Conex DIA-2Q Instrument amplifier and controller - GRUNDFOS ALLDOS INSTRUCTIONS
←
→
Page content transcription
If your browser does not render page correctly, please read the page content below
GRUNDFOS ALLDOS INSTRUCTIONS Conex® DIA-2Q Instrument amplifier and controller Installation and operating instructions
Declaration of Conformity We Grundfos Alldos declare under our sole responsibility that the product Conex® DIA-2Q, to which this declaration relates, is in conformity with the Council Directives on the approximation of the laws of the EC Member States relating to – Electromagnetic compatibility (89/336/EEC). Standards used: EN 61000-3-2: 1995 + A1 + A2: 1998, EN 61000-3-3: 1995, EN 61326: 1997 + A1: 1998 + A2: 2001 Class B, EN 61326: 1997 / A1: 1998 + A2: 2001. – Electrical equipment designed for use within certain voltage limits (73/23/EEC) [95]. Standard used: EN 61010-1: 2002. Pfinztal, 15th April 2008 W. Schwald Ulrich Stemick Managing Director Technical Director 2
CONTENTS Warning Page These complete installation and 1. Symbols used in this document 3 operating instructions are also 2. A few words in advance 4 available on www.Grundfosalldos.com. 3. Installation data 4 Prior to installation, read these installation and operating instructions. 4. Installation sketch 5 Installation and operation must comply 5. Instrument settings 6 with local regulations and accepted 6. General information 8 codes of good practice. 6.1 Warranty 8 7. Applications 8 1. Symbols used in this document 8. Safety 8 8.1 Obligations of the owner/ Warning operations manager 8 If these safety instructions are not 8.2 Avoidance of danger 8 observed, it may result in personal 9. Identification 9 injury! 9.1 Nameplate 9 9.2 Type key, Conex® DIA-2Q controllers 10 If these safety instructions are not 9.3 Type key, Conex® DIA-2Q preassembled Caution observed, it may result in malfunction systems 10 or damage to the equipment! 10. Technical data 12 Notes or instructions that make the job 10.1 Design / enclosure class 12 Note easier and ensure safe operation. 10.2 General data 12 10.3 Electronic data and functions 12 10.4 Measuring ranges 14 10.5 Dimensions 14 11. Installation 15 11.1 Transport and storage 15 11.2 Unpacking 15 11.3 Installation requirements 15 11.4 Installation in control panel 15 11.5 Installation of wall-mounted enclosure 16 12. Commissioning / electrical connections 17 12.1 Terminals 18 12.2 Power supply connection 20 12.3 Relay outputs 20 12.4 Current output 21 12.5 Connections of controller stop, sample-water sensor and temperature sensor 21 12.6 Connection of measuring cells 22 13. Operation 24 13.1 Control and display elements 24 13.2 Display elements during initial commissioning 24 13.3 Software overview 26 13.4 Main menu 27 13.5 Setup 28 13.6 Selection, configuration and parameterisation of the controller 33 13.7 "Alarm" menu 39 13.8 Checking the settings in the "service" menu 41 13.9 Calibration 43 13.10 Manual operation 48 14. Fault finding 51 15. Maintenance 52 16. Disposal 52 3
2. A few words in advance 3. Installation data The Conex® DIA-2Q (Dosing Instrumentation Please fill in the data below after Advanced, with 2 inputs, with flow measurement) commissioning. It will help you and multipurpose measuring amplifier and controller you Note your Grundfos Alldos service partner have just purchased uses the company’s leading- make subsequent adjustments to the edge measuring technology. installation. The Conex® DIA-2Q is a multipurpose device Owner: designed to carry out high-precision measurements and controls of the following: • a value from parameter group 1: Grundfos Alldos customer number: – chlorine, chlorine dioxide, ozone, hydrogen peroxide or peracetic acid Order number: • a value from parameter group 2: – pH value, redox potential or fluoride. Product number: The integrated controller, the high-resolution graphics display and the multilingual plain-text user Product serial number: interface make complicated measuring and control tasks in water chemistry much easier. Put into service on: Just a few button inputs lead you to your goal. The potentiostat helps save even more time, being automatically matched to the various input variables. Location of product: The safety standard of the dosing process is raised by the automatic open-circuit monitoring of the Used for: current outputs. Properties of the Conex® DIA-2Q measuring amplifier and controller include the following: • all control functions including PID and continuous-action controls • proportional control • setpoint control • combined setpoint control with external disturbance variable feedforward • manual or automatic temperature compensation • logbook function: chronological recording of calibration values with date and time • user coding function as a means of protection against access by unauthorised persons and for system administration • error message function for indication of non- functioning sensors. 4
4. Installation sketch 5
5. Instrument settings Parameter 1 Setup Controller Alarm Proportional Xp Alarm Parameter Controller On:_ % Off:_ Proport. contrl. Chlorine:_ Current input Reset time TN Chlorine dioxide:_ Alarm value 1 Proport. variable (PI/PID control) Switching point Ozone:_ Off:_ Peroxide:_ 0 - 20 mA:_ Peracetic acid:_ 4 - 20 mA:_ sec. Others:_ ________ Deriv. action Tv Alarm value 1 Temp. meas. Interpulse ctrl.:_ (PID control) Switching direction Yes:_ Pulse freq. ctrl:_ Upward violation:_ No:_ Cont. controller:_ sec. Downward viol.:_ Setpoint contrl. Combined contrl. Int.pulse period Alarm value 2 Temp. comp. Off:_ (interpulse ctrl.) Switching point Yes:_ Limit:_ No:_ Interpulse ctrl.:_ sec. Pulse freq. ctrl:_ Cont. controller:_ Min. ON time Alarm value 2 pH compensation Limit contact (interpulse ctrl.) Switching direction Yes:_ Downward viol.:_ Upward violation:_ No:_ Upward violation:_ sec. Downward viol.:_ Max. frequency Measuring ranges Ctrl. direction Hysteresis (pulse freq. ctrl.) Downward control:_ mg/l (ppm) Upward control:_ n/min Type of control Setpoint Current output 1 Alarm delay P:_ (limit contact) PI:_ mA sec. PID:_ Current input Disturbance variable Constant load Dos. time monit. Off:_ Measuring cell On:_ 0 - 20 mA:_ % Off:_ 4 - 20 mA:_ Others:_ ________ Max. dosing flow Dos. time monit. Measuring cell type Max. dosing time % min Limit Cleaning motor (limit contact) Monitoring on:_ Monitoring off:_ Hysteresis Water sensor (limit monitor) On:_ Off:_ 6
