closely monitored.
                   it has not been without controversy. Inadvertent spills or
                                                                                 The high levels of dissolved salts leached from bedrock
                   the dumping of fracking flowback (fracking solution that
                                                                                 are a challenge to analyzing fracking flowback solutions.
                   returns to the surface via the well bore) into unlined
                                                                                 Injection of high ionic strength solutions can exceed the
                   collection ponds can contaminate ground water.4
                                                                                 capacity of columns used in ion chromatography,
                   Additionally, high levels of chloride and bromide found       resulting in poor chromatography, peak suppression
                   in fracking wastewater can result in effluent discharged      (i.e., poor recovery), and inaccurate reporting. Additionally,
                   by wastewater treatment facilities that exceed the levels     high concentrations may exceed the linear calibration
                   allowed in the facility’s discharge permits. 5 Chlorination   range for a particular analyte. These samples require
                   or ozonation, as part of the purification process, can        dilution prior to injection, which results in an additional
                   produce disinfection byproducts (DBP) such as bromate,
                   chlorate, and haloacetic acids. These DBPs are carcinogens
                   or interfere with human development and, consequently,
                   are regulated by the U.S. Environmental Protection
                   Agency’s National Primary Drinking Water Regulations.
2

                          sample preparation step and the possibility of errors. To                     Reagents and Standards
                          minimize errors and reduce labor costs, there are a number                    • 18 MW-cm resistivity degassed deionized (DI) water
                          of options available to automate the determination of
                                                                                                        • T hermo Scientific Dionex 1000 µS conductivity
                          the need for and then the loading of less sample:
                                                                                                           standard (P/N 080172)
                          •T
                            he AutoDilution feature of Thermo Scientific™                              • U ltra Scientific 1000 mg/L Certified IC Standards or
                           Dionex™ Chromeleon™ Chromatography Data System                                  Fisher Scientific reagents, ACS grade
                           (CDS) software. If a chromatogram has an analyte
                           peak height or area that exceeds a pre-set limit, a                          Samples
                           smaller amount of sample is re-injected using either a:                      Fracking flowback solutions from the Marcellus Shale
                           a. Partial loop                                                              (F1–10) were collected from successively later times
                           b. Smaller loop                                                              during the fracking process.
                           c. V
                               ial to vial dilution using a Thermo Scientific
                                                                                                        Conditions
                              Dionex AS-AP Autosampler
                                                                                                        Chromatographic
                          • Measurement of sample conductivity prior to injection.
                                                                                                        Columns:	Thermo Scientific™ Dionex™ IonPac™
                             This can be done a) manually, which is tedious, labor
                                                                                                                  AG18/AS18, 4 mm
                             intensive, and prone to error, or b) automatically,
                             using the Thermo Scientific Dionex AS-AP Sample                            Eluent Source:                 Thermo Scientific Dionex EGC III KOH Cartridge
                             Conductivity and pH Accessory. If the conductivity                         Eluent:                        23 mM or 39 mM KOH
                             exceeds a specified amount, Chromeleon CDS can
                                                                                                        Flow Rate:                     1.0 mL/min
                             trigger the loading of less sample as per a–c above.
                                                                                                        Column Temp.                   30 °C
                          These options are depicted in Figure 1 and are described
                          in greater detail in Technical Note 138.7                                     Inj. Volume:                   25 µL
                                                                                                        Detection: Suppressed conductivity, Thermo Scientific™
                                       Dionex AS-AP
                                                                                                        	         Dionex™ ASRS™ 300 Anion Self-Regenerating
                                       Autosampler                                                                 Suppressor*, recycle mode, 57 mA (23 mM
    Fracking Sample
                                                                                                                   KOH); 97 mA (39 mM KOH)
                                     Sample Conductivity
                                      and pH Accessory                                                  Background                     < 2 µS
                                                                                                        Conductance:
     Centrifugation
                                                                                          IC System     Noise:                         < 3 nS
                                        Does conductivity No
                                         exceed cutoff?                                                 System Backpressure: 2100 psi
       Filtration
                                                 Yes                                     Chromatogram   * E quivalent or improved results can be obtained using the Thermo Scientific™ Dionex™
                                                                                                          AERS™ 500 Anion Electrolytically Regenerated Suppressor.
                                        Sample Dilution
                                                               Yes                  No
                                                                     Does peak area
                                                                                            Report      Dionex AS-AP Autosampler Dilution
                                                                     or height exceed
                                                                          cutoff?                       Syringe:                       1 mL
                                                               Chromeleon AutoDilution
                                                                   option required
                                                                                                        Buffer Line Volume:            1.2 mL
                                                                                                        Wash Solution: 	18 MW-cm resistivity degassed DI water
Figure 1. Automated sample analysis options available using the Dionex AS-AP Autosampler and
                                                                                                                         from autosampler wash liquid reservoir
Chromeleon software.
                                                                                                        Vials:                         10 mL
                          This Technical Note describes the analysis of fracking
                                                                                                        Dilution Volumes:              20 µL sample into 1980 µL DI water
                          flowback solutions using ion chromatography following
                          in-line determination of sample conductivity and                              Draw/Dispense Speed: 50/25 µL/sec (water)
                          automated dilution of high ionic strength samples.                                                 10/5         (sample)
                                                                                                        Mix by Shake Iterations: 10
                          Equipment
                                                                                                        Carousel Speed:                30,000 encoder counts (maximum, default)
                          • T hermo Scientific™ Dionex™ ICS-2100 Integrated
                            Reagent-Free™ IC (RFIC ™) System with an additional
                                                                                                        Dionex AS-AP Sample Conductivity and pH Accessory
                             10-port valve (P/N 069473)
                                                                                                        Volume Drawn:                  600 µL
                          • Dionex AS-AP Autosampler
                                                                                                        Draw/Dispense Speed: 20/10 µL/sec
                            – Dionex AS-AP Sample Conductivity and pH
                                                                                                        Part numbers of the consumables used in this document
                               Accessory (P/N 074923) with the included pH
                                                                                                        are listed in Table 1.
                               electrode port plug and seal installed

