Determination of Anions in Fracking Flowback Water From the Marcellus Shale Using Automated Dilution and Ion Chromatography
←
→
Page content transcription
If your browser does not render page correctly, please read the page content below
Determination of Anions in Fracking Te ch ni cal N ote 1 3 9 Flowback Water From the Marcellus Shale Using Automated Dilution and Ion Chromatography Carl Fisher, Richard Jack, and Linda Lopez Thermo Fisher Scientific, Sunnyvale, CA, USA Key Words Dionex AS-AP Autosampler, Dionex IonPac AS18, Hydraulic Fracturing, In-line Conductivity Goal Demonstrate the use of in-line conductivity measurement to trigger automated dilution prior to separation and quantification of inorganic ions and organic acids in Marcellus Shale fracking flowback samples using ion chromatography. Introduction The rapid increase in U.S. natural gas production in recent years has been propelled by the extensive use of hydraulic fracturing (also known as fracking). This process extracts natural gas by drilling into bedrock (primarily shale) and then injecting fluid under high Organic acids, such as acetate and formate, are added to pressure causing cracks in bedrock, thereby releasing fracking fluid to adjust the pH.6 The pH of fracking fluid trapped gas that can be captured.1 Fracking fluid contains is maintained within a narrow range to ensure the approximately 85% water and 13% sand (proppant), the effectiveness of surfactant additives that facilitate the latter to prop open cracks, which facilitates the flow of penetration of fracking solution into bedrock. These gas. The remainder consists of chemical additives such organic acids can act as a carbon source to promote the as friction reducers, anti-bacterial agents, and corrosion growth of anaerobic bacteria, resulting in the production inhibitors. 2 of corrosive and toxic hydrogen sulfide. Additionally, Fracking has been in use since the 1940s primarily for because not all of the fracking solution is recovered, oil extraction, however recent refinements to this process bacteria can grow within the interstitial cracks following have enabled the exploitation of energy resources that had a fracking event and block gas flow over the long term, previously been inaccessible. 3 While fracking provides inhibiting well production. Thus, the concentration of financial benefits to both local and national economies, organic acids should be closely monitored. it has not been without controversy. Inadvertent spills or The high levels of dissolved salts leached from bedrock the dumping of fracking flowback (fracking solution that are a challenge to analyzing fracking flowback solutions. returns to the surface via the well bore) into unlined Injection of high ionic strength solutions can exceed the collection ponds can contaminate ground water.4 capacity of columns used in ion chromatography, Additionally, high levels of chloride and bromide found resulting in poor chromatography, peak suppression in fracking wastewater can result in effluent discharged (i.e., poor recovery), and inaccurate reporting. Additionally, by wastewater treatment facilities that exceed the levels high concentrations may exceed the linear calibration allowed in the facility’s discharge permits. 5 Chlorination range for a particular analyte. These samples require or ozonation, as part of the purification process, can dilution prior to injection, which results in an additional produce disinfection byproducts (DBP) such as bromate, chlorate, and haloacetic acids. These DBPs are carcinogens or interfere with human development and, consequently, are regulated by the U.S. Environmental Protection Agency’s National Primary Drinking Water Regulations.
2 sample preparation step and the possibility of errors. To Reagents and Standards minimize errors and reduce labor costs, there are a number • 18 MW-cm resistivity degassed deionized (DI) water of options available to automate the determination of • T hermo Scientific Dionex 1000 µS conductivity the need for and then the loading of less sample: standard (P/N 080172) •T he AutoDilution feature of Thermo Scientific™ • U ltra Scientific 1000 mg/L Certified IC Standards or Dionex™ Chromeleon™ Chromatography Data System Fisher Scientific reagents, ACS grade (CDS) software. If a chromatogram has an analyte peak height or area that exceeds a pre-set limit, a Samples smaller amount of sample is re-injected using either a: Fracking flowback solutions from the Marcellus Shale a. Partial loop (F1–10) were collected from successively later times b. Smaller loop during the fracking process. c. V ial to vial dilution using a Thermo Scientific Conditions Dionex AS-AP Autosampler Chromatographic • Measurement of sample conductivity prior to injection. Columns: Thermo Scientific™ Dionex™ IonPac™ This can be done a) manually, which is tedious, labor AG18/AS18, 4 mm intensive, and prone to error, or b) automatically, using the Thermo Scientific Dionex AS-AP Sample Eluent Source: Thermo Scientific Dionex EGC III KOH Cartridge Conductivity and pH Accessory. If the conductivity Eluent: 23 mM or 39 mM KOH exceeds a specified amount, Chromeleon CDS can Flow Rate: 1.0 mL/min trigger the loading of less sample as per a–c above. Column Temp. 30 °C These options