Parameter 2 Setup Controller Alarm Proportional Xp Alarm Parameter Controller On:_ % Off:_ Proport. contrl. Current input Reset time TN pH:_ Alarm value 1 Proport. variable (PI/PID control) Switching point Redox:_ Off:_ Fluoride:_ 0 - 20 mA:_ 4 - 20 mA:_ sec. Others:_ _______ Deriv. action TV Alarm value 1 Temp. meas. Interpulse ctrl.:_ (PID control) Switching direction Yes:_ Pulse freq. ctrl:_ Upward violation:_ No:_ Cont. controller:_ sec. Downward viol.:_ Setpoint contrl. Combined contrl. Int.pulse period Alarm value 2 Temp. comp. Off:_ (interpulse ctrl) Switching point Yes:_ Limit:_ No:_ Interpulse ctrl.:_ sec. Pulse freq. ctrl:_ Cont. controller:_ Min. ON time Alarm value 2 pH compensation Limit contact (interpulse ctrl) Switching direction Yes:_ Downward viol.:_ Upward violation:_ No:_ Upward violation:_ sec. Downward viol.:_ Max. frequency Measuring ranges Ctrl. direction Hysteresis (pulse freq. ctrl) Downward control:_ mg/I (ppm) Upward control:_ n/min Current output 1 Type of control Alarm delay P:_ Setpoint PI:_ (limit contact) mA PID:_ sec. Current input Disturbance variable Constant load Dos. time monit. Off:_ On:_ 0 - 20 mA:_ % Off:_ 4 - 20 mA:_ Others:_ _______ Dos. time monit. Max. dosing flow Max. dosing time % min Limit (limit contact) Hysteresis (limit contact) 7
6. General information 8. Safety These installation and operating instructions contain all information important for users of the 8.1 Obligations of the owner/operations Conex ® DIA-2Q: manager • technical data The owner/operations manager must ensure that • instructions on commissioning, use and persons working with the Conex® DIA-2Q maintenance instrument amplifier and controller fulfil these • safety information. requirements: Should you require further information or should you • They are acquainted with the regulations encounter problems that are not handled in sufficient concerning working safety and accident depth in this manual, please contact Grundfos prevention. Alldos. We shall be pleased to support you with our • They have been trained in use of the device. comprehensive know-how in the fields of measuring • They have read and understood the warning and control technology as well as water treatment. information and handling symbols. We always welcome suggestions on how to optimise The owner/operations manager is also responsible our installation and operating instructions to satisfy for ensuring that this manual is kept in the immediate our customers. vicinity of the device and is always available for the operating personnel. 6.1 Warranty A warranty claim in accordance with our general 8.2 Avoidance of danger terms of sale and delivery is only valid if these requirements are complied with: Warning • The device has been used in accordance with the Installation and connection of the information in this manual. device and the associated supplementary components must only • The device has not been dismantled or handled be carried out by authorised personnel! incorrectly. The local safety regulations must be observed! 7. Applications The Conex® DIA-2Q instrument amplifier and Warning controller is suitable for measuring chlorine (Cl2), Switch off the power supply before chlorine dioxide (ClO2), ozone (O3), hydrogen connecting the power supply cable and peroxide (H 2O2), peracetic acid, fluoride, pH or relay contacts! redox potential and for controlling these variables using appropriate actuators within the applications Do not dismantle the device! described in this manual. Maintenance and repair must only be carried out by authorised personnel! Warning Other applications are not approved The mounting location must be and not permitted. Grundfos Alldos selected so that the housing is not cannot be held liable for any damage Caution subjected to mechanical loading. resulting from incorrect use. Check that all settings are correct before starting up the device! 8
9. Identification 9.1 Nameplate 1 DIA-2Q, 1-D/HP/PA 2-P/R/F, Q-W-G 2 352-2200-10001 12 S/N: 08/12345 3 Conex DIA-2Q 4 230/240V 50/60Hz, 15 VA, 5 10 IP 65 11 96609140P1108120808565 TM04 0455 0708 6 7 8 9 Fig. 1 Nameplate, Conex® DIA-2Q Pos. Description 1 Type designation 2 Model 3 Product name 4 Voltage [V] 5 Frequency [Hz] 6 Product number 7 Country of origin 8 Year and week of production 9 Marks of approval, CE mark, etc. 10 Power consumption [VA] 11 Enclosure class 12 Serial number 9
9.2 Type key, Conex® DIA-2Q controllers Type key example: DIA-2Q, 1-D/HP/PA 2-P/R/F, Q-W-G Example: DIA -2Q 1-D/HP/PA 2-P/R/F Q -W -G Measuring amplifier and controller Dosing Instrumentation Advanced with 2 inputs + DIA-2Q flow measurement Input parameter 1 D Chlorine (Cl2), chlorine dioxide (ClO2) or ozone (O3) HP Hydrogen peroxide (H2O2) PA Peracetic acid (PAA) Input parameter 2 P pH R Redox (ORP) F Fluoride (F–) Variant Q Flow compensation Mounting W Wall-mounted P Panel-mounted Voltage G 1 x 230 V, 50/60 Hz H 1 x 120 V, 50/60 Hz I 24 VDC 9.3 Type key, Conex® DIA-2Q preassembled systems Type key example: DIA-2Q-A, D2-X-AU-PCX-QS-T, W-G Example: DIA -2Q -A D2 -X -AU -PCX -QS -T W -G Units for measurement and control Dosing Instrumentation Advanced, with 2 inputs, DIA-2Q with flow measurement Assembly A Preassembled Cell type D1 Pressure-proof, with cleaning motor D2 Pressure-proof, with hydro-mechanical cleaning D3 Pressureless, with hydro-mechanical cleaning D4 For total chlorine measurement D5 For free chlorine measurement with buffer dosing P/R pH or redox (ORP) only P With pressure retention valve X Without pressure retention valve (to be continued) 10
Example: DIA -2Q -A D2 -X -AU -PCX -QS -T W -G Disinfection electrodes AU Gold PT Platinum X No disinfection measuring Other electrodes PCB pH, ceramic diaphragm, incl. buffer solution PTB pH, PTFE diaphragm, incl. buffer solution PKB pH, KCl filling, incl. buffer solution PGB pH, gel filling incl. buffer solution PCX pH, ceramic diaphragm, excl. buffer solution PTX pH, PTFE diaphragm, excl. buffer solution PKX pH, KCL filling, excl. buffer solution PGX pH, gel filling, excl. buffer solution RCB Redox (ORP), ceramic diaphragm, incl. buffer solution RTB Redox (ORP), PTFE diaphragm, incl. buffer solution RCX Redox (ORP), ceramic diaphragm, excl. buffer solution RTX Redox (ORP), PTFE diaphragm, excl. buffer solution F Fluoride PA Peracetic acid HP Hydrogen peroxide X No electrode Flow sensor QS Flow sensor integrated X No flow sensor Temperature sensor T With Pt100 X No temperature sensor Mounting of controller W Wall-mounted P Panel-mounted Voltage G 1 x 230 V, 50/60 Hz H 1 x 120 V, 50/60 Hz I 24 VDC 11