                          • Chromeleon CDS, version 6.8 (release 6.80 DU11c
                             or later) or 7 (release 7.1 SR1 or later)
3

Table 1. Consumables list.                                                                                                  Dionex ICS-2100 IC
                                                                                                                            Reagent-Free System
                                                         Part    Dionex AS-AP Autosampler
     Product Name                 Description                                                                                      Regen
                                                        Number   Fresh 18 MΩ-cm
                                                                 Deionized Water
 Dionex EGC III KOH          Eluent generator
                                                        074532                                               Eluent In
                                                                                                                          ERS 500
                                                                                                                                               CD
 cartridge                   cartridge                                                                                   Suppressor

 Thermo Scientific           Continuously regenerated
                                                        060477
 Dionex CR-ATC column        anion trap column                       Syringe
                                                                                             Needle               Columns
 Dionex IonPac AS18          Separation column,
                                                        060549
 column                      4 mm i.d.                                Tray Carousel

 Dionex IonPac AG18                                                                                                 Heat                                          Fresh 18 MΩ-cm
                             Guard column, 4 mm i.d.    060551                                                      Exchange                                      Deionized Water
 column                                                                                                             Tubing                       Regen
 Thermo Scientific™
                             Suppressor cartridge,                                                                             Degas
                                                                                                                                               CR-TC       EGC III        Pump*
 Dionex™ AERS™ 500                                      082540                                                                 Module
                             4 mm i.d. columns                                        Sample Loop
 suppressor                                                                                                       Backpressure
                                                                                                                      Coil
                             Polystyrene vials with                                                                                              Waste
 Dionex AS-AP
                             caps and blue septa,       074228   Figure 2. Flow diagram for the Dionex ICS-2100 IC Reagent-Free System with 10-port valve.
 Autosampler vials
                             10 mL, package of 100

Sample and Standard Preparation
                                                                                              Wash Liquid Line
Samples were centrifuged at 10,000 × g, 10 min to pellet
solid particles, which facilitated filtration. The resultant          To                            Buffer Line
                                                                 Syringe
supernatant was filtered with a 0.2 µm polyethersulfone                                      Syringe Waste Line
(PES) syringe filter. Samples were diluted as required                                                                                                          Wash
                                                                                                                                                               Reservoir
with degassed 18 MW-cm resistivity DI water.
                                                                                           Sample Conductivity
                                                                               Eluent                                    Eluent
It is important to use 18 MW-cm resistivity, DI water for                                   and pH Accessory
                                                                                   In                                    Out
eluent and autosampler flush solutions to avoid system
contamination, decreased sensitivity, and poor calibration.
Degassing the DI water by vacuum filtration prior to use                                   Accessory
                                                                                            Inlet Line
is a good laboratory practice.
                                                                                                    Auxiliary Valve                                               3       2
Standards were prepared by diluting Certified IC Standard                                             Bypass Line
                                                                                                                                                              4               1
stock solutions or by dissolving ACS grade reagents with                                               Accessory
                                                                                                       Waste Line                                                 5       6
DI water.
                                                                                                                                                             Auxiliary Valve
Instrument Setup and Installation                                                                     Sampling Needle                                         (B Position)