are depicted in Figure 1 and are described in greater detail in Technical Note 138.7 Inj. Volume: 25 µL Detection: Suppressed conductivity, Thermo Scientific™ Dionex AS-AP Dionex™ ASRS™ 300 Anion Self-Regenerating Autosampler Suppressor*, recycle mode, 57 mA (23 mM Fracking Sample KOH); 97 mA (39 mM KOH) Sample Conductivity and pH Accessory Background < 2 µS Conductance: Centrifugation IC System Noise: < 3 nS Does conductivity No exceed cutoff? System Backpressure: 2100 psi Filtration Yes Chromatogram * E quivalent or improved results can be obtained using the Thermo Scientific™ Dionex™ AERS™ 500 Anion Electrolytically Regenerated Suppressor. Sample Dilution Yes No Does peak area Report Dionex AS-AP Autosampler Dilution or height exceed cutoff? Syringe: 1 mL Chromeleon AutoDilution option required Buffer Line Volume: 1.2 mL Wash Solution: 18 MW-cm resistivity degassed DI water Figure 1. Automated sample analysis options available using the Dionex AS-AP Autosampler and from autosampler wash liquid reservoir Chromeleon software. Vials: 10 mL This Technical Note describes the analysis of fracking Dilution Volumes: 20 µL sample into 1980 µL DI water flowback solutions using ion chromatography following in-line determination of sample conductivity and Draw/Dispense Speed: 50/25 µL/sec (water) automated dilution of high ionic strength samples. 10/5 (sample) Mix by Shake Iterations: 10 Equipment Carousel Speed: 30,000 encoder counts (maximum, default) • T hermo Scientific™ Dionex™ ICS-2100 Integrated Reagent-Free™ IC (RFIC ™) System with an additional Dionex AS-AP Sample Conductivity and pH Accessory 10-port valve (P/N 069473) Volume Drawn: 600 µL • Dionex AS-AP Autosampler Draw/Dispense Speed: 20/10 µL/sec – Dionex AS-AP Sample Conductivity and pH Part numbers of the consumables used in this document Accessory (P/N 074923) with the included pH are listed in Table 1. electrode port plug and seal installed • Chromeleon CDS, version 6.8 (release 6.80 DU11c or later) or 7 (release 7.1 SR1 or later)
3 Table 1. Consumables list. Dionex ICS-2100 IC Reagent-Free System Part Dionex AS-AP Autosampler Product Name Description Regen Number Fresh 18 MΩ-cm Deionized Water Dionex EGC III KOH Eluent generator 074532 Eluent In ERS 500 CD cartridge cartridge Suppressor Thermo Scientific Continuously regenerated 060477 Dionex CR-ATC column anion trap column Syringe Needle Columns Dionex IonPac AS18 Separation column, 060549 column 4 mm i.d. Tray Carousel Dionex IonPac AG18 Heat Fresh 18 MΩ-cm Guard column, 4 mm i.d. 060551 Exchange Deionized Water column Tubing Regen Thermo Scientific™ Suppressor cartridge, Degas CR-TC EGC III Pump* Dionex™ AERS™ 500 082540 Module 4 mm i.d. columns Sample Loop suppressor Backpressure Coil Polystyrene vials with Waste Dionex AS-AP caps and blue septa, 074228 Figure 2. Flow diagram for the Dionex ICS-2100 IC Reagent-Free System with 10-port valve. Autosampler vials 10 mL, package of 100 Sample and Standard Preparation Wash Liquid Line Samples were centrifuged at 10,000 × g, 10 min to pellet solid particles, which facilitated filtration. The resultant To Buffer Line Syringe supernatant was filtered with a 0.2 µm polyethersulfone Syringe Waste Line (PES) syringe filter. Samples were diluted as required Wash Reservoir with degassed 18 MW-cm resistivity DI water. Sample Conductivity Eluent Eluent It is important to use 18 MW-cm resistivity, DI water for and pH Accessory In Out eluent and autosampler flush solutions to avoid system contamination, decreased sensitivity, and poor calibration. Degassing the DI water by vacuum filtration prior to use Accessory Inlet Line is a good laboratory practice. Auxiliary Valve 3 2 Standards were prepared by diluting Certified IC Standard Bypass Line 4 1 stock solutions or by dissolving ACS grade reagents with Accessory Waste Line 5 6 DI water. Auxiliary Valve Instrument Setup and Installation Sampling Needle (B Position) Install, hydrate, and condition the Dionex EGC III KOH L eluent generator cartridge. This cartridge requires at least Carousel 5 4 2000 psi of backpressure for optimal operation. Install W 6 3 C and hydrate the Dionex CR-ATC trap column and the 12 Dionex AERS 500 Anion Electrolytically Regenerated L P Suppressor. Figure 2 outlines the flow path of the standard Injection Valve Dionex ICS-2100 system setup with a Dionex AS-AP Injection Port Sample Transfer Line (Inject Position) Autosampler. Hydrate the devices and finish the system setup according to the product manuals and the Dionex Figure 3. Dionex AS-AP plumbing schematic for Push Mode with a Sample Conductivity and pH Accessory. ICS-2100 Operator’s Manual.8–12 Install and condition the column using the conditions described in that column’s Quality Assurance Report (QAR) for 30 min prior to installing them in-line with the suppressor. If total system backpressure is < 2000 psi, install one or more backpressure coils between the Dionex CR-ATC Continuously Regenerating Anion Trap Column and the injection valve to increase the system pressure to 2000–3000 psi. Install the Dionex AS-AP Sample Conductivity and pH Accessory in the Dionex AS-AP Autosampler as outlined in the Accessory Setup and Operation Guide.13 Figure 3 depicts the setup required for Push mode with this accessory.