10. Technical data 10.3 Electronic data and functions 10.3.1 Electronics 10.1 Design / enclosure class Electronics 16-bit microprocessor Wall-mounted enclosure High-resolution graphics LCD Display (distance from IP65 with background light sensors up to three 1 alarm relay, 2 controller metres) Potential-free relay relays (250 V/6 A, maximum outputs IP54 (front) / 550 VA) IP65 (sensor interface) Controller stop; proportional Control panel (distance from sensor Signal inputs value, disturbance variable 0/ enclosure interface up to 100 metres, 4 to 20 mA; water sensor including separate distance from sensor potentiostat 4 analog outputs 0/4 to interface to sensors up to three metres) Signal outputs 20 mA, freely adjustable, maximum load 500 Ω 10.2 General data • Parameter group 1: chlorine, chlorine dioxide, Input power Approximately 15 VA ozone, peroxide, peracetic Permissible acid ambient 0 °C to +50 °C • Parameter group 2: pH, temperature redox or fluoride Permissible • For temperature or Freely adjustable continuous control storage –20 °C to +65 °C analog outputs for (0/4 to 20 mA) of the temperature measured values parameters from group 2 Maximum relative (pH, redox or fluoride) 90 % (non-condensing) humidity • For continuous control Weight 1.5 kg (0/4 to 20 mA) of the parameters from group 1 Plastic (control panel (chlorine, chlorine dioxide, Enclosure enclosure: noryl, wall- ozone, peroxide, peracetic mounted enclosure: ABS) acid) • 230/240 V (50/60 Hz) (standard model) 10.3.2 Functions of the instrument amplifier Power supply versions • 115/120 V (50/60 Hz) Measured-value display: • 24 V DC measured value with its unit, Display mode temperature display: in °C or °F Temperature Manual or automatic with compensation Pt100 (–5 °C to +120 °C) Manual calibration, or with Calibration automatic recognition of buffer solution 12
10.3.3 Proportional controller functions 0 to 20 mA / 4 to 20 mA or Basic load 0 to 50 % of dosing rate Proportional value other value (freely adjustable Maximum dosing Value (basic load + 1) to input between 0 mA and 20 mA) rate 100 % of dosing rate Proportional or inversely Adjustable to upward or Controller output Control direction proportional downward control Interpulse output, pulse Controller stop Adjustable to NC or NO Controller output frequency output, continuous controller output 10.3.5 Combined controller functions 1 to 100 seconds (only with Interpulse period Minimum ON time 0.1 to 10.0 seconds (only with interpulse controller) Tmin interpulse controller) Minimum ON time 0.1 to 10.0 seconds (only with 1 to 180 pulses per minute Tmin interpulse controller) Maximum (only with pulse frequency frequency 1 to 180 pulses per minute controller) Maximum (only with pulse frequency frequency Basic load 0 to 50 % controller) Maximum dosing Value (basic load + 1) to Basic load 0 to 100 % of dosing rate rate 100 % of dosing rate 0 to 100 % of dosing rate Maximum dosing Adjustable to upward or (value for maximum dosing Control direction rate downward control rate < basic load is possible!) Controller stop Adjustable to NC or NO 0.1 to 10 (weighting of Dosing factor D.F. proportional value) 0 to 20 mA / 4 to 20 mA or Disturbance other (freely adjustable Controller stop Adjustable to NC or NO variable input between 0 and 20 mA) 10.3.4 Setpoint controller functions 0.1 to 10.0 (weighting of Dosing factor D.F. disturbance variable) Limit monitor, interpulse controller (P, PI, PID), pulse Controller output frequency controller (P, PI, PID), continuous controller (P, PI, PID) 0 to 100 % of measuring range, adjustable in the unit Limit of the measured value (only with limit monitor) 0 to 50 % of full-scale value, adjustable in the unit of the Hysteresis measured value (only with limit monitor) 0 to 100 % of measuring Setpoint input range, adjustable in the unit of the measured value Proportional 0.1 to 3000.0 % band Xp 1 to 3000 seconds, resolution Reset time Tn 1 second (only with PI, PID controllers) 0 to 1000 seconds, resolution Derivative action 1 second (only with PID time Tv controller) 1 to 100 seconds (only with Interpulse period interpulse controller) Minimum ON time 0.1 to 10.0 seconds (only with Tmin interpulse controller) 1 to 180 pulses per minute Maximum (only with pulse frequency frequency controller) 13
10.4 Measuring ranges Peracetic CI2 CIO2 O3 H2O2 pH Redox (ORP) Fluoride acid mg/l mg/l mg/l mg/l mg/l pH mV mg/l 0.00 - 0.50 0.00 - 0.50 0.00 - 0.50 0-100 0-100 0.00 - 14.00 –1500 to +1500 0.00 - 2.00 0.00 - 1.00 0.00 - 1.00 0.00 - 1.00 0-500 0-500 2.00 - 12.00 0-1000 0.00 - 5.00 0.00 - 2.00 0.00 - 2.00 0.00 - 2.00 0-1000 0-1000 5.00 - 9.00 0.00 - 5.00 0.00 - 5.00 0.00 - 5.00 0-2000 0-2000 0.00 - 10.00 0.00 - 10.0 0.00 - 20.00 The measuring ranges are also freely adjustable (within the above-mentioned range limits). Note In the case of chlorine and chlorine dioxide, the free adjustment of the measuring range can be extended up to a maximum of 50.0 mg/l. 10.5 Dimensions 84 10 180 27 125 110 184.5 145 59.5 TM03 6687 4506 Ø 4.5 59 198 212 Fig. 2 Wall-mounted enclosure Conex® DIA-2Q 158 18 90 96 90 TM03 6688 4506 90 166 96 Fig. 3 Control panel enclosure Conex® DIA-2Q 14
11. Installation 11.4 Installation in control panel 11.1 Transport and storage > 20 • Transport the device carefully, do not drop! • Store at dry and cool location. 92+0.8 11.2 Unpacking • Check the device for damage. 92+0.8 Install as soon as possible after unpacking. • Do not install or connect damaged devices! > 20 Retain the packing material or dispose Note TM03 6689 4506 of it according to local regulations. 11.3 Installation requirements • Dry room • Room temperature: 0 °C to 50 °C • Vibration-free location. Fig. 4 Control panel enclosure Conex® DIA-2Q If you do not observe the installation requirements, the device may be Caution damaged! The measurements may not be correct! TM03 6690 4506 Fig. 5 Sensor interface 1. Make an opening of 92 + 0.8 mm x 92 + 0.8 mm in the control panel. 2. Slip on the supplied gasket. 3. Insert the Conex® DIA-2Q into the opening from the front. Do not damage the gasket! Caution The gasket must be fitted exactly! 1. Hook the clamps into the tightening cones on the sides at the top and bottom. 2. Secure the device from the rear using a screwdriver. 3. Install a separate sensor interface near the sensors (maximum distance of three metres). 15
11.5 Installation of wall-mounted enclosure 27 Warning Switch off the power supply before installing! Enclosure class IP65 is only 145 guaranteed if the terminal cover is correctly sealed, if the front panel of the terminal enclosure is closed and TM03 6691 4506 the appropriate cable glands or dummy caps fitted. Do not damage the terminal cover Caution gasket! The terminal cover gasket must 10.5 198 fit exactly! Fig. 6 Wall-mounted enclosure Conex® DIA-2Q 1. Drill three holes (∅8 mm) as shown in the diagram, and insert the supplied dowels. 2. Screw the screw (A) into the top centre dowel until it projects by approximately 1 cm. See fig. 7. 3. Loosen the fastening screws of the front panel, and remove the front panel. 4. Hang the instrument onto the screw (A). 5. Tighten the instrument with the two screws (B). 6. Mount the front panel of the enclosure. A B B TM03 6692 4506 Fig. 7 Mounting drawing 16