Install, hydrate, and condition the Dionex EGC III KOH
                                                                                                                                                                              L
eluent generator cartridge. This cartridge requires at least                             Carousel                                                                     5   4
2000 psi of backpressure for optimal operation. Install
                                                                                                                                                          W 6           3 C
and hydrate the Dionex CR-ATC trap column and the
                                                                                                                                                                   12
Dionex AERS 500 Anion Electrolytically Regenerated                                                                                                                L  P
Suppressor. Figure 2 outlines the flow path of the standard                                                                                                 Injection Valve
Dionex ICS-2100 system setup with a Dionex AS-AP                                                              Injection Port          Sample Transfer Line (Inject Position)

Autosampler. Hydrate the devices and finish the system
setup according to the product manuals and the Dionex            Figure 3. Dionex AS-AP plumbing schematic for Push Mode with a Sample Conductivity and pH Accessory.
ICS-2100 Operator’s Manual.8–12 Install and condition the
column using the conditions described in that column’s
Quality Assurance Report (QAR) for 30 min prior to
installing them in-line with the suppressor. If total
system backpressure is < 2000 psi, install one or more
backpressure coils between the Dionex CR-ATC
Continuously Regenerating Anion Trap Column and
the injection valve to increase the system pressure to
2000–3000 psi.

Install the Dionex AS-AP Sample Conductivity and
pH Accessory in the Dionex AS-AP Autosampler as
outlined in the Accessory Setup and Operation Guide.13
Figure 3 depicts the setup required for Push mode with
this accessory.
4

    For the experiments presented here, the 10-port valve of        Columns:         Dionex IonPac AG18/AS18, 4 mm
                                                                    Eluent Source:   Dionex EGC III KOH cartridge
    the Dionex ICS-2100 system was used as the auxiliary            Eluent:          39 mM KOH
    valve. Alternatively, if the optional high-pressure valves      Flow Rate:       1 mL/min
                                                                    Inj. Volume:     25 µL
    are installed in the Dionex AS-AP Autosampler, they can
                                                                    Col. Temp.:      30 °C
    be used to divert sample to the Dionex AS-AP Sample             Detection:       Suppressed conductivity, Dionex ASRS 300 Suppressor,
    Conductivity and pH Accessory. Install the pH electrode                          recycle mode
                                                                    Sample:          100-fold fracking flowback, filtered, 0.2 µm
    port plug, which reduces the amount of injected sample
    from 2000 to 600 µL for conductivity determination. Prior                        Peaks:               Measured         Undiluted
                                                                                     1. Acetate          < 0.05 mg/L
5

   Column:        Dionex IonPac AG18/AS18, 4 mm
   Eluent:        23 mM KOH                                                                                         350
   Flow Rate:     1 mL/min
                                                                                                                    300
   Inj. Volume:   25 µL
   Col. Temp.:    30 °C                                                                                             250

                                                                                       Area ( µS*min)
   Detection:     Suppressed conductivity, Dionex ASRS 300 Suppressor,
                  recycle mode                                                                                      200
   Sample:        A. Standard                                                                                                                                                 2
                                                                                                                    150                                                       r = 1.00
                  B–E. 100-fold fracking flowback F1, F2, F5, and F10, filtered,
                  0.2 µm                                                                                                                                                      %RSD= 1.12
                                                                                                                    100
                  Peaks:                      A
                  1. Fluoride                0.5 mg/L                                                                 50
                  2. Acetate                 2.5
                                                                                                                       0
                  3. Propionate/Formate       –
                  4. Formate                 1.0                                                                            0    200    400    600    800 1000 1200 1400 1600 1800 2000
  4.8                                                                                                                                                 Concentration (mg/L)
                                                                                        Figure 6. Chloride standard curve.
                                  1