4 For the experiments presented here, the 10-port valve of Columns: Dionex IonPac AG18/AS18, 4 mm Eluent Source: Dionex EGC III KOH cartridge the Dionex ICS-2100 system was used as the auxiliary Eluent: 39 mM KOH valve. Alternatively, if the optional high-pressure valves Flow Rate: 1 mL/min Inj. Volume: 25 µL are installed in the Dionex AS-AP Autosampler, they can Col. Temp.: 30 °C be used to divert sample to the Dionex AS-AP Sample Detection: Suppressed conductivity, Dionex ASRS 300 Suppressor, Conductivity and pH Accessory. Install the pH electrode recycle mode Sample: 100-fold fracking flowback, filtered, 0.2 µm port plug, which reduces the amount of injected sample from 2000 to 600 µL for conductivity determination. Prior Peaks: Measured Undiluted 1. Acetate < 0.05 mg/L
5 Column: Dionex IonPac AG18/AS18, 4 mm Eluent: 23 mM KOH 350 Flow Rate: 1 mL/min 300 Inj. Volume: 25 µL Col. Temp.: 30 °C 250 Area ( µS*min) Detection: Suppressed conductivity, Dionex ASRS 300 Suppressor, recycle mode 200 Sample: A. Standard 2 150 r = 1.00 B–E. 100-fold fracking flowback F1, F2, F5, and F10, filtered, 0.2 µm %RSD= 1.12 100 Peaks: A 1. Fluoride 0.5 mg/L 50 2. Acetate 2.5 0 3. Propionate/Formate – 4. Formate 1.0 0 200 400 600 800 1000 1200 1400 1600 1800 2000 4.8 Concentration (mg/L) Figure 6. Chloride standard curve. 1 2 3 Change in Ion Concentrations During Fracking 4 As fracking progresses, the level of anions in flowback water changes. As can be seen in Figures 7 and 8, chloride and bromide concentrations increased ~10-fold from the first to the second fraction and then, in subsequent fractions, showed a slower, but steady increase. This result suggests µS that, as fracking proceeds, progressively more of these ions are dissolved into the flowback solution from the bedrock. A Conversely, the concentration of acetate decreased sharply B and then remained at low levels from fraction two onwards. These elevated concentrations would increase the amount C of processing required by water treatment facilities. In a D similar manner, what appears to be propionate (or a mixture with formate) in the first fraction is no longer E present in measureable quantities in subsequent fractions. 0 In contrast to the other analytes quantified, the level of 2.5 3 3.5 4 sulfate remained relatively constant, averaging 13 mg/L. Minutes 100,000 Figure 5. Comparison of early eluting inorganic anion and organic acids in 90,000 fracking flowback water. 80,000 70,000 Concentration (mg/L) Chloride Standard Curve Determination 60,000 One concern with using concentrated samples is that the 50,000 conductivity measured for an analyte will be outside of 40,000 the linear quantification range. For the wastewater samples 30,000 20,000 analyzed in this Technical Note, the primary analyte for 10,000 which this situation might arise is chloride. To confirm 0 that the concentrations of diluted samples were within F1 F2 F3 F4 F5 F6 F7 F8 F9 F10 the linear range for chloride, a standard curve was Fraction generated. As can be seen in Figure 6, the standard curve Figure 7. Concentration of chloride in fracking flowback samples. obtained was linear from 1–2000 mg/L with an r2 of 1.00 1,000 and %RSD of 1.12. This result indicates that even a sample solution containing 200,000 mg/L of chloride, if 900 diluted 1:100, would still fall within the range for which 800 an accurate calculation of concentration can be obtained 700 for this analyte. 600 Concentration (mg/L) 500 Bromide 400 Acetate 300 Sulfate 200 Formate 100 0 F1 F2 F3 F4 F5 F6 F7 F8 F9 F10 Fraction Figure 8. Concentration of inorganic anions and organic acids in fracking flowback samples.