12. Commissioning / electrical connections Warning Switch off the power supply before installing! Enclosure class IP65 is only guaranteed with the front panel of the terminals enclosure closed and with appropriate cable glands or dummy caps. Warning Switch off the power supply before connecting the power supply cable and the relay contacts! For safety reasons, the protective conductor must be connected correctly! Observe the local safety regulations! Protect the cable connections and plugs against corrosion and humidity. Before connecting the power supply cable, check that the supply voltage specified on the nameplate corresponds to the local conditions! An incorrect supply voltage may destroy the device! To guarantee electromagnetic Caution compatibility (EMC), the input and current output cables must be screened. Connect the screening to the screen ground on one side. Refer to the wiring diagram! Route the input, current output and power supply cables in separate cable channels. Enclosure class IP65 is only guaranteed if the terminal cover is correctly sealed! Do not damage the Caution gasket on the terminal cover! The gasket on the terminal cover must be positioned precisely! Do not damage the gasket! Note Unused terminals must remain open. 1. Remove the terminal cover on the front of the device. 2. Use the appropriate cable feedthroughs and tighten the screws carefully. 3. Connect the cables used to the terminals according to the Conex® DIA-2Q terminal assignment. 4. Close the terminal cover again with correctly positioned gasket. 17
18 15 (white) 16 (brown) Fig. 8 12.1 Terminals + - L N PE 1 3 5 1 3 5 7 9 11 13 15 17 19 21 23 25 27 29 31 33 35 37 38 39 40 41 42 2 4 6 2 4 6 8 10 12 14 16 18 20 22 24 26 28 30 32 34 36 Cells / electrodes N.O15 (wh) 37 = 39 = B/R + - . L N PE 16 (br) 38 = M 40 = G/C 33/35 1 2 N.C. Cells 23 - 30 External 41/42 = +/– mV 17 - 22 32,34,36 Relays Alarm PAA Sensors Outputs 1 4 input pH – 115/120 H 2 O2 mA F Terminals of wall-mounted enclosure 24 24 V/DC V/DC 115/120VV mA 12.1.1 Wall-mounted enclosure Conex ® DIA-2Q 230/240 230/240 VV +– External mA input 15 17 19 21 23 25 27 29 31 33 35 15 17 19 21 23 25 27 29 31 33 35 16 18 20 22 24 26 28 30 32 34 36 16 18 20 22 24 26 28 30 32 34 36 Outputs 1 4 Sensors Cl2,ClO 2 , O3, mA 21/22 = Pt 100 H2O2, PAA P1 1 +– 19 = + H2O, 20 = – H2O (water) – P2 mA pH, mV, F 2 +– 17/18 = Controller stop mA T / P2 3 +– mA P1 4 +– TM03 6693 4506
1/1 (brown), 1/1 (br), 1/21/2 (white) (wh) Outputs mA Cells: PAA 2 /1, 2/2 = Controller stop Fig. 9 Cl2 , ClO2 , H2O2 2/3 (+), 2/4 (–) = H2O (water) 1 4 O3 , H2O2 , 1 PAA 1/1 2/1 5/1 6/1 5/1 1/2 2/2 5/2 6/2 1 1/3 2/3 5/3 6/3 5/2 Relays pH, mV, F – 2 1/4 2/4 5/4 6/4 2 5/3 1/5 5/5 5/1 - 5/5 1/6 3/1 5/4 T 3 3/2 7/1 N.O. 1/7 5/5 1/8 3/3 7/2 Alarm 1/9 3/4 7/3 N.C. 5/1 - 5/5 Terminals of control panel enclosure 4 1/10 12.1.2 Control panel enclosure Conex® DIA-2Q 4/1 8/1 PE 8/1 1/11 4/2 8/2 N 8/2 - 1/12 4/3 8/3 L 8/3 + 1/13 External input (24 1/14 4/4 Conex DIA-2Q 115/120 115/120 VV 24V/DC V/DC mA –+ 230/240 230/240 VV Connection to Sensor interface 1/11 -1/14 4/1 - 4/4 3/1 - 3/4 Conex® DIA-2Q Pt 100 Cells Electrodes Cl 2, ClO 2, pH, mV, F – A 1/1 1/1(brown) (br) DIP O3, PAA, 1/2 B 1/2(white) (wh) H2O2 PAA off on Cells Jumper H2O2 Cells / electrodes A 2 /3 (+) M B/R G/C Pt 100 1/11 1/13 4/1 4/3 3/1 3/3 A B 2 /4 (-) + – H 2O (water) 1/12 1/14 4/2 4/4 3/2 3/4 B pH, mV, F – TM03 6694 4506 19
Legend of terminals 12.2 Power supply connection 1. Control panel enclosure: Plug the plug strip into Pos. Description the corresponding terminal strip at the rear side of the device. Ensure correct orientation. Relays Relay 1 + 2 2. Connect the protective earth conductor (PE) to Alarm relay terminal 5 (wall-mounted enclosure) or terminal 8/ Alarm – N.O.: normally open 1 (control panel enclosure). – N.C.: normally closed 3. Connect the neutral conductor (N) (or the – Pt100 Temperature sensor conductor with 24 V version) to terminal 3 (wall- mounted enclosure) or terminal 8/2 (control panel H2O Water sensor enclosure). Controller stop 4. Connect phase (L1) (or the + conductor with 24 V Stop (NO or NC selectable) version) to terminal 1 (wall-mounted enclosure) or 8/3 (control panel enclosure). Outputs Current outputs [mA] Switch the device on and off by switching the power Cl2 (chlorine), ClO2 (chlorine supply on and off accordingly. The device itself is not dioxide), O3 (ozone), H2O2 equipped with a separate on/off switch. 1 (hydrogen peroxide) or PAA (peracetic acid) 12.3 Relay outputs 2 pH, mV (redox), F– (fluoride) The connection of the relay outputs T/P2: temperature / continuous depends on the application and the 3 controller parameter 2 final control elements used. Therefore Note P1: output of continuous controller the connections described below 4 should only be considered as parameter 1 guidelines. Inputs External inputs [mA] With inductive loads (also relays and contactors), Measuring cells, electrodes and interference suppression is necessary. If this is not Electrodes single-rod measuring chains possible, protect the relay contacts using a M Measuring electrode suppressor circuit as described below. • With AC voltage: B/R Reference electrode G/C Counter electrode Current up to Capacitor C Resistor R Earth 60 mA 10 nF, 275 V 390 Ω, 2 W mV Redox electrode 70 mA 47 nF, 275 V 22 Ω, 2 W F– Fluoride electrode 150 mA 100 nF, 275 V 47 Ω, 2 W screen 1.0 A 220 nF, 275 V 47 Ω, 2 W Control panel enclosure Conex® DIA-2Q • With DC voltage: Connect the free-wheeling • Conex® DIA-2Q: for installation in the control diode in parallel to relay or contactor. panel. • Sensor interface: for installation near the sensors. Caution Provide the relay outputs with a Use the following cable to connect the sensor corresponding backup fuse! interface to the Conex® DIA-2Q: shielded cable, 4 x 0.5 mm2. DC AC + TM03 7209 4506 R C - Fig. 10 Suppressor circuit, DC/AC 20
12.4 Current output Output 4: continuous control parameter 1 This current output shows the calculated actuating Make sure that the polarity of the variable signal as an analog current signal. Caution current output is correct! Use of actuating variable signal: Maximum load: 500 Ω. • as input signal for a continuous final control The current output can be set to one of the two element. standard ranges "0-20 mA" or "4-20 mA", or it can be 1. Connect the + conductor to terminal 29 (wall- freely adjusted. mounted enclosure) or terminal 1/8 (control panel • Connect the screen to earth (PE) at one end. enclosure). Output 1: chlorine, chlorine dioxide, ozone, 2. Connect the – conductor to terminal 30 (wall- hydrogen peroxide or peracetic acid mounted enclosure) or terminal 1/7 (control panel (parameter 1) enclosure). This current output shows the displayed measured 12.5 Connections of controller stop, value as an analog current signal. sample-water sensor and temperature Use of current signal for measured values: sensor • as input signal for another indicator • as input signal for an external controller. Connecting the controller stop 1. Connect the + conductor to terminal 23 (wall- 1. Connect the + conductor to terminal 17 (wall- mounted enclosure) or terminal 1/3 (control panel mounted enclosure) or terminal 2/1 (control panel enclosure). enclosure). 