                                             2           3                              Change in Ion Concentrations During Fracking
                                                              4                         As fracking progresses, the level of anions in flowback water
                                                                                        changes. As can be seen in Figures 7 and 8, chloride and
                                                                                        bromide concentrations increased ~10-fold from the first
                                                                                        to the second fraction and then, in subsequent fractions,
                                                                                        showed a slower, but steady increase. This result suggests
   µS                                                                                   that, as fracking proceeds, progressively more of these ions
                                                                                        are dissolved into the flowback solution from the bedrock.
           A
                                                                                        Conversely, the concentration of acetate decreased sharply
           B                                                                            and then remained at low levels from fraction two onwards.
                                                                                        These elevated concentrations would increase the amount
           C
                                                                                        of processing required by water treatment facilities. In a
           D                                                                            similar manner, what appears to be propionate (or a
                                                                                        mixture with formate) in the first fraction is no longer
           E
                                                                                        present in measureable quantities in subsequent fractions.
    0                                                                                   In contrast to the other analytes quantified, the level of
     2.5                      3                         3.5                        4    sulfate remained relatively constant, averaging 13 mg/L.
                                        Minutes
                                                                                                                      100,000
Figure 5. Comparison of early eluting inorganic anion and organic acids in                                             90,000
fracking flowback water.                                                                                               80,000
                                                                                                                       70,000
                                                                                               Concentration (mg/L)

Chloride Standard Curve Determination                                                                                  60,000
One concern with using concentrated samples is that the                                                                50,000
conductivity measured for an analyte will be outside of                                                                40,000
the linear quantification range. For the wastewater samples                                                            30,000
                                                                                                                       20,000
analyzed in this Technical Note, the primary analyte for
                                                                                                                       10,000
which this situation might arise is chloride. To confirm
                                                                                                                            0
that the concentrations of diluted samples were within                                                                           F1    F2     F3     F4   F5     F6    F7    F8   F9    F10
the linear range for chloride, a standard curve was                                                                                                       Fraction
generated. As can be seen in Figure 6, the standard curve                               Figure 7. Concentration of chloride in fracking flowback samples.
obtained was linear from 1–2000 mg/L with an r2 of 1.00
                                                                                                                       1,000
and %RSD of 1.12. This result indicates that even a
sample solution containing 200,000 mg/L of chloride, if                                                                    900

diluted 1:100, would still fall within the range for which                                                                 800

an accurate calculation of concentration can be obtained                                                                   700
for this analyte.                                                                                                          600
                                                                                             Concentration (mg/L)

                                                                                                                           500                                                     Bromide
                                                                                                                           400                                                     Acetate
                                                                                                                           300                                                     Sulfate
                                                                                                                           200                                                     Formate
                                                                                                                           100
                                                                                                                            0
                                                                                                                                 F1    F2     F3     F4   F5     F6    F7    F8   F9   F10
                                                                                                                                                          Fraction
                                                                                        Figure 8. Concentration of inorganic anions and organic acids in fracking
                                                                                        flowback samples.
One explanation for the apparent decrease in acetate                                                            References