One explanation for the apparent decrease in acetate References Te ch ni cal N ote 1 3 9 and propionate/formate is that, at high chloride 1. A merican Water Works Association (AWWA). Water concentrations, the capacity of the column for these and Hydraulic Fracturing: A White Paper from the analytes is diminished. To investigate this possibility, American Water Works Association, American acetate was added to 100-fold dilutions of fractions five Water Works Association, http://www.awwa.org/ and ten at a concentration 50% of that already present. fracturing. Denver, CO, 2013. As shown in Table 2, the recovery of acetate was 114 2. Bomgardner, M. Cleaner Fracking. C&EN, 2012, and 103%, for fractions 5 and 10, respectively. If high 90(142), 13–16. chloride levels had decreased the binding of acetate to 3. U.S. Environmental Protection Agency (EPA). Study the column, a lower recovery would have been expected, of the Potential Impacts of Hydraulic Fracturing on particularly for fraction 10, which had the highest Drinking Water Resources: Progress Report; chloride concentration. Since this was not the case, the EPA/601/R-12/011; U.S. Environmental Protection large drop in acetate measured indicates a decrease in Agency, Office of Research and Development: the amount present in fracking flowback water samples Washington, DC, 2012. after the first fraction. 4. O nishi, N. Fracking Tests Ties Between California ‘Oil and Ag’ Interests. New York Times, June 2, Table 2. Recovery of acetate in fracking flowback water. 2013, A14. Amount Acetate Amount Acetate 5. K iparsky, M. and Hein, J. F. Regulation of Hydraulic Fraction Measured Added Recovery (%) Fracturing in California: A Wastewater and Water (µg/L) (µg/L) Quality Perspective. Wheeler Inst. For Water Law F5 230 110 114 and Policy, UC Berkeley School of Law, Berkeley, F10 160 80 103 CA, 2013. 6. Frac Focus Chemical Disclosure Registry, What Chemicals are Used Page. http://fracfocus.org/ Conclusion chemical-use/what-chemicals-are-used (accessed This Technical Note demonstrates that by using the Aug 22, 2013). Dionex AS-AP Sample Conductivity and pH Accessory 7. T hermo Fisher Scientific. Thermo Scientific Technical in combination with Chromeleon CDS, samples Note 138, Accurate and Precise Automated Dilution containing high levels of anions, such as fracking flowback and In-line Conductivity Measurement Using wastewater, can be automatically diluted prior to the AS-AP Autosampler Prior to Analysis by Ion injection if a preset conductivity maximum is exceeded. Chromatography, TN70764_E, Sunnyvale, CA, 2013. This prescreening eliminates repeat analysis, thereby 8. T hermo Fisher Scientific. Product Manual for Eluent reducing eluent waste, extending column life, and Generator Cartridges. Doc. No. 065018-04, providing fast, accurate results. Sunnyvale, CA, 2012. 9. T hermo Fisher Scientific. Product Manual for the Dionex CR-TC. Doc. No. 031910-03, Sunnyvale, CA, 2010. 10. T hermo Fisher Scientific. Product Manual for the Dionex ERS 500 Suppressor. Doc. No. 031956-08, Sunnyvale, CA, 2013. 11. T hermo Fisher Scientific. Dionex ICS-2100 Ion Chromatography System Operator’s Manual. Doc. No. 065291, Sunnyvale, CA, 2012. 12. T hermo Fisher Scientific. Dionex AS-AP Autosampler Operator’s Manual. Doc. No. 065361, Sunnyvale, CA, 2012. 13. T hermo Fisher Scientific. Dionex AS-AP Sample Conductivity and pH Accessory Setup and Operation Guide. Doc. No. 065470, Sunnyvale, CA, 2012. www.thermoscientific.com/dionex ©2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an Thermo Fisher Scientific, example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any Sunnyvale, CA USA is manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. ISO 9001:2008 Certified. Not all products are available in all countries. Please consult your local sales representative for details. Australia +61 3 9757 4486 Finland +358 9 3291 0200 Korea +82 2 3420 8600 Taiwan +886 2 8751 6655 Austria +43 810 282 206 France +33 1 60 92 48 00 Netherlands +31 76 579 55 55 UK/Ireland +44 1442 233555 Belgium +32 53 73 42 41 Germany +49 6103 408 1014 Norway +46 8 556 468 00 USA and Canada +847 295 7500 Brazil +55 11 3731 5140 India +91 22 6742 9494 Singapore +65 6289 1190 China +852 2428 3282 Italy +39 02 950 591 Sweden +46 8 556 468 00 Denmark +45 70 23 62 60 Japan +81 6 6885 1213 Switzerland +41 61 716 77 00 TN70773_E 09/13M