2. Connect the – conductor to terminal 24 (wall- 2. Connect the – conductor to terminal 18 (wall- mounted enclosure) or terminal 1/4 (control panel mounted enclosure) or terminal 2/2 (control panel enclosure). enclosure). Output 2: pH, redox or fluoride (parameter 2) Connecting the sample-water sensor This current output shows the displayed measured Cable colours and marking: See connections of value as an analog current signal. measuring-cell types AQC-D1/AQC-D2/AQC-D3. Use of current signal for measured values: 1. Connect the + conductor to terminal 19 (wall- mounted enclosure) or terminal 2/3 (control panel • as input signal for another indicator enclosure). • as input signal for an external controller. 2. Connect the – conductor to terminal 20 (wall- 1. Connect the + conductor to terminal 25 (wall- mounted enclosure) or terminal 2/4 (control panel mounted enclosure) or terminal 1/5 (control panel enclosure). enclosure). 2. Connect the – conductor to terminal 26 (wall- When using measuring cell AQC-D2, mounted enclosure) or terminal 1/4 (control panel Note the water sensor must always be enclosure). connected and activated! Output 3: continuous controller parameter 2 (or Connecting the Pt100 temperature sensor temperature) 1. Connect the + conductor to terminal 21 (wall- Shows the calculated actuating variable signal of the mounted enclosure) or terminal 1/11 (control controller as an analog current signal. panel enclosure). Use of actuating variable signal: 2. Connect the – conductor to terminal 22 (wall- • as input signal for a final control element with mounted enclosure) or terminal 1/12 (control current input panel enclosure). or temperature output: Shows the temperature measured by the optional temperature sensor as an analog current signal. Use of temperature current signal: • as input signal for another indicator. 1. Connect the + conductor to terminal 27 (wall- mounted enclosure) or terminal 1/6 (control panel enclosure). 2. Connect the – conductor to terminal 28 (wall- mounted enclosure) to terminal 1/7 (control panel enclosure). 21
12.6 Connection of measuring cells 37 38 39 40 41 42 Jumper setting • All measuring values except for peroxide: + - All cell types: position 1 (standard). 7 TM03 6967 4506 • Measuring value peroxide: 13 Cell type HP (peroxide): position 1 (standard). • Measuring value peroxide: Cell types AQC-D1/AQC-D3: position 2 (move the wire strap of the jumper (blue) to the higher position). Fig. 14 Connection to single-rod measuring TM03 6696 4506 chains for pH, redox and fluoride Peroxide with AQC-D1/AQC- 2 2: D2/AQC-D3 12.6.2 Connection of control panel enclosure 1 Conex ® DIA-2Q 1: Standard Jumper M B/R G/C Pt 100 Fig. 11 Jumper setting 1 2 3 4 12.6.1 Connection of wall-mounted enclosure 5 Conex ® DIA-2Q 7 6 15 17 19 21 23 25 27 29 31 33 35 37 38 39 40 41 42 16 18 20 22 24 26 28 30 32 34 36 5 1 8 1 2 4 4 3 6 9 TM03 5871 1106 7 Conex 2/ 2/ 10 Sensor interface Sensor-Interface +3 - 4 8 1/ 1/ 4/ 4/ 3/ 3/ 10 A 1 11 13 1 3 1 3 1/ 1/ 4/ 4/ 3/ 3/ 9 TM03 5872 1106 2 B 4 12 14 4 2 4 Fig. 15 Connection to measuring cells AQC-D1/AQC-D2/AQC-D3 M B/R G/C 1/11 1/13 4/1 4/3 3/1 3/3 A Fig. 12 Connection to measuring cells 1/12 1/14 4/2 4/4 3/2 3/4 B AQC-D1/AQC-D2/AQC-D3 15 17 19 21 23 25 27 29 31 33 35 37 38 39 40 41 42 12 11 16 18 20 22 24 26 28 30 32 34 36 1 2 1/1 1/2 - + TM03 6968 4506 11 12 2 1 TM03 6966 4506 Fig. 16 Connection to measuring cells PA (peracetic acid) / HP (peroxide) Fig. 13 Connection to measuring cells PA (peracetic acid) / HP (peroxide) 22
Peculiarities for the connection at the control + - panel enclosure of Conex® DIA-2Q 1. First connect the measuring cells, the single-rod measuring chains and the water sensor (if present) to the separate sensor interface. 7 2. Secondly, connect the sensor interface to the TM03 6969 4506 13 Conex® DIA-2Q control panel enclosure. Connecting the sensor interface to the Conex® DIA-2Q (see back side of device) • Measuring cells AQC-D1/AQC-D2/AQC-D3, PA (peracetic acid) / HP (peroxide) Fig. 17 Connection to single-rod measuring Connect terminals 4/1 to 4/14 of the sensor chains for pH, redox and fluoride interface to the corresponding terminals of the Conex® DIA-2Q. • Measuring cells HP (peroxide) / Pos. Description PA (peracetic acid) 1 Brown In addition, connect terminal A to terminal 1/1 and terminal B to terminal 1/2 of the Conex® DIA-2Q. 2 White • Single-rod measuring chains 3 Black (pH, redox, fluoride) Connect terminals 3/1 to 3/4 of the sensor 4 Blue interface to the corresponding terminals of the 5 Screen Conex® DIA-2Q. 6 Outer conductor (screen) • Sample-water deficiency sensor Connect terminal A to terminal 2/3 and terminal B 7 Inner conductor to terminal 2/4 of the Conex® DIA-2Q. 8 Reference electrode When using measuring cell AQC-D2, Measuring electrode Note the water sensor must always be 9 Counter electrode connected and activated! Pt100 temperature sensor DIP switch • If necessary, set the DIP switch for the desired 10 Water sensor measured value to "On" (move the switch to the 11 Yellow right, top switch 1 is not used). 12 Green DIP TM03 6699 4506 13 Outer conductor OFF On 1 2 ClO2, O3 3 H2O2 4 Cl2 Fig. 18 DIP switch 23
13. Operation 13.1 Control and display elements 2 7 8 1 TM03 6970 4506 1 6 5 4 3 2 7 8 6 5 4 3 Fig. 19 Wall-mounted enclosure and control panel enclosure Pos. Description 13.2 Display elements during initial Control elements commissioning When connected to the power supply on the device’s Red alarm LED initial commissioning and following the start-up 1 • Flashes in case of faults or incorrect indication, the display shows the "Sprache/language" entries. menu. 2 Display Language/Langue... Operating buttons Deutsch [OK] button English 3 • Enters the selected menu. • Confirms the selected line or value. Français [Down] button Nederlands • Moves one line downward (the selected 4 line is displayed inversely). • The word "language" in the header is displayed in • Decreases values. all available languages successively at intervals of approximately one second. [Up] button Start the Conex® DIA-2Q by selecting the desired • Moves one line upward (the selected line 5 operating language: is displayed inversely). • Increases values. 1. Select the desired language using the [Up] and [Down] buttons. [Esc] button – The selected language is displayed inversely • Returns to the previous menu. (white letters on black background). 6 – The data which were entered last are 2. Press [OK] to confirm. not changed. The selected language is stored and Function buttons with yellow LED will remain the operating language [Cal] button when the device is rebooted. If • Switches between calibration mode and Note necessary, the operating language can 7 measuring mode. be changed in the "language" submenu in the "setup" menu. – In calibration mode, the corresponding LED illuminates. See section 13.5 Setup. [Man] button • Switches between automatic and manual 8 modes. – In manual mode, the corresponding LED illuminates. 24