                                                                                                                                                                               Te ch ni cal N ote 1 3 9
and propionate/formate is that, at high chloride                                                                  1. A merican Water Works Association (AWWA). Water
concentrations, the capacity of the column for these                                                                  and Hydraulic Fracturing: A White Paper from the
analytes is diminished. To investigate this possibility,                                                              American Water Works Association, American
acetate was added to 100-fold dilutions of fractions five                                                             Water Works Association, http://www.awwa.org/
and ten at a concentration 50% of that already present.                                                               fracturing. Denver, CO, 2013.
As shown in Table 2, the recovery of acetate was 114                                                              2. Bomgardner, M. Cleaner Fracking. C&EN, 2012,
and 103%, for fractions 5 and 10, respectively. If high                                                              90(142), 13–16.
chloride levels had decreased the binding of acetate to
                                                                                                                  3. U.S. Environmental Protection Agency (EPA). Study
the column, a lower recovery would have been expected,                                                                of the Potential Impacts of Hydraulic Fracturing on
particularly for fraction 10, which had the highest                                                                   Drinking Water Resources: Progress Report;
chloride concentration. Since this was not the case, the                                                              EPA/601/R-12/011; U.S. Environmental Protection
large drop in acetate measured indicates a decrease in                                                                Agency, Office of Research and Development:
the amount present in fracking flowback water samples                                                                 Washington, DC, 2012.
after the first fraction.
                                                                                                                  4. O
                                                                                                                      nishi, N. Fracking Tests Ties Between California
                                                                                                                     ‘Oil and Ag’ Interests. New York Times, June 2,
Table 2. Recovery of acetate in fracking flowback water.
                                                                                                                     2013, A14.
                    Amount Acetate Amount Acetate                                                                 5. K iparsky, M. and Hein, J. F. Regulation of Hydraulic
  Fraction            Measured        Added                                     Recovery (%)                          Fracturing in California: A Wastewater and Water
                       (µg/L)         (µg/L)                                                                          Quality Perspective. Wheeler Inst. For Water Law
       F5                    230                           110                          114                           and Policy, UC Berkeley School of Law, Berkeley,
      F10                    160                            80                          103                           CA, 2013.
                                                                                                                  6. Frac Focus Chemical Disclosure Registry, What
                                                                                                                      Chemicals are Used Page. http://fracfocus.org/
Conclusion                                                                                                            chemical-use/what-chemicals-are-used (accessed
This Technical Note demonstrates that by using the                                                                   Aug 22, 2013).
Dionex AS-AP Sample Conductivity and pH Accessory                                                                 7. T hermo Fisher Scientific. Thermo Scientific Technical
in combination with Chromeleon CDS, samples                                                                          Note 138, Accurate and Precise Automated Dilution
containing high levels of anions, such as fracking flowback                                                          and In-line Conductivity Measurement Using
wastewater, can be automatically diluted prior to                                                                    the AS-AP Autosampler Prior to Analysis by Ion
injection if a preset conductivity maximum is exceeded.                                                               Chromatography, TN70764_E, Sunnyvale, CA, 2013.
This prescreening eliminates repeat analysis, thereby                                                             8. T
                                                                                                                      hermo Fisher Scientific. Product Manual for Eluent
reducing eluent waste, extending column life, and                                                                    Generator Cartridges. Doc. No. 065018-04,
providing fast, accurate results.                                                                                    Sunnyvale, CA, 2012.
                                                                                                                  9. T hermo Fisher Scientific. Product Manual for the
                                                                                                                      Dionex CR-TC. Doc. No. 031910-03, Sunnyvale,
                                                                                                                      CA, 2010.
                                                                                                                10. T hermo Fisher Scientific. Product Manual for the
                                                                                                                     Dionex ERS 500 Suppressor. Doc. No. 031956-08,
                                                                                                                     Sunnyvale, CA, 2013.
                                                                                                                11. T
                                                                                                                     hermo Fisher Scientific. Dionex ICS-2100 Ion
                                                                                                                    Chromatography System Operator’s Manual. Doc.
                                                                                                                    No. 065291, Sunnyvale, CA, 2012.
                                                                                                                12. T
                                                                                                                     hermo Fisher Scientific. Dionex AS-AP Autosampler
                                                                                                                    Operator’s Manual. Doc. No. 065361, Sunnyvale,
                                                                                                                    CA, 2012.
                                                                                                                13. T hermo Fisher Scientific. Dionex AS-AP Sample
                                                                                                                    Conductivity and pH Accessory Setup and Operation
                                                                                                                    Guide. Doc. No. 065470, Sunnyvale, CA, 2012.

www.thermoscientific.com/dionex
©2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization.
All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an
                                                                                                                                           Thermo Fisher Scientific,
example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any             Sunnyvale, CA USA is
manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change.           ISO 9001:2008 Certified.
Not all products are available in all countries. Please consult your local sales representative for details.

Australia +61 3 9757 4486                 Finland +358 9 3291 0200                      Korea +82 2 3420 8600         Taiwan +886 2 8751 6655
Austria +43 810 282 206                   France +33 1 60 92 48 00                      Netherlands +31 76 579 55 55 UK/Ireland +44 1442 233555
Belgium +32 53 73 42 41                   Germany +49 6103 408 1014                     Norway +46 8 556 468 00       USA and Canada +847 295 7500
Brazil +55 11 3731 5140                   India +91 22 6742 9494                        Singapore +65 6289 1190
China +852 2428 3282                      Italy +39 02 950 591                         Sweden +46 8 556 468 00
Denmark +45 70 23 62 60                   Japan +81 6 6885 1213                         Switzerland +41 61 716 77 00                  TN70773_E 09/13M
You can also read