The translations for the words "setup" and See sections 10.4 Measuring ranges and "language" into the currently available languages 13.5.4 Setting the measuring ranges for chlorine, are: chlorine dioxide, ozone, peroxide, peracetic acid, pH, redox and fluoride. Deutsch 13.2.1 Display modes Hauptmenü Grundeinstellung Sprache Switching between display modes 1 and 2: English 1. Use the [Up] and [Down] buttons to switch main menu setup language between the two display modes: Français Display mode 1 with two parameters Menu principal Paramétrage Langue Nederlands ClO2 pH 0.30 hoofdmenu basisinstelling taal TM03 6971 4506 Español mg/l 4.00 Menu principal Ajuste básico Idioma Italiano pH Menu principale programm. fond. lingua The large display shows the measured values for Português the two parameters (for example ClO2 and pH). Menu principal Posição inicial Língua The inverted font in the header displays the two parameters left-justified and right-justified. Pусский The measured value for the left-justified parameter 1 главное меню станд. настройка язык is displayed in the first line, and the measured value Polski for the right-justified parameter 2 is displayed in the second line underneath the header. The measured manu główne ustawienia język value for a possibly active temperature measurement is not output in this display mode. chlorine Display mode 2 with two parameters ClO2 pH 0.43 mg/l 0.30 TM03 6972 4506 mg/l • After selecting and confirming the operating 4.00 pH c language by pressing [OK] during initial 100.0 °C c commissioning, the display shows "chlorine" as the value to be measured. The large display shows the measured values for – On subsequent start-ups, the display will show parameter 1, and the small display shows the the measured value selected the previous time. measured values for parameter 2 and the possibly active temperature measurement. "c" is displayed if the compensation functions are active. chlorine 0.43 mg / l TM03 6704 4506 5.20 pH c 22 °C c • On upward or downward violation of the selected measuring range, the display will show the upper or lower limit and flash. 25
Display with one parameter 13.3 Software overview Provided that the code numbers for ClO2 access right have been set, some menus (and submenus) as well as the 0.30 TM03 6973 4506 Note functions Cal and Man are protected mg/l against unauthorised access. The protected menus are marked with "C" (calibration rights) or "F" (full rights). 100.0 °C c Main menus • Controller 1 / 2, see section Controller 1/2. The large display shows the measured values, and the selected parameter is displayed in the header. • Alarm 1 / 2, see section Alarm 1/2. The measured value for a possibly active • Service, see section Service. temperature measurement is output in the second • Setup, see section Setup. line. "c" is displayed if the compensation functions are active. Function buttons • Cal, see section Calibration function. Alarm signals • Man, see section Manual operation. A parameter-based alarm refers to the flashing parameter in the header. Parameter-independent Controller 1/2 alarms are displayed in the bottom line, but the "F" (full rights) display does not flash for this alarm. Setting the parameters If several alarms are present, they can be scrolled • Relay 1/2 using the [Up] and [Down] buttons. They are displayed in chronological order. The selected alarm – setpoint signal is displayed with inverted font. Subsequently, – controller parameters. the display is switched to the last display mode set. • Continuous controller Display mode with active alarms and two – setpoint parameters – controller parameters. Alarm 1/2 ClO2 pH "F" (full rights) • Alarm settings 0.30 4.00 TM03 6974 4506 – alarm 1/2 values (switching points) – effective direction alarm line 1 – hysteresis alarm line 2 – alarm delay. • Dosing time monitoring The parameters are displayed left-justified and right- – maximum dosing time (at constant dosing flow justified in the header, and the measured values of level of 100 %). the parameters are displayed directly underneath. The measured value for a possibly active Service temperature measurement and the display "c", if the • Calibration logbook 1/2 compensation functions are active, are suppressed. – the last 10 calibrations. Display mode with active alarms and one • Calling up the controller 1/2 settings parameter • Test of functioning of current outputs • Test of functioning of relays ClO2 • Test of functioning of display. 0.30 TM03 6975 4506 mg/l alarm line 1 alarm line 2 The large display shows the measured values, and the selected parameter is displayed in the header. The measured value for a possibly active temperature measurement and the display "c", if the compensation functions are active, are suppressed. 26
Setup Selecting the functions "calibration" and "F" (full rights) "manual operation" • Selecting language See the buttons [Cal] and [Man] to the right of the display. • Parameter 1/2: selecting measured values • Calibration: Press [Cal] to switch to the • Selecting measuring cell calibration menu (the yellow LED illuminates). • Selecting measuring range • Manual operation: Press [Man] to switch to • Configuration of controller 1/2 manual operation (the yellow LED illuminates). • Switching on/off water sensor • Time setting Switching to manual operation is only possible if a controller type has been • Code function selected in the "setup" menu. The Note • Display contrast functions "Cal" and "Man" can only be • Calling up factory settings accessed by persons with calibration • Current output: assignment of current outputs to rights or full rights. measuring ranges • Calling up program version. Calibration function "C" (calibration rights) • Calibration of selected parameters – chlorine, chlorine dioxide, ozone, peroxide, peracetic acid, pH, fluoride. Manual operation "C" (calibration rights) • Running the configured controllers manually. 13.4 Main menu 1. Switch to "main menu" by pressing [OK] or, if necessary, by pressing [Esc] several times. Options in "main menu" main menu controller alarm service setup • "controller" In this submenu, a controller can be parameterised. This option will only be displayed if a type of controller has been selected in the "setup" menu. • "alarm" In this submenu, the measured value is compared with the permitted value, and an alarm is triggered, if necessary. • "service" Diagnosis submenu. The measured values and the last 10 alterations of calibration data can be viewed. In this mode, values cannot be altered. If measuring cell AQC-D2 is selected, the data of the water sensor can be viewed. • "setup" In this submenu, the settings for language, parameters, measuring ranges, controllers, etc. can be adjusted. 27
13.5 Setup All standard settings of the device can be defined in The "setup" menu can only be accessed by persons the "setup" menu. During initial commissioning, basic having full rights. functions are configured which after that should only See section 13.5.8 Code function. be altered rarely or even not at all. Enter the code number. 13.5.1 Selection of measured values for chlorine, chlorine dioxide, ozone, peroxide, peracetic acid, pH, redox and fluoride setup parameter 1 off chlorine chlorine dioxide temp. meas. temp. comp. ozone on on peroxide off off peracetic acid parameter 2 pH-comp. chlorine * on off off temp. meas. temp. meas. pH redox on on fluoride off off TM03 6733 4506 Fig. 20 Selection of measured value in menus "parameter 1" and "parameter 2" * pH-comp. chlorine is only available if chlorine and pH have been selected as parameters. 1. Select the line "parameter 1" / "parameter 2" using the [Up] and [Down] buttons, and press [OK] to switch to the corresponding menu. 2. Use the [Up] and [Down] buttons to select the measured variable (parameter). Press [OK] to confirm. 3. Use the [Up] and [Down] buttons to select further settings. Press [OK] to confirm. 28
13.5.2 pH and temperature compensation The measuring cell that corresponds to the selected measured value can be selected: pH compensation with chlorine measurement • AQC-D1/AQC-D2/AQC-D3 with chlorine, chlorine The dissociation of hypochlorous acid (HOCl) into dioxide or ozone the hypochlorit-anion (OCl–) leads to a loss of slope (loss of sensitivity) of the sensor for chlorine • AQC-D1/AQC-D3, HP with peroxide. measuring. The measuring cell PA (peracetic acid) is selected Using a mathematical compensation function, this automatically if "peracetic acid" is selected as the loss of sensitivity can be eliminated. measured value. Temperature compensation Monitoring the cleaning motor Using a compensation function, the influence of (measuring cells AQC-D1) temperature on the measurement can be eliminated Having selected measuring cell type AQC-D1, press mathematically. [OK] to get to the "cleaning motor" submenu. Control field for pH value, pH compensation, 1. Switch the monitoring of the cleaning motor on/ temperature and temperature compensation in off. the display of measured value Measuring cell AQC-D2 / -D3 selected chlorine chlorine 0.43 mg/l 0.43 mg/l TM03 6708 4506 TM03 6703 4506 5.20 pH c 22 °C c Fig. 21 Chlorine measurement with pH and Fig. 22 Chlorine measurement with AQC-D2 / - temperature compensation D3 • Press [Esc] once or twice to change to the display A black bar on the display indicates the actual data with the measured value. of the water sensor. See sections 13.2 Display elements during initial See installation and operating instructions for the commissioning and 13.6.3 Controller control fields potentiostatic measuring cell AQC-D2 / -D3. on the display. 13.5.4 Setting the measuring ranges for chlorine, In addition to the current measured value, the chlorine dioxide, ozone, peroxide, current pH value and the temperature of the sample peracetic acid, pH, redox and fluoride water are indicated as well. In the "measuring ranges" menu, the corresponding With activated pH compensation or temperature measuring ranges are assigned to the measured compensation, the letter "c" for "compensation" is values which have been selected in the "parameters" displayed at the end of the line. menu. With deactivated temperature measurement, the • In the "measuring ranges" menu, the following water temperature for compensation should be options are available: entered manually after calibration. See section 13.9.2 Calibrating the pH value. – standard measuring ranges – "others." 13.5.3 Selection of measuring cell See section 10.4 Measuring ranges. The "measuring cell" submenu is • The measuring ranges for temperature can be shown only if the measured values defined in °C (Celsius) or in °F (Fahrenheit). Note chlorine, chlorine dioxide, ozone, peroxide or peracetic acid were The "temperature" submenu is only selected during the parameterisation. shown if the temperature measurement Note has been switched on in the Names of the measuring cells contained in the "parameter" menu, or if parameter pH software or fluoride has been selected. • AQC-D1 13.5.5 Controller parameters 1/2 • AQC-D2 See section 13.6 Selection, configuration and • AQC-D3 parameterisation of the controller in order to adjust • HP (peroxide). the controllers. 29
13.5.6 Water sensor 13.5.8 Code function In the "wat. def. sensor" menu, the water sensor can The code function is designed to protect the device be switched on/off. from unauthorised access. In the event of sample-water deficiency, the following Entering the four-digit code number and will occur: changing it • The malfunction-indicating relay is activated, and 1. Select the line "code function" in the "setup" the alarm message "fault water sensor" is menu using the [Up] and [Down] buttons, and triggered. press [OK] to switch to the corresponding menu. • With devices with control functions, the actuators There are two types of access rights: are switched to passive or closed. Calibration rights: The owner of calibration rights The water sensor must be switched on has access to the calibration menu and to manual Caution operation. for the measuring cell AQC-D2! The owner of calibration rights may do the following: 13.5.7 Setting the current time (date/time/summer • alter calibration data time) • operate control relays manually. 1. Switch from the "setup" menu to the "date/time" menu. Full rights: The owner of full rights has full access to all settings including calibration data. The owner of full rights may do the following: time set • alter calibration data • alter setup • parameterise the controller 01 : 18 : 40 pm • alter alarm settings TM03 6709 4506 • run the device manually. cal right time code 0000 Fig. 23 Time setting 2. Set the device to the current time in the "time set" old code menu in the format hh.mm.ss. – Press [OK] to switch to the next number field. 2. Select the desired type of access rights, and The number field which is currently activated press [OK] to get to the corresponding menu. flashes. 3. If an old code exists, confirm the old code number – Alter the settings by pressing the [Up] and first (on first use, confirm code 0000). [Down] buttons. 4. Select "change". – Pressing the buttons continuously makes the 5. Enter the old code in the submenu (see comment adjustment dynamically faster. line at the bottom of the display) using the [Up] 3. Set the device to the current date in the "date" and [Down] buttons, and press [OK] to confirm menu in the format mm.dd.yyyy with English as and switch to the query "new code" (see the operating language or the format dd.mm.yyyy for comment line at the bottom of the display). all other languages similarly to the adjustment of 6. Enter a new four-digit code number using the [Up] time. The corresponding day of the week and [Down] buttons. Press [OK] to confirm. (Mo...Su) is displayed automatically. 7. Repeat to confirm: Set a new four-digit code 4. If necessary, enter the beginning and end date of number using the [Up] and [Down] buttons, and the summer time, the kind of time shift and the press [OK] to confirm. number of hours of the time shift. 8. Access rights can be cancelled by deleting the corresponding code number in the "delete" submenu. • Entering a valid four-digit code number gives the operator access to the corresponding functions for a limited time of 60 minutes. • If no code was entered previously, the desired menu can be accessed without restrictions. The code number 0000 will not be displayed / the operator will not be asked to enter it. 30
13.5.11 Factory setting reset If a wrong code is entered, access to the corresponding menus is denied. An In the "factory setting" submenu, the Note error message is triggered Conex® DIA-2Q can also be reset to the factory (duration: five seconds), and the device setting using code 6742. will return to "main menu". Only use this function in an emergency. Reset function All device settings are lost and must be • Entering the code number "1998" deletes all Caution re-entered! previously entered code numbers. Do not disconnect the device from the • All prior access codes are deleted and reset to power supply during the reset! "0000". Resetting the Conex® DIA-2Q to the factory 13.5.9 Adjusting display contrast setting In the "display" menu, the contrast of the display can • Reset be adjusted. - Code: 0000 is displayed. 1. Switch from the "setup" menu to the "display" - Set code 6742 using the [Up] and [Down] menu using the [Up] and [Down] buttons, and buttons, and press [OK] to confirm. press [OK] to confirm. The device is now returned to the original factory setting. If the contrast setting is too high, it may not be possible to read the dis- Before subsequent start-up: Note play. Hold down the [Down] button Caution Check all parameters and set the device until the contrast setting is low enough again according to your application! to be read again. General factory settings 13.5.10 Saving / accessing user settings • Language**: The Conex® DIA-2Q is waiting for In the "factory setting" submenu, the current device language input. The languages "Deutsch/English setting can be saved so that it can be reactivated Français...." are indicated alternately. later, or a saved setting can be activated. • Summer time: off. • Setup • Code numbers: 0000 calibration rights, 0000 full - Save setup: Saves all current device settings rights. from the menus (not just the "setup" settings). • Current outputs 1-4: 0-20 mA. - Activate setup: Resets the device to the last • Dosing monitoring: off. saved setup. • Parameter 1: chlorine. In this menu, save your device settings • Parameter 2: off. once you have set all the values in the ** For the first calibration, the default value = "setup", "controller" and "alarm" last setting. Note menus. You can then activate them again at any time (even after a factory reset)! Default values for parameter group 1 Cl2 ClO2/O3 PAA H2O2 Measuring range [mg/l] 0.00 - 2.00 0.00 - 1.00 0-100 0-100 PA Measuring cell AQC-D1 AQC-D1 AQC-D1 (peracetic acid) Temperature measurement off off off off Water sensor off off off off Default values for parameter group 2 pH Redox Fluoride Measuring range 0.00 - 14.00 0-1000 0.00 - 2.00 Temperature measurement off off off Water sensor off off off 31
13.5.12 Current outputs In the "current output" menu, the operator can assign • Enter the associated start-of-scale current value the selected values to the current outputs and then using the [Up] button. Press [OK] to confirm. assign the current outputs to the selected measuring – An additional line is displayed with the scan for ranges. the full-scale value of the range "20.00 mg/l". In the "current output" menu, the selected • Enter the full-scale value of the range using the parameters are listed in the following order: [Down] button. Press [OK] to confirm. – In the same line, "= 20 mA" is displayed. current output chlorine chlorine pH 0.20 mg/l = 0 mA temperature 15.00 mg/l = 20 mA • Output 1: parameter group 1: chlorine, chlorine dioxide, ozone, hydrogen, peroxide and peracetic • Enter the associated full-scale current value using acid the [Down] button. Press [OK] to confirm and • Output 2: parameter group 2: pH, redox, fluoride return to the "current output" menu. • Output 3: temperature or continuous controller for parameters from group 2 13.5.13 Program version • Output 4: continuous controller for parameters from group 1. program version Assignment of the current outputs to the measuring range (example: chlorine Conex DIA-2Q measurement) There are two standard ranges, "0-20 mA" and "4-20 mA" which can be assigned to the current v0.20.1 20030304 outputs. They are assigned to the measuring range linearly. In case problems make it necessary to contact our The start-of-scale and full-scale values can be freely service department, the "program version" menu defined within 0 to 20 mA in an optional setting for enables the operator to easily identify the software the current outputs 1 and 2 as well as for current components of the actual device. output 3 with temperature measurements. • Version, for example v0.20.1. • Last update: yyyymmdd. chlorine 0-20 mA 4-20 mA others • In the corresponding submenu (for example chlorine), select the line "others" using the [Up] and [Down] buttons. Press [OK] to confirm. • In the subsequent display "0.00 mg/l", enter the start-of-scale value of the range using the [Up] button. Press [OK] to confirm. – In the same line, "= 0 mA" is displayed. chlorine 0.20 mg/l = 0 mA 32
13.6 Selection, configuration and parameterisation of the controller It takes two steps to set up the controller: • First: selection and configuration of the controller type in the "setup" menu, "controller" submenu. • Secondly: parameterisation of the selected controller type in "main menu", "controller" submenu. 13.6.1 Selection and configuration of the controller setup controller parameter 1 current input off prop. controller parameter 1 proport. contrl. 0-20 mA interpulse ctrl. 4-20 mA pulse freq. ctrl. others low mA cont. controller high mA setpt controller Limit contact off downward viol. limit contact upward violation setpoint interpulse ctrl. contrl. pulse freq. ctrl. cont. controller ctrl. direction ctrl. charact. downward control P upward control PI PID combined comp. contrl. controller ctrl. direction ctrl. charact. current off downward input P interpulse ctrl. control PI 0-20 mA low mA pulse freq. ctrl. upward control PID 4-20 mA high mA controller cont. controller others parameter 2 for mA, select values parameter 2 with [Up] and [Down] TM03 6740 4506 as controller buttons parameter 1 Fig. 24 Selection and configuration of the controller • In the "setup" menu, select the line "controller parameter 1" or "controller parameter 2" using the [Up] and [Down] buttons, and select the menu with the same name using [OK]. 33
